Summary

Desarrollo de una<em> In Vitro</em> Plataforma ocular para probar las lentes de contacto

Published: April 06, 2016
doi:

Summary

Los modelos actuales in vitro para la evaluación de las lentes de contacto (CLS) y otras aplicaciones relacionadas con el ojo son muy limitadas. La plataforma ocular presentado simula fisiológica flujo lagrimal, volumen de lágrimas, la exposición al aire y el desgaste mecánico. Este sistema es muy versátil y se puede aplicar a diversos análisis in vitro con CLS.

Abstract

Currently, in vitro evaluations of contact lenses (CLs) for drug delivery are typically performed in large volume vials,1-6 which fail to mimic physiological tear volumes.7 The traditional model also lacks the natural tear flow component and the blinking reflex, both of which are defining factors of the ocular environment. The development of a novel model is described in this study, which consists of a unique 2-piece design, eyeball and eyelid piece, capable of mimicking physiological tear volume. The models are created from 3-D printed molds (Polytetrafluoroethylene or Teflon molds), which can be used to generate eye models from various polymers, such as polydimethylsiloxane (PDMS) and agar. Further modifications to the eye pieces, such as the integration of an explanted human or animal cornea or human corneal construct, will permit for more complex in vitro ocular studies. A commercial microfluidic syringe pump is integrated with the platform to emulate physiological tear secretion. Air exposure and mechanical wear are achieved using two mechanical actuators, of which one moves the eyelid piece laterally, and the other moves the eyeballeyepiece circularly. The model has been used to evaluate CLs for drug delivery and deposition of tear components on CLs.

Introduction

Dos importantes áreas de interés dentro de la arena de lentes de contacto (CL) incluyen malestar y el desarrollo de nuevas aplicaciones CL. Elucidar los mecanismos subyacentes CL malestar es un tema que ha eludido el campo durante décadas. 8 El desarrollo de la novela, CL funcionales, tales como dispositivos de liberación de fármacos 1,3,9 y biosensores, 10-12 es un área de creciente interés, con los mercados potenciales sustanciales. En ambas circunstancias, un sofisticado modelo in vitro podría proporcionar información relevante para ayudar en la selección de materiales de lentes o características de diseño apropiado durante la fase de desarrollo. Por desgracia, la corriente en modelos in vitro para la evaluación de las NFT y otras aplicaciones relacionadas con el ojo son relativamente crudo y poco sofisticado. Tradicionalmente, se llevan a cabo en los estudios in vitro que evalúan CL deposición de la película lagrimal o la administración de fármacos en viales de volumen estáticas, grandes que contienen un volumen de fluido fijo, que GREAtly excede cantidades fisiológicas. Además, este modelo simple no tiene la componente de flujo de lágrima natural y el reflejo del parpadeo, ambos de los cuales son la definición de factores del entorno ocular.

El desarrollo de un "modelo" sofisticado, fisiológicamente relevante ojo requerirá un enfoque multidisciplinar y requieren sustancial en la validación in vivo. Por estas razones, el marco fundamental para nuestro modelo de ojo in vitro es altamente versátil, de tal manera que el modelo puede ser mejorado continuamente a través de las actualizaciones futuras y modulaciones. Hasta la fecha, el modelo es capaz de simular el volumen de lágrimas, el flujo de lágrimas, desgaste mecánico y la exposición al aire. El objetivo es crear un modelo in vitro que proporcionará resultados significativos, que es predictivo y conexión a in vivo y ex vivo observaciones.

