Summary

Medición de la extracelular Ion fundentes Uso del referencias Ion selectivo microelectrodos Técnica

Published: May 03, 2015
doi:

Summary

Transporters in cell membranes allow differential segregation of ions across cell membranes or cell layers and play crucial roles during tissue physiology, repair and pathology. We describe the ion-selective self-referencing microelectrode that allows the measurement of specific ion fluxes at single cells and tissues in vivo.

Abstract

Las células de animales, plantas y células individuales están encerradas por una barrera llamada la membrana celular que separa el citoplasma desde el exterior. Capas de células tales como epitelios también forman una barrera que separa el interior del exterior o diferentes compartimentos de organismos multicelulares. Una característica clave de estas barreras es la distribución diferencial de iones a través de membranas celulares o capas de células. Dos propiedades permiten esta distribución: 1) membranas y epitelios presentan permeabilidad selectiva a iones específicos; 2) iones son transportados a través de bombas a través de las membranas celulares y capas de células. Estas propiedades juegan un papel crucial en el mantenimiento de la fisiología del tejido y actúan como una señal de señales después de los daños, durante la reparación, o bajo condición patológica. La referencia a sí misma microelectrodo selectiva de iones permite mediciones de flujos específicos de iones tales como calcio, potasio o sodio en los niveles de células y tejidos individuales. El microelectrodo contiene un cóctel ionóforo que espermeable selectivamente a un ion específico. La solución de relleno interna contiene una concentración conjunto de los iones de interés. El potencial eléctrico de la microelectrodo se determina por la concentración fuera del ión. Como la concentración de iones varía, el potencial de la microelectrodo cambia como una función del logaritmo de la actividad de los iones. Cuando se mueve hacia atrás y adelante cerca de una fuente o sumidero del ion (es decir, en un gradiente de concentración debido a flujo de iones) el potencial microelectrodo fluctúa a una amplitud proporcional al flujo de iones / gradiente. El amplificador amplifica la señal de microelectrodos y la salida se registra en el ordenador. El flujo de iones se puede calcular entonces por la ley de difusión de Fick usando la fluctuación electrodo de potencial, la excursión de microelectrodo, y otros parámetros tales como la movilidad de iones específico. En este trabajo se describe en detalle la metodología para medir los flujos de iones extracelulares utilizando la autorreferencia microelectrodo un ion-selectivod presentar algunos resultados representativos.

Introduction

Todas las células animales están rodeadas por una membrana de bicapa lipídica que separa el citoplasma del ambiente exterior. La célula mantiene un potencial de membrana eléctrica, negativa en el interior, por el transporte activo de iones 1. El potencial de membrana es una fuente de energía almacenada que la célula puede utilizar para operar diversos dispositivos moleculares en la membrana 2. Las neuronas y otras células excitables tienen grandes potenciales de membrana. Apertura rápida de los canales de sodio se colapsa el potencial de membrana (despolarización) y produce el potencial de acción que es transportado a lo largo de la longitud de la neurona 2. Aparte de estos rápidos cambios eléctricos, muchos tejidos y órganos generan y mantienen los potenciales eléctricos significativos a largo plazo. Por ejemplo, la piel y el epitelio corneal generar y mantener los potenciales de trans-epiteliales y corrientes eléctricas extracelulares por bombeo direccional de iones (principalmente sodio y cloruro) 3.

tienda de campaña "> Mientras que las mediciones de corriente eléctrica extracelular endógena usando la sonda vibrante 4-6 y mediciones de la membrana o trans-epiteliales potenciales utilizando el sistema de microelectrodo 7-10 permitir la medición de los parámetros eléctricos de las membranas celulares y capas de células epiteliales, no dan indicación de las especies de iones implicados.

