Published: November 7, 2014 doi: 10.3791/51878

Summary

This protocol is used to evaluate spatial and temporal gait variables of neurological/orthopedic patients and older persons by means of a recently-introduced floor-based photocell system.

Abstract

Spatial and temporal characteristics of human walking are frequently evaluated to identify possible gait impairments, mainly in orthopedic and neurological patients1-4, but also in healthy older adults5,6. The quantitative gait analysis described in this protocol is performed with a recently-introduced photoelectric system (see Materials table) which has the potential to be used in the clinic because it is portable, easy to set up (no subject preparation is required before a test), and does not require maintenance and sensor calibration. The photoelectric system consists of series of high-density floor-based photoelectric cells with light-emitting and light-receiving diodes that are placed parallel to each other to create a corridor, and are oriented perpendicular to the line of progression7. The system simply detects interruptions in light signal, for instance due to the presence of feet within the recording area. Temporal gait parameters and 1D spatial coordinates of consecutive steps are subsequently calculated to provide common gait parameters such as step length, single limb support and walking velocity8, whose validity against a criterion instrument has recently been demonstrated7,9. The measurement procedures are very straightforward; a single patient can be tested in less than 5 min and a comprehensive report can be generated in less than 1 min.

Introduction

Walking is one of the most important physical activities in everyday life, and is a main determinant of the quality of life for elderly and patient populations who may present with gait deteriorations. Clinical evaluation of gait function is therefore important to reveal potential alterations induced by aging and/or neurological/orthopedic pathologies, but also to prove the functional benefits of a treatment. Different instruments have been developed for the quantitative assessment of gait parameters, e.g., force plates, video-based 3D motion analysis, body-mounted accelerometers10,11, and instrumented walkway mats or treadmills12. However, these systems are mainly used for research studies rather than for clinical purposes because they are complex to operate, have low accessibility, and fragile sensors.

A floor-based photoelectric system has recently been introduced, which is able to provide a valid calculation of temporal features and 1D spatial coordinates of walking steps. This measuring instrument has several advantages compared to pre-existing systems: it is easy to handle, data are collected very quickly, it is simple to create a detailed report and it is a modular system which means that the length of the system can be changed. Thus, it can be used with confidence to measure within-group changes in longitudinal assessments and between-group differences in cross-sectional comparisons. The goals of the described protocol are to focus on the equipment and its installation, and to objectively and straightforwardly describe the assessment procedures for evaluating spatiotemporal gait parameters in elderly and patient populations.

Subscription Required. Please recommend JoVE to your librarian.

Protocol

The protocol follows the guidelines of the local human Ethics Committee in Zurich (KEK Zurich).

1. Hardware Installation (Figure 1)

  1. Use two 10-m sets of floor-based bars and place them parallel to each other (and to the line of progression) to create a corridor with an inter-set distance of approximately 1 m.
    NOTE: This distance can be increased up to 8 m. Each bar has a length of 1 m and consists of 96 light diodes.
  2. Make a distinction between light-transmitting (T) and light-receiving (R) units for the installation of the bars by placing the light-transmitting (T) units on the right side and the light-receiving (R) units on the left side with respect to the walking direction.
    NOTE: The first meter bars (both T and R) have silver drums. The T and R bars disposed on the remaining 9 m are all equal and interchangeable.
  3. Connect all bars of a row with caps (wireless). Use 2 power supplies: one for each set of bars (T and R).
  4. Connect the first R bar to the laptop with a USB cable.
  5. Position the camera next to the first bar for offline verifications (e.g., starting foot), and connect it to the laptop with a USB cable.
  6. Put a mark 2 m before and after the beginning and end of the track.
  7. Switch on the photoelectric device using the on-off switch of the first R and T bar.
  8. Check that the control LEDs located on all R bars are green.
    NOTE: If so, the system is correctly positioned and testing can start; however, if one or more of the control LEDs are red the system is not correctly positioned and/or connected. Control all caps by checking that they are completely clicked into place and then turn the system off and on again.

Figure 1
Figure 1. The photoelectric system consists of light-transmitting (T) and light-receiving (R) units that are placed parallel to each other with a distance of approximately 1 m. The camera is installed close to the start area for control purposes. The laptop is connected with USB cables to the first R bar and to the camera. Please click here to view a larger version of this figure.

2. Software Installation and Preparation of a Test

  1. Load the software operating the photoelectric system from www.optogait.com/Support/Downloads. Note: This protocol is described by using the 1.8.1 version.
  2. If the software is used for the first time for gait analysis create a new test as follows (otherwise proceed to step 2.3):
    1. Select Test then click on Define / Modify tests. Now click on Gait test and then select Duplicate test. Click on Confirm in the pop-up window so that the test is duplicated.
    2. Double click on the duplicated test to modify the name (e.g., Gait Test 10 bars) and select 10 for the number of bars. Use the standard parameters for the gait test, which are presented in Figure 2. Finally, select Save to save all modifications.
  3. Add a new patient to the database. Select Patients, click on Insert / Modify patient, and then click on New patient to enter the data. Then Save the data.

Figure 2
Figure 2. Standard settings for a gait test with 10 bars, as described in the present protocol. These settings have to be defined when the photoelectric system is used for the first time. In this protocol the starting foot is not defined. The system starts measuring when the patient enters the recording area and stops measuring when the patient leaves the measuring units. Please click here to view a larger version of this figure.

