Summary

生体内 、経皮、ニードルベース、腎大衆の光コヒーレンストモグラフィー

Published: March 30, 2015
doi:

Summary

Optical coherence tomography (OCT) is a high resolution imaging technique that allows analysis of tissue specific optical properties providing the means for tissue differentiation. We developed needle based OCT, providing real-time imaging combined with on-the-spot tumor differentiation. This publication describes a method for percutaneous, needle based OCT of renal masses.

Abstract

Optical coherence tomography (OCT) is the optical equivalent of ultrasound imaging, based on the backscattering of near infrared light. OCT provides real time images with a 15 µm axial resolution at an effective tissue penetration of 2-3 mm. Within the OCT images the loss of signal intensity per millimeter of tissue penetration, the attenuation coefficient, is calculated. The attenuation coefficient is a tissue specific property, providing a quantitative parameter for tissue differentiation.

Until now, renal mass treatment decisions have been made primarily on the basis of MRI and CT imaging characteristics, age and comorbidity. However these parameters and diagnostic methods lack the finesse to truly detect the malignant potential of a renal mass. A successful core biopsy or fine needle aspiration provides objective tumor differentiation with both sensitivity and specificity in the range of 95-100%. However, a non-diagnostic rate of 10-20% overall, and even up to 30% in SRMs, is to be expected, delaying the diagnostic process due to the frequent necessity for additional biopsy procedures.

We aim to develop OCT into an optical biopsy, providing real-time imaging combined with on-the-spot tumor differentiation. This publication provides a detailed step-by-step approach for percutaneous, needle based, OCT of renal masses.

Introduction

過去数十年、腎大衆1,2の発生率の着実な増加を示している。今まで、腎腫治療の決定は、主にMRIとCT撮像特性、年齢および併存疾患に基づいて行われてきた。しかしこれらの診断法と臨床パラメーターは本当に腎質量の悪性の可能性を検出するためのフィネスを欠いている。病理学的評価のための十分な組織(診断)とコア生検または細針吸引95〜100%3の範囲の感度と特異性の両方を持つ客観的な腫瘍の分化を提供します。そのため生検は不審な腎大衆4,5の評価で受け入れを集めています。しかし、十分な組織生検なしでは、遅延、診断、または腎実質(非診断)法線が全体の10〜20%の割合で発生し、小さな腎腫瘤では30%までで(<4センチのSRM)を確立する追加のための頻繁な必要性に起因する診断プロセス生検手順3,5。

光コヒーレンストモグラフィ(OCT)は、腎腫分化における上記の障害を克服する可能性を有する新規なイメージングモダリティである。近赤外光の後方散乱に基づいて、10月2~3程度の効果的な組織浸透に15μmの距離分解能を有する画像を提供する( 図1,2)。ファーバーによって記載されるように6:組織浸透のミリメートル当たり信号強度の損失、組織特異的な光散乱の結果は、減衰係数(ミリメートル-1μOCT)として表される組織学的特徴は、組織分化の定量的パラメータ( 図3)を提供するのOCT値をμと相関することができる。

発癌中は、悪性細胞が増加した数、とのより大きく、より不規則な形の核を表示高い屈折率およびより活性なミトコンドリア。による細胞成分のこの過剰発現、μ10月の変化が良性腫瘍または非罹患組織7に悪性腫瘍を比較する際に予想される。

最近では、良性と悪性の腎腫瘤8,9を区別する表面的な10月の能力を研究した。 16人の患者では、腫瘍組織の術中のOCT測定は、外部に配置されたOCTプローブを用いて得た。同じ患者の非罹患組織のOCT測定値からなる制御アーム。正常組織は、腫瘍の分化のために、OCTの可能性を確認し、悪性組織と比較して有意に低い中央値の減衰係数を示した。この定量分析は、尿路上皮癌10,11及び外陰上皮新形成分化12などの悪性組織のグレード他のタイプと同様の方法で適用されている。

と診断された患者でENT ">我々は、現場での腫瘍分化と組み合わせたリアルタイムイメージングを提供し、光学的生検に10月を開発することを目指しています。現在の研究の目的は、経皮、針ベースを記述することで、10月のアプローチ腎臓の質量を強化する固体。この方法の説明には、我々の知る限り、腎腫瘍の10月に基づく針の可能性を評価する最初のものです。

Protocol

提示手順は、学術医療センターアムステルダム、登録番号NL41985.018の治験審査委員会によって承認された研究プロトコルの下で行われます。書面によるインフォームドコンセントは、すべての参加者から要求されている。 1.システムこの実験のために、1,280-1,350波長帯13で動作する、フーリエドメインOCTシステムを使用する。フーリエ領域低コヒーレンス?…

Representative Results

最初の25の腫瘍(23例)のうち、24成功したOCTの手順の合計を行った。あるケースでは、プローブの故障は、OCTスキャンを取得することができないことにつながった。二つの有害事象(AE)は、議論の項で詳細に記載されており、発生した。一般的な患者の特徴を表1に記載されている。 10月コンソールは、取得したデータセットの即時定性分析のために、リア…

Discussion

この公報では、腎臓の10​​月、ベースの経皮的、針の実現可能性について報告する。これは、「光学的生検」と呼ば腫瘍分化のための臨床的に適用可能な技術、に10月の発展に不可欠な最初のステップです。私たちの最初の25人の患者が簡単で安全手順であることが、経皮10月を示している。光学的生検は、従来のコア生検上の2つの利点がある。まず、OCTデータのリアルタイム取得および分?…

Declarações

The authors have nothing to disclose.