Microelectrodos con ionóforo selectivo pueden medir la concentración de iones específicos en solución. Gradientes de iones o de flujo podrían ser medidos con dos o más electrodos en diferentes posiciones. Sin embargo, la variación de la tensión intrínseca de cada sonda sería diferente, provocando mediciones inexactas o incluso la detección de un gradiente de que no estaba presente. Un único electrodo utilizado en el modo de "auto-referenciación" por la cual se mueve a baja frecuencia entre dos puntos resuelve este problema. Ahora el flujo de iones puede ser visto en el contexto de una deriva de señal relativamente lento y estable (véase la Figura 3B). </p>

El sistema de medición sensible a los iones utiliza microelectrodos de autorreferencia selectivos de iones para detectar pequeños flujos extracelulares de iones cerca de los tejidos o células individuales. El sistema consta de un amplificador que procesa la señal desde el microelectrodo y un motor paso a paso y micro controlador para controlar el movimiento de la microelectrodo. El microelectrodo selectiva de iones y el electrodo de referencia que cerrar el circuito están conectados al amplificador a través de un cabezal de la platina pre-amplificador (Figura 1A). Software de ordenador determina los parámetros del movimiento microelectrodo (frecuencia, distancia) y también registra la salida del amplificador. El motor paso a paso controla el movimiento a través de un microelectrodo microposicionador tridimensional. Una baja frecuencia de vibración de microelectrodos selectivo de iones fue desarrollado por primera vez en 1990 para medir el flujo de calcio específico 11. Así como el calcio, cócteles ionóforos comercialmente accesibles ahora están disponibles para hacer microelectrodes sensible al sodio, cloruro, potasio, hidrógeno, magnesio, nitrato, amonio, fluoruro, litio o mercurio.

Básicamente, la técnica de microelectrodo selectiva de iones de autorreferencia convierte la actividad de un ion específico disuelto en una solución en un potencial eléctrico, que puede ser medida por un voltímetro. El cóctel de ionóforo es un líquido inmiscible fase (orgánica, lipófilo) con propiedades de intercambio iónico. El ionóforo complejos selectivamente (se une) iones específicos reversible y los transfiere entre la solución acuosa contenida en el microelectrodo (electrolito) y la solución acuosa en la que se sumerge el microelectrodo (Figura 1D). Esta transferencia de iones conduce a un equilibrio electroquímico y una variación del potencial eléctrico entre el microelectrodo y el electrodo de referencia se mide por el voltímetro. El voltaje es proporcional al logaritmo de la actividad del ión específico de acuerdo con la Nernst equation permitiendo el cálculo de la concentración de iones (Figura 2A y B).

En la actualidad, varios sistemas permiten la medición del flujo de iones usando un concepto o principio similar. Por ejemplo, la técnica de electrodo de exploración de ion selectivo (SIET) 12,13 o la técnica de microelectrodos Ion Flux Estimation (MIFE) desarrollado por Newman y Shabala 14-16 están comercialmente disponibles y ampliamente utilizados por la comunidad de investigación con el fin de determinar ion específica fundentes que ocurre en la membrana celular y el tejido a través de una variedad de modelos de células vivas individuales animal, vegetal y. Microelectrodos selectivos de iones se han utilizado para medir hidrógeno, potasio y flujo de calcio a través de raíces de la planta 17, cloruro de flujo en las arterias cerebrales de rata 18 y 19 en los tubos de polen, flujo de hidrógeno en células de la retina del patín 20, el flujo de calcio en el hueso del ratón 21, vario ion flujos de hifas fúngicas 22 y en ren la córnea 23, y finalmente el flujo de calcio en las células sola herida de curación 12,24. Ver también la siguiente revisión para obtener información detallada sobre autorreferencia microelectrodos selectivos de iones 25.

El siguiente artículo describe con detalle cómo preparar y llevar a cabo la medición de los flujos de iones extracelulares endógenos mediante la técnica de microelectrodos autorreferencia selectivo de iones a nivel de células individuales.

