Summary

Fuldt automatiseret Centrifugal mikrofluidanordning for Ultrasensitive Protein Detection fra fuldblod

Published: April 16, 2016
doi:

Summary

This protocol demonstrates how to achieve femto molar detection sensitivity of proteins in 10 µL of whole blood within 30 min. This can be achieved by using electrospun nanofibrous mats integrated in a lab-on-a-disc, which offers high surface area as well as effective mixing and washing for enhanced signal-to-noise ratio.

Abstract

Enzyme-linked immunosorbent assay (ELISA) is a promising method to detect small amount of proteins in biological samples. The devices providing a platform for reduced sample volume and assay time as well as full automation are required for potential use in point-of-care-diagnostics. Recently, we have demonstrated ultrasensitive detection of serum proteins, C-reactive protein (CRP) and cardiac troponin I (cTnI), utilizing a lab-on-a-disc composed of TiO2 nanofibrous (NF) mats. It showed a large dynamic range with femto molar (fM) detection sensitivity, from a small volume of whole blood in 30 min. The device consists of several components for blood separation, metering, mixing, and washing that are automated for improved sensitivity from low sample volumes. Here, in the video demonstration, we show the experimental protocols and know-how for the fabrication of NFs as well as the disc, their integration and the operation in the following order: processes for preparing TiO2 NF mat; transfer-printing of TiO2 NF mat onto the disc; surface modification for immune-reactions, disc assembly and operation; on-disc detection and representative results for immunoassay. Use of this device enables multiplexed analysis with minimal consumption of samples and reagents. Given the advantages, the device should find use in a wide variety of applications, and prove beneficial in facilitating the analysis of low abundant proteins.

Introduction

Flere platforme til sygdomsdiagnose er blevet udviklet baseret på nanoskala materialer 1,2 såsom nanotråde, 3 nanopartikler, 4 nanorør, 5 og nanofibre (NFS) 6-8. Disse nanomaterialer tilbyder fremragende udsigter i udformningen af ​​nye teknologier til meget følsomme bioassays grund af deres unikke fysisk-kemiske egenskaber. For eksempel har mesoporøse zinkoxid nanofibre blevet anvendt til femto-molære sensitiv påvisning af brystkræft biomarkører. 9 nylig nanomaterialer på basis af titandioxid (TiO2) er blevet udforsket for bioanalytiske applikationer 10 tilgodese deres kemiske stabilitet, 11 ubetydelig proteindenaturering , 12 og biokompatibilitet. 13 Endvidere hydroxylgrupperne på overfladen af TiO2 lette kemisk modifikation og den kovalente binding af biomolekyler. 14,15 Mønstrede TiO2 thin film 16 eller TiO2 nanorør 17 er blevet anvendt til at forbedre påvisningen følsomhed et målprotein ved at øge overfladearealet; imidlertid fremstillingsprocessen er temmelig kompleks og kræver dyrt udstyr. På den anden side er elektrospundne forbund modtager opmærksomhed på grund af deres høje overfladeareal samt ligetil og billig fremstillingsproces, 18,19 endnu, den skrøbelige eller løse karakteristisk for elektrospundne TiO2 NF måtten gør det vanskeligt at håndtere og integrere med mikrovæskeanordninger. 6,20 Derfor blev TiO 2 NF måtter sjældent anvendes i bioanalytiske anvendelser, især dem, der kræver barske vaskebetingelser.

