Summary

脈絡叢上皮細胞の繊毛運動の観測<em>エクスビボ</em

Published: July 13, 2015
doi:

Summary

In this study, a detailed light microscopic technique was optimized for real-time observation and analysis of the motion of CPEC cilia ex vivo together with an electron microscopic method for ultrastructural analysis.

Abstract

The choroid plexus is located in the ventricular wall of the brain, the main function of which is believed to be production of cerebrospinal fluid. Choroid plexus epithelial cells (CPECs) covering the surface of choroid plexus tissue harbor multiple unique cilia, but most of the functions of these cilia remain to be investigated. To uncover the function of CPEC cilia with particular reference to their motility, an ex vivo observation system was developed to monitor ciliary motility during embryonic, perinatal and postnatal periods. The choroid plexus was dissected out of the brain ventricle and observed under a video-enhanced contrast microscope equipped with differential interference contrast optics. Under this condition, a simple and quantitative method was developed to analyze the motile profiles of CPEC cilia for several hours ex vivo. Next, the morphological changes of cilia during development were observed by scanning electron microscopy to elucidate the relationship between the morphological maturity of cilia and motility. Interestingly, this method could delineate changes in the number and length of cilia, which peaked at postnatal day (P) 2, while the beating frequency reached a maximum at P10, followed by abrupt cessation at P14. These techniques will enable elucidation of the functions of cilia in various tissues. While related techniques have been published in a previous report1, the current study focuses on detailed techniques to observe the motility and morphology of CPEC cilia ex vivo.

Introduction

Cilia are hair-like projections on the surface of most vertebrate cells, which have attracted attention by medical researchers because of a class of diseases termed ciliopathies24. Despite the ubiquitous expression of the organelle, a wide variety of ciliary functions have been reported, including motility and biosensing. For example, motile cilia on the mucoepithelial surface transport mucus5 and epithelial debris to the outlet of tracts, thereby preventing disease by clearing the surface of epithelia. Moreover, during early developmental periods and embryonic stages, cilia regulate the proliferation of stem cells6, and are involved in the determination of left–right asymmetry of the vertebrate body7.

Choroid plexus epithelial cells (CPECs) are derivatives of neuroepithelial cells that cover the surface of the choroid plexus tissue in the brain, which play important roles in maintaining homeostasis of the intracranial environment by production of cerebrospinal fluid (CSF). It has been previously demonstrated that CPECs have multiple non-motile cilia that regulate the production of CSF through G-protein-coupled receptors that are specifically concentrated on the cilia8. Although these cilia had been regarded as quiescent non-motile cilia, it was discovered that some CPEC cilia exhibit transient motility during the neonatal period1. This finding was quite important because it revealed that so-called non-motile cilia are not necessarily immotile from the beginning of development and might display transient motility during specific time windows, possibly in response to specific physiological demands and functions9. To precisely describe the motile nature of CPEC cilia, it is necessary to develop an ex vivo observation system that encompasses analysis of the kinetic profiles unique to CPEC cilia.

With respect to motility, although several technical reports have described observations of the motile cilia of the tracheal epithelium5,10, motile single-cell flagella11, so-called conventional motile cilia12, and nodal cilia13, detailed analytical methods applicable to relatively undulated structures such as the choroid plexus have not been well documented so far. Moreover, a high time resolution is required to analyze the ciliary movement of CPECs, in which expensive high-speed cameras are indispensable. To circumvent this necessity and simplify monitoring the ciliary motility of various cell types, a low cost, high-speed camera has been introduced, and an easily accessible and convenient method to record the motility of motile cilia, especially to describe the speed and pattern of motion of each cilium, has been developed1. Moreover, original image analysis software “TI Workbench” has been used here to facilitate detailed analysis of motility. Collectively, this method provides a new concise strategy to analyze ciliary motion together with correlative scanning electron microscopy (SEM), which can be adopted in a wide range of cilium research.