Protocol

1. Ion selectivo autorreferencia Preparación microelectrodos Preparación de microelectrodos selectivo de iones Heat tire delgadas capilares de borosilicato de paredes sin filamento (1,5 mm de diámetro exterior, 1,12 mm de diámetro interior) con un extractor de microelectrodos. Nota: Esto le da consejos de 3-4 m de diámetro. Consejos más pequeñas tienen mayor resistencia que hace microelectrodos más susceptibles al ruido electrónico y también está asociado con una respuesta más lent…

Representative Results

Hemos demostrado anteriormente que la entrada de calcio aparece después de una sola célula hiriendo a 24. Por lo tanto, preguntamos si otros flujos de iones se producen en una sola herida celular. Utilizamos la X. laevis ovocito, un modelo bien establecido para celular herida única curación 30-34 y registro electrofisiológico 24,35-39. Curiosamente, los iones de potasio son más concentrada dentro de X. laevis (aproximadamente 110 mm) 40 que en la soluci…

Discussion

Los pasos más críticos para la medición exitosa de los flujos de iones extracelulares in vivo son: la reducción del ruido, la fabricación correcta de los microelectrodos y referencia selectivos de iones del electrodo, y el posicionamiento de la muestra y ambos electrodos.

Con el fin de minimizar el ruido, el sistema de grabación debe estar en una puesta a tierra (puesta a tierra) jaula de Faraday preferiblemente con una mesa (aislamiento de vibraciones) de metal-rematada que t…

Divulgations

The authors have nothing to disclose.

Acknowledgements

This work was supported by National Science Foundation grant MCB-0951199, and in part by the NIH grant EY01910, California Institute of Regenerative Medicine grants RB1-01417 and by the Fundação para a Ciência e Tecnologia (FCT) grant SFRH/BD/87256/2012.

Materials

IonAmp   BioCurrents Research Center, Woods Hole, MA, USA none amplifier created by the BioCurrents Research Center, Woods Hole, MA, USA; Similar system can be purchased from “XBL function matters” (http://www.xuyue.org/) or from “YoungerUSA” (http://www.youngerusa.com/) or from Applicable Electronics(http://www.applicableelectronics.com/)
IonAmp32   BioCurrents Research Center, Woods Hole, MA, USA none software created by the BioCurrents Research Center, Woods Hole, MA, USA; Similar system can be purchased from “XBL function matters” (http://www.xuyue.org/) or from “YoungerUSA” (http://www.youngerusa.com/) or from Applicable Electronics(http://www.applicableelectronics.com/)
Headstage pre-amplifier  BioCurrents Research Center, Woods Hole, MA, USA INA116 BSR Voltage Follower INA116, designed by the BioCurrents Research Center, Woods Hole, MA, USA; Similar system can be purchased from “XBL function matters” (http://www.xuyue.org/) or from “YoungerUSA” (http://www.youngerusa.com/) or from Applicable Electronics(http://www.applicableelectronics.com/)
