Summary

Medição de Extracelular Ion Fluxos Usando a técnica microeletrodos de auto-referência íon-seletivo

Published: May 03, 2015
doi:

Summary

Transporters in cell membranes allow differential segregation of ions across cell membranes or cell layers and play crucial roles during tissue physiology, repair and pathology. We describe the ion-selective self-referencing microelectrode that allows the measurement of specific ion fluxes at single cells and tissues in vivo.

Abstract

As células de animais, plantas e células individuais são envolvidas por uma barreira chamada a membrana celular que separa o citoplasma a partir do exterior. Camadas de células, tais como os epitélios também formam uma barreira que separa o interior a partir do exterior ou diferentes compartimentos de organismos multicelulares. Uma característica chave destas barreiras é a distribuição diferencial de íons através das membranas celulares ou camadas de células. Duas propriedades permitem que esta distribuição: 1) membranas e epitélios exibir permeabilidade seletiva de íons específicos; 2) são transportados através de bombas de iões através das membranas celulares e camadas de células. Estas propriedades desempenham um papel crucial na manutenção da fisiologia do tecido e agir como sinalização sinais de avaria, durante o reparo, ou sob condição patológica. A auto-referenciar microeléctrodo selectivo de iões permite medições de fluxos de iões específicos, tais como cálcio, potássio ou sódio em níveis de células e tecidos individuais. O microeléctrodo contém um ionóforo que é cocktailselectivamente permeável a um ião específico. A solução de enchimento interno contém um conjunto de concentração do íon de interesse. O potencial eléctrico do microeléctrodo é determinada pela concentração do ião exterior. À medida que a concentração de ião varia, o potencial do microeléctrodo muda como uma função do log da actividade de iões. Quando mudou-se para trás e perto de uma fonte ou sumidouro de íon (ou seja, em um gradiente de concentração devido ao fluxo de iões) o potencial flutua microeletrodos a uma amplitude proporcional ao fluxo de iões / gradiente. O amplificador amplifica o sinal de microeléctrodos e a produção é registada no computador. O fluxo de iões pode então ser calculado pela lei de Fick da difusão utilizando a flutuação potencial de eléctrodo, a excursão de microeléctrodo, e outros parâmetros, tais como a mobilidade de iões específica. Neste artigo, descreve em detalhes a metodologia para medir fluxos de iões extracelulares utilizando a auto-referência de microeletrodos um íon-seletivod apresentar alguns resultados representativos.

Introduction

Todas as células animais estão rodeados por uma membrana de bicamada lipídica que separa o citoplasma a partir do ambiente exterior. A célula mantém um potencial de membrana elétrica, negativa no interior, por transporte ativo de íons um. O potencial de membrana é uma fonte de energia que a célula pode utilizar para operar vários dispositivos moleculares na membrana 2. Neurônios e outras células excitáveis ​​têm grandes potenciais de membrana. Rápida abertura dos canais de sódio recolhe o potencial de membrana (despolarização) e produz o potencial de acção que é transportado ao longo do comprimento do neurónio 2. Aparte destas alterações eléctricos rápidos, muitos tecidos e órgãos gerar e manter potenciais eléctricos significativos a longo prazo. Por exemplo, a pele e os epitélios da córnea gerar e manter potenciais trans-epiteliais e correntes eléctricas extracelulares por bombagem de iões direccional (principalmente cloreto de sódio e) 3.

tenda "> Enquanto as medições de corrente eléctrica extracelular endógena utilizando a sonda de vibração 4-6 e medições de membrana ou trans-epitelial potenciais utilizando o sistema de microeléctrodos 7-10 permitem a medição dos parâmetros eléctricos de membranas celulares e camadas de células epiteliais, eles não dão indicação da espécie iónicos envolvidos.

