Summary

Herstellung und Charakterisierung eines Conformal-Haut-like Electronic System for Quantitative, kutane Wundmanagement

Published: September 02, 2015
doi:

Summary

This article presents methods to fabricate and characterize a conformal, skin-like electronic system and protocols for the use in clinical applications, particularly on cutaneous wound management.

Abstract

Recent advances in the development of electronic technologies and biomedical devices offer opportunities for non-invasive, quantitative assessment of cutaneous wound healing on the skin. Existing methods, however, still rely on visual inspections through various microscopic tools and devices that normally include high-cost, sophisticated systems and require well trained personnel for operation and data analysis. Here, we describe methods and protocols to fabricate a conformal, skin-like electronics system that enables conformal lamination to the skin surface near the wound tissues, which provides recording of high fidelity electrical signals such as skin temperature and thermal conductivity. The methods of device fabrication provide details of step-by-step preparation of the microelectronic system that is completely enclosed with elastomeric silicone materials to offer electrical isolation. The experimental study presents multifunctional, biocompatible, waterproof, reusable, and flexible/stretchable characteristics of the device for clinical applications. Protocols of clinical testing provide an overview and sequential process of cleaning, testing setup, system operation, and data acquisition with the skin-like electronics, gently mounted on hypersensitive, cutaneous wound and contralateral tissues on patients.

Introduction

In der klinischen Studie und der biomedizinischen Forschung, hat die Überwachung der Wundheilung auf ein invasives Verfahren, das auf der histologische Auswertung von Gewebe morphologischen Veränderung in Wunden 1,2 basierend fokussiert ist. Kürzlich, schnelle Fortschritte in der elektronischen Technologien ermöglichen die Entwicklung von hochgenauen Abbildung und Analyse-Tools, die eine Sichtprüfung kann die Wundheilung via Digital-Imaging-3,4 oder der konfokalen Scanmikroskopie und Spektroskopie 4,5. Jedoch brauchen Abbildungs ​​Ansätze erfordern hohe Kosten, komplizierte optische Werkzeuge und Operationen und vor allem müssen die Patienten während des Tests immobilisiert werden. Daher besteht ein Bedarf für neue Geräte und Systeme die quantitative, non-invasive, einfach zu bedienende, kostengünstige und multifunktionale genauere Wundversorgung anzubieten.

Hier wird eine hautartige elektronisches System, das eine präzise Echtzeitzuordnung von Temperatur und Wärme condu bietet stellen wirctivity und liefert eine genaue Höhe der Heizung an Wundstellen über konforme Laminierung des Geräts nicht-invasiv. Diese Vorrichtung stellt eine Klasse von Technologien, haut montiert epidermalen elektronische Systeme, die ausgelegt sind, um mechanische und den Materialeigenschaften entsprechen (Gesamtdicke, Biegesteifigkeit wirksam Moduli und Massendichte) der Oberhaut 6-9.

Das Gerät ist in einem biokompatiblen, hautfreundlichen, wasserdicht gestaltet, und wiederverwendbare Form, gewaschen und für klinische Anwendungen am Patienten 10 desinfiziert werden kann. Die konforme elektronische Gerät in der Nähe der Wundgewebe montiert fängt die Entzündungsphase (einer der Wundheilung), durch erhöhte Durchblutung und enzymatische Reaktionen auf die Wunde 11,12 verursacht, durch die quantitative Erfassung der Temperatur 8 und Wärmeleitfähigkeit 13, um Trink korreliert . Experimentelle und theoretische Studien zu bestimmen, eine optimale Mechanik-Design, um Unterkunftaß natürlichen Bewegungen und angewandte Stämme ohne mechanische Bruch und erfassen die zugrunde liegende Physik der Dehnung Mechanik der hautähnlichen Elektronik, die konform auf der Hautoberfläche, die Übernahme von High-Fidelity-Signale bietet Laminate.

Die in diesem Artikel beschriebenen Protokolle stellen die Verfahren der Mikrofabrikation für hautartige elektronische Systeme, Testvorbereitung einschließlich Gerätereinigung Geräteanordnung in einer klinischen Umgebung und klinischen Anwendungen zur quantitativen Überwachung der Temperatur und der thermischen Leitfähigkeit auf Hautwunden.