Protocol

プロトコルおよび実験動物の使用は、山梨と早稲田大学大学の制度的動物のケアと使用委員会によって承認されました。動物のケアは、機関のガイドラインに従って行いました。 1. CPEC準備次の装置と材料を準備します。実体顕微鏡下から照明を送信できることが好ましいです。時計屋のピンセット(デュモン#3または#4)、火炎滅菌し、ストレート操作はさみ…

Representative Results

ワークフローの概要は、装置の画像を含む、 図1に示されています。 CPECsのライブの動きの観察 映画1は、周産期のマウスから単離したCPECsの映画を示しており、 ムービー2ムービー1の画像の拡大図を示している。それは、個々の繊毛の先端は映画のものと比較して、静止画ではあまり明確であることに留意すべ?…

Discussion

この方法の展望

ここに記載された技術は、以前に公開された方法よりも繊毛のより詳細な分析を提供していませんが、この技術の重要性は、簡単に繊毛運動ex vivoでのあらゆる種類のスクリーニングに適用することができるシステムと費用対効果のシンプルさにあります。具体的には、TI Workbenchは、より簡単に繊毛運動を観察し、分析する研究者を可能にシンプル?…

Divulgazioni

The authors have nothing to disclose.

Acknowledgements

This work was supported by a Project for Private Universities: matching fund subsidy from the Ministry of Education, Culture, Sports, Science and Technology (MEXT) of Japan (T.I.) and Grants-in-Aid for Scientific Research (C) from MEXT (S.T. and K.N).

Materials

for both live imaging and SEM preparation
stereo microscope Olympus SZX7
flat paper towel
Φ10-cm plastic dish
100-mL beaker
straight operating scissors Sansyo S-2B
watchmaker forceps Dumont No.DU-3 or -4, INOX
for live cell imaging
glass bottom dish Matsunami Glass D110300
for SEM preparation
alminum foil
5-mL glass vial with a polyethylene cap Nichiden Rika-Glass PS-5A
transfer pipette Samco Scientific SM251-1S for specimen tranfer
toothpick for specimen transfer
ion sputter with gold-palladium Hitachi E-1030
critical point dryer Hitachi HCP-2
for live cell imaging
inverted microscope Olympus IX81
100 W mercury lump housing and power supply Olympus U-ULH and BH2-RFL-T3
100 W mercury lamp Ushio USH103D
DIC condenser, n.a. 0.55 Olympus IX-LWUCD
electrrical shutter Vincent Associates VS35S22M1R3-24 and VMM-D1 manual shutter can be used.
band-pass filter (400-700 nm, Φ45 mm) Koshin Kagaku C10-110621-1
ND filter (Φ45 mm) Olympus 45ND6, 45ND25 combination of 25% and 6% ND filters are used
objective lens (water immersion) with DIC element Olympus UApo 40XW/340, n.a., 1.15 with IX-DPAO40
high-speed video camera Allied Vision Technologies GE680 >= 200 Hz frame rate and 1 msec expose time
image acquisition / analysis software in-hous software TI Workbench capable of acquisition at high frame rates.
PC for camera control / analysis Apple Mac Pro
vibration isolation table Meiritsu Seiki AD0806
weight for tissue Warner Instruments slice anchor kits It can be made with nylon mesh glued to a U-shape squashed Φ0.5mm platinum wire.
for SEM
inverted microscope Olympus IX81
scanning electron microscope JEOL JSM-6510
for live cell imaging
ethanol  Wako Chemicals 057-00456
Leibovitz L-15 medium Life Technologies 11415-064
for SEM preparation
ethanol  Wako Chemicals 057-00456
Hank's balanced salt solution Life Technologies 14170112
paraformaldehyde  Merck 1040051000
glutaraldehyde  Nacalai tesque 17003-05
isoamyl acetate Nacalai tesque 02710-95
Molecular Sieves 4A 1/8  Wako Chemicals 130-08655 for preparation of anhydrous ethanol
phosphate buffer saline (PBS) Sigma-Aldrich D1408
phosphate buffer, 0.1 M To make 100 ml, mix 19.0 ml of 0.1 M NaH2PO4 and 81.0 ml of 0.1 M Na2HPO4 
monosodium phosphate (dihydrate) Nacalai tesque 31718-15
disodium phosphate (anhydrous) Nacalai tesque 31801-05
suclose Nacalai tesque 30406-25
osmium tetroxide Nisshin EM 300
dry ice

