Summary

Entièrement automatisé centrifuge microfluidique Dispositif pour Ultrasensitive Protein Détection de sang total

Published: April 16, 2016
doi:

Summary

This protocol demonstrates how to achieve femto molar detection sensitivity of proteins in 10 µL of whole blood within 30 min. This can be achieved by using electrospun nanofibrous mats integrated in a lab-on-a-disc, which offers high surface area as well as effective mixing and washing for enhanced signal-to-noise ratio.

Abstract

Enzyme-linked immunosorbent assay (ELISA) is a promising method to detect small amount of proteins in biological samples. The devices providing a platform for reduced sample volume and assay time as well as full automation are required for potential use in point-of-care-diagnostics. Recently, we have demonstrated ultrasensitive detection of serum proteins, C-reactive protein (CRP) and cardiac troponin I (cTnI), utilizing a lab-on-a-disc composed of TiO2 nanofibrous (NF) mats. It showed a large dynamic range with femto molar (fM) detection sensitivity, from a small volume of whole blood in 30 min. The device consists of several components for blood separation, metering, mixing, and washing that are automated for improved sensitivity from low sample volumes. Here, in the video demonstration, we show the experimental protocols and know-how for the fabrication of NFs as well as the disc, their integration and the operation in the following order: processes for preparing TiO2 NF mat; transfer-printing of TiO2 NF mat onto the disc; surface modification for immune-reactions, disc assembly and operation; on-disc detection and representative results for immunoassay. Use of this device enables multiplexed analysis with minimal consumption of samples and reagents. Given the advantages, the device should find use in a wide variety of applications, and prove beneficial in facilitating the analysis of low abundant proteins.

Introduction

Plusieurs plates – formes pour le diagnostic de la maladie ont été élaborés à base de matériaux à l' échelle nanométrique 1,2 tels que des nanofils, 3 nanoparticules, 4 nanotubes, 5 et nanofibres (SNF) 6-8. Ces nanomatériaux offrent d'excellentes perspectives dans la conception de nouvelles technologies pour les essais biologiques très sensibles en raison de leurs propriétés physico-chimiques uniques. Par exemple, les nanofibres d'oxyde de zinc mésoporeuse ont été utilisées pour la femtoseconde molaire détection sensible des marqueurs biologiques du cancer du sein. 9 Récemment, des nanomatériaux à base de dioxyde de titane (TiO 2) ont été étudiées pour des applications bioanalytiques 10 compte tenu de leur stabilité chimique, 11 négligeable dénaturation des protéines 12 et de la biocompatibilité. 13 En outre, les groupes hydroxyle sur la surface de TiO 2 faciliter la modification chimique et la liaison covalente de biomolécules. 14,15 modelée TiO2 thin films 16 ou 17 nanotubes de TiO 2 ont été utilisés pour améliorer la sensibilité de la détection d'une protéine cible en augmentant la zone de surface; Cependant, le procédé de fabrication est relativement complexe et nécessite un équipement coûteux. D'autre part, FNs électrofilées bénéficient d'une attention en raison de leur grande surface ainsi que processus simple et la fabrication à faible coût; 18,19 encore, la caractéristique fragile ou lâche du électrofilés TiO 2 mat NF rend difficile à manipuler et intégrer avec des dispositifs microfluidiques. 6,20 par conséquent, les TiO 2 tapis NF ont rarement été utilisés dans des applications bioanalytiques, en particulier celles qui nécessitent des conditions de lavage difficiles.

