Vue d'ensemble

Fonte: Laboratorio del Dr.B. Jill Venton – Università della Virginia

L’obiettivo di molte analisi chimiche è un’analisi quantitativa, in cui viene determinata la quantità di una sostanza in un campione. Per calcolare con precisione la concentrazione di uno sconosciuto da un campione, è fondamentale un’attenta preparazione del campione. Ogni volta che un campione viene maneggiato o trasferito, parte del campione può andare perso. Esistono tuttavia strategie per ridurre al minimo la perdita di campioni. Esistono anche strategie per far fronte alla perdita del campione e fare ancora misurazioni accurate della concentrazione.

Per ridurre al minimo la perdita di campioni, l’ideale è ridurre al minimo il numero di passaggi di gestione e trasferimento del campione. Ad esempio, la massa di un campione solido direttamente in un pallone in cui verrà realizzata una soluzione riduce una fase di trasferimento. Se è necessario trasferire da un pallone all’altro e si sta facendo una diluizione, il triplo risciacquo della vetreria aiuta a garantire che tutto il campione venga trasferito. Altre strategie sono più specifiche per il campione. Ad esempio, i campioni che assorbono al vetro, come le proteine, potrebbero essere gestiti meglio in tubi monouso in polipropilene. I tubi non sono idrofili, quindi se una piccola quantità di campione deve essere pipettata in acqua, è meglio aver già aggiunto l’acqua al tubo, in modo che il campione possa essere pipettato direttamente nel solvente. Potrebbe essere meglio concentrarsi, piuttosto che asciugare completamente un campione, a causa delle perdite da insolubilità dopo la reidratazione.

Un’altra fonte di perdita del campione è attraverso manipolazioni incomplete del campione. Ad esempio, se viene utilizzata una procedura di derivatizzazione e la derivatizzazione è incompleta, l’intera quantità di campione non viene osservata. Errori come questo sono errori sistematici e possono essere risolti correggendo il problema, ad esempio modificando la procedura di derivatizzazione. Un’altra causa di errore sistematico nelle misurazioni sono gli effetti della matrice. Questi possono interferire con la misurazione di determinate sostanze e l’esecuzione di calibrazioni nella stessa matrice del campione può ridurre questo effetto.

L’analisi quantitativa viene in genere eseguita utilizzando standard esterni o interni. Per gli standard esterni, viene effettuata una curva di calibrazione misurando diverse concentrazioni note dell’analita di interesse. Quindi, l’esempio viene eseguito separatamente dallo standard. Per gli standard interni, lo standard si trova nello stesso campione dell’analita di interesse, consentendo di eseguire la misurazione contemporaneamente. Tipicamente, viene aggiunta una specie diversa chiamata standard interno e viene calcolato il rapporto tra la risposta per quello standard interno e l’analita. L’idea è che il rapporto della risposta, chiamato fattore di risposta, sia proporzionale alla loro concentrazione. Mentre il metodo deve essere in grado di distinguere tra l’analita di interesse e lo standard interno, eventuali perdite di campione che si verificano dopo l’aggiunta dello standard interno dovrebbero essere simili per entrambe le sostanze e quindi il rapporto della risposta rimane lo stesso. Un caso speciale di utilizzo di standard interni è il metodo delle aggiunte standard, in cui quantità crescenti di analita vengono aggiunte alla soluzione e la quantità originale di analita viene retro-calcolata. Gli standard interni possono essere utilizzati in cromatografia, elettrochimica e spettroscopia.

Principles

Procédure

1. Corretta gestione del campione: creazione di una soluzione Prendi un becher pulito e ammassa la quantità corretta di campione in esso. Registrare la massa effettiva utilizzata. In questo esempio, una soluzione di adenina viene realizzata in un pallone volumetrico per l’uso come standard interno per l’analisi successiva. La massa di adenina è di 100 mg. Non ammassare direttamente in un matraccio volumetrico perché ha un collo lungo e l’adenina non può essere facilmente aggiunta o rimossa. Aggiun…

Applications and Summary

Internal standards are used in many fields, including spectroscopy and chromatography. In spectroscopy, internal standards can help correct for random errors due to changes in light source intensity. If a lamp or other light source has variable power, it will affect the absorption and consequently, emission of a sample. However, the ratio of an internal standard to analyte will stay constant, even if the light source does not. One example of this is using lithium (Li) as an internal standard for the analysis of sodium in a blood sample by flame spectroscopy. Li is chemically similar to sodium but is not natively found in blood.

For chromatography, internal standards are often used in both gas chromatography and liquid chromatography. For applications with mass spectrometry as the detector, the internal standard can be an isotopically-labeled analyte, so that the molecular weight (MW) will be different than the analyte of interest. Internal standards are commonly used in pharmaceutical or environmental analyses.

Transcription

Sample loss can occur every time a sample is handled or transferred, thereby making accurate calculations of concentration difficult.

To ensure accuracy, the effects of sample loss must be minimized using careful sample preparation and by limiting the number of sample handling and transfer steps. However, sample loss can also occur due to systematic errors, such as incomplete sample manipulation, matrix effects, and variations in analytic procedure.

These sources of loss can be accounted for by adding a known concentration of a species similar, but not identical, to the compound of interest. This is called an internal standard. Any sample losses that occur to the internal standard should be similar for the analyte, allowing for the concentration to be accurately calculated.