Microeletrodos com ionóforo seletivo pode medir a concentração de íon específico em solução. Gradientes de iões ou de fluxo podia ser medido com dois ou mais eléctrodos em diferentes posições. No entanto, o desvio de tensão intrínseco de cada sonda seria diferente, fazendo medições imprecisas ou mesmo a detecção de um gradiente que não estava presente. Um único eletrodo usado em modo "auto-referência", pelo qual ele se move em baixa frequência entre dois pontos resolve este problema. Agora o fluxo de iões pode ser visto no contexto de um relativamente lento e estável desvio de sinal (veja a Figura 3B). </p>

O sistema de medição sensível-ion utiliza microeletrodos de auto-referência de íon-seletivo para detectar pequenos fluxos extracelulares de íons perto de tecidos ou células individuais. O sistema é composto por um amplificador que processa o sinal a partir do microeléctrodos e um motor passo a passo e micro controlador para controlar o movimento do microeléctrodo. O microeléctrodo selectivo de iões e do eléctrodo de referência que fecha o circuito está ligado ao amplificador de andar de entrada por meio de um pré-amplificador (Figura 1A). O software de computador determina os parâmetros do movimento de microeléctrodos (frequência, distância) e também regista a saída do amplificador. O motor de passo controla o movimento através de um microeléctrodo microposicionador tridimensional. A baixa freqüência de vibração de microeletrodos íon-seletivo foi desenvolvido pela primeira vez em 1990 para medir o fluxo de cálcio específica 11. Bem como cálcio, cocktails ionóforos comercialmente acessíveis estão agora disponíveis para fazer microelectrodes sensíveis ao sódio, cloreto, potássio, hidrogénio, magnésio, nitrato, amónio, fluoreto, lítio ou mercúrio.

Basicamente, o ião-selectivo técnica de microeléctrodo auto-referenciar converte a actividade de um ião específico dissolvido numa solução para um potencial eléctrico, que pode ser medida por um voltímetro. O coquetel de ionóforo é um líquido imiscível (lipofílico orgânico,) fase com propriedades de permuta iónica. O ionóforo de complexos (se liga selectivamente iões específicos reversivelmente) e transfere-os entre a solução aquosa contida no microeléctrodo (electrólito) e a solução aquosa em que o microeléctrodo é imerso (Figura 1D). Esta transferência de iões leva a um equilíbrio electroquímica e uma variação do potencial eléctrico entre os microeléctrodos e o eléctrodo de referência é medido pelo voltímetro. A tensão é proporcional ao logaritmo da actividade específica de iões de acordo com a Nernst equation permitindo o cálculo da concentração de iões (Figura 2A e B).

Actualmente, vários sistemas de permitir a medição do fluxo de iões utilizando um conceito ou princípio semelhante. Por exemplo, a técnica do eletrodo íon-seletivo Scanning (SIET) 12,13 ou o microeletrodos Ion Flux Estimação (MIFE) técnica desenvolvida por Newman e Shabala 14-16 estão comercialmente disponíveis e amplamente utilizado pela comunidade de investigação a fim de determinar íon específico fundentes ocorrendo a membrana da célula e do tecido através de uma variedade de animais, plantas e modelos de células de estar solteiro. Microeletrodos íon-seletivo têm sido usados ​​para medir hidrogênio, potássio e fluxo de cálcio através de raízes de plantas 17, cloreto de fluxo em artérias cerebrais de ratos 18 e em tubos polínicos 19, fluxo de hidrogênio em células da retina de skate 20, o fluxo de cálcio no osso do rato 21, vário íon fluxos em 22 de hifas fúngicas e no ra córnea 23, e, finalmente, o fluxo de cálcio durante a cura da ferida célula única 12,24. Veja também a seguinte análise para obter informações detalhadas sobre auto-referência microeletrodos íon-seletivo 25.

O artigo a seguir descreve em detalhes como preparar e realizar a medição de fluxos de iões extracelulares endógenos utilizando a auto-referência técnica microeletrodos íon-seletivo no nível da célula única.

