Summary

Completamente dispositivo automatizzato centrifuga Microfluidic per il rilevamento di proteine ​​Ultrasensitive da sangue intero

Published: April 16, 2016
doi:

Summary

This protocol demonstrates how to achieve femto molar detection sensitivity of proteins in 10 µL of whole blood within 30 min. This can be achieved by using electrospun nanofibrous mats integrated in a lab-on-a-disc, which offers high surface area as well as effective mixing and washing for enhanced signal-to-noise ratio.

Abstract

Enzyme-linked immunosorbent assay (ELISA) is a promising method to detect small amount of proteins in biological samples. The devices providing a platform for reduced sample volume and assay time as well as full automation are required for potential use in point-of-care-diagnostics. Recently, we have demonstrated ultrasensitive detection of serum proteins, C-reactive protein (CRP) and cardiac troponin I (cTnI), utilizing a lab-on-a-disc composed of TiO2 nanofibrous (NF) mats. It showed a large dynamic range with femto molar (fM) detection sensitivity, from a small volume of whole blood in 30 min. The device consists of several components for blood separation, metering, mixing, and washing that are automated for improved sensitivity from low sample volumes. Here, in the video demonstration, we show the experimental protocols and know-how for the fabrication of NFs as well as the disc, their integration and the operation in the following order: processes for preparing TiO2 NF mat; transfer-printing of TiO2 NF mat onto the disc; surface modification for immune-reactions, disc assembly and operation; on-disc detection and representative results for immunoassay. Use of this device enables multiplexed analysis with minimal consumption of samples and reagents. Given the advantages, the device should find use in a wide variety of applications, and prove beneficial in facilitating the analysis of low abundant proteins.

Introduction

Diverse piattaforme per la diagnosi della malattia sono stati sviluppati sulla base di materiali su scala nanometrica 1,2, come nanofili, 3 nanoparticelle, nanotubi 4, 5 e nanofibre (NFS) 6-8. Questi nanomateriali offrono ottime prospettive nella progettazione di nuove tecnologie per saggi biologici altamente sensibili per le loro proprietà fisico-chimiche uniche. Ad esempio, mesoporosa zinco nanofibre ossido sono stati utilizzati per la rivelazione sensibile femto-molare di biomarcatori cancro al seno. 9 Recentemente, nanomateriali a base di biossido di titanio (TiO 2) sono stati esplorati per applicazioni bioanalitici 10 considerando la loro stabilità chimica, 11 denaturazione proteica trascurabile , 12 e biocompatibilità. 13 Inoltre, i gruppi idrossilici sulla superficie di TiO 2 facilitare modificazione chimica e l'attacco covalente di biomolecole. 14,15 modellato TiO 2 thin film 16 o TiO 2 nanotubi 17 sono stati utilizzati per migliorare la sensibilità di rilevazione di una proteina bersaglio, aumentando l'area superficiale; Tuttavia, il processo di fabbricazione è piuttosto complessa e richiede attrezzature costose. D'altra parte, NFS elettrofilate stanno ricevendo attenzione a causa della loro elevata area superficiale, così come processo semplice e di fabbricazione a basso costo; 18,19 ancora, la caratteristica fragile o allentato del NF mat elettrofilate TiO 2 rende difficile da gestire e integrazione con dispositivi microfluidici. 6,20 Pertanto, le TiO2 tappetini NF sono stati raramente utilizzati in applicazioni bioanalitiche, in particolare quelle che richiedono condizioni di lavaggio difficili.