I denne undersøgelse for at overvinde disse begrænsninger har vi udviklet en ny teknologi til at overføre elektrospindes NF-måtter på overfladen af ​​enhver mål-substratet ved anvendelse af en tynd polydimethylsiloxan (PDMS) klæbelag. Furthermore, har vi med succes viste integrationen af elektrospundne TiO2 NF-måtter på en centrifugal mikrofluidanordning af polycarbonat (PC). Ved anvendelse af denne anordning, var en høj-følsom, fuldautomatisk, og integreret påvisning af C-reaktivt protein (CRP) samt kardial troponin I (cTnI) opnås inden for 30 min fra kun 10 pi helblod. 21 På grund af den kombinerede fordele ved egenskaberne af NFS og den centrifugale platform, assayet udviste et bredt dynamikområde på seks størrelsesordener fra 1 pg / ml (~ 8 FM) til 100 ng / ml (~ 0,8 pm) med en nedre detektionsgrænse på 0,8 pg / ml (~ 6 FM) for CRP og et dynamisk område fra 10 pg / ml (~ 0,4 pm) til 100 ng / ml (~ 4 nm) med en detektionsgrænse på 37 pg / ml (~ 1,5 pm) for cTnI. Disse detektionsgrænser er ~ 300 og ~ 20 gange lavere sammenlignet med deres tilsvarende konventionelle ELISA-resultater. Denne teknik kunne anvendes til påvisning af eventuelle målproteiner med passende antistoffer. Samlet set denne enhed could i høj grad bidrage til in vitro-diagnostik og biokemiske analyser, da det kan detektere sjældne mængder af målproteiner med stor følsomhed selv fra meget små mængder af biologiske prøver, fx 10 pi fuldblod. Selvom vi kun demonstreret serumproteinet påvisning ved ELISA i denne undersøgelse, kan overførslen og integration teknologi af elektrospundet NF'er med mikrofluidenheder være mere udstrakt anvendelse i andre biokemiske reaktioner, som kræver et stort overfladeareal til høj detekteringsfølsomhed.

Protocol

BEMÆRK: Blod blev udtaget fra sunde individer og blev opsamlet i en blodopsamlingsrør. Skriftligt informeret samtykke blev opnået fra alle frivillige. 1. Fremstilling af TiO 2 NF Mat Fremstilling af precursor-opløsning 22 Opløs 1,5 g titantetraisopropoxid (TTIP) i en blanding af ethanol (99,9%, 3 ml) og iseddikesyre (3 ml) og bland opløsningen ved stuetemperatur (25 ° C) i 30 minutter på en magnetomrører. <…

Representative Results

Ved anvendelse af denne protokol, et fuldt automatiseret centrifugal mikrofluidanordning for protein detektion fra fuldblod med høj følsomhed blev fremstillet. De TiO 2 NF måtter fremstillet ved processer elektrospinning og kalcinering. For at fremstille NFS af ønskede diameter, morfologi og tykkelse, elektrospinding tilstande såsom strømningshastighed, spænding, og spinding tid blev optimeret. Når betingelserne ikke var optimeret, kvaliteten af ​​NFS dannede var d…

Discussion

Assayet på TiO2 NF integreret disk er en hurtig, billig og bekvem teknik til ultrasensitiv påvisning af lave rigelige proteiner til stede i meget lave blodvolumen. Denne teknik har den fordel at bruge små prøvevolumener (10 pi) og er modtagelig for analyse af flere prøver samtidigt. Dette tilvejebringer et stort potentiale som en multiplexing immunoassay enhed. Indretningen har den fordel, at det ikke kræver prøveforbehandling trin som plasma separation, som er nødvendige i konventionelle ELISA'er…

Declarações

The authors have nothing to disclose.

Acknowledgements

Dette arbejde blev støttet af National Research Foundation Korea (NRF) tilskud (2013R1A2A2A05004314, 2012R1A1A2043747), et tilskud fra den koreanske Health Technology R & D Project, Ministeriet for Sundhed og Velfærd (A121994) og IBS-R020-D1 finansieret af den koreanske regering.

