Summary

In the protocol, we present a method to manufacture a small caliber stent-graft by sandwiching a balloon expandable stent between two electrospun nanofibrous polyurethane layers.

Abstract

Stent-grafts are widely used for the treatment of various conditions such as aortic lesions, aneurysms, emboli due to coronary intervention procedures and perforations in vasculature. Such stent-grafts are manufactured by covering a stent with a polymer membrane. An ideal stent-graft should have a biocompatible stent covered by a porous, thromboresistant, and biocompatible polymer membrane which mimics the extracellular matrix thereby promoting injury site healing. The goal of this protocol is to manufacture a small caliber stent-graft by encapsulating a balloon expandable stent within two layers of electrospun polyurethane nanofibers. Electrospinning of polyurethane has been shown to assist in healing by mimicking native extracellular matrix, thereby promoting endothelialization. Electrospinning polyurethane nanofibers on a slowly rotating mandrel enabled us to precisely control the thickness of the nanofibrous membrane, which is essential to achieve a small caliber balloon expandable stent-graft. Mechanical validation by crimping and expansion of the stent-graft has shown that the nanofibrous polyurethane membrane is sufficiently flexible to crimp and expand while staying patent without showing any signs of tearing or delamination. Furthermore, stent-grafts fabricated using the methods described here are capable of being implanted using a coronary intervention procedure using standard size guide catheters.

Introduction

कोरोनरी हस्तक्षेप प्रक्रियाओं पट्टिका और पोत दीवार के विघटन के कारण महत्वपूर्ण पोत दीवार चोट के कारण। इस restenosis, नस ग्राफ्ट में परिधीय आवेश, और कोरोनरी लुमेन 1-4 से अलगाव में परिणाम है। इन जटिलताओं से बचने के लिए एक आशाजनक रणनीति एंजियोप्लास्टी साइट है, जो संभवतः restenosis को रोकना होगा, पोत लुमेन के अलगाव से जोखिमों को कम करने, और परिधीय का आवेश को रोकने में संवहनी सतह को कवर करने के लिए किया जाएगा। पिछले अध्ययनों स्टेंट-ग्राफ्ट को स्टेंट-ग्राफ्ट 5 के लिए सकारात्मक परिणामों के साथ नंगे धातु स्टेंट की तुलना में है। शोधकर्ताओं ने कई सामग्री का इस्तेमाल किया है स्टेंट को कवर करने के लिए झिल्ली का निर्माण करने के लिए। इस पॉलीथीन tetraphthalate (पीईटी), polytetrafluoroethylene (PTFE), polyurethane (पु), और सिलिकॉन या ऑटोलॉगस पोत ऊतक कवर स्टेंट 6-9 निर्माण करने के लिए की तरह सिंथेटिक सामग्री भी शामिल है। एक आदर्श भ्रष्टाचार स्टेंट को कवर करने के लिए इस्तेमाल सामग्री thromboresistant किया जाना चाहिए, गैर biodegradable, और अत्यधिक प्रसार और सूजन 10 बिना मूल ऊतक के साथ एकीकृत करना चाहिए। भ्रष्टाचार स्टेंट को कवर करने के लिए इस्तेमाल सामग्री भी स्टेंट भ्रष्टाचार की चिकित्सा को बढ़ावा देने चाहिए।