In questo studio, per superare queste limitazioni, abbiamo sviluppato una nuova tecnologia per il trasferimento del elettrofilate NF stuoie sulla superficie di un substrato bersaglio utilizzando uno strato adesivo polidimetilsilossano sottile (PDMS). Furthermore, abbiamo dimostrato con successo l'integrazione di elettrofilate TiO 2 NF stuoie su un dispositivo microfluidico centrifuga in policarbonato (PC). Utilizzando questo dispositivo, una rivelazione sensibile alto, completamente automatizzato e integrato di proteina C-reattiva (CRP) e troponina I cardiaca (cTnI) è stata raggiunta entro 30 minuti da solo 10 ml di sangue intero. 21 A causa della combinata vantaggi delle proprietà della FN e la piattaforma centrifuga, il saggio esposti un'ampia gamma dinamica di sei ordini di grandezza da 1 pg / ml (~ 8 fM) a 100 ng / ml (~ 0,8 pm) con un limite inferiore di rilevamento di 0,8 pg / ml (~ 6 fM) per CRP e una gamma dinamica di 10 pg / ml (~ 0,4 pM) a 100 ng / ml (~ 4 nM) con un limite di rilevazione di 37 pg / ml (~ 1,5 pM) per cTnI. Questi limiti di rilevabilità sono ~ 300 ~ e 20 volte inferiore rispetto ai loro corrispondenti risultati convenzionali ELISA. Questa tecnica può essere applicata per la rilevazione di eventuali proteine ​​bersaglio, con anticorpi appropriati. Nel complesso, questo dispositivo could contribuiscono notevolmente alla diagnostica in vitro e saggi biochimici poiché può rilevare rare quantità di proteine ​​bersaglio con grande sensibilità anche da piccole quantità di campioni biologici; ad esempio, 10 ml di sangue intero. Sebbene abbiamo dimostrato solo la rilevazione di proteine ​​del siero utilizzando ELISA in questo studio, la tecnologia di trasferimento e l'integrazione dei elettrofilate NFs con dispositivi microfluidici potrebbe essere più ampiamente applicato in altre reazioni biochimiche che richiedono una grande superficie per alta sensibilità di rilevazione.

Protocol

NOTA: Il sangue è stato prelevato da individui sani ed è stato raccolto in un tubo di raccolta del sangue. Consenso informato scritto è stato ottenuto da tutti i volontari. 1. Realizzazione di TiO2 NF Mat Preparazione della soluzione di precursore 22 Sciogliere 1.5 g di titanio tetraisopropoxide (TTIP) in una miscela di etanolo (99,9%, 3 ml) e acido acetico glaciale (3 ml) e miscelare la soluzione a temperatura a…

Representative Results

Usando questo protocollo, un dispositivo di microfluidica centrifuga completamente automatizzato per il rilevamento di proteine ​​dal sangue intero ad alta sensibilità è stato preparato. TIO 2 tappetini NF sono stati preparati da processi di elettrospinning e calcinazione. Al fine di realizzare NFS di diametro desiderato, la morfologia e lo spessore, condizioni quali la portata, la tensione, e il tempo spinning electrospinning sono stati ottimizzati. Se le condizioni non…

Discussion

Il saggio su TiO 2 NF disco integrato è una tecnica rapida, economica e conveniente per la rilevazione ultrasensibile di basse proteine ​​abbondanti presenti in molto basso volume di sangue. Questa tecnica ha il vantaggio di utilizzare piccoli volumi di campione (10 microlitri) ed è suscettibile di analisi di campioni multipli simultaneamente. Questo fornisce un grande potenziale come dispositivo di multiplexing immunologico. Il dispositivo ha il vantaggio che non richiede Campione fasi di pretrattament…

Divulgations

The authors have nothing to disclose.

Acknowledgements

Questo lavoro è stato sostenuto dalla Fondazione per la Ricerca Nazionale di Corea (NRF) sovvenzioni (2013R1A2A2A05004314, 2012R1A1A2043747), una sovvenzione da parte della Corea Salute Technology R & S del progetto, Ministero della Salute e del Welfare (A121994) e IBS-R020-D1 finanziata dal governo coreano.