Acknowledgements

This work is funded by the Cure for Cancer Foundation, Dutch Technology Foundation (STW) and The Netherlands Organisation for Health Research and Development (ZonMw).

Materials

15G / 7.5cm Co-Axial Introducer Needle Angiotech, Gainesville, USA MCXS1612SX
18G / 20cm Trocar Needle Cook medical, Bloomington, USA DTN-18-20.0-U
16G / 20cm Quick-Core Biopsy Gun Cook Medical, Bloomington, USA G07827
Ilumien Optis PCI Optimization System (OCT & FFR) St. Jude medical, St. Paul, USA C408650 Part of Dragonfly Kit.
St. Jude medical, St. Paul, USA. (C4088643)
Dragonfly Duo Imaging Catheter LightLab Imaging, Westford, USA C408644 Part of Dragonfly Kit.
St. Jude medical, St. Paul, USA. (C4088643)
Sterile Dock Cover CFI Med. Solutions, Fenton, USA 200-700-00 Part of Dragonfly Kit.
St. Jude medical, St. Paul, USA. (C4088643)
5ml Luer-lock Syringe Merit Med. Syst., South Jordan, USA C408647
10ml Syringe BD, Franklin Lakes, USA 300912
18G Blunt Fill Needle BD, Franklin Lakes, USA 305180
21G Injection Needle BD, Franklin Lakes, USA 301155
Sterile scalpel BD, Franklin Lakes, USA 372611
NaCl 0,9% solution Braun, Melsungen AG, Germany 222434
Lidocaïne HCl 2% (20mg/ml) solution Braun, Melsungen AG, Germany 3624480
Sterile Ultrasound Gel, Aquasonic 100 Parker Lab. Inc., Fairfield, USA GE424609
Sterile Ultrasound Cover Microtek Med., Alpharetta, USA PC1289EU
Pathology Container
AMIRA software package FEI Visualization Sciences Group, Hillsboro, USA Software platform for 3D data analysis
FIJI software package (open source) Open source, http://fiji.sc/Fiji Open source image processing software

Referências

  1. Jemal, A., Siegel, R., Xu, J., Ward, E. Cancer statistics, 2010. CA Cancer J. Clin. 60, 277-300 (2010).
  2. Mathew, A., Devesa, S. S., Fraumeni, J. F., Chow, W. H. Global increases in kidney cancer incidence, 1973-1992. Eur. J. Cancer Prev. 11, 171-178 (2002).
  3. Volpe, A., et al. Contemporary management of small renal masses. Eur. Urol. 60, 501-515 (2011).
  4. Ljungberg, B., et al. EAU guidelines on renal cell carcinoma: the 2010 update. Eur. Urol. 58, 398-406 (2010).
  5. Donat, S. M., et al. Follow-up for Clinically Localized Renal Neoplasms. AUA Guideline, J. Urol. 190, 407-416 (2013).
  6. Faber, D. J., van der Meer, F. J., Aalders, M. C. G., van Leeuwen, T. G. Quantitative measurement of attenuation coefficients of weakly scattering media using optical coherence tomography. Optics Express. 12, 4353-4365 (2004).
  7. Xie, T. Q., Zeidel, M. L., Pan, Y. T. Detection of tumorigenesis in urinary bladder with optical coherence tomography: optical characterization of morphological changes. Optics Express. 10, 1431-1443 (2002).
  8. Barwari, K., et al. Differentiation between normal renal tissue and renal tumours using functional optical coherence tomography: a phase I in vivo human study. BJU. Int. 110, E415-E420 (2012).
  9. Barwari, K., et al. Advanced diagnostics in renal mass using optical coherence tomography: a preliminary report. J. Endourol. 25, 311-315 (2011).
  10. Cauberg, E. C., et al. Quantitative measurement of attenuation coefficients of bladder biopsies using optical coherence tomography for grading urothelial carcinoma of the bladder. J. Biomed. Opt. 15, 066013 (2010).
  11. Bus, M. T., et al. Volumetric in vivo visualization of upper urinary tract tumors using optical coherence tomography: a pilot study. J. Urol. 190, 2236-2242 (2013).
  12. Wessels, R., et al. Optical coherence tomography in vulvar intraepithelial neoplasia. Journal of Biomedical Optics. 17, (2012).
  13. Yun, S. H., Tearney, G. J., de Boer, J. F., Iftimia, N., Bouma, B. E. High-speed optical frequency-domain imaging. Optics Express. 11, 2953-2963 (2003).
  14. Kodach, V. M., Kalkman, J., Faber, D. J., van Leeuwen, T. G. Quantitative comparison of the OCT imaging depth at 1300 nm and 1600 nm. Biomed. Opt. Express. 1, 176-185 (2010).
  15. Kinkelder, R., de Bruin, D. M., Verbraak, F. D., van Leeuwen, T. G., Faber, D. J. Comparison of retinal nerve fiber layer thickness measurements by spectral-domain optical coherence tomography systems using a phantom eye model. J. Biophotonics. 6, 314-320 (2013).
  16. Baxter, G. M., Sihdu, P. S. . Ultrasound of the Urogenital System. , (2006).

Play Video

Citar este artigo
Wagstaff, P. G., Swaan, A., Ingels, A., Zondervan, P. J., van Delden, O. M., Faber, D. J., van Leeuwen, T. G., de la Rosette, J. J., de Bruin, D. M., Laguna Pes, M. P. In Vivo, Percutaneous, Needle Based, Optical Coherence Tomography of Renal Masses. J. Vis. Exp. (97), e52574, doi:10.3791/52574 (2015).