资料来源:哈姆娜·库雷希和克雷格·戈尔根,珀杜大学韦尔登生物医学工程学院,西拉斐特,印第安纳州

在这里,我们将强调人类和啮齿动物之间非侵入性血压测量技术的主要相似性和差异,并研究控制血压的工程原理。还将讨论控制当前袖口技术以获得收缩和舒张压力的原理。

与移动设备连接的商用袖口通常结构紧凑且便携,因此几乎可以在任何地方进行测量。非侵入性便携式血压袖口对于高血压和其他心血管问题患者特别有用,需要仔细监测和及早发现血压的任何变化。

同样,非侵入性血压测量系统也可用于啮齿动物。该技术用于实验室设置,并可用于在整个研究中监测动物健康。虽然放射性遥测是啮齿动物血压测量的黄金标准,但这种技术具有侵入性,如果操作不当,可能导致动物死亡。因此,非侵入性方法便于在动物身上进行测量,因为它们可以提供有价值的数据,而无需植入设备。一个商用系统将被用来演示如何在临床环境以外的人身上测量血压。这种技术允许患者定期监测自己的血压,而不必每次需要进行这些测量时访问诊所。

此处描述的方法利用压力传感器和遮挡袖口利用流经啮齿动物尾部的血流。无论是人类移动血压袖口还是啮齿动物非侵入性尾袖口方法,都利用类似的血液动力学原理获取血压测量,为包括临床医生、研究人员和患者。

Princípios

Procedimento

1. 无线血压袖口 打开袖口,并将其与移动设备配对,并将其输出结果。配对可以通过蓝牙完成。 将袖口绑在上臂上,以便血压监测仪对内二头肌进行放置。Ssrap 应紧紧地固定在手臂上。 打开移动设备上的袖口应用程序。确保袖口与移动设备配对。 定位自己,让你坐在你的手臂休息,使袖口是在相同的水平,你的心。确保您处于平静的环境中,以便在收集数据时保?…

Resultados

A typical graph obtained from taking rodent blood pressure using a noninvasive blood pressure system is shown in Figure 1. There are two curves on the graph: one for the VPR sensor and one for the occlusion cuff. The inflection points of the VPR sensor curve are where systolic and diastolic blood pressures are determined, as depicted in Figure 1. A status is indicated at the bottom of the figure that communicates whether this measurement is deemed acceptable by the system. In order to obtain acceptable readings, blood pressures are taken multiple times per trial.

Figure 1
Figure 1. Systolic and diastolic blood pressures obtained by a noninvasive measurement system.

Figure 2
Figure 2. Screenshot of systolic and diastolic blood pressure.

For a commercially available cuff for humans, the results of the blood pressure measurement are outputted on a screen as shown in Figure 2. Users can see their systolic and diastolic blood pressure measurements using this technology. Blood pressure can be taken multiple times so that the results output an average of the data that was collected.

Applications and Summary

Both commercial cuffs available for at-home use and noninvasive methods for animal research studies take advantage of automated technologies. By tracking changes in blood flow with sensors, algorithms detect the systolic and diastolic pressures and then communicate these results to users. While technologically similar, there are some key differences between the two measurement techniques that should be noted. When taking blood pressures in human subjects, the environment is typically relaxed as the subject is unrestrained and fairly comfortable. Conversely, animals must be restrained or anesthetized. Restraining can make the animal anxious and therefore skew blood pressure measurements to be higher, whereas anesthesia can lead to hypothermia and vasoconstriction, making reliable measurements even harder to acquire. This is why it is preferable to obtain rodent blood pressures in conscious animals that are in a calm, dimly lit environment where they can be kept warm with no startling noises. Vasodilation through heating is also used when rodents are having their blood pressure measured to promote blood flow through their tail. Acclimating animals to the holder and cuffs before acquiring data can help reduce discomfort and allow collection of more accurate results. Also, if a user is collecting measurements from the same animal on multiple days, the results will be more meaningful if acquired at roughly the same time each day.

