Encyclopedia of Experiments
Biological Techniques
JoVE 비디오를 활용하시려면 도서관을 통한 기관 구독이 필요합니다.  전체 비디오를 보시려면 로그인하거나 무료 트라이얼을 시작하세요.
Encyclopedia of Experiments Biological Techniques
Mitochondrial Calcium Uptake Assay: A Plate Reader-Based Method for Measuring Calcium Uptake in Isolated Mitochondria

Mitochondrial Calcium Uptake Assay: A Plate Reader-Based Method for Measuring Calcium Uptake in Isolated Mitochondria

내레이션 대본

To start mitochondrial calcium uptake measurement using a plate reader, program the reader to perform a kinetic read of calcium green-5N fluorescence every second, for a total time of 1,000 seconds.

Program the reagent injectors to dispense 5 microliters of calcium chloride solution at various time points. Then, prime the injectors with the calcium chloride solution that will be used.

Next, add 200 micrograms of isolated mitochondria to one well of a 96-well plate, and add the appropriate volume of potassium chloride buffer to achieve a total volume of 197 microliters. Then, add 1 microliter of 1 molar pyruvate, 1 microliter of 500 millimolar malate, and 1 microliter of 1 millimolar calcium green-5N stock, and mix gently by pipetting.

Incubate protected from light at room temperature for 2 minutes, for mitochondria to become energized. Finally, start the pre-programmed kinetic protocol, and monitor calcium green-5N fluorescence.

Related Videos

Read Article