Protocol

Die Experimente zur Bauelementherstellung, Haut Laminierung und Charakterisierung in den 1, 2 gezeigt, und 4 beteiligten zwei Freiwillige, alle in der Bio-Schnittstelle Nanoengineering Laboratory an der Virginia Commonwealth University (VCU), Richmond, VA, USA durchgeführt. Diese Studie wurde von der VCU Institutional Review Board genehmigt (Protokollnummer: HM20001454) und folgte den Forschungs Richtlinien der Human Research VCU. (: STU69718 Nummer) vom Institutional Review Board, der…

Representative Results

Abbildung 1 gibt einen Überblick über die Eigenschaften der konformen, hautartigen elektronischen Systems für die quantitative, kutaner Wundbehandlung an Patienten ausgelegt. Das multifunktionale elektronische Gerät besteht aus Mikro fraktale Strukturen 3,14 und fadenförmigen Serpentinenspuren 9,17 auf einer dünnen Elastomermembran, die außergewöhnliche mechanische Dehnbarkeit und Biegefähigkeit bietet. Die kompatiblen Gerät, das vollständig von Silikonschichten eingesch…

Discussion

This article highlights the methods and protocols to fabricate a conformal, skin-like electronics system that enables conformal lamination near the wound tissues, which offers quantitative measurement of skin temperature and thermal conductivity mapping on the skin.

The key features include the utilization of novel techniques of materials transfer printing and hard-soft materials integration to design and develop the flexible/stretchable, soft electronic device. The use of biocompatible, elec…

Disclosures

The authors have nothing to disclose.

Acknowledgements

Diese Arbeit wurde von der Startup-Finanzierung von der School of Engineering, Virginia Commonwealth University und einige elektronische Geräte wurden in den Mikrofabrikationsanlagen in der Wright Virginia Microelectronics Center vorbereitet unterstützt. Wir erkennen die Forscher, die Beiträge für das Gerät und klinischen Daten (3 und 5 in diesem Papier), aus der veröffentlichten Artikels 10 erworben werden. W.-HY dank Yoshiaki Hattori für die maßgeschneiderte, Daten-Recording-Software.

Materials

3" Silicon wafer University Wafer, USA N/A Use as carrier to fabricate the device
Acetone Fisher Scientific, USA A18-1 Use to clean a wafer and to remove photoresist
Isopropanol (IPA) Fisher Scientific, USA A459-1 Use to clean a wafer
AZ4620 photoresist AZ Electrionic Materials, USA N/A Use to make patterns on metals and polymers
AZ400K developer AZ Electrionic Materials, USA N/A Use to develop AZ4620 photoresist
Chromium etchant Transene, USA 1020AC Use to etch Cr layer of device
Copper etchant Transene, USA ASP-100 Use to etch Cu layer of device
Sylgard 184 Silicone Elastomer Kit (PDMS) Dow Corning, USA 39100000 Use as a substrate for 'dry' retrieval
PI2545 polyimide HD MicroSystem, USA N/A Use to encapsulate metal layer
Solaris Smooth-On, USA N/A Use as substrate and to encapsulate device
Petridish Carolina, USA 741255 Use as mold to make substrate
Water-Soluble Wave Solder Tape 5414 3M, USA AM000000217 Use to retrive device from PDMS layer
High Activity Liquid Stainless Steel Flux Worthington, USA 331929 Use to remove oxidation layer on Cu
Flexible, micro-film cable Elform, USA N/A (customized) Use to make the electrical connection between the electronic device and the data acquisition system
pH Neutral Cleaner Australian Gold, USA N/A Use as disinfectant solution to clean device in clinical testing
Solder Kester, USA 24-6337-9703 Use as material to solder hard wires
Ultraviolet lamp Cole-Parmer, USA 97600-00 Use to activate PDMS layer as hydrophilic surface
Multiplexer FixYourBoard, USA U802 Use to acquire measurements from six sensing components 
DC/AC current source Keithley, USA 6221 Use to supply current
SMD Digital Hot Air Rework Station Aoyue, China 968A+ Use to solder hard wires, to electrically connect between the device and external instruments
Infrared camera FLIR, USA 435-0001-01-00 Use to take infrared images in experiment
Digital multimeter Fluke, USA 117 Use to check electrical connection
Lock-in amplifier Stanford Research System, USA SR830 Use to perform four-point-probe-measurement
Electron beam evaporator 9 scale Vacuum Products, USA N/A (customized) Use to deposit thin films (Cu and SiO2)

