Here we describe, step by step, a detailed protocol for performing echocardiography in the rabbit model. We show how to correctly obtain the different echocardiographic views and imaging planes, as well as the different imaging modes available in a clinical echocardiography system routinely used in human and veterinary patients.
Large animal models such as the rabbit are valuable for translational preclinical research. Rabbits have a similar cardiac electrophysiology compared to that of humans and that of other large animal models such as dogs and pigs. However, the rabbit model has the additional advantage of lower maintenance costs compared to other large animal models. The longitudinal evaluation of cardiac function using echocardiography, when appropriately implemented, is a useful methodology for preclinical assessment of novel therapies for heart failure with reduced ejection fraction (e.g. cardiac regeneration). The correct use of this non-invasive tool requires the implementation of a standardized examination protocol following international guidelines. Here we describe, step by step, a detailed protocol supervised by veterinary cardiologists for performing echocardiography in the rabbit model, and demonstrate how to correctly obtain the different echocardiographic views and imaging planes, as well as the different imaging modes available in a clinical echocardiography system routinely used in human and veterinary patients.
Longitudinal evaluation of cardiac function in large animal models is a robust research methodology commonly used for the assessment of the effects of novel therapies for treating ischemic and non-ischemic cardiomyopathy. Amongst the several cardiovascular imaging techniques available for preclinical research, echocardiography has been used extensively because of its non-invasive and portable characteristics. In experienced hands, echocardiography is also a very reproducible imaging technique to study cardiac anatomy as well as systolic and diastolic function of the heart.
Large preclinical animal models such as pigs, dogs and rabbits, are paramount for preclinical translational research1,2,3. Indeed, the potential benefit of novel therapies such as cardiac regenerative medicine in the setting of cardiomyopathy requires extensive hypothesis testing in large preclinical models before they can be considered for human use2,4. Compared to other large preclinical models, the rabbit model offers some advantages, including its low maintenance cost, which is comparable to that of mice and rats. However, in contrast to mice and rats, the Ca+2 transport system and cardiac electrophysiology are similar in rabbits as those of humans, and those of other large animal models such as dogs and pigs, thus increasing the translational potential of the rabbit model1,5. Therefore, the rabbit, as a large experimental preclinical model, has an exceptional balance of cost and reproducibility for preclinical translational research.
The rabbit has the additional benefit of its amenability for echocardiographic imaging using clinical ultrasound units routinely used in human and veterinary patients, thus taking advantage of the superiority of harmonic imaging and state-of-the-art technology. For this, sector transducers (also known as phase array) of relatively high frequency (up to 12 MHz), such as those used in neonatal/pediatric cardiology, are preferred. Echocardiographic examination in the rabbit preclinical model allows the complete evaluation of systolic and diastolic function using multiple views and different modes available in modern echocardiographic units (e.g. continuous wave Doppler (CWD), pulsed-wave Doppler (PWD), and Tissue Doppler imaging (TDI)).
Echocardiography is an operator-dependent technique and therefore requires extensive training and core knowledge of the technique in accord with international guidelines. Part of this training can be facilitated with the visualization of videos explaining in detail how different echocardiographic views can be obtained. The achievement of high competency in echocardiographic imaging, as well as development of a standardized protocol and correct technique, are essential to minimize the influence of the operator and to generate reliable quantitative data, as required in rigorous scientific research.
Some considerations are necessary regarding the system and laboratory setup used for echocardiography in rabbits and other large animal models. For a standard transthoracic echocardiographic evaluation of cardiac function, the ultrasound system must include the following modalities: bi-dimensional mode (B-mode or 2D), motion mode (M-mode), color Doppler, as well as CWD, PWD and TDI. Moreover, the machine should have full cardiac analysis and measurement software installed, as well as sufficient internal hard drive space to store enough high quality digital still images and video loops for offline analysis. Some systems use linear array transducers; however, for the best imaging of the heart, phased array sector transducers with a small scan head diameter are preferred, because these allow an easier passage of the ultrasound waves through the narrow intercostal spaces. For rabbits, we use relatively high frequency transducers (up to 12 MHz). The position of the animal for imaging is of utmost importance to acquire good quality images. Thus, both right and left lateral recumbent positions are recommended to obtain all standard imaging planes during an echocardiographic examination. For this, a table with a notch that coincides with the cardiac area of the chest is advisable (Figure 1A). This notched table facilitates the access with the transducer to the area of the chest that will be scanned, and therefore allows free mobility of the hand of the operator whist maintaining the best scanning position of the animal. Positioning the animal in a lateral recumbent position results in a fall of the heart towards the transducer and elevation of the lungs, as well as widening the access window of the ultrasound beam through the intercostal spaces, thus improving overall imaging quality (Figure 1A). The echocardiographic examination should be performed in a blinded fashion and following the guidelines of the Echocardiography Committee of the American College of Veterinary Internal Medicine and the American Society of Echocardiography/European Association for Cardiovascular Imaging6,7,8.