Materials

Si wafer LG SILTRON Polished Wafer, test grade Dia. (mm) = 150, orientation = <100>, dopant = boron, RES(Ohm-cm) = 1 – 30, thickness (μm) = 650 – 700
Polycarbonate (PC)  Daedong Plastic PCS#6900 Thickness (mm) = 1 and 5 
Titanium tetraisopropoxide, 98%, Sigma-Aldrich 205273
Polyvinylpyrrolidone, Mw = 1,300,000 Sigma-Aldrich 437190
Acetic acid Sigma-Aldrich 320099
Anhydrous ethanol Sigma-Aldrich 459836
Tridecafluoro-1,1,2,2-tetrahydrooctyl)-1-trichlorosilane Sigma-Aldrich 448931
PDMS and curing agent Dow Corning SYLGARD 184
GPDES Gelest Inc SIG5832.0 
Ethanol J T Baker
FE-SEM FEI Nova NanoSEM
X-ray photoelectron spectroscopy ThermoFisher K-alpha
3D modeling machine M&I CNC Lab, Korea CNC milling machine
Wax-dispensing machine Hanra Precision Eng. Co. Ltd., Korea Customized
Double-sided adhesive tape FLEXcon, USA DFM 200 clear 150 POLY H-9 V-95
Cutting plotter Graphtec Corporation, Japan Graphtec CE3000-60 MK2
Spin coater MIDAS SPIN-3000D
Furnace (calcination) R. D. WEBB COMPANY WEBB 99
Rheometer (Tack test) Thermo Scientific Haake MARS III – ORM Package
Oxygen plasma system FEMTO CUTE
Monoclonal mouse antihuman hsCRP Hytest Ltd., Finland 4C28 (clone # C5)
Monoclonal mouse anti-cTnI Hytest Ltd., Finland 4T21 (clone # 19C7)
HRP conjugated goat polyclonal anti-hsCRP Abcam plc., MA ab19175
HRP conjugated mouse monoclonal anti-cTnI Abcam plc., MA ab24460 (clone # 16A11)
hsCRP Abcam plc., MA ab111647
cTnI Fitzgerald, MA 30-AT43
Bovine Albumin Sigma-Aldrich A7906
PBS Amresco Inc E404
Blood collection tubes BD vacutainer 367844 K2 EDTA 7.2 mg plus blood
collection tubes
SuperSignal ELISA femto Invitrogen 37074
Modular multilabel plate reader Perkin Elmer Envision 2104
Disc operating machine Hanra Precision Eng. Co. Ltd., Korea Customized
Photomultiplier tube (PMT) Hamamatsu Photonics H1189-210
AutoCAD AutoDesk Version 2012 Design software
SolidWorks 3D CAD software  SOLIDWORKS Corp. Version 2013 3D Design software,
Edgecam Vero software version 2009.01.06928 Code generating software
DeskCNC Carken Co. version 2.0.2.18 CNC milling machine software