MicroStep Driver  BioCurrents Research Center, Woods Hole, MA, USA none three MicroStep drivers are required for X, Y and Z-positioning; created by the BioCurrents Research Center, Woods Hole, MA, USA; Similar system can be purchased from “XBL function matters” (http://www.xuyue.org/) or from “YoungerUSA” (http://www.youngerusa.com/) or from Applicable Electronics(http://www.applicableelectronics.com/)
Manual micropositioner   World Precision Instruments  Model KITE-R Similar system can be purchased from Applicable Electronics(http://www.applicableelectronics.com/)
Magnetic stand    World Precision Instruments Model M10 Similar system can be purchased from Applicable Electronics(http://www.applicableelectronics.com/)
Vibration isolation table   Newport Inc.      Model VW-3036-OPT-023040 Similar system can be purchased from Applicable Electronics(http://www.applicableelectronics.com/)
Part of three dimentional micropositioner: angle bracket, 90°, slotted faces Newport Inc.      Model 360-90 Assemblage of the three dimantionnal micropositionner requires also Three electric rotary motors for X, Y, Z control, MPH-1 mounting arm with MCA-2 adjustable-angle post and Various Newport connectors and screws to bolt onto vibration table
Part of three dimentional micropositioner: Peg-Joining Dovetail Stage 0.5 inch X Travel Newport Inc.      460PD-X none
Part of three dimentional micropositioner: Quick-Mount Linear Stage, 0.5 inch XY Travel Newport Inc. 460A-XY none
Kwik-Fil thin walled borosilicate glass capillaries without filament  World Precision Instruments  TW150-4 none
Electrode puller  Narishige  PC-10 none
Metal rack Made in-house none Metal electrode holder made in-house by drilling 2 mm wide holes half centimeter spaced in a 10cm by 15cm rectangular base of steel
Oven QL Model 10 Lab Oven none
Silanization solution I  Sigma-Aldrich 85126 Hazardous, handle as recommended by provider 
Glass Petri dish; Pyrex Fisher Scientific 316060 none
Electrode/micropipette storage jar World Precision Instruments  E215 none
Glass dessicator Fisher Scientific 08-595E Contains Drierite dessicant (W.A. Hammond Drierite Co. Ltd, Xenia, OH, USA). Place petroleum jelly on the seal to make it airtight.
Plastic Pasteur pipette  Fisher Scientific 11597722 none
Bunsen burner Fisher Scientific S97329 none
Microscope slide Sigma-Aldrich S8902 none
Straight microelectrode holder Warner Instruments QSW-A15P with a gold 1 mm male connector and Ag/AgCl wire
Straight microelectrode holder  World Precision Instruments MEH3S with a AgCl(Ag+)pellet inside and a gold 2 mm male connector 
6 cm Petri dish VWR 60872-306 none
Nitex mesh Dynamic Aqua-Supply Ltd. NTX750 none
Glue; Loctite epoxy VWR 500043-451 Mix glue and hardener in equal parts in a plastic weighing boat and mix thoroughly. Sets quickly but leave at RT for 24 h for full curing
Deionized water  Sigma-Aldrich 99053 none
Sodium Chloride Sigma-Aldrich S7653 none
Potassium Chloride Sigma-Aldrich P9333 none
Calcium Chloride Sigma-Aldrich C1016 none
Magnesium Chloride Sigma-Aldrich M8266 none
Hepes Sigma-Aldrich H3375 none
Sodium Hydroxyde Sigma-Aldrich S8045 none
Potassium Acetate Sigma-Aldrich P1190 none
Agarose Sigma-Aldrich A9539 none