Part of our scientific team is associated with the Cardiology Service of a Veterinary Teaching Hospital that attends daily to veterinary patients (e.g. dogs and cats), for which it has the relevant training and accreditation in veterinary cardiology and echocardiography, and its different imaging modalities, as well as extensive experience in imaging different sizes of animal patients and thoracic conformations with this technique. In addition, we commonly use echocardiography for longitudinal evaluation of cardiac function in a rabbit model of cardiomyopathy induced by anthracyclines9. Here, we describe a step by step echocardiography protocol for evaluation of cardiac function using a clinical ultrasound unit in a large preclinical model such as the rabbit. This protocol is adapted for current international guidelines8, and includes practical recommendations based on our own experiences in clinical and experimental settings.
The experiments described herein were approved by the Ethical Research Committee of the University of Murcia, Spain, and were performed in accordance with Directive 2010/63/EU of the European Commission. The steps described were performed under standard operating protocols that were part of the plan of work and have not been performed solely for the purpose of filming the accompanying video to this paper.
1. Preparation of the rabbit
Figure 1. Preparation and positioning of the rabbit for echocardiography. (A) Table with notch that coincides with the cardiac area to be imaged. (B) Remove hair from the chest. (C) Attach ECG electrodes to monitor the heart. (D) Positioning of the operator whilst preforming echocardiographic examination. Please click here to view a larger version of this figure.
2. Parasternal long axis (sagittal) view of the heart
Figure 2. How to obtain a PSLAX view of the heart. (A- B) Positioning of the transducer to obtain the two different planes of the PSLAX view of the heart (see description in the text). Please click here to view a larger version of this figure.
3. Parasternal short axis view of the heart
Figure 3. How to obtain a PSSAX view and its different imaging planes. (A) Position of the transducer to obtain a PSSAX view at the level of the papillary muscles. (B) Demonstration of the role of the left hand to help in rotating the transducer when switching from a PSLAX to a PSSAX view. (C) Location of the cursor of M-mode in the papillary muscles plane of the PSSAX view. (D) Position of the transducer to obtain a PSSAX view of the heart at the mitral valve plane. (E) Location of the cursor of the M-mode in the MV plane of the PSSAX view. (F) Position of the transducer to obtain the AV plane in the PSSAX view. (G) Demonstration of color Doppler and positioning of the PWD sample volume to evaluate the outflow of the PV. (H) Location of the cursor of the M-Mode in the AoV plane of the PSSAX view. LV = Left ventricle; RV = right ventricle; FW = LV free wall; AoV = aortic valve; RVOT = right ventricular outflow track; PV = pulmonary valve; PA = pulmonary artery; LA = Left atrium; RA = right atrium. Please click here to view a larger version of this figure.
4. Apical 4 chambers view of the heart
Figure 4. How to obtain the AP4C and AP5C views of the heart. (A) Positioning of the rabbit in left lateral decubitus for an AP4C view of the heart. (B) Position of the transducer to obtain an AP4C view of the heart. (C) Location of the sample volume at the MV leaflet tips to evaluate MV inflow. (D) Location of the sample volume for TDI analysis of myocardial velocities at the lateral side of the MV annulus. (E) Position of the transducer to obtain an AP5C view of the heart. (F) Location of the sample volume for PWD analysis of the outflow across the AoV. LV = Left ventricle; RV = right ventricle; MV = mitral valve; LA = left atrium; RA = right atrium; AoV= Aortic valve. Please click here to view a larger version of this figure.