Protocol

Todos los experimentos se realizaron de acuerdo y cumpliendo con todas las directrices pertinentes señalados por la Universidad de Waterloo comité de ética de la investigación con animales. Los ojos bovinos son donados generosamente de un matadero local. 1. Modelo de ojos Diseño y producción de moldes 13 Diseño de los modelos de ojo de acuerdo con las dimensiones fisiológicas promedio de los ojos humanos adultos. 13 Deje un espacio de 250 m entre el glo…

Representative Results

Los moldes ojo sintetizados obtenidos a partir de la tienda de la máquina y de la impresión 3-D se muestran en la Figura 1. Estos moldes se pueden utilizar con una variedad de polímeros, tales como PDMS y agarosa, para producir los oculares con las propiedades deseadas. El conjunto de señas de la plataforma de modelo de ojo con una bomba de jeringa de microfluidos se muestra en la Figura 2. La plataforma simula el desgaste mecánico a través de la r…

Discussion

Hay tres pasos críticos dentro del protocolo que requieren una atención especial: diseño y producción de moldes (sección 1.1), conjunto de plataforma (sección 2.2.1-2.2.3), y el seguimiento de la ejecución experimental (sección 2.2.4-2.2.7 ). En términos del diseño y la producción de moldes (sección 1.1), la pieza de globo ocular debe ser diseñado de acuerdo con las dimensiones de una córnea humana. Sin embargo, se puede requerir múltiples prototipos del molde antes de una pieza globo ocular puede ser cre…

Divulgazioni

The authors have nothing to disclose.

Acknowledgements

Los autores desean reconocer nuestra fuente de financiación CRSNG 20/20 Red para el Desarrollo de Materiales Avanzados oftálmicos.

Materials

Arduino Uno R3 (Atmega328 – assembled) Adafruit 50 Board
Stepper motor Adafruit 324 Motor and Motor shield
Equal Leg Coupler 1.6mm 1/16" VWR CA11009-280 50 pcs of tube connector
Tubing PT/SIL 1/16"x1/8" VWR 16211-316 Case of 50feet
PDMS Dow Corning Sylgard 184 Solar Cell Encapsulation
Agarose, Type 1-A, low EEO Sigma-Aldrich A0169-25G
PHD UltraTM Harvard Apparatus 703006 MicroFluidic Pump
Bovine cornea Cargill, Guelph/ON
Soldidworks Dassault Systemes Software
3-D printing University of Waterloo – 3D Print Centre
Dissection tools Fine Science Tools General dissection tools
Medium 199 Sigma-Aldrich Culture medium storage for cornea
Fetal bovine serum Thermo Fisher Add to culture medium, 3% total volume

