Cardiac Magnetic Resonance Imaging
  • 00:07Vue d'ensemble
  • 01:51Principles of Cardiac MRI
  • 04:14MRI Imaging Preparation
  • 06:13Cardiac MRI Imaging
  • 08:40Résultats
  • 10:04Applications
  • 11:04Summary

心脏磁共振成像

English

Diviser

Vue d'ensemble

资料来源:弗雷德里克·达门和克雷格·戈尔根,珀杜大学韦尔登生物医学工程学院,西拉斐特,印第安纳州

在本视频中,通过生理监测,演示了高场、小孔磁共振成像(MRI),以获取鼠心血管系统的门控膜环。此过程为评估左心室功能、可视化血管网络和量化因呼吸引起的器官运动提供了基础。类似的小型动物心血管成像模式包括高频超声和微计算机断层扫描(CT);但是,每种模式都与应考虑的权衡相关。虽然超声波确实提供高空间和时间分辨率,但成像伪像很常见。例如,密集组织(即胸骨和肋骨)可以限制成像穿透深度,而气体和液体(即肺周围的胸膜)之间的超呼信号可以模糊附近组织的对比度。相比之下,微CT不会遭受尽可能多的平面内伪影,但具有较低的时间分辨率和有限的软组织对比度。此外,微型CT使用X射线辐射,并经常要求使用造影剂来可视化血管,这两种物质已知在高剂量下造成副作用,包括辐射损伤和肾损伤。心血管MRI通过否定电离辐射的需要和为用户提供无反光剂成像的能力(尽管造影剂常用于MRI),在这些技术之间提供了很好的折衷。

这些数据是通过触发快速低角度热(FLASH)MRI序列获得的,该序列在心脏周期的R峰和呼吸中的呼气中呼退的高原上被封闭。这些生理事件通过皮下电极和固定在腹部的压力敏感枕头进行监测。为确保鼠标正确加热,插入了直肠温度探头,用于控制 MRI 安全加热风扇的输出。一旦动物入 MRI 扫描仪的孔和导航序列运行以确认定位,门控 FLASH 成像平面被规定和获取数据。总体而言,高场核磁共振成像是一种强大的研究工具,可以为研究小动物疾病模型提供软组织对比。

Principles

Procédure

1. 动物准备 识别在笼子里成像的小鼠并将其转移到麻醉感应室。 使用除胶对鼠标进行麻醉,并使用脚趾捏技术确认击倒。抓住拇指和食指之间的爪子,用力捏住以检查反应。如果动物收回他们的脚,你应该等待或重新剂量与麻醉根据批准的协议。 检查所有进入成像设施的员工都 MR 安全。这包括去除任何磁性服装/配件,确认没有磁性植入物或心脏起搏器,以及去除含有穿?…

Résultats

Figure 1 shows a cine loop of a short-axis view of the left ventricle, which is directly perpendicular to the base-apex axis of the heart and at a position that includes the papillary muscles.

Figure 1
Figure 1: Bright blood cine imaging of a mouse heart with 14 short-axis view snapshots across the cardiac cycle, including end diastole (t = 8) and peak systole (t = 13). The regions of dropout signal within the lumen of the left ventricle indicates fast blood moving, which was originally out of plane and not tagged by the RF wave excitation.

The second representative image shows a 4-chamber view of the heart with bright-blood inflow through the mitral and tricuspid valves, which then flows out through the aortic and pulmonary valves, respectively.

Figure 2
Figure 2: Bright blood cine imaging of a mouse heart with a four-chamber view showing end diastole (left) and peak systole (right). The regions of dropout signal within the lumen of the left ventricle indicates fast blood moving, which was originally out of plane and not tagged by the RF wave excitation.

Finally, a third representative result is a maximum intensity projection (MIP) that shows how the multiple slices can be spatially combined to visualize the cardiovascular system of the whole mouse body.

Figure 3
Figure 3: Maximum intensity projection of a three-dimensional stack of time-synced two-dimensional bright blood images, showing the thoracic and abdominal regions of a mouse. The heart, inferior vena cava, and small abdominal aortic aneurysm (red circle) can be seen from this view.