स्टेंट-ग्राफ्ट व्यापक रूप से महाधमनी निसंकुचन, मन्या धमनी के छद्म aneurysms, धमनी fistulae के इलाज के लिए उपयोग किया जाता है, ग्राफ्ट नस degenerated, और विशाल मस्तिष्क aneurysms करने के लिए बड़ी। लेकिन छोटे कैलिबर स्टेंट-ग्राफ्ट का विकास कम प्रोफ़ाइल और लचीलापन है, जो स्टेंट-ग्राफ्ट 11-14 की तैनाती में एड्स बनाए रखने की क्षमता सीमित है। पु अच्छा यांत्रिक शक्ति है जो एक कम प्रोफ़ाइल और अच्छा लचीलापन 15,16 प्राप्त करने के लिए एक वांछित विशेषता है के साथ एक elastomeric बहुलक है। अच्छा deliverability होने के अलावा, स्टेंट-ग्राफ्ट भी तेजी से उपचार और endothelialization को बढ़ावा देना चाहिए। पु कवर स्टेंट-ग्राफ्ट बेहतर biocompatibility प्रदर्शन किया और बढ़ाया endothelialization 17 है। शोधकर्ताओं ने किया हैपहले से पु उन्हें endothelial कोशिकाओं 17 के साथ बोने के द्वारा कवर स्टेंट-ग्राफ्ट endothelialize करने की कोशिश की। पु के Electrospinning nanofiber मैट्रिक्स बनाने के लिए नाड़ी के उत्पादन के लिए 18,19 grafts एक महत्वपूर्ण तकनीक होना दिखाया गया है। Nanofibers कि देशी बाह्य मैट्रिक्स की वास्तुकला की नकल के अस्तित्व को भी endothelial सेल प्रसार 20,21 बढ़ावा देने के लिए जाना जाता है। Electrospinning भी सामग्री 22 की मोटाई पर नियंत्रण के लिए अनुमति देता है। छोटे कैलिबर संवहनी पु के बने ग्राफ्ट ऐसी सतह कोटिंग्स, विरोधी coagulants, और सेल प्रसार दबाने के रूप में संशोधनों का उपयोग करके चिकित्सा को बढ़ावा देने के लिए अध्ययन किया गया है। इन सभी संशोधनों मेजबान स्वीकृति मध्यस्थता और भ्रष्टाचार को बढ़ावा देने के उपचार के 23 करने के लिए तैयार कर रहे हैं।

हमारे समूह के एक गुब्बारा विस्तार योग्य नंगे धातु स्टेंट जो पशु मॉडल 24-26 में तैनात किया जा सकता विकसित की है। एक Electrospun polyurethane जाल और एक गेंद के संयोजनoon विस्तार योग्य स्टेंट हमें सक्षम है छोटे कैलिबर गुब्बारा विस्तार योग्य स्टेंट-ग्राफ्ट उत्पन्न करते हैं। वर्तमान में उपलब्ध स्टेंट-ग्राफ्ट से अधिकांश एक हस्तक्षेप की प्रक्रिया के दौरान और्विक धमनी के माध्यम से पेश कर रहे हैं, लेकिन केवल कुछ ही वाणिज्यिक कवर स्टेंट पेश किया जा सकता 1 फ्रेंच आकार एक संयुक्त राष्ट्र फुलाया गुब्बारे 27 के लिए आवश्यक है कि अधिक से बड़ा है। इस अध्ययन में हम Electrospun पु की दो परतों जो एक कोरोनरी धमनी की एक percutaneous हस्तक्षेप प्रक्रिया में एक मानक 8-9 फ्रेंच गाइड कैथेटर का उपयोग करने के लिए दिया जा सकता है के बीच एक गुब्बारा विस्तार योग्य स्टेंट encapsulating द्वारा एक छोटे कैलिबर संवहनी स्टेंट भ्रष्टाचार विकसित किया है।

Protocol

खराद का धुरा कलेक्टर पर polyurethane से 1. Electrospinning Electrospinning के लिए खराद का धुरा तैयार एक स्नातक की उपाधि प्राप्त सिलेंडर में biocompatible, भोजन ग्रेड, पानी में घुलनशील समर्थन सामग्री के लगभग 8 मिलीलीटर पिघला (लग?…

Representative Results

हमारे electrospinner सेटअप (चित्रा 1) उच्च गुणवत्ता polyurethane nanofibers (चित्रा 2) में हुई है। एक स्टेंट भ्रष्टाचार एक खराद का धुरा पर polyurethane से एक भीतरी परत electrospinning, इस परत पर एक नंगे धातु स्टेंट फिसल, और (चित्रा 3…

Discussion

We have developed a fabrication technique for a small caliber stent-graft which can be deployed using a standard percutaneous coronary intervention (PCI) procedure. Stent-grafts currently available are limited in their ability to maintain a low profile and flexibility for deployment. Bare metal stents developed by our group in our previous studies have proven to assist in rapid healing of the stented artery24,26. Various polymers have been electrospun by other groups and polyurethane has been proven biostable …

Declarações

The authors have nothing to disclose.