References

  1. Weber, W. M., Liebold, K. M., Clauss, W. Amiloride-sensitive Na+ conductance in native Xenopus oocytes. Biochimica et biophysica acta. 1239, 201-206 (1995).
  2. McCaig, C. D., Song, B., Rajnicek, A. M. Electrical dimensions in cell science. Journal of cell science. 122, 4267-4276 (2009).
  3. Zhao, M. Electrical fields in wound healing-An overriding signal that directs cell migration. Seminars in cell & developmental biology. 20, 674-682 (2009).
  4. Jaffe, L. F., Nuccitelli, R. An ultrasensitive vibrating probe for measuring steady extracellular currents. The Journal of cell biology. 63, 614-628 (1974).
  5. Reid, B., Nuccitelli, R., Zhao, M. Non-invasive measurement of bioelectric currents with a vibrating probe. Nature protocols. 2, 661-669 (2007).
  6. Reid, B., Zhao, M. Measurement of bioelectric current with a vibrating probe. Journal of visualized experiments : JoVE. , (2011).
  7. Neher, E., Sakmann, B. Single-channel currents recorded from membrane of denervated frog muscle fibres. Nature. 260, 799-802 (1976).
  8. Moore, J. W. The patch clamp: single-channel recording. Science. 224, 50-51 (1984).
  9. Brown, A. L., Johnson, B. E., Goodman, M. B. Patch clamp recording of ion channels expressed in Xenopus oocytes. Journal of visualized experiments. , (2008).
  10. McCaig, C. D., Robinson, K. R. The ontogeny of the transepidermal potential difference in frog embryos. Developmental biology. 90, 335-339 (1982).
  11. Kuhtreiber, W. M., Jaffe, L. F. Detection of extracellular calcium gradients with a calcium-specific vibrating electrode. The Journal of cell biology. 110, 1565-1573 (1990).
  12. Cai, G., Cresti, M., Moscatelli, A. The use of the vibrating probe technique to study steady extracellular currents during pollen germination and tube growth. Fertilisation in Higher Plants: molecular and cytological aspects. , 235-252 (1999).
  13. Kunkel, J. G., Xu, Y., Shipley, A. M., Feijó, J. A. The use of non-invasive ion-selective microelectrode techniques for the study of plant development. Plant Electrophysiology – Theory and Methods. (ed Volkov AG. , 109-137 (2006).
  14. Ordonez, N. M., Shabala, L., Gehring, C., Shabala, S. Noninvasive microelectrode ion flux estimation technique (MIFE) for the study of the regulation of root membrane transport by cyclic nucleotides. Methods in molecular biology. 1016, 95-106 (2013).
  15. Tegg, R. S., Melian, L., Wilson, C. R., Shabala, S. Plant cell growth and ion flux responses to the streptomycete phytotoxin thaxtomin A: calcium and hydrogen flux patterns revealed by the non-invasive MIFE technique. Plant & cell physiology. 46, 638-648 (2005).
  16. Newman, I. A. Ion transport in roots: measurement of fluxes using ion-selective microelectrodes to characterize transporter function. Plant, cell & environment. 24, 1-14 (2001).
  17. Kochian, L. V., Shaff, J. E., Kuhtreiber, W. M., Jaffe, L. F., Lucas, W. J. Use of an extracellular, ion-selective, vibrating microelectrode system for the quantification of K(+), H (+), and Ca (2+) fluxes in maize roots and maize suspension cells. Planta. 188, 601-610 (1992).
  18. Doughty, J. M., Langton, P. D. Measurement of chloride flux associated with the myogenic response in rat cerebral arteries. The Journal of physiology. 534, 753-761 (2001).
  19. Messerli, M. A., Smith, P. J., Lewis, R. C., Robinson, K. R. Chloride fluxes in lily pollen tubes: a critical reevaluation. The Plant journal : for cell and molecular biology. 40, 799-812 (2004).
  20. Molina, A. J., et al. Neurotransmitter modulation of extracellular H+ fluxes from isolated retinal horizontal cells of the skate. The Journal of physiology. 560, 639-657 (2004).
  21. Marenzana, M., Shipley, A. M., Squitiero, P., Kunkel, J. G., Rubinacci, A. Bone as an ion exchange organ: evidence for instantaneous cell-dependent calcium efflux from bone not due to resorption. Bone. 37, 545-554 (2005).
  22. Lew, R. R. Ionic currents and ion fluxes in Neurospora crassa hyphae. Journal of experimental botany. 58, 3475-3481 (2007).
  23. Vieira, A. C., et al. Ionic components of electric current at rat corneal wounds. PloS one. 6, e17411 (2011).
  24. Luxardi, G., Reid, B., Maillard, P., Zhao, M. Single cell wound generates electric current circuit and cell membrane potential variations that requires calcium influx. Integrative biology : quantitative biosciences from nano to macro. 6, 662-672 (2014).
  25. Smith, P. J. S., Sanger, R. H., Messerli, M. A., Michael, A. C., Borland, L. H. . Electrochemical Methods for Neuroscience. , (2007).