Protocol

1. íon-seletivo de auto-referência microeletrodos Preparação Preparação de microeléctrodos selectivo de iões Calor puxar finas capilares de borosilicato murados sem filamentos (diâmetro externo 1,5 milímetros, 1,12 mm de diâmetro interno), utilizando um extrator de microeletrodos. Nota: Isto dá dicas de 3-4 m de diâmetro. Dicas menores têm maior resistência que torna mais microeléctrodos sensíveis ao ruído electrónico e também está associada com uma resposta mais lenta par…

Representative Results

Temos anteriormente demonstrado que o influxo de cálcio aparece depois de uma única célula ferindo 24. Nós, portanto, perguntado se outros fluxos de iões ocorrer após ferimento única célula. Utilizou-se o X. laevis oócito, um modelo bem estabelecido para única célula a cicatrização de feridas 30-34 e gravação eletrofisiológico 24,35-39. Curiosamente, íons de potássio são mais concentrados dentro X. oócitos laevis (cerca de 110 mM) de 40…

Discussion

Os passos mais críticos para a medição de fluxos de iões bem sucedida extracelulares in vivo são os seguintes: a redução do ruído, o fabrico correcto do micro-eléctrodos selectivos de iões e do eléctrodo de referência, e o posicionamento da amostra e ambos os eléctrodos.

A fim de minimizar o ruído, o sistema de gravação deve estar em uma ligação à terra (aterrada) gaiola de Faraday de preferência com um (isolamento de vibração) mesa com tampo de metal que tamb?…

Divulgazioni

The authors have nothing to disclose.

Acknowledgements

This work was supported by National Science Foundation grant MCB-0951199, and in part by the NIH grant EY01910, California Institute of Regenerative Medicine grants RB1-01417 and by the Fundação para a Ciência e Tecnologia (FCT) grant SFRH/BD/87256/2012.

Materials

IonAmp   BioCurrents Research Center, Woods Hole, MA, USA none amplifier created by the BioCurrents Research Center, Woods Hole, MA, USA; Similar system can be purchased from “XBL function matters” (http://www.xuyue.org/) or from “YoungerUSA” (http://www.youngerusa.com/) or from Applicable Electronics(http://www.applicableelectronics.com/)
IonAmp32   BioCurrents Research Center, Woods Hole, MA, USA none software created by the BioCurrents Research Center, Woods Hole, MA, USA; Similar system can be purchased from “XBL function matters” (http://www.xuyue.org/) or from “YoungerUSA” (http://www.youngerusa.com/) or from Applicable Electronics(http://www.applicableelectronics.com/)
Headstage pre-amplifier  BioCurrents Research Center, Woods Hole, MA, USA INA116 BSR Voltage Follower INA116, designed by the BioCurrents Research Center, Woods Hole, MA, USA; Similar system can be purchased from “XBL function matters” (http://www.xuyue.org/) or from “YoungerUSA” (http://www.youngerusa.com/) or from Applicable Electronics(http://www.applicableelectronics.com/)