Acknowledgements

We would like to thank the Division of Engineering, Mayo Clinic for their technical support. This study was financially supported by European Regional Development Fund – FNUSA-ICRC (No. CZ.1.05/1.100/02.0123), National Institutes of Health (T32 HL007111), American Heart Association Scientist Development Grant (AHA #06-35185N), and The Grainger Innovation Fund – Grainger Foundation.

Materials

Glass syringe Air Tite 7.140-33 Syringe for spinneret
Graduated cylinder 5 mL Fisher Scientific 08-552-4G 5 mL pyrex graduated cylinder about 9mm diameter and 11 cm long
High voltage generator Bertan Accociates, Inc. 205A-30P Used to apply voltage difference across spinneret and collector
Laboratory mixer with rpm control Scilogex SCI-84010201 Available from various laboratory equipment suppliers
Polyurethane DSM BioSpan SPU Biospan Segmented Polyurethane
Rubber sheet McMaster Carr 1370N11 Used to insulate syringe during electrospinning
Stainless steel mandrel N/A N/A Manufactured 
Stainless steel needle Hamilton 91018 Used as spinneret in electrospinning
Support material EnvisionTec B04-HT-DEMOMAT Biocompatible water soluble material
Syringe Pump Harvard Apparatus 55-3333

Referências

  1. Elsner, M., et al. Coronary stent grafts covered by a polytetrafluoroethylene membrane. Am. J. Cardiol. 84 (3), 335-338 (1999).
  2. Störger, H., Haase, J. Polytetrafluoroethylene-Covered Stents: Indications, Advantages, and Limitations. J. Interv. Cardiol. 12 (6), 451-456 (1999).
  3. Moreno, P. R., et al. Macrophage infiltration predicts restenosis after coronary intervention in patients with unstable angina. Circulation. 94 (12), 3098-3102 (1996).
  4. Briguori, C., Sarais, C., Colombo, A. The polytetrafluoroethylene-covered stent: a device with multiple potential advantages. Int. J. Cardiovasc. Interv. 4 (3), 145-149 (2001).
  5. Qureshi, M. A., Martin, Z., Greenberg, R. K. Endovascular management of patients with Takayasu arteritis: stents versus stent grafts. Semin. Vasc. Surg. 24 (1), 44-52 (2011).
  6. Ahmadi, R., Schillinger, M., Maca, T., Minar, E. Femoropopliteal arteries: immediate and long-term results with a Dacron-covered stent-graft. Radiology. 223 (2), 345-350 (2002).
  7. Geremia, G., et al. Experimental arteriovenous fistulas: treatment with silicone-covered metallic stents. AJNR. Am. J. Neuroradiol. 18 (2), 271-277 (1997).
  8. Saatci, I., et al. Treatment of internal carotid artery aneurysms with a covered stent: experience in 24 patients with mid-term follow-up results. AJNR. Am. J. Neuroradiol. 25 (10), 1742-1749 (2004).
  9. Stefanadis, C., et al. Stents Wrapped in Autologous Vein: An Experimental Study1. J. Am. Coll. Cardiol. 28 (4), 1039-1046 (1996).
  10. Palmaz, J. C. Review of polymeric graft materials for endovascular applications. J. Vasc. Interv. Radiol. 9, 7-13 (1998).
  11. Bruckheimer, E., Dagan, T., Amir, G., Birk, E. Covered Cheatham-Platinum stents for serial dilation of severe native aortic coarctation. Catheter Cardiovasc. Interv. 74 (1), 117-123 (2009).
  12. Tzifa, A., et al. Covered Cheatham-platinum stents for aortic coarctation: early and intermediate-term results. J. Am. Coll. Cardiol. 47 (7), 1457-1463 (2006).