5. Apical 5 chambers view of the heart
Parasternal long axis view of the heart
Figure 5A shows an imaging plane of the right PSLAX view where the 4 chambers of the heart are clearly distinguished. You can identify in this view the right ventricle (RV), tricuspid valve (TV), IVS, LV, FW, as well as the mitral valve (MV). When the apex is clearly visible on the left side of the image in this view and the LV is not foreshortened, it is possible to estimate accurately the LV volume using the biplane method of disks (modified Simpson's rule) as shown in Figure 5B,C8, which for accuracy should be combined with a similar measurement of the LV volume in the AP4C view, especially if the rabbit model used presents with wall motion abnormalities. Figure 5D shows the other imaging plane of the right PSLAX where the LVOT and the Aorta (Ao) also come into view. The location for placement of the calipers for accurate measurement of the LVOT is also shown in Figure 5D.
Figure 5. Imaging planes obtained in a PSLAX view of the heart. (A) Imaging plane demonstrating the 4 chambers of the heart. (B) End diastolic and (C) end systolic images, demonstrating Simpson's method for analysis of the LV. (D) Imaging plane where the LVOT and aorta come into view in the PSLAX view of the heart. LV = Left ventricle; RV = right ventricle; IVS = interventricular septum; Ao = aorta; LVOT = left ventricular outflow track; LA = Left atrium; RA = right atrium; MV = mitral valve; TV = tricuspid valve; FW = free wall of the LV; PC = pericardium. Please click here to view a larger version of this figure.
Parasternal short axis view of the heart
In Figure 6A, a right PSSAX view of the heart at the level of the papillary muscles and chordae tendineae plane is shown. It is possible identify in this view the RV, IVS, LV, and FW, as well as the anterolateral (AL) and posteromedial (PM) papillary muscles (Figure 6A). In this view, the area trace tool is used to measure the circumferential area in end-diastole (CAd) (Figure 6B), and in end-systole (CAs) (Figure 6C), which allows the calculation of the total circumferential shortening area (CSA) by using the formula:
CSA=CAd-CAs/CAd×100.
An example of an M-mode trace in the PSSAX at the papillary muscles level is shown in Figure 6D, where the placement of calipers, leading edge to leading edge, for the different measurements of the structures of the LV is also demonstrated. These measurements provide useful information regarding size of the LV structures. Thus, measuring the LV end-diastolic diameter (LVDd) and LV end-systolic diameter (LVDs) from three consecutive heart beats allows the calculation of the LV shortening fraction (%SF), using the formula:
SF%= LVDd-LVDs/LVDd
as well as the LV systolic and diastolic volumes (LVVd, LVVs), using the Teichholz formula:
(7×(LVD)3)/(2.4+LVD)
The LV ejection fraction (LVEF (%)) is subsequently calculated according the formula LVEF=(LVVd-LVVs)/(LVVd×100).
An M-mode trace at the level of the MV plane in PSSAX view is shown in Figure 6E, where the location of the calipers for measurement of the E-point to septal separation (EPSS) of the mitral valve is also shown. An example of a PSSAX view of the heart at the AoV plane level is shown in Figure 6F, where the location of the calipers for measurement of the Aortic root diameter (AoD), as well as the left atrial dimension (LAD) are demonstrated.
An example of the PV outflow analysis using both color Doppler and pulsed wave Doppler is shown in Figure 6G. Note the blue colored outflow through the PV with color Doppler, which indicates that the flow observed is moving away from the transducer. Examples of how to quantitate the pre-ejection period of the PV (PEP PV), as well as the PV outflow using the volume time integral (VTI), are shown in Figure 6H.