MicroStep Driver  BioCurrents Research Center, Woods Hole, MA, USA none three MicroStep drivers are required for X, Y and Z-positioning; created by the BioCurrents Research Center, Woods Hole, MA, USA; Similar system can be purchased from “XBL function matters” (http://www.xuyue.org/) or from “YoungerUSA” (http://www.youngerusa.com/) or from Applicable Electronics(http://www.applicableelectronics.com/)
Manual micropositioner   World Precision Instruments  Model KITE-R Similar system can be purchased from Applicable Electronics(http://www.applicableelectronics.com/)
Magnetic stand    World Precision Instruments Model M10 Similar system can be purchased from Applicable Electronics(http://www.applicableelectronics.com/)
Vibration isolation table   Newport Inc.      Model VW-3036-OPT-023040 Similar system can be purchased from Applicable Electronics(http://www.applicableelectronics.com/)
Part of three dimentional micropositioner: angle bracket, 90°, slotted faces Newport Inc.      Model 360-90 Assemblage of the three dimantionnal micropositionner requires also Three electric rotary motors for X, Y, Z control, MPH-1 mounting arm with MCA-2 adjustable-angle post and Various Newport connectors and screws to bolt onto vibration table
Part of three dimentional micropositioner: Peg-Joining Dovetail Stage 0.5 inch X Travel Newport Inc.      460PD-X none
Part of three dimentional micropositioner: Quick-Mount Linear Stage, 0.5 inch XY Travel Newport Inc. 460A-XY none
Kwik-Fil thin walled borosilicate glass capillaries without filament  World Precision Instruments  TW150-4 none
Electrode puller  Narishige  PC-10 none
Metal rack Made in-house none Metal electrode holder made in-house by drilling 2 mm wide holes half centimeter spaced in a 10cm by 15cm rectangular base of steel
Oven QL Model 10 Lab Oven none
Silanization solution I  Sigma-Aldrich 85126 Hazardous, handle as recommended by provider 
Glass Petri dish; Pyrex Fisher Scientific 316060 none
Electrode/micropipette storage jar World Precision Instruments  E215 none
Glass dessicator Fisher Scientific 08-595E Contains Drierite dessicant (W.A. Hammond Drierite Co. Ltd, Xenia, OH, USA). Place petroleum jelly on the seal to make it airtight.
Plastic Pasteur pipette  Fisher Scientific 11597722 none
Bunsen burner Fisher Scientific S97329 none
Microscope slide Sigma-Aldrich S8902 none
Straight microelectrode holder Warner Instruments QSW-A15P with a gold 1 mm male connector and Ag/AgCl wire
Straight microelectrode holder  World Precision Instruments MEH3S with a AgCl(Ag+)pellet inside and a gold 2 mm male connector 
6 cm Petri dish VWR 60872-306 none
Nitex mesh Dynamic Aqua-Supply Ltd. NTX750 none
Glue; Loctite epoxy VWR 500043-451 Mix glue and hardener in equal parts in a plastic weighing boat and mix thoroughly. Sets quickly but leave at RT for 24 h for full curing
Deionized water  Sigma-Aldrich 99053 none
Sodium Chloride Sigma-Aldrich S7653 none
Potassium Chloride Sigma-Aldrich P9333 none
Calcium Chloride Sigma-Aldrich C1016 none
Magnesium Chloride Sigma-Aldrich M8266 none
Hepes Sigma-Aldrich H3375 none
Sodium Hydroxyde Sigma-Aldrich S8045 none
Potassium Acetate Sigma-Aldrich P1190 none
Agarose Sigma-Aldrich A9539 none