Applications and Summary

Here, cardiac MRI is used in conjunction with cardiac- and respiration-gating to acquire cine loop data of the murine heart. While the heart was the focus of demonstration, additional regions of the cardiovascular system can be imaged following the same methodology. Even though MRI does not suffer from the same artifacts commonly seen with other imaging modalities, there is a noticeable trade-off with spatial resolution achieved per acquisition duration. This trade-off is of concern when the mouse cannot withstand longer durations of anesthesia, such as in severe disease models. Still, MRI has the advantage of visualizing the underlying tissue without the risk of ionizing radiation damage inherent to micro-CT. Using MRI techniques, an in vivo assessment of the cardiovascular can be performed, laying the foundation for longitudinal studies of both disease progression and associated therapy response in small animal models.

As an extension of the described technique, this technology can be used to compare the kinematics of healthy versus diseased hearts. Murine models of cardiac dysfunction can be far more controlled than those found in the clinic, allowing researchers to identify specific factors contributing to heart disease as well as study the remodeling process after mechanical injury. Furthermore, a comparable research endeavor can be performed with a vascular focus such as that with abdominal aortic aneurysm (AAA) formation. Given that blood gives a high intensity signal under the described methods, the contrast can be exploited to assess the expansion of a AAA and measure changes to the vessel's biomechanical properties. Finally, studies looking at vascularization of the brain can be conducted to compare angiogenic responses to traumatic brain injury or stroke. Ideally, as with most pre-clinical imaging, techniques such as high field cardiovascular MRI can further our understanding of human disease processes as well as spark innovation towards the next generation of diagnostic technology.

Transcription

High-field small-bore magnetic resonance imaging, or cardiac MRI, assesses cardiovascular function without the use of ionizing radiation or contrast agents.

Comparable cardiovascular imaging modalities include high frequency ultrasound, which emits a beam of acoustic waves from a transducer and records the echoes created as the waves reflect to generate live images. It provides high spatial and temporal resolution images; however, imaging artifacts can be observed due to the limited penetration depth in dense tissue.

Another imaging technique is micro-CT, which takes a series of X-ray projections to create 3D cross sections. It has a lower temporal resolution and limited soft tissue contrast, and often requires the use of contrast agents to visualize vascular structures. These are known to cause radiation damage and renal failure at high doses.

Alternatively, MRI uses strong electromagnets to image tissues in the body based on their magnetic properties. In cardiac MRI, conventional MRI sequences are gated off of R peaks in the cardiac cycle and expiratory plateaus in respiration to assess cardiovascular function.

This video will illustrate how to gather MRI data with a triggering fast low angle shot, or FLASH MRI sequence. This technique provides high quality soft tissue contrast for the study of small animal disease models.

Magnetic resonance imaging is a technique that uses the paramagnetic properties of tissue to visualize soft tissue contrast. The bore of an MRI machine is conventionally wrapped using a solenoid coil that provides a constant homogeneous magnetic field, B-zero, when an electric current is applied.

In high-field mirroring imaging, a 7-Tesla magnetic field strength can be employed, approximately 140,000 times that of Earth’s magnetic field, and more than double the common clinical 3-Tesla and 1.5-Tesla scanner field strengths. This homogeneous magnetic field causes the hydrogen protons inherent to almost all living tissues to align their axes of rotation. These spins can then be tipped using radiofrequency, or RF waves, to a certain angle relative to the axis of rotation, also known as the flip angle.

As the protons then attempt to relax back to their original orientation, the component of their spin perpendicular to their main axis induces a detectable electrical signal, resulting in an image. Furthermore, magnetic gradients can be applied that perturb the main magnetic field and allow for spatially isolated RF excitations to localize the received signal. Specific to the methods described in this video is the FLASH sequence, which uses low flip angle RF excitations that are rapidly repeated to induce a steady state pattern in the proton motion. The repetition time is much shorter than the typical proton relaxation time.

When unexcited hydrogen, such as that in blood, enters the imaging frame, a relatively high signal is produced. This allows the cardiovascular system to be imaged rapidly and provide stable snapshots within the cardiac cycle. Through triggering the FLASH sequence with physiologic signals, images of the cardiovascular system can be acquired that highlight cardiac, vascular, and respiratory motion.

Having reviewed the main principles of cardiac MRI, let us now walk through the step-by-step procedure to prepare and image an animal.

First, identify the mouse to be imaged, then transfer the mouse to the knockdown chamber. Then, anesthetize the animal using isoflurane and confirm knockdown using a toe-pinch technique. Next, open the isoflurane flow to the nose cone in the MRI room and close the isoflurane flow to the knockdown chamber. This primes the longer tubing with anesthetic.