  26. Smith, P. J., Hammar, K., Porterfield, D. M., Sanger, R. H., Trimarchi, J. R. Self-referencing, non-invasive, ion selective electrode for single cell detection of trans-plasma membrane calcium flux. Microscopy research and technique. 46, 398-417 (1999).
  27. Messerli, M. A., Smith, P. J. Construction theory, and practical considerations for using self-referencing of Ca(2+)-selective microelectrodes for monitoring extracellular Ca(2+) gradients. Methods in cell biology. 99, 91-111 (2010).
  28. Chambers, J., Hastie, T., Pregibon, D., Momirović, K., Mildner, V. Ch. 48. Compstat. , 317-321 (1990).
  29. Chambers, J. M., Cleveland, W. S., Kleiner, B., Tukey, P. A. . Graphical methods for data analysis. , (1983).
  30. Burkel, B. M., Benink, H. A., Vaughan, E. M., von Dassow, G., Bement, W. M. A Rho GTPase signal treadmill backs a contractile array. Developmental cell. 23, 384-396 (2012).
  31. Bement, W. M., Mandato, C. A., Kirsch, M. N. Wound-induced assembly and closure of an actomyosin purse string in Xenopus oocytes. Current biology : CB. 9, 579-587 (1999).
  32. Mandato, C. A., Bement, W. M. Contraction and polymerization cooperate to assemble and close actomyosin rings around Xenopus oocyte wounds. The Journal of cell biology. 154, 785-797 (2001).
  33. Benink, H. A., Bement, W. M. Concentric zones of active RhoA and Cdc42 around single cell wounds. The Journal of cell biology. 168, 429-439 (2005).
  34. Simon, C. M., Vaughan, E. M., Bement, W. M., Edelstein-Keshet, L. Pattern formation of Rho GTPases in single cell wound healing. Molecular biology of the cell. 24, 421-432 (2013).
  35. Petersen, C. C., Dupont, G. The initiation of a calcium signal in Xenopus oocytes. Cell calcium. 16, 391-403 (1994).
  36. Horisberger, J. D., Lemas, V., Kraehenbuhl, J. P., Rossier, B. C. Structure-function relationship of Na,K-ATPase. Annual review of physiology. 53, 565-584 (1991).
  37. Miledi, R. A calcium-dependent transient outward current in Xenopus laevis oocytes. Proceedings of the Royal Society of London. Series B, Containing papers of a Biological character. Royal Society. , 491-497 (1982).
  38. Miledi, R., Parker, I. Chloride current induced by injection of calcium into Xenopus oocytes. The Journal of physiology. 357, 173-183 (1984).
  39. Parker, I., Miledi, R. A calcium-independent chloride current activated by hyperpolarization in Xenopus oocytes. Proceedings of the Royal Society of London. Series B, Containing papers of a Biological character. Royal Society. 233, 191-199 (1988).
  40. Costa, P. F., Emilio, M. G., Fernandes, P. L., Ferreira, H. G., Ferreira, K. G. Determination of ionic permeability coefficients of the plasma membrane of Xenopus laevis oocytes under voltage clamp. The Journal of physiology. 413, 199-211 (1989).
  41. Adams, D. S., Levin, M. General principles for measuring resting membrane potential and ion concentration using fluorescent bioelectricity reporters. Cold Spring Harbor protocols. 2012, 385-397 (2012).
  42. Porterfield, D. M. Measuring metabolism and biophysical flux in the tissue, cellular and sub-cellular domains: recent developments in self-referencing amperometry for physiological sensing. Biosensors. 22, 1186-1196 (2007).
  43. McLamore, E. S., et al. A self-referencing glutamate biosensor for measuring real time neuronal glutamate flux. Journal of neuroscience methods. 189, 14-22 (2010).
  44. Yin, M., et al. Highly sensitive and fast responsive fiber-optic modal interferometric pH sensor based on polyelectrolyte complex and polyelectrolyte self-assembled nanocoating. Analytical and bioanalytical chemistry. 399, 3623-3631 (2011).
  45. Chatni, M. R., Porterfield, D. M. Self-referencing optrode technology for non-invasive real-time measurement of biophysical flux and physiological sensing. The Analyst. 134, 2224-2232 (2009).
  46. McLamore, E. S., Jaroch, D., Chatni, M. R., Porterfield, D. M. Self-referencing optrodes for measuring spatially resolved, real-time metabolic oxygen flux in plant systems. Planta. 232, 1087-1099 (2010).
  47. McLamore, E. S., et al. A self referencing platinum nanoparticle decorated enzyme-based microbiosensor for real time measurement of physiological glucose transport. Biosensors & bioelectronics. 26, 2237-2245 (2011).
check_url/fr/52782?article_type=t

Play Video

Citer Cet Article
Luxardi, G., Reid, B., Ferreira, F., Maillard, P., Zhao, M. Measurement of Extracellular Ion Fluxes Using the Ion-selective Self-referencing Microelectrode Technique. J. Vis. Exp. (99), e52782, doi:10.3791/52782 (2015).

View Video