Riferimenti

  1. Weber, W. M., Liebold, K. M., Clauss, W. Amiloride-sensitive Na+ conductance in native Xenopus oocytes. Biochimica et biophysica acta. 1239, 201-206 (1995).
  2. McCaig, C. D., Song, B., Rajnicek, A. M. Electrical dimensions in cell science. Journal of cell science. 122, 4267-4276 (2009).
  3. Zhao, M. Electrical fields in wound healing-An overriding signal that directs cell migration. Seminars in cell & developmental biology. 20, 674-682 (2009).
  4. Jaffe, L. F., Nuccitelli, R. An ultrasensitive vibrating probe for measuring steady extracellular currents. The Journal of cell biology. 63, 614-628 (1974).
  5. Reid, B., Nuccitelli, R., Zhao, M. Non-invasive measurement of bioelectric currents with a vibrating probe. Nature protocols. 2, 661-669 (2007).
  6. Reid, B., Zhao, M. Measurement of bioelectric current with a vibrating probe. Journal of visualized experiments : JoVE. , (2011).
  7. Neher, E., Sakmann, B. Single-channel currents recorded from membrane of denervated frog muscle fibres. Nature. 260, 799-802 (1976).
  8. Moore, J. W. The patch clamp: single-channel recording. Science. 224, 50-51 (1984).
  9. Brown, A. L., Johnson, B. E., Goodman, M. B. Patch clamp recording of ion channels expressed in Xenopus oocytes. Journal of visualized experiments. , (2008).
  10. McCaig, C. D., Robinson, K. R. The ontogeny of the transepidermal potential difference in frog embryos. Developmental biology. 90, 335-339 (1982).
  11. Kuhtreiber, W. M., Jaffe, L. F. Detection of extracellular calcium gradients with a calcium-specific vibrating electrode. The Journal of cell biology. 110, 1565-1573 (1990).
  12. Cai, G., Cresti, M., Moscatelli, A. The use of the vibrating probe technique to study steady extracellular currents during pollen germination and tube growth. Fertilisation in Higher Plants: molecular and cytological aspects. , 235-252 (1999).
  13. Kunkel, J. G., Xu, Y., Shipley, A. M., Feijó, J. A. The use of non-invasive ion-selective microelectrode techniques for the study of plant development. Plant Electrophysiology – Theory and Methods. (ed Volkov AG. , 109-137 (2006).
  14. Ordonez, N. M., Shabala, L., Gehring, C., Shabala, S. Noninvasive microelectrode ion flux estimation technique (MIFE) for the study of the regulation of root membrane transport by cyclic nucleotides. Methods in molecular biology. 1016, 95-106 (2013).
  15. Tegg, R. S., Melian, L., Wilson, C. R., Shabala, S. Plant cell growth and ion flux responses to the streptomycete phytotoxin thaxtomin A: calcium and hydrogen flux patterns revealed by the non-invasive MIFE technique. Plant & cell physiology. 46, 638-648 (2005).
  16. Newman, I. A. Ion transport in roots: measurement of fluxes using ion-selective microelectrodes to characterize transporter function. Plant, cell & environment. 24, 1-14 (2001).
  17. Kochian, L. V., Shaff, J. E., Kuhtreiber, W. M., Jaffe, L. F., Lucas, W. J. Use of an extracellular, ion-selective, vibrating microelectrode system for the quantification of K(+), H (+), and Ca (2+) fluxes in maize roots and maize suspension cells. Planta. 188, 601-610 (1992).
  18. Doughty, J. M., Langton, P. D. Measurement of chloride flux associated with the myogenic response in rat cerebral arteries. The Journal of physiology. 534, 753-761 (2001).
  19. Messerli, M. A., Smith, P. J., Lewis, R. C., Robinson, K. R. Chloride fluxes in lily pollen tubes: a critical reevaluation. The Plant journal : for cell and molecular biology. 40, 799-812 (2004).
  20. Molina, A. J., et al. Neurotransmitter modulation of extracellular H+ fluxes from isolated retinal horizontal cells of the skate. The Journal of physiology. 560, 639-657 (2004).
  21. Marenzana, M., Shipley, A. M., Squitiero, P., Kunkel, J. G., Rubinacci, A. Bone as an ion exchange organ: evidence for instantaneous cell-dependent calcium efflux from bone not due to resorption. Bone. 37, 545-554 (2005).
  22. Lew, R. R. Ionic currents and ion fluxes in Neurospora crassa hyphae. Journal of experimental botany. 58, 3475-3481 (2007).
  23. Vieira, A. C., et al. Ionic components of electric current at rat corneal wounds. PloS one. 6, e17411 (2011).
  24. Luxardi, G., Reid, B., Maillard, P., Zhao, M. Single cell wound generates electric current circuit and cell membrane potential variations that requires calcium influx. Integrative biology : quantitative biosciences from nano to macro. 6, 662-672 (2014).
  25. Smith, P. J. S., Sanger, R. H., Messerli, M. A., Michael, A. C., Borland, L. H. . Electrochemical Methods for Neuroscience. , (2007).