This video will illustrate the use of an internal standard and proper lab technique to account for sample loss when determining the concentration of an unknown.

An internal standard is a substance added in a known amount to standards, samples, and blanks during an analysis.

In chromatography and spectroscopy, the ratio of the signal for the internal standard and the analyte is calculated. This ratio, called the response factor, is proportional to the ratio of the analyte and standard concentrations.

Response factor, R, can be expressed by the following equation, where A represents the analytical signals of the sample and internal standard and C represents the concentrations of the sample and internal standard.

An internal standard can compensate for both systematic and random errors. For example, random errors—such as inconsistencies when measuring a sample—will be the same for both the internal standard and the analyte. Therefore, the ratio of their signals will not change.

For systematic errors, such as matrix effects in solution, the ratio will be unaffected as long as the matrix effect is equal for both the standard and the analyte.

While internal standards provide great benefit, it can be difficult to choose one that is suitable. An internal standard must have a signal that is similar, but not identical, to the analyte. It also cannot affect the measurement of the analyte in any way.

Finally, the concentration must be well known. This is achieved by ensuring that the internal standard is not natively present in the sample; thus, the only source of it in solution is the known concentration added.

In the following experiment, the concentration of caffeine in an unknown sample will be determined by gas chromatography.

This is achieved by creating a calibration curve using known caffeine solutions, with adenine as the internal standard. The slope of the calibration curve is equal to the response factor.

Once the response factor is known, the concentration of the unknown can be calculated from its measured chromatogram area ratio.

Now that you understand the basics of internal standards, let’s take a look at the procedure.

To begin the procedure, accurately weigh 100 mg of the internal standard, adenine, into a clean beaker.

Next, dissolve it in roughly 20 mL of dimethyl sulfoxide, and mix the solution.

Once the adenine has dissolved, pour the solution into a 50-mL volumetric flask.

Rinse the beaker and stir bar with 10 mL of DMSO, and pour the rinse into the flask. Repeat this rinse twice, to ensure proper solution transfer. Fill to the calibration mark, resulting in an internal standard with a concentration of 2 mg/mL.

Next, weigh 100 mg of caffeine into a beaker to prepare a stock solution. Dissolve the caffeine with a small amount of methanol. Then, use 3 rinses to transfer this solution to a fresh 25 mL volumetric flask. This is the 4 mg/mL stock solution. Use it to create 3 caffeine standards.

Next, add 0.2 mL of the internal standard, adenine, to each flask. Fill each to the final volume with methanol. Transfer each solution to a sample vial.

Run each caffeine standard through a gas chromatograph. Calculate the ratio of peak areas for the caffeine versus the adenine standard.

First, weigh 2 g of coffee into a 100-mL beaker, and record the weight.

Next, add 20 mL of methanol to extract the caffeine from the coffee. Allow the solution to stir for 20 min.

Using a Büchner funnel, filter out the coffee grounds. Rinse the beaker with a small amount of methanol, and pour this rinse into the funnel. Repeat the rinse twice.

Measure the final volume of the filtrate; it should be approximately 35 mL.

To prepare the sample for analysis, add 1 mL of the coffee extract to a sample vial. Then, add 0.2 mL of the adenine internal standard, and place the vial into the instrument’s auto-sampler rack.

Run a gas chromatography analysis of the sample, ensuring that the conditions are such that the caffeine and adenine are separate.

After completing the analysis, compute the peak area for both the internal standard and the analyte.

Once all the samples have been analyzed, the standard calibration curve can be determined for the caffeine/adenine solutions by plotting the ratios of the peak areas versus the ratios of the concentrations. The slope of this line, which represents the response factor, was 1.8.

Next, the GC data from the extracted coffee sample is analyzed. The ratio of the peak areas was calculated to be 1.78. Using the response factor and the known concentration of the internal standard, adenine, the concentration of caffeine in the unknown sample was calculated to be 0.33 mg/mL.

Many different types of reactions, across various scientific disciples, utilize internal standards to minimize the effects of errors and sample loss.

The effects of sample loss encountered during sample preparation can be minimized using internal standards, keeping their concentration ratio nearly constant.

In this example, bioactive lipids were extracted from lysed cells using a liquid-liquid extraction process. Stable isotope internal standards were added at the beginning of extraction to account for errors during sample preparation.

Internal standards were not only critical for the preparation of the bioactive lipids, but also for the analysis. The lipids were separated using high-performance liquid chromatography, and analyzed via mass spectrometry.

In spectroscopy, internal standards can help correct for random errors due to changes in light source intensity. If a lamp or other light source has variable power, it will affect the absorption and consequently, emission of a sample. However, the ratio of an internal standard to analyte will stay constant, even if the light source does not.

In chromatography, one of the largest sources of error is the injection. Auto-samplers help minimize this, but error can still be 1–2% relative standard deviation.

In this example, vapor standards containing an internal standard were analyzed using gas chromatography to establish a calibration curve. Once this was complete, the unknown sample could then be measured and the losses due to volatility of the sample accounted for.

You’ve just watched JoVE’s introduction to internal standards. You should now understand best practices for minimizing sample loss, internal standards, and response factors.

Thanks for watching!