Dans cette étude, pour surmonter ces limitations, nous avons développé une nouvelle technologie pour le transfert du électrofilés NF Tapis sur la surface d'un substrat cible en utilisant un polydiméthylsiloxane mince (PDMS) couche adhésive. Furthermore, nous avons démontré avec succès l'intégration de électrofilée TiO2 NF mats sur un dispositif microfluidique centrifuge en polycarbonate (PC). En utilisant ce dispositif, une détection de haute sensibilité, entièrement automatisé et intégré de la protéine C-réactive (CRP), ainsi que la troponine I cardiaque (cTnI) a été réalisée dans les 30 minutes à partir de seulement 10 pi de sang total. 21 En raison de la combinaison avantages des propriétés de la FN et la plate-forme centrifuge, l'essai ont montré une large plage dynamique de six ordres de grandeur de 1 pg / ml (~ 8 fM) à 100 ng / ml (~ 0,8 pM) avec une limite inférieure de détection de 0,8 pg / ml (~ 6 fM) pour CRP et une gamme dynamique de 10 pg / ml (~ 0,4 pM) à 100 ng / ml (~ 4 nM) avec une limite de détection de 37 pg / ml (~ 1,5 pM) pour cTnI. Ces limites de détection sont ~ 300 et ~ 20 fois plus faible par rapport à leurs résultats ELISA classiques correspondants. Cette technique pourrait être appliquée à la détection de toutes les protéines cibles, avec des anticorps appropriés. Dans l'ensemble, cette coopération de l'appareilULD contribuent grandement au diagnostic in vitro et des essais biochimiques , car il peut détecter des quantités rares de protéines cibles avec une grande sensibilité même de très petites quantités d'échantillons biologiques, par exemple, 10 pl de sang total. Bien que nous ne démontré la détection des protéines sériques par ELISA dans cette étude, la technologie de transfert et l'intégration des électrofilés FNs avec des dispositifs microfluidiques pourrait être appliquée plus largement dans d'autres réactions biochimiques qui nécessitent une grande surface pour une sensibilité de détection élevée.

Protocol

REMARQUE: On prélève du sang d'individus sains et a été recueilli dans un tube de collecte de sang. Le consentement éclairé écrit a été obtenu à partir de tous les bénévoles. 1. Fabrication de TiO 2 Mat NF Préparation de la solution de précurseur 22 Dissoudre 1,5 g de tétraisopropylate de titane (TTIP) dans un mélange d'éthanol (99,9%, 3 ml) et de l'acide acéti…

Representative Results

En utilisant ce protocole, un dispositif microfluidique centrifuge entièrement automatisé pour la détection des protéines du sang entier avec une sensibilité élevée a été préparé. Les TiO 2 tapis NF ont été préparés par des procédés de électrofilage et calcination. Afin de fabriquer le FNs de diamètre désiré, la morphologie et l'épaisseur, des conditions telles que le débit, la tension et le temps filer électrofilage ont été optimisés. Lorsque le…

Discussion

Le dosage de TiO 2 NF disque intégré est une technique rapide, économique et pratique pour la détection ultrasensible de faible abondance des protéines présentes dans un volume très faible de sang. Cette technique présente l'avantage d'utiliser des volumes d'échantillons de petite taille (10 pi) et se prête à l'analyse de plusieurs échantillons simultanément. Cela fournit un grand potentiel en tant que dispositif de multiplexage de dosage immunologique. Le dispositif a l'avant…

Divulgations

The authors have nothing to disclose.

Acknowledgements

Ce travail a été soutenu par le National Research Foundation de Corée (NRF) subventions (2013R1A2A2A05004314, 2012R1A1A2043747), une subvention de la Korean Health Technology R & D Project, Ministère de la Santé et Bien-être (A121994) et IBS-R020-D1 financé par le gouvernement coréen.