Being able to take blood pressure measurements with a user-friendly, compact cuff is an extremely useful tool for patients as it allows patients to monitor their blood pressures at their own convenience. Commercially available blood pressure cuffs, however, are not meant to replace clinically obtained measurements; rather these cuffs are meant to help patients monitor their systemic blood pressure between visits. Commercial cuffs are also subject to noise and do not always collect accurate results. Similarly, noninvasive measurements of rodent blood pressure are also noisy, especially when the animal moves. As such, it is important to be judicious when looking at data from these systems.

Portable noninvasive blood pressure measurement methods are useful for a variety of applications. Commercial cuffs allow hypertensive patients to monitor their health between visits to the clinic. This method can help patients become more conscious of their dietary and exercise decisions, as they can see the direct impact that their lifestyle has on their blood pressure. Commercial cuffs can also help pregnant women track their blood pressures in the final months of their pregnancies. Pregnant women are at risk for developing preeclampsia. Therefore, it is important for these women to know if they are at risk for this condition during the latter half of gestation. Using a portable cuff helps these women track their blood pressure in the critical months before the baby is born.

For animal use, noninvasive blood pressure measurements are valuable for research purposes, especially those regarding cardiovascular health. Taking blood pressures of animals after procedures and studies, such as those inducing hypertension in rodents, can show the effects of these actions on an animal's health. While other invasive techniques can provide more consistent data, the noninvasive nature of tail-cuff systems make them an attractive method for those using animal models of cardiovascular disease.

Transcrição

Blood pressure measurements are often used to monitor cardiovascular health. Blood pressure is the lateral pressure produced by circulating blood on vessel walls. Close monitoring is important to detect and treat hypotension, low blood pressure, and hypertension, high blood pressure, both of which can impair the function of vital organs, such as the heart, kidneys, and brain.

Blood pressure is measured manually by auscultation, where the clinician occludes the artery using an inflatable cuff and then listens for the sound of blood flow through the brachial artery as it deflates. Blood pressure can also be measured automatically by oscillometry, which allows patients to measure their blood pressure at home. The inflatable cuff contains a pressure transducer that converts the arterial vibrations to systolic and diastolic pressures based on hemodynamic principles.

This video will illustrate the principles of blood pressure, demonstrate the current methods to measure the systolic and diastolic blood pressures of humans and small animals, and discuss some applications.

Blood pressure measurements consist of systolic and diastolic components, both of which are important indicators of cardiovascular health.

The systolic blood pressure is a measure of the maximum intraluminal pressure exerted against the artery during heart contraction when your heart beats, and the diastolic pressure is a measure of the minimum intraluminal pressure, or the pressure in between beats when the heart relaxes and the aortic valve is closed. To measure blood pressure, a cuff is placed on the patient’s upper arm and then inflated to the point where blood flow is completely occluded. As the cuff deflates, the clinician auscultates while observing the manometer readings.

The first audible Korotkoff sound corresponds to the systolic pressure, the pressure at which blood begins to flow again through the open vessel. As the cuff continues to deflate, blood flows more freely as the vessel remains open. Diastolic pressure is recorded at the point when the clinician no longer hears the beating sound or the self-monitoring device detects a leveling-off of oscillations.

A similar method can be used to monitor rodent blood pressure. A cuff is inflated and deflated around the tail artery, while a pressure transducer detects systolic and diastolic pressure. Now that you understand the difference between systolic and diastolic pressure, let’s demonstrate how to measure blood pressure using a wireless cuff for humans and a noninvasive technique for rodents.

Turn on the wireless cuff and the mobile device and open the cuff application. Then, pair the devices via Bluetooth and open the cuff application on the mobile device. Then, strap the cuff onto your upper arm. Make sure that the blood pressure monitor is placed against your inner bicep. The strap should be secured tightly on your arm. Sit in a position so that the cuff is at the same level as your heart. Make sure you are calm and relaxed before starting.