Materials

Si wafer LG SILTRON Polished Wafer, test grade Dia. (mm) = 150, orientation = <100>, dopant = boron, RES(Ohm-cm) = 1 – 30, thickness (μm) = 650 – 700
Polycarbonate (PC)  Daedong Plastic PCS#6900 Thickness (mm) = 1 and 5 
Titanium tetraisopropoxide, 98%, Sigma-Aldrich 205273
Polyvinylpyrrolidone, Mw = 1,300,000 Sigma-Aldrich 437190
Acetic acid Sigma-Aldrich 320099
Anhydrous ethanol Sigma-Aldrich 459836
Tridecafluoro-1,1,2,2-tetrahydrooctyl)-1-trichlorosilane Sigma-Aldrich 448931
PDMS and curing agent Dow Corning SYLGARD 184
GPDES Gelest Inc SIG5832.0 
Ethanol J T Baker
FE-SEM FEI Nova NanoSEM
X-ray photoelectron spectroscopy ThermoFisher K-alpha
3D modeling machine M&I CNC Lab, Korea CNC milling machine
Wax-dispensing machine Hanra Precision Eng. Co. Ltd., Korea Customized
Double-sided adhesive tape FLEXcon, USA DFM 200 clear 150 POLY H-9 V-95
Cutting plotter Graphtec Corporation, Japan Graphtec CE3000-60 MK2
Spin coater MIDAS SPIN-3000D
Furnace (calcination) R. D. WEBB COMPANY WEBB 99
Rheometer (Tack test) Thermo Scientific Haake MARS III – ORM Package
Oxygen plasma system FEMTO CUTE
Monoclonal mouse antihuman hsCRP Hytest Ltd., Finland 4C28 (clone # C5)
Monoclonal mouse anti-cTnI Hytest Ltd., Finland 4T21 (clone # 19C7)
HRP conjugated goat polyclonal anti-hsCRP Abcam plc., MA ab19175
HRP conjugated mouse monoclonal anti-cTnI Abcam plc., MA ab24460 (clone # 16A11)
hsCRP Abcam plc., MA ab111647
cTnI Fitzgerald, MA 30-AT43
Bovine Albumin Sigma-Aldrich A7906
PBS Amresco Inc E404
Blood collection tubes BD vacutainer 367844 K2 EDTA 7.2 mg plus blood
collection tubes
SuperSignal ELISA femto Invitrogen 37074
Modular multilabel plate reader Perkin Elmer Envision 2104
Disc operating machine Hanra Precision Eng. Co. Ltd., Korea Customized
Photomultiplier tube (PMT) Hamamatsu Photonics H1189-210
AutoCAD AutoDesk Version 2012 Design software
SolidWorks 3D CAD software  SOLIDWORKS Corp. Version 2013 3D Design software,
Edgecam Vero software version 2009.01.06928 Code generating software
DeskCNC Carken Co. version 2.0.2.18 CNC milling machine software