Materials

Si wafer LG SILTRON Polished Wafer, test grade Dia. (mm) = 150, orientation = <100>, dopant = boron, RES(Ohm-cm) = 1 – 30, thickness (μm) = 650 – 700
Polycarbonate (PC)  Daedong Plastic PCS#6900 Thickness (mm) = 1 and 5 
Titanium tetraisopropoxide, 98%, Sigma-Aldrich 205273
Polyvinylpyrrolidone, Mw = 1,300,000 Sigma-Aldrich 437190
Acetic acid Sigma-Aldrich 320099
Anhydrous ethanol Sigma-Aldrich 459836
Tridecafluoro-1,1,2,2-tetrahydrooctyl)-1-trichlorosilane Sigma-Aldrich 448931
PDMS and curing agent Dow Corning SYLGARD 184
GPDES Gelest Inc SIG5832.0 
Ethanol J T Baker
FE-SEM FEI Nova NanoSEM
X-ray photoelectron spectroscopy ThermoFisher K-alpha
3D modeling machine M&I CNC Lab, Korea CNC milling machine
Wax-dispensing machine Hanra Precision Eng. Co. Ltd., Korea Customized
Double-sided adhesive tape FLEXcon, USA DFM 200 clear 150 POLY H-9 V-95
Cutting plotter Graphtec Corporation, Japan Graphtec CE3000-60 MK2
Spin coater MIDAS SPIN-3000D
Furnace (calcination) R. D. WEBB COMPANY WEBB 99
Rheometer (Tack test) Thermo Scientific Haake MARS III – ORM Package
Oxygen plasma system FEMTO CUTE
Monoclonal mouse antihuman hsCRP Hytest Ltd., Finland 4C28 (clone # C5)
Monoclonal mouse anti-cTnI Hytest Ltd., Finland 4T21 (clone # 19C7)
HRP conjugated goat polyclonal anti-hsCRP Abcam plc., MA ab19175
HRP conjugated mouse monoclonal anti-cTnI Abcam plc., MA ab24460 (clone # 16A11)
hsCRP Abcam plc., MA ab111647
cTnI Fitzgerald, MA 30-AT43
Bovine Albumin Sigma-Aldrich A7906
PBS Amresco Inc E404
Blood collection tubes BD vacutainer 367844 K2 EDTA 7.2 mg plus blood
collection tubes
SuperSignal ELISA femto Invitrogen 37074
Modular multilabel plate reader Perkin Elmer Envision 2104
Disc operating machine Hanra Precision Eng. Co. Ltd., Korea Customized
Photomultiplier tube (PMT) Hamamatsu Photonics H1189-210
AutoCAD AutoDesk Version 2012 Design software
SolidWorks 3D CAD software  SOLIDWORKS Corp. Version 2013 3D Design software,
Edgecam Vero software version 2009.01.06928 Code generating software
DeskCNC Carken Co. version 2.0.2.18 CNC milling machine software