Now, begin the blood pressure measurement by pressing start. Each measurement takes about one to three minutes. Repeat the measurement several times to obtain an average value in the results. The result will be displayed on the screen of the mobile device. After saving the results, unstrap the cuff from your arm and shut off the device.

To measure the blood pressure of a rodent, first turn on the desktop computer, blood pressure system, and heated stage. Then, attach the VPR cuff and occlusion cuff to the blood pressure system. Use software to confirm that the cuffs are properly calibrated and working as expected. Choose the appropriate parameters for data acquisition. Parameters include deflation time, the number of sets per cycle, occlusion pressure, and the delay between sets. Occlusion pressure is typically set to 250 millimeters of mercury, and there are usually 20 sets per cycle. Other parameters can be specified based on user needs.

Calm the animal to prevent it from experiencing unnecessary stress. The environment should be quiet, with minimal noise. Then, place the animal in an appropriately sized holder. Make sure the animal’s paws do not get stuck in the attachments of the holder and that the tail is outside of the restrainer. Now, place the holder on the heated stage. Slide the occlusion cuff, followed by the VPR cuff onto the base of the tail, approximately one to two millimeters apart. Once the cuffs are in place, tape them to the stage. This will restrict tail movement.

Then, heat the tail to 32 to 34 degrees Celsius to promote vasodilation and blood flow through the tail. As the stage is warming, choose the type of animal on the computer whose blood pressure you will be acquiring and then enter its animal ID into the system so you can access its data later. Once the stage is sufficiently heated, start acquiring blood pressure data by pressing start. Try to maintain a calm environment. Keep the position of the cuff constant and make a note if the animal moves substantially. If the cuffs slide down the tail, or if the animal moves inside the restrainer, press the pause button, then make adjustments after the current measurement is completed, and resume testing.

Check to make sure the blood pressure data are being acquired as expected. Collect 20 to 40 measurements. Once all data has been collected, remove the animal from the holder and place it back into a cage. Then turn off the equipment and clean the surfaces that were in contact with the animal.

The results using the blood pressure monitor for humans are outputted on a screen as shown. Using this technology, systolic and diastolic blood pressure measurements are displayed and multiple readings are averaged.

Healthy blood pressures in humans are typically below 120 millimeters of mercury for the systolic pressure and below 80 millimeters of mercury for the diastolic pressure. Blood pressures above these values could indicate that a patient is starting to develop high blood pressure, which can lead to further health complications.

Here is a typical graph that is obtained when taking rodent blood pressure. The two curves on the graph represent data collected from the VPR sensor, the blue line, and the occlusion cuff, the red line. The inflection points of the blue curve are used to calculate the systolic and diastolic pressure. The measurements are listed at the bottom of the graph along with the OK status, which indicates that this measurement is deemed acceptable by the system.

Typical blood pressure measurements for mice are around 120 over 70 millimeters of mercury, whereas rats have slightly higher blood pressures at around 130 over 90 millimeters of mercury.

Now, let’s take a look at how blood pressure is noninvasively measured in the public and scientific community.

Portable noninvasive blood pressure cuffs allow hypertensive patients to monitor their health between visits to the clinic. Pregnant women at risk for developing pre-eclampsia can monitor their risk for this condition in the critical months before the baby is born.

Research on the development of new therapeutic agents requires frequent blood pressure assessment. While invasive techniques such as subcutaneous telemetry and carotid artery cannulation in mice provide long-term blood pressure data, noninvasive tail-cuff systems are an attractive method to monitor the blood pressure in animal models of cardiovascular disease.

You’ve just watched JoVE’s introduction to blood pressure measurement techniques. You should now understand how noninvasive measurement methods are used for both humans and rodents and how the technology is applied to improve health monitoring for the community. Thanks for watching!