Figure 6. Imaging planes obtained in the PSSAX view. (A) Representative image of a PSSAX view at the papillary muscles plane. (B) End diastolic and (C) end systolic tracing of the endocardial border to measure the total CSA. (D) M-mode trace obtained in a PSSAX view at the level of the papillary muscles. (E) An example of M-mode trace obtained in a PSSAX view at the level of the MV. (F) Representative 2D image of a PSSAX vie in the plane of the AV. (G) Color Doppler-guided PWD tracing of the PV outflow. (H) Demonstration of a VTI tracing using the PWD signal obtained from the PV outflow. LV = Left ventricle; RV = right ventricle; IVS = interventricular septum; FW = free wall of the LV; AL = anterolateral papillary muscle; PM = posteromedial papillary muscle; LVDd = left ventricular diameter at end-diastole; LVDs = left ventricular diameter at end-systole; PC = pericardium; EPSS = E-point to septal separation; AoD = aortic root diameter; LAD = left atrial dimension; MV = mitral valve; TV = tricuspid valve; PEP PV = pre-ejection period of the pulmonary valve; ET PV = ejection time of the pulmonary valve; VTI PV = volume time integral of the pulmonary valve. Please click here to view a larger version of this figure.
Apical 4 chambers view
An example of MV inflow using color Doppler in an AP4C view is shown in Figure 7A. Note the predominant red color of the MV inflow indicating that the flow is moving towards the transducer. Thus, a useful mnemonic to describe and learn how blood flows across the structures of the heart is the acronym BART (Blue Away, Red Towards the transducer). Using PWD, the MV inflow spectrum can be assessed as shown in Figure 7B, where the early (E) and late (A) filling waves during diastole are easily differentiated. Examples of myocardial tissue velocities of the MV annulus as assessed by TDI at both the lateral and septal walls are shown in Figure 7C and Figure 7D, respectively. The systolic component is denoted by the S wave, whilst the E' and A' waves correspond with myocardial movement of the mitral valve annulus during early filling (E') and late filling (A') components of diastole.
Apical 5 chambers view
Figure 7E shows an example of color Doppler positioned at the LVOT in an apical 5 chambers view. Note that, in line with the BART mnemonic described above, the blue color observed indicates that blood flow is moving away from the transducer. Figure 7F shows an example of how to quantitate the AoV outflow using PWD signal to evaluate the VTI of the AoV, systolic ejection time (ET) and pre-ejection period of the AoV (PEP AoV).
Figure 7. The AP4C and AP5C views. (A) An example of color Doppler in an AP4C view. (B) Representative image of the PWD signal of the MV inflow in an AP4C, where E wave corresponds with early diastolic filling and A corresponds with atrial contraction component during diastole. (C–D) Representative images of myocardial velocity signals obtained from the lateral (C) and septal (D) segments of the MV annulus using TDI in an AP4C view. S corresponds with systole, whilst E' corresponds with early filling phase and A' with late filling phase during diastole. (E) An example of color Doppler signal obtained from the AoV in an AP5C view. (F) Demonstration of a VTI tracing using the PWD signal obtained from the AoV outflow. AoV = Aortic valve; VTI = volume time integral; PEP = pre-ejection period; ET = ejection time. Please click here to view a larger version of this figure.
We have described a protocol for the echocardiographic examination of cardiac function parameters in the rabbit, representing a large preclinical model1,2,3. The step by step methodology described herein should be considered guidance, which with a complementary study of the basic principles of echocardiography, and a basic knowledge of ultrasound imaging, will help the researcher to obtain, through practice and complementary and expert guidance, good quality data in a relative short period of time.
There are several critical steps to increase the value and reproducibility of the results whilst using the echocardiography protocol described here. First, ensure the skin of the thorax is hair free and clean; for this we recommend cleaning the skin with ethanol to remove excess of natural skin grease before applying ultrasound gel. Next, whilst it is possible to image the chest in a supine position, the lungs tend to inflate and reduce an already difficult to image chest wall with poor echogenicity, thus, a left or right recumbent position of the rabbit and the application of the transducer to the chest through the cut-out notch of a purpose built imaging table is the best way to improve overall imaging quality. Then, the researcher operating the ultrasound system should spend some time creating cardiac imaging presets with optimized imaging settings, which are essential to improve overall imaging quality in all views and will also shorten your imaging time at future imaging sessions. Some of the most important control settings to master are total gain and time-gain compensation, given the poor imaging of the chest of the rabbit (see step 2.4.2). It is also important to be systematic and always perform the echocardiographic examination in an orderly fashion. For this, getting into the habit of acquiring all the imaging views and imaging planes in the same sequence will avoid missing important information whilst performing the study. Furthermore, during imaging analysis it is recommended to perform all measurements in at least three consecutive cardiac cycles in the acquired images for each modality. Finally, the blinding of the observer during imaging as well as during the offline analysis is important to avoid bias and increase the value of the results for translational medicine. Taking into account all of the above considerations, together with the application of the principles of imaging and analysis according to current guidelines7,8, will ensure the reproducibility of the research using longitudinal evaluation of cardiac function via echocardiography in a large animal model such as the rabbit.