Riferimenti

  1. Phan, C. M., Subbaraman, L. N., Jones, L. In vitro drug release of natamycin from beta-cyclodextrin and 2-hydroxypropyl-beta-cyclodextrin-functionalized contact lens materials. J Biomater Sci Polym Ed. 25, 1907-1919 (2014).
  2. Peng, C. C., Kim, J., Chauhan, A. Extended delivery of hydrophilic drugs from silicone-hydrogel contact lenses containing vitamin E diffusion barriers. Biomaterials. 31, 4032-4047 (2010).
  3. Hui, A., Willcox, M., Jones, L. In vitro and in vivo evaluation of novel ciprofloxacin-releasing silicone hydrogel contact lenses. Invest Ophthalmol Vis Sci. 55, 4896-4904 (2014).
  4. Boone, A., Hui, A., Jones, L. Uptake and release of dexamethasone phosphate from silicone hydrogel and group I, II, and IV hydrogel contact lenses. Eye Contact Lens. 35, 260-267 (2009).
  5. Lorentz, H., Heynen, M., Trieu, D., Hagedorn, S. J., Jones, L. The impact of tear film components on in vitro lipid uptake. Optom Vis Sci. 89, 856-867 (2012).
  6. Hall, B., Phan, C. M., Subbaraman, L., Jones, L. W., Forrest, J. Extraction versus in situ techniques for measuring surface-adsorbed lysozyme. Optom Vis Sci. 91, 1062-1070 (2014).
  7. Mishima, S., Gasset, A., Klyce, S. D., Baum, J. L. Determination of tear volume and tear flow. Invest Ophthalmol Vis Sci. 5, 264-276 (1966).
  8. Nichols, J. J., et al. The TFOS international workshop on contact lens discomfort: executive summary. Invest Ophthalmol Vis Sci. 54, 7-13 (2013).
  9. Peng, C. C., Burke, M. T., Carbia, B. E., Plummer, C., Chauhan, A. Extended drug delivery by contact lenses for glaucoma therapy. J Control Release. 162, 152-158 (2012).
  10. Faschinger, C., Mossbock, G. Continuous 24 h monitoring of changes in intraocular pressure with the wireless contact lens sensor Triggerfish. First results in patients. Der Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft. 107, 918-922 (2010).
  11. Shaw, A. J., Davis, B. A., Collins, M. J., Carney, L. G. A technique to measure eyelid pressure using piezoresistive sensors. IEEE transactions on bio-medical engineering. 56, 2512-2517 (2009).
  12. Liao, Y. T., Yao, H. F., Lingley, A., Parviz, B., Otis, B. P. A 3-mu W CMOS glucose sensor for wireless contact-lens tear glucose monitoring. Ieee J Solid-St Circ. 47, 335-344 (2012).
  13. Coster, D. J. . Cornea. , (2002).
  14. Parekh, M., et al. A simplified technique for in situ excision of cornea and evisceration of retinal tissue from human ocular globe. Journal of visualized experiments : JoVE. , e3765 (2012).
  15. Way, S. Gear and pinion. US patent. , (1942).
  16. Lorentz, H., et al. Contact lens physical properties and lipid deposition in a novel characterized artificial tear solution. Molecular vision. 17, 3392-3405 (2011).
  17. Furukawa, R. E., Polse, K. A. Changes in tear flow accompanying aging. American journal of optometry and physiological optics. 55, 69-74 (1978).
  18. Bajgrowicz, M., Phan, C. M., Subbaraman, L., Jones, L. Release of ciprofloxacin and moxifloxacin from daily disposable contact lenses from an in vitro eye model. Invest Ophthalmol Vis Sci. , (2015).
  19. Luensmann, D., Zhang, F., Subbaraman, L., Sheardown, H., Jones, L. Localization of lysozyme sorption to conventional and silicone hydrogel contact lenses using confocal microscopy. Current eye research. 34, 683-697 (2009).
  20. Tieppo, A., Pate, K. M., Byrne, M. E. In vitro controlled release of an anti-inflammatory from daily disposable therapeutic contact lenses under physiological ocular tear flow. Eur J Pharm Biopharm. 81, 170-177 (2012).
  21. Ali, M., et al. Zero-order therapeutic release from imprinted hydrogel contact lenses within in vitro physiological ocular tear flow. J Control Release. 124, 154-162 (2007).
  22. White, C. J., McBride, M. K., Pate, K. M., Tieppo, A., Byrne, M. E. Extended release of high molecular weight hydroxypropyl methylcellulose from molecularly imprinted, extended wear silicone hydrogel contact lenses. Biomaterials. 32, 5698-5705 (2011).
  23. Kaczmarek, J. C., Tieppo, A., White, C. J., Byrne, M. E. Adjusting biomaterial composition to achieve controlled multiple-day release of dexamethasone from an extended-wear silicone hydrogel contact lens. J Biomater Sci Polym Ed. 25, 88-100 (2014).
  24. Mohammadi, S., Postnikoff, C., Wright, A. M., Gorbet, M. Design and development of an in vitro tear replenishment system. Ann Biomed Eng. 42, 1923-1931 (2014).
  25. Lorentz, H., Heynen, M., Khan, W., Trieu, D., Jones, L. The impact of intermittent air exposure on lipid deposition. Optom Vis Sci. 89, 1574-1581 (2012).
  26. Peng, C. C., Fajardo, N. P., Razunguzwa, T., Radke, C. J. In vitro spoilation of silicone-hydrogel soft contact lenses in a model-blink cell. Optom Vis Sci. 92, 768-780 (2015).
  27. Liu, P., et al. Dissolution studies of poorly soluble drug nanosuspensions in non-sink conditions. AAPS PharmSciTech. 14, 748-756 (2013).
check_url/it/53907?article_type=t

Play Video

Citazione di questo articolo
Phan, C., Walther, H., Gao, H., Rossy, J., Subbaraman, L. N., Jones, L. Development of an In Vitro Ocular Platform to Test Contact Lenses. J. Vis. Exp. (110), e53907, doi:10.3791/53907 (2016).

View Video