Referências

  1. Zhang, Y., et al. Nanomaterials for Ultrasensitive Protein Detection. Adv. Mater. 25 (28), 3802-3819 (2013).
  2. Hu, W., Li, C. M. Nanomaterial-based advanced immunoassays. Wiley Interdiscip. Rev. Nanomed. Nanobiotechnol. 3 (2), 119-133 (2011).
  3. Yang-Kyu, C., Chang-Hoon, K. Silicon Nanowire Biosensor for Cancer Markers. Biosensors and Cancer. , 164-183 (2012).
  4. Baltazar, R., Vistas, C. R., Ferreira, G. M. Biosensing Applications Using Nanoparticles. Nanocomposite Particles for Bio-Applications. , 265-282 (2011).
  5. Roy, P., Berger, S., Schmuki, P. TiO2 Nanotubes: Synthesis and Applications. Angew. Chem. Int. Ed. 50 (13), 2904-2939 (2011).
  6. Yang, D., et al. Electrospun Nanofibrous Membranes: A Novel Solid Substrate for Microfluidic Immunoassays for HIV. Adv. Mater. 20 (24), 4770-4775 (2008).
  7. Chantasirichot, S., Ishihara, K. Electrospun phospholipid polymer substrate for enhanced performance in immunoassay system. Biosens. Bioelectron. 38 (1), 209-214 (2012).
  8. Zhang, N., et al. Electrospun TiO2 Nanofiber-Based Cell Capture Assay for Detecting Circulating Tumor Cells from Colorectal and Gastric Cancer Patients. Adv. Mater. 24 (20), 2756-2760 (2012).
  9. Ali, M. A., Mondal, K., Singh, C., Dhar Malhotra, B., Sharma, A. Anti-epidermal growth factor receptor conjugated mesoporous zinc oxide nanofibers for breast cancer diagnostics. Nanoscale. 7 (16), 7234-7245 (2015).
  10. Mondal, K., Ali, M. A., Agrawal, V. V., Malhotra, B. D., Sharma, A. Highly Sensitive Biofunctionalized Mesoporous Electrospun TiO2 Nanofiber Based Interface for Biosensing. ACS Appl. Mater. Interfaces. 6 (4), 2516-2527 (2014).
  11. Tu, W., Dong, Y., Lei, J., Ju, H. Low-Potential Photoelectrochemical Biosensing Using Porphyrin-Functionalized TiO2 Nanoparticles. Anal. Chem. 82 (20), 8711-8716 (2010).
  12. Liu, S., Chen, A. Coadsorption of Horseradish Peroxidase with Thionine on TiO2 Nanotubes for Biosensing. Langmuir. 21 (18), 8409-8413 (2005).
  13. Portan, D. V., Kroustalli, A. A., Deligianni, D. D., Papanicolaou, G. C. On the biocompatibility between TiO2 nanotubes layer and human osteoblasts. J.Biomed.Mater.Res. Part A. 100 (10), 2546-2553 (2012).
  14. Dettin, M., et al. Covalent surface modification of titanium oxide with different adhesive peptides: Surface characterization and osteoblast-like cell adhesion. J. Biomed. Mater. Res. Part A. 90 (1), 35-45 (2009).
  15. Kim, W. -. J., et al. Enhanced Protein Immobilization Efficiency on a TiO2 Surface Modified with a Hydroxyl Functional Group. Langmuir. 25 (19), 11692-11697 (2009).
  16. Son, K. J., Ahn, S. H., Kim, J. H., Koh, W. -. G. Graft Copolymer-Templated Mesoporous TiO2 Films Micropatterned with Poly(ethylene glycol) Hydrogel: Novel Platform for Highly Sensitive Protein Microarrays. ACS Appl. Mater. Interfaces. 3 (2), 573-581 (2011).
  17. Kar, P., Pandey, A., Greer, J. J., Shankar, K. Ultrahigh sensitivity assays for human cardiac troponin I using TiO2 nanotube arrays. Lab Chip. 12 (4), 821-828 (2012).
  18. Agarwal, S., Wendorff, J. H., Greiner, A. Use of electrospinning technique for biomedical applications. Polymer. 49 (26), 5603-5621 (2008).
  19. Ding, B., Wang, M., Wang, X., Yu, J., Sun, G. Electrospun nanomaterials for ultrasensitive sensors. Mater. Today. 13 (11), 16-27 (2010).
  20. Liu, Y., Yang, D., Yu, T., Jiang, X. Incorporation of electrospun nanofibrous PVDF membranes into a microfluidic chip assembled by PDMS and scotch tape for immunoassays. ELECTROPHORESIS. 30 (18), 3269-3275 (2009).
  21. Lee, W. S., Sunkara, V., Han, J. -. R., Park, Y. -. S., Cho, Y. -. K. Electrospun TiO2 nanofiber integrated lab-on-a-disc for ultrasensitive protein detection from whole blood. Lab Chip. 15 (2), 478-485 (2015).
  22. Li, D., Xia, Y. Fabrication of Titania Nanofibers by Electrospinning. Nano Lett. 3 (4), 555-560 (2003).
  23. Lombard, M. . SolidWorks 2013 BIBLE. , (2013).
  24. Tickoo, S. . EdgeCAM 11.0 for Manufacturers. , (2007).
  25. Zhu, R., et al. Improved adhesion of interconnected TiO2 nanofiber network on conductive substrate and its application in polymer photovoltaic devices. Appl. Phys. Lett. 93 (1), 013102 (2008).
  26. Song, M. Y., Ahn, Y. R., Jo, S. M., Kim, D. Y., Ahn, J. -. P. TiO2 single-crystalline nanorod electrode for quasi-solid-state dye-sensitized solar cells. Appl. Phys. Lett. 87 (11), 113113 (2005).
  27. Katsuhiro, O., et al. Electrospinning processed nanofibrous TiO2 membranes for photovoltaic applications. Nanotechnology. 17 (4), 1026-1031 (2006).

Play Video

Citar este artigo
Park, Y., Sunkara, V., Kim, Y., Lee, W. S., Han, J., Cho, Y. Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood. J. Vis. Exp. (110), e54143, doi:10.3791/54143 (2016).

View Video