References

  1. Dargaville, T. R., et al. Sensors and imaging for wound healing: A review. Biosens Bioelectron. 41, 30-42 (2013).
  2. Panuncialman, J., Hammerman, S., Carson, P., Falanga, V. Wound edge biopsy sites in chronic wounds heal rapidly and do not result in delayed wound healing. J Invest Dermatol. 129, S47-S47 (2009).
  3. Hess, C. T., Kirsner, R. S. . Orchestrating Wound Healing: Assessing and Preparing the Wound Bed. Adv Skin Wound Care. 16 (5), 246-257 (2003).
  4. Lange-Asschenfeldt, S., et al. Applicability of confocal laser scanning microscopy for evaluation and monitoring of cutaneous wound healing. J Biomed Opt. 17 (7), (2012).
  5. Crane, N. J., Elster, E. A. Vibrational spectroscopy: a tool being developed for the noninvasive monitoring of wound healing. J Biomed Opt. 17 (7), (2012).
  6. Jeong, J. W., et al. Materials and Optimized Designs for Human-Machine Interfaces Via Epidermal Electronics. Adv Mater. 25 (47), 6839-6846 (2013).
  7. Kim, D. H., et al. Epidermal Electronics. Science. 333 (6044), 838-843 (2011).
  8. Webb, R. C., et al. Ultrathin conformal devices for precise and continuous thermal characterization of human skin (vol 12, pg 938). Nat Mater. 12, 1078-1078 (2013).
  9. Yeo, W. H., et al. Multifunctional Epidermal Electronics Printed Directly Onto the Skin. Adv Mater. 25 (20), 2773-2778 (2013).
  10. Hattori, Y., et al. Multifunctional skin-like electronics for quantitative, clinical monitoring of cutaneous wound healing. Adv Healthc Mater. 3 (10), 1597-1607 (2014).
  11. Guo, S., DiPietro, L. A. Factors Affecting Wound Healing. J Dent Res. 89 (3), 219-229 (2010).
  12. Matzeu, G., et al. Skin temperature monitoring by a wireless sensor. Ieee Ind Elec. , 3533-3535 (2011).
  13. Cahill, D. G. Thermal-Conductivity Measurement from 30-K to 750-K – the 3-Omega Method. Rev Sci Instrum. 61 (2), 802-808 (1990).
  14. Fan, J. A., et al. Fractal Design Concepts for Stretchable Electronics. Nature Commun. 5 (3266), (2013).
  15. Song, Y. M., et al. Digital cameras with designs inspired by the arthropod eye. Nature. 497 (7447), 95-99 (2013).
  16. Jeong, J. W., et al. Capacitive Epidermal Electronics for Electrically Safe, Long-Term Electrophysiological Measurements. Adv Healthc Mater. 3 (5), 642-648 (2013).
  17. Zhang, Y. H., et al. Experimental and Theoretical Studies of Serpentine Microstructures Bonded To Prestrained Elastomers for Stretchable Electronics. Adv Funct Mater. 24 (14), 2028-2037 (2014).
  18. Carlson, A., Bowen, A. M., Huang, Y. G., Nuzzo, R. G., Rogers, J. A. Transfer Printing Techniques for Materials Assembly and Micro/Nanodevice Fabrication. Adv Mater. 24 (39), 5284-5318 (2012).
  19. Yeo, W. H., Webb, R. C., Lee, W., Jung, S., Rogers, J. A. Bio-integrated electronics and sensor systems. Proc Spie. 8725, (2013).
  20. Chung, H. J., et al. Fabrication of Releasable Single-Crystal Silicon–Metal Oxide Field-Effect Devices and Their Deterministic Assembly on Foreign Substrates. Adv Funct Mater. 21 (16), 3029-3036 (2011).
  21. Kim, H. S., et al. Unusual strategies for using indium gallium nitride grown on silicon (111) for solid-state lighting. Proc Natl Acad Sci U.S.A. 108 (25), 10072-10077 (2011).
  22. Padilla-Medina, J. A., et al. Assessment technique for acne treatments based on statistical parameters of skin thermal images. J Biomed Opt. 19 (4), 046019-046019 (2014).

Play Video

Cite This Article
Lee, W., Kwon, O., Lee, D. S., Yeo, W. Fabrication and Characterization of a Conformal Skin-like Electronic System for Quantitative, Cutaneous Wound Management. J. Vis. Exp. (103), e53037, doi:10.3791/53037 (2015).

View Video