  13. Kuraishi, K., et al. Development of nanofiber-covered stents using electrospinning: in vitro and acute phase in vivo experiments. J. Biomed. Mater. Res. Part B Appl. Biomater. 88 (1), 230-239 (2009).
  14. Pant, S., Bressloff, N. W., Limbert, G. Geometry parameterization and multidisciplinary constrained optimization of coronary stents. Biomech. Model Mechanobiol. 11 (1-2), 61-82 (2012).
  15. Muller-Hulsbeck, S., et al. Experience on endothelial cell adhesion on vascular stents and stent-grafts: first in vitro results. Invest. Radiol. 37 (6), 314-320 (2002).
  16. Sarkar, S., Salacinski, H. J., Hamilton, G., Seifalian, A. M. The mechanical properties of infrainguinal vascular bypass grafts: their role in influencing patency. Eur. J. Vasc. Endovasc. Surg. 31 (6), 627-636 (2006).
  17. Shirota, T., Yasui, H., Shimokawa, H., Matsuda, T. Fabrication of endothelial progenitor cell (EPC)-seeded intravascular stent devices and in vitro endothelialization on hybrid vascular tissue. Biomaterials. 24 (13), 2295-2302 (2003).
  18. Grasl, C., et al. Electrospun polyurethane vascular grafts: in vitro mechanical behavior and endothelial adhesion molecule expression. J. Biomed. Mater. Res. A. 93 (2), 716-723 (2010).
  19. Kidoaki, S., Kwon, I. K., Matsuda, T. Structural features and mechanical properties of in situ-bonded meshes of segmented polyurethane electrospun from mixed solvents. J. Biomed. Mater. Res. Part B Appl. Biomater. 76 (1), 219-229 (2006).
  20. Stegemann, J. P., Kaszuba, S. N., Rowe, S. L. Review: advances in vascular tissue engineering using protein-based biomaterials. Tissue Eng. 13 (11), 2601-2613 (2007).
  21. Sankaran, K. K., Subramanian, A., Krishnan, U. M., Sethuraman, S. Nanoarchitecture of scaffolds and endothelial cells in engineering small diameter vascular grafts. Biotechnol. J. 10 (1), 96-108 (2015).
  22. Gibson, P., Schreuder-Gibson, H., Rivin, D. Transport properties of porous membranes based on electrospun nanofibers. Colloid Surf., A. 187, 469-481 (2001).
  23. Zdrahala, R. J. Small caliber vascular grafts. Part II: Polyurethanes revisited. J. Biomater. Appl. 11 (1), 37-61 (1996).
  24. Uthamaraj, S., et al. Design and validation of a novel ferromagnetic bare metal stent capable of capturing and retaining endothelial cells. Ann. Biomed. Eng. 42 (12), 2416-2424 (2014).
  25. Tefft, B. J., et al. Cell Labeling and Targeting with Superparamagnetic Iron Oxide Nanoparticles. J. Vis. Exp. (105), e53099 (2015).
  26. Uthamaraj, S., et al. Ferromagnetic Bare Metal Stent for Endothelial Cell Capture and Retention. J. Vis. Exp. (103), e53100 (2015).
  27. de Giovanni, J. V. Covered stents in the treatment of aortic coarctation. J. Interv. Cardiol. 14 (2), 187-190 (2001).
  28. Hans, F. J., et al. Treatment of wide-necked aneurysms with balloon-expandable polyurethane-covered stentgrafts: experience in an animal model. Acta. Neurochir. (Wien). 147 (8), 871-876 (2005).
  29. Hasan, A., et al. Electrospun scaffolds for tissue engineering of vascular grafts. Acta. Biomater. 10 (1), 11-25 (2014).
check_url/pt/54731?article_type=t

Play Video

Citar este artigo
Uthamaraj, S., Tefft, B. J., Jana, S., Hlinomaz, O., Kalra, M., Lerman, A., Dragomir-Daescu, D., Sandhu, G. S. Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents. J. Vis. Exp. (116), e54731, doi:10.3791/54731 (2016).

View Video