  26. Smith, P. J., Hammar, K., Porterfield, D. M., Sanger, R. H., Trimarchi, J. R. Self-referencing, non-invasive, ion selective electrode for single cell detection of trans-plasma membrane calcium flux. Microscopy research and technique. 46, 398-417 (1999).
  27. Messerli, M. A., Smith, P. J. Construction theory, and practical considerations for using self-referencing of Ca(2+)-selective microelectrodes for monitoring extracellular Ca(2+) gradients. Methods in cell biology. 99, 91-111 (2010).
  28. Chambers, J., Hastie, T., Pregibon, D., Momirović, K., Mildner, V. Ch. 48. Compstat. , 317-321 (1990).
  29. Chambers, J. M., Cleveland, W. S., Kleiner, B., Tukey, P. A. . Graphical methods for data analysis. , (1983).
  30. Burkel, B. M., Benink, H. A., Vaughan, E. M., von Dassow, G., Bement, W. M. A Rho GTPase signal treadmill backs a contractile array. Developmental cell. 23, 384-396 (2012).
  31. Bement, W. M., Mandato, C. A., Kirsch, M. N. Wound-induced assembly and closure of an actomyosin purse string in Xenopus oocytes. Current biology : CB. 9, 579-587 (1999).
  32. Mandato, C. A., Bement, W. M. Contraction and polymerization cooperate to assemble and close actomyosin rings around Xenopus oocyte wounds. The Journal of cell biology. 154, 785-797 (2001).
  33. Benink, H. A., Bement, W. M. Concentric zones of active RhoA and Cdc42 around single cell wounds. The Journal of cell biology. 168, 429-439 (2005).
  34. Simon, C. M., Vaughan, E. M., Bement, W. M., Edelstein-Keshet, L. Pattern formation of Rho GTPases in single cell wound healing. Molecular biology of the cell. 24, 421-432 (2013).
  35. Petersen, C. C., Dupont, G. The initiation of a calcium signal in Xenopus oocytes. Cell calcium. 16, 391-403 (1994).
  36. Horisberger, J. D., Lemas, V., Kraehenbuhl, J. P., Rossier, B. C. Structure-function relationship of Na,K-ATPase. Annual review of physiology. 53, 565-584 (1991).
  37. Miledi, R. A calcium-dependent transient outward current in Xenopus laevis oocytes. Proceedings of the Royal Society of London. Series B, Containing papers of a Biological character. Royal Society. , 491-497 (1982).
  38. Miledi, R., Parker, I. Chloride current induced by injection of calcium into Xenopus oocytes. The Journal of physiology. 357, 173-183 (1984).
  39. Parker, I., Miledi, R. A calcium-independent chloride current activated by hyperpolarization in Xenopus oocytes. Proceedings of the Royal Society of London. Series B, Containing papers of a Biological character. Royal Society. 233, 191-199 (1988).
  40. Costa, P. F., Emilio, M. G., Fernandes, P. L., Ferreira, H. G., Ferreira, K. G. Determination of ionic permeability coefficients of the plasma membrane of Xenopus laevis oocytes under voltage clamp. The Journal of physiology. 413, 199-211 (1989).
  41. Adams, D. S., Levin, M. General principles for measuring resting membrane potential and ion concentration using fluorescent bioelectricity reporters. Cold Spring Harbor protocols. 2012, 385-397 (2012).
  42. Porterfield, D. M. Measuring metabolism and biophysical flux in the tissue, cellular and sub-cellular domains: recent developments in self-referencing amperometry for physiological sensing. Biosensors. 22, 1186-1196 (2007).
  43. McLamore, E. S., et al. A self-referencing glutamate biosensor for measuring real time neuronal glutamate flux. Journal of neuroscience methods. 189, 14-22 (2010).
  44. Yin, M., et al. Highly sensitive and fast responsive fiber-optic modal interferometric pH sensor based on polyelectrolyte complex and polyelectrolyte self-assembled nanocoating. Analytical and bioanalytical chemistry. 399, 3623-3631 (2011).
  45. Chatni, M. R., Porterfield, D. M. Self-referencing optrode technology for non-invasive real-time measurement of biophysical flux and physiological sensing. The Analyst. 134, 2224-2232 (2009).
  46. McLamore, E. S., Jaroch, D., Chatni, M. R., Porterfield, D. M. Self-referencing optrodes for measuring spatially resolved, real-time metabolic oxygen flux in plant systems. Planta. 232, 1087-1099 (2010).
  47. McLamore, E. S., et al. A self referencing platinum nanoparticle decorated enzyme-based microbiosensor for real time measurement of physiological glucose transport. Biosensors & bioelectronics. 26, 2237-2245 (2011).

Play Video

Citazione di questo articolo
Luxardi, G., Reid, B., Ferreira, F., Maillard, P., Zhao, M. Measurement of Extracellular Ion Fluxes Using the Ion-selective Self-referencing Microelectrode Technique. J. Vis. Exp. (99), e52782, doi:10.3791/52782 (2015).

View Video