Riferimenti

  1. Nonami, Y., Narita, K., Nakamura, H., Inoue, T., Takeda, S. Developmental changes in ciliary motility on choroid plexus epithelial cells during the perinatal period. Cytoskeleton. 70 (12), 797-803 (2013).
  2. Bisgrove, B. W., Yost, H. J. The roles of cilia in developmental disorders and disease. Development. 133 (21), 4131-4143 (2006).
  3. Fliegauf, M., Benzing, T., Omran, H. When cilia go bad: cilia defects and ciliopathies. Nature Reviews. Molecular Cell Biology. 8 (11), 880-893 (2007).
  4. Oh, E. C., Katsanis, N. Cilia in vertebrate development and disease. Development. 139 (3), 443-448 (2012).
  5. Shah, A. S., Ben-Shahar, Y., Moninger, T. O., Kline, J. N., Welsh, M. J. Motile cilia of human airway epithelia are chemosensory. Science(New York, N.Y). 325 (5944), 1131-1134 (2009).
  6. Kiprilov, E. N., et al. Human embryonic stem cells in culture possess primary cilia with hedgehog signaling machinery). The Journal of Cell Biology. 180 (5), 897-904 (2008).
  7. Hirokawa, N., Tanaka, Y., Okada, Y., Takeda, S. Nodal flow and the generation of left-right asymmetry. Cell. 125 (1), 33-45 (2006).
  8. Narita, K., Kozuka-Hata, H., et al. Proteomic analysis of multiple primary cilia reveals a novel mode of ciliary development in mammals. Biology Open. 1 (8), 815-825 (2012).
  9. Takeda, S., Narita, K. Structure and function of vertebrate cilia, towards a new taxonomy. Differentiation; Research in Biological Diversity. 83 (2), S4-S11 (2012).
  10. Ikegami, K., Sato, S., Nakamura, K., Ostrowski, L. E., Setou, M. Tubulin polyglutamylation is essential for airway ciliary function through the regulation of beating asymmetry. Proceedings of the National Academy of Sciences of the United States of America. 107 (23), 10490-10495 (2010).
  11. Foster, K. W. Analysis of the ciliary/flagellar beating of Chlamydomonas. Methods in Cell Biology. 91, 173-239 (2009).
  12. Lechtreck, K. -. F., Sanderson, M. J., Witman, G. B. High-speed digital imaging of ependymal cilia in the murine brain. Methods in Cell Biology. 91, 255-264 (2009).
  13. Okada, Y., Hirokawa, N. Observation of nodal cilia movement and measurement of nodal flow. Methods in Cell Biology. 91, 265-285 (2009).
  14. Chilvers, M. A., Rutman, A., O’Callaghan, C. Ciliary beat pattern is associated with specific ultrastructural defects in primary ciliary dyskinesia. The Journal of Allergy and Clinical Immunology. 112 (3), 518-524 (2003).
  15. Takeda, S., et al. Left-right asymmetry and kinesin superfamily protein KIF3A: new insights in determination of laterality and mesoderm induction by kif3A-/- mice analysis. The Journal of Cell Biology. 145 (4), 825-836 (1999).
  16. Karnovsky, M. J. A formaldehyde-glutaraldehyde fixative of high osmolarity for use in electron microscopy. The Journal of Cell Biology. 27, 137-138A (1965).
check_url/it/52991?article_type=t

Play Video

Citazione di questo articolo
Inoue, T., Narita, K., Nonami, Y., Nakamura, H., Takeda, S. Observation of the Ciliary Movement of Choroid Plexus Epithelial Cells Ex Vivo. J. Vis. Exp. (101), e52991, doi:10.3791/52991 (2015).

View Video