Ensure that all personnel are MR safe, then transfer the mouse to the imaging stage and secure the nose cone around the animal. Position the mouse such that its heart is approximately aligned with a center of the RF coil. Next, reconfirm knockdown using the toe-pinch technique. Then, insert the three electrocardiogram leads subcutaneously. Place one lead each to the left and right of the heart and one at the base of the left hind limb.

Insert the rectal thermometer probe using a sterile probe sheath and lubricant. Then, place a pillow respiration sensor on the epigastric regions of the abdomen, and secure it in place using cardboard to acquire pressure sensitive signaling.

Confirm that all physiological signals are being acquired through the monitoring software outside the scanner room. Next, set up the heating module and fan to begin warming the airflow to the mouse. Secure the air tubing in place such that the warm air blows towards the mouse, starting just past the tip of its tail. Finally, place the RF coil over the mouse and make sure all cables and tubing are secure.

Let us now review the step-by-step protocol to perform cardiac MRI on the anesthetized mouse.

First, tune and match the RF coil outside the bore of the magnet to ensure maximum signal detection. This is indicated by a narrow valley at zero hertz for each component of the RF coil. Next, slowly insert the imaging stage into the bore of the magnet. Ensure that the mouse is positioned directly in the center of the bore and the gradient coil has equal spacing along all radial directions. This position ensures a homogeneous main magnetic field.

Next, run a navigation scan to locate the mouse within the scanner. Confirm if some segment of the heart is visualized within all three planes, namely axial, sagittal, and coronal. Then, set the parameters for the FLASH sequence and select the external triggering to be on. On the monitoring software, configure the external triggers such that the MRI sequences are serially run only on R-peaks in cardiac cycles during stable respiration in the expiratory phase.

Next, prescribe the initial FLASH sequence by setting the parameters and positioning an imaging plane rectangle in the coronal view. Then press continue to run it such that the slice plane follows the axis from the apex of the heart through the aortic valve. This initial cine loop will provide a two-chamber view of the heart.

Then, while referencing the results from the two-chamber view, prescribe and run a new FLASH sequence along the apex aortic valve axis to visualize a four-chamber view.

Finally, prescribe a short axis slice that is perpendicular to the apex aortic valve axis approximately halfway through the heart. The papillary muscles should be distinctly visible within the cine loop output at this location. Once imaging is completed, transfer acquired data to an appropriate location for analysis, then retract the imaging stage from the bore of the magnet and remove the gradient coil and all probes from the animal before transferring the animal from the scanner bed.

Now that we’ve obtained a cardiac MRI in a mouse, let us review the results of the scans. This figure shows the cine loop of a short axis view of the left ventricle, directly perpendicular to the base apex axis of the heart and at a position that includes the papillary muscles.

Here, we see the blood cine imaging of a mouse heart with 14 short axis view snapshots across the cardiac cycle, including end diastole and peak systole. The regions of dropout signal within the lumen of the left ventricle indicates fast blood moving, which was originally out of plane and not tagged by the RF wave excitation.

This image shows a four-chamber view of the heart with bright blood inflow through the mitral and tricuspid valves, and then out through the aortic and pulmonary valves, respectively.

Finally, here is a maximum intensity projection that shows how multiple slices can be spatially combined to visualize the cardiovascular system of the whole mouse. The figure shows a 3-dimensional stack of time-synced, bright, 2-dimensional blood images showing the thoracic and abdominal regions of a mouse.

Let us now look at some other applications of this MRI technique. As an extension of the described technique, we can use this technology to compare the kinematics of healthy versus diseased hearts. Murine models of cardiac dysfunction can be far more controlled than those found in the clinic. This allows researchers to identify specific factors contributing to heart disease as well as study the remodeling process after injury.

A comparable research endeavor can be performed with a vascular focus, such as that with abdominal aortic aneurysm formation. Blood gives a high intensity signal using the high-field small-bore MRI method described here. This increase in contrast can be exploited to assess the expansion of an abdominal aortic aneurysm and measure changes to the vessel’s biomechanical properties.

You’ve just watched JoVE’s introduction to cardiovascular magnetic resonance imaging.

You should now know how to perform cardiac imaging and how to acquire cine loop data of the murine heart using standard bright blood FLASH MRI sequences synced with cardiac and respiration signals. Finally, you should also know how to identify cardiac structures in these images. Thanks for watching!