References

  1. Zhang, Y., et al. Nanomaterials for Ultrasensitive Protein Detection. Adv. Mater. 25 (28), 3802-3819 (2013).
  2. Hu, W., Li, C. M. Nanomaterial-based advanced immunoassays. Wiley Interdiscip. Rev. Nanomed. Nanobiotechnol. 3 (2), 119-133 (2011).
  3. Yang-Kyu, C., Chang-Hoon, K. Silicon Nanowire Biosensor for Cancer Markers. Biosensors and Cancer. , 164-183 (2012).
  4. Baltazar, R., Vistas, C. R., Ferreira, G. M. Biosensing Applications Using Nanoparticles. Nanocomposite Particles for Bio-Applications. , 265-282 (2011).
  5. Roy, P., Berger, S., Schmuki, P. TiO2 Nanotubes: Synthesis and Applications. Angew. Chem. Int. Ed. 50 (13), 2904-2939 (2011).
  6. Yang, D., et al. Electrospun Nanofibrous Membranes: A Novel Solid Substrate for Microfluidic Immunoassays for HIV. Adv. Mater. 20 (24), 4770-4775 (2008).
  7. Chantasirichot, S., Ishihara, K. Electrospun phospholipid polymer substrate for enhanced performance in immunoassay system. Biosens. Bioelectron. 38 (1), 209-214 (2012).
  8. Zhang, N., et al. Electrospun TiO2 Nanofiber-Based Cell Capture Assay for Detecting Circulating Tumor Cells from Colorectal and Gastric Cancer Patients. Adv. Mater. 24 (20), 2756-2760 (2012).
  9. Ali, M. A., Mondal, K., Singh, C., Dhar Malhotra, B., Sharma, A. Anti-epidermal growth factor receptor conjugated mesoporous zinc oxide nanofibers for breast cancer diagnostics. Nanoscale. 7 (16), 7234-7245 (2015).
  10. Mondal, K., Ali, M. A., Agrawal, V. V., Malhotra, B. D., Sharma, A. Highly Sensitive Biofunctionalized Mesoporous Electrospun TiO2 Nanofiber Based Interface for Biosensing. ACS Appl. Mater. Interfaces. 6 (4), 2516-2527 (2014).
  11. Tu, W., Dong, Y., Lei, J., Ju, H. Low-Potential Photoelectrochemical Biosensing Using Porphyrin-Functionalized TiO2 Nanoparticles. Anal. Chem. 82 (20), 8711-8716 (2010).
  12. Liu, S., Chen, A. Coadsorption of Horseradish Peroxidase with Thionine on TiO2 Nanotubes for Biosensing. Langmuir. 21 (18), 8409-8413 (2005).
  13. Portan, D. V., Kroustalli, A. A., Deligianni, D. D., Papanicolaou, G. C. On the biocompatibility between TiO2 nanotubes layer and human osteoblasts. J.Biomed.Mater.Res. Part A. 100 (10), 2546-2553 (2012).
  14. Dettin, M., et al. Covalent surface modification of titanium oxide with different adhesive peptides: Surface characterization and osteoblast-like cell adhesion. J. Biomed. Mater. Res. Part A. 90 (1), 35-45 (2009).
  15. Kim, W. -. J., et al. Enhanced Protein Immobilization Efficiency on a TiO2 Surface Modified with a Hydroxyl Functional Group. Langmuir. 25 (19), 11692-11697 (2009).
  16. Son, K. J., Ahn, S. H., Kim, J. H., Koh, W. -. G. Graft Copolymer-Templated Mesoporous TiO2 Films Micropatterned with Poly(ethylene glycol) Hydrogel: Novel Platform for Highly Sensitive Protein Microarrays. ACS Appl. Mater. Interfaces. 3 (2), 573-581 (2011).
  17. Kar, P., Pandey, A., Greer, J. J., Shankar, K. Ultrahigh sensitivity assays for human cardiac troponin I using TiO2 nanotube arrays. Lab Chip. 12 (4), 821-828 (2012).
  18. Agarwal, S., Wendorff, J. H., Greiner, A. Use of electrospinning technique for biomedical applications. Polymer. 49 (26), 5603-5621 (2008).
  19. Ding, B., Wang, M., Wang, X., Yu, J., Sun, G. Electrospun nanomaterials for ultrasensitive sensors. Mater. Today. 13 (11), 16-27 (2010).
  20. Liu, Y., Yang, D., Yu, T., Jiang, X. Incorporation of electrospun nanofibrous PVDF membranes into a microfluidic chip assembled by PDMS and scotch tape for immunoassays. ELECTROPHORESIS. 30 (18), 3269-3275 (2009).
  21. Lee, W. S., Sunkara, V., Han, J. -. R., Park, Y. -. S., Cho, Y. -. K. Electrospun TiO2 nanofiber integrated lab-on-a-disc for ultrasensitive protein detection from whole blood. Lab Chip. 15 (2), 478-485 (2015).
  22. Li, D., Xia, Y. Fabrication of Titania Nanofibers by Electrospinning. Nano Lett. 3 (4), 555-560 (2003).
  23. Lombard, M. . SolidWorks 2013 BIBLE. , (2013).
  24. Tickoo, S. . EdgeCAM 11.0 for Manufacturers. , (2007).
  25. Zhu, R., et al. Improved adhesion of interconnected TiO2 nanofiber network on conductive substrate and its application in polymer photovoltaic devices. Appl. Phys. Lett. 93 (1), 013102 (2008).
  26. Song, M. Y., Ahn, Y. R., Jo, S. M., Kim, D. Y., Ahn, J. -. P. TiO2 single-crystalline nanorod electrode for quasi-solid-state dye-sensitized solar cells. Appl. Phys. Lett. 87 (11), 113113 (2005).
  27. Katsuhiro, O., et al. Electrospinning processed nanofibrous TiO2 membranes for photovoltaic applications. Nanotechnology. 17 (4), 1026-1031 (2006).

Play Video

Citer Cet Article
Park, Y., Sunkara, V., Kim, Y., Lee, W. S., Han, J., Cho, Y. Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood. J. Vis. Exp. (110), e54143, doi:10.3791/54143 (2016).

View Video