References

  1. Zhang, Y., et al. Nanomaterials for Ultrasensitive Protein Detection. Adv. Mater. 25 (28), 3802-3819 (2013).
  2. Hu, W., Li, C. M. Nanomaterial-based advanced immunoassays. Wiley Interdiscip. Rev. Nanomed. Nanobiotechnol. 3 (2), 119-133 (2011).
  3. Yang-Kyu, C., Chang-Hoon, K. Silicon Nanowire Biosensor for Cancer Markers. Biosensors and Cancer. , 164-183 (2012).
  4. Baltazar, R., Vistas, C. R., Ferreira, G. M. Biosensing Applications Using Nanoparticles. Nanocomposite Particles for Bio-Applications. , 265-282 (2011).
  5. Roy, P., Berger, S., Schmuki, P. TiO2 Nanotubes: Synthesis and Applications. Angew. Chem. Int. Ed. 50 (13), 2904-2939 (2011).
  6. Yang, D., et al. Electrospun Nanofibrous Membranes: A Novel Solid Substrate for Microfluidic Immunoassays for HIV. Adv. Mater. 20 (24), 4770-4775 (2008).
  7. Chantasirichot, S., Ishihara, K. Electrospun phospholipid polymer substrate for enhanced performance in immunoassay system. Biosens. Bioelectron. 38 (1), 209-214 (2012).
  8. Zhang, N., et al. Electrospun TiO2 Nanofiber-Based Cell Capture Assay for Detecting Circulating Tumor Cells from Colorectal and Gastric Cancer Patients. Adv. Mater. 24 (20), 2756-2760 (2012).
  9. Ali, M. A., Mondal, K., Singh, C., Dhar Malhotra, B., Sharma, A. Anti-epidermal growth factor receptor conjugated mesoporous zinc oxide nanofibers for breast cancer diagnostics. Nanoscale. 7 (16), 7234-7245 (2015).
  10. Mondal, K., Ali, M. A., Agrawal, V. V., Malhotra, B. D., Sharma, A. Highly Sensitive Biofunctionalized Mesoporous Electrospun TiO2 Nanofiber Based Interface for Biosensing. ACS Appl. Mater. Interfaces. 6 (4), 2516-2527 (2014).
  11. Tu, W., Dong, Y., Lei, J., Ju, H. Low-Potential Photoelectrochemical Biosensing Using Porphyrin-Functionalized TiO2 Nanoparticles. Anal. Chem. 82 (20), 8711-8716 (2010).
  12. Liu, S., Chen, A. Coadsorption of Horseradish Peroxidase with Thionine on TiO2 Nanotubes for Biosensing. Langmuir. 21 (18), 8409-8413 (2005).
  13. Portan, D. V., Kroustalli, A. A., Deligianni, D. D., Papanicolaou, G. C. On the biocompatibility between TiO2 nanotubes layer and human osteoblasts. J.Biomed.Mater.Res. Part A. 100 (10), 2546-2553 (2012).
  14. Dettin, M., et al. Covalent surface modification of titanium oxide with different adhesive peptides: Surface characterization and osteoblast-like cell adhesion. J. Biomed. Mater. Res. Part A. 90 (1), 35-45 (2009).
  15. Kim, W. -. J., et al. Enhanced Protein Immobilization Efficiency on a TiO2 Surface Modified with a Hydroxyl Functional Group. Langmuir. 25 (19), 11692-11697 (2009).
  16. Son, K. J., Ahn, S. H., Kim, J. H., Koh, W. -. G. Graft Copolymer-Templated Mesoporous TiO2 Films Micropatterned with Poly(ethylene glycol) Hydrogel: Novel Platform for Highly Sensitive Protein Microarrays. ACS Appl. Mater. Interfaces. 3 (2), 573-581 (2011).
  17. Kar, P., Pandey, A., Greer, J. J., Shankar, K. Ultrahigh sensitivity assays for human cardiac troponin I using TiO2 nanotube arrays. Lab Chip. 12 (4), 821-828 (2012).
  18. Agarwal, S., Wendorff, J. H., Greiner, A. Use of electrospinning technique for biomedical applications. Polymer. 49 (26), 5603-5621 (2008).
  19. Ding, B., Wang, M., Wang, X., Yu, J., Sun, G. Electrospun nanomaterials for ultrasensitive sensors. Mater. Today. 13 (11), 16-27 (2010).
  20. Liu, Y., Yang, D., Yu, T., Jiang, X. Incorporation of electrospun nanofibrous PVDF membranes into a microfluidic chip assembled by PDMS and scotch tape for immunoassays. ELECTROPHORESIS. 30 (18), 3269-3275 (2009).
  21. Lee, W. S., Sunkara, V., Han, J. -. R., Park, Y. -. S., Cho, Y. -. K. Electrospun TiO2 nanofiber integrated lab-on-a-disc for ultrasensitive protein detection from whole blood. Lab Chip. 15 (2), 478-485 (2015).
  22. Li, D., Xia, Y. Fabrication of Titania Nanofibers by Electrospinning. Nano Lett. 3 (4), 555-560 (2003).
  23. Lombard, M. . SolidWorks 2013 BIBLE. , (2013).
  24. Tickoo, S. . EdgeCAM 11.0 for Manufacturers. , (2007).
  25. Zhu, R., et al. Improved adhesion of interconnected TiO2 nanofiber network on conductive substrate and its application in polymer photovoltaic devices. Appl. Phys. Lett. 93 (1), 013102 (2008).
  26. Song, M. Y., Ahn, Y. R., Jo, S. M., Kim, D. Y., Ahn, J. -. P. TiO2 single-crystalline nanorod electrode for quasi-solid-state dye-sensitized solar cells. Appl. Phys. Lett. 87 (11), 113113 (2005).
  27. Katsuhiro, O., et al. Electrospinning processed nanofibrous TiO2 membranes for photovoltaic applications. Nanotechnology. 17 (4), 1026-1031 (2006).
check_url/fr/54143?article_type=t

Play Video

Citer Cet Article
Park, Y., Sunkara, V., Kim, Y., Lee, W. S., Han, J., Cho, Y. Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood. J. Vis. Exp. (110), e54143, doi:10.3791/54143 (2016).

View Video