Given the variability in body size and fat composition at different ages of the rabbits and the particular experimental settings, some variations of the technique will be required, such as subtle movements of the transducer (e.g., sweeping, rotation) relative to the intercostal space, in order to achieve the desired imaging planes. Therefore, the protocol described here must be interpreted as a starting point that should be adapted to the particular objectives of the research program involving this technique.
Whilst clinical echocardiography systems are widely available in most research centers, there are some limitations to the technique described herein. Indeed, the quality of the images obtained from echocardiographic studies depends to a large extent on the sophistication and technology of the ultrasound machine, the skills and expertise of the operator, and the individual patient characteristics. The minimum technical characteristics that the ultrasound equipment must meet were described in the introduction. Thus, inadequate equipment (e.g., a linear array transducer) constitutes a fundamental limitation for the use of the echocardiographic technique in the rabbit model. In addition, the echocardiographic technique and its results are strongly influenced by the operator. Therefore, an operator without enough experience and practical training could dramatically limit the obtaining of standardized images of appropriate quality. Similarly, inexperienced operators could also make mistakes in obtaining measurements even if they are performed on echocardiographic images of excellent technical quality. Furthermore, as mentioned above, some of the limitations are inherent to the rabbit model, such as age and, more specifically, by the size and body fat composition of the rabbits studied via echocardiography. In our experience, young rabbits weighing up to 2.5 kg have low subcutaneous and intra-thoracic fatty deposits. This phenotypic stage provides the best acoustic windows and offers crisper and sharper echocardiographic images and very few artefacts. As the size and body fat composition increase, the quality and accuracy of the echocardiographic study becomes limited, and the skills of the operator will ultimately play a fundamental role in achieving the best possible imaging under these circumstances.
We currently use echocardiography for longitudinal evaluation of cardiac function in a rabbit model of cardiomyopathy induced by anthracyclines and to test stem cell therapies for this condition9,12,13. The technique described here could also be used in other preclinical studies involving ischemia or valvular heart disease.
Another cardiovascular imaging technique is cardiac magnetic resonance (CMR), whose main advantage is better endocardial-myocardial definition, which translates into a more accurate estimation of LV volumes and systolic function14. However, CMR is limited by its high cost and lack of portability and therefore its limited availability in most research centers. Similarly, CMR has relative poor performance for the analysis of diastolic function, thus making echocardiography a better overall choice for longitudinal evaluation of systolic and diastolic function of the heart15.
In our experience, the anesthetic regime used in the protocol described herein is safe and achieves reproducible results without significant depression of myocardial function attributable to the anesthesia9. However, it is important to standardize the anesthetic regime in each laboratory to ensure reproducible results for your particular experimental settings. After inducing anesthesia, in experienced hands the echocardiographic examination can be completed within 15 min.
The authors have nothing to disclose.
This work was supported in part by: Fundación Séneca, Agencia de Ciencia y Tecnología, Región de Murcia, Spain (JT) (Grant number: 11935/PI/09) and the University of Reading, United Kingdom (AG, GB) (Central Funding). The funders had no role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Bluesensor | Medicotest | 13BY1062 | Disposable adhesive ECG lectrodes |
Domtor (Medetomidine) | Esteve | CN 570686.3 | Veterinary prescription is necessary |
HD11 XE Ultrasound System | Philips | 10670267 | Echocardiography system. |
Heating Pad | Solac | CT8632 | |
Imalgene (Ketamine) | Merial | RN 9767 | Veterinary prescription is necessary |
Omnifix-F 1 ml syringe | Braun | 9161406V | |
S12-4 | Philips | B01YgG | 4-12 MHz phase array transducer |
Ultrasound Transmision Gel (Aquasone) | Parker laboratories Inc. | N 01-08 |