3. Testing Procedures

  1. Always give the same instructions to the patient13.
    1. Instruct the patient to walk with flat-soled shoes along the 10-m walkway at two different velocities: normal (“walk at a pace that is comfortable for you”), and faster than normal (“walk at a pace that is faster than you would normally walk”).
    2. Ask the patient to look straight ahead during the walking trials.
    3. Ask the patient to initiate the first step with the same foot to better standardize the test conditions.
    4. Ask the patient to start walking 2 m before the first photoelectric bar and to conclude each trial 2 m after the last bar in order to maintain constant gait velocity13.
  2. Demonstrate one trial at normal velocity to the patient.
  3. Request the patient to perform three familiarization trials followed by one experimental trial at each velocity. Always complete normal velocity trials first.
  4. To get ready with the software, click on Test and then Execute to start the measurements with the created test.
    1. Select the patient by clicking on Select, choosing the patient and then clicking on Confirm.
    2. Select the test by clicking on Select, and choosing the test e.g., Gait Test 10 bars. Check that only this test is selected for the measurement.
    3. Place the camera so that it can record the entire walk. Change the position of the camera while checking the live picture on the screen of your laptop.
    4. Finally, click on Execute again.
      NOTE: Now the software is ready to measure. As soon as the patient enters the bars, the system starts to measure and a pop-up window appears asking for the starting foot.
  5. Click on the appropriate foot so that the gait parameters will be correctly calculated.
    NOTE: The camera records automatically once the test is started.
  6. Save the test.

4. Data Analysis

  1. Click on Results to display the completed trials. Then click on the arrow next to the test of interest to transfer the test from the Test list to the Test analysis section. Now click on View to display the selected test. Refer to the Results section for all tests performed in this study.
    NOTE: A window with all test data appears (Figure 3). On the left side of the window there are some command buttons for activating various functions. The other part of the window presents 4 types of information regarding the current test. Each set of information can be shown/hidden using the configuration commands. From top to bottom the items are the following: video, charts displaying the results, table with numerical data, and photoelectric bars.
  2. Click on Gait data to display a gait report (Figure 3).
  3. Click on Print (a window with the report appears) to print out the report.
    NOTE: The report can be printed as it is or can be modified. For the different spatiotemporal parameters, the following results are presented in the report: mean values ± standard deviation (SD) of the left and right side, coefficients of variation (CV) expressing gait variability14,15, and percent difference between the left and the right side (asymmetry).
    1. If required, modify the report as follows: open the report as described above (step 4.3). Click on the buttons Show or Hide on the left side of the screen to adapt the presented data and charts of the report.
    2. To change the logo and/or the footer on the report, click on the respective button (Change logo or Change footer) to modify these parameters.
  4. To compare two or more tests performed on different occasions, e.g., before and after an intervention, select the different tests in the results section by clicking on the arrow next to the tests and then click on the Compare button.
    NOTE: A report with the direct comparison of all parameters of the different tests is displayed.

Figure 3
Figure 3. Screenshot of all test data. The command buttons appear on the left side of the window (e.g., by clicking on Print a report is generated, which can be eventually modified). The other part of the window presents the following information regarding the current test, from top to bottom: video, charts displaying the results, table with numerical data, and photoelectric bars. These details can be shown/hidden using the Configure button on the right side. The actual view of the data can be changed by clicking on Gait data or Gait report, respectively. Please click here to view a larger version of this figure.

Subscription Required. Please recommend JoVE to your librarian.

Representative Results

A recent study demonstrated the validity of the photoelectric system against a criterion instrument (a validated electronic walkway) for the assessment of spatiotemporal gait parameters in orthopedic patients and healthy elderly controls7. The same between-group differences in gait variables were detected by the two systems. Although concurrent validity was excellent, with intraclass correlation coefficients ranging between 0.933 and 0.999 (p < 0.001), a systematic bias (p < 0.001) was observed between the two measuring instruments. Stance time and cycle time were significantly longer while swing time and step length were shorter for the photoelectric system than for the electronic walkway. In the same way, walking speed and cadence were slightly (1-2%) but significantly lower for the photoelectric system.

Data from a representative report are presented in Figure 4. The report shows the results of a walking trial conducted at normal velocity with 12 steps (6 left and 6 right). Spatiotemporal gait parameters of this trial are presented as mean ± SD and CV for the left and right side. Furthermore the percent difference between the left and right side (Diff.) is presented. Data of the left and right side are presented in purple and turquoise, respectively. The most common gait parameters such as step length, stance phase, swing phase, single support, step time, cadence, and speed are instantaneously (on-line) calculated and presented on the screen during the actual trials (Figure 3). The same values are presented in the off-line gait report (Figure 4). Percent difference between the two sides expresses the so-called side-to-side (or bilateral) asymmetry that is a good indicator of gait recovery, e.g., before and after an intervention. Recovery of symmetrical gait function is one of the primary goals in patients rehabilitation so to regain independence in daily activities. CV is used as an indicator of gait variability, which is generally increased in patients with clinically-relevant syndromes such as falling and neuro-degenerative diseases16,17 and therefore it is a relevant outcome measure for neurological patients and for subjects with mild cognitive impairments and dementia.

Figure 4
Figure 4. Representative gait report of a walking trail conducted at the normal velocity by an orthopedic patient. The figure on the top presents a simplified gait cycle with the different temporal gait parameters (per foot). The table presents averaged test data of consecutive steps for the selected test. For the different spatiotemporal gait parameters the following results are presented: mean ± standard deviation (SD) for left and right side, coefficient of variation (CV) for left and right side, and percent difference (Diff.) between the left and right side. Please click here to view a larger version of this figure.

Subscription Required. Please recommend JoVE to your librarian.

Discussion

The protocol presented here can be used to evaluate spatial and temporal gait parameters of patients (orthopedic, neurological, cardiorespiratory, etc.) and healthy older adults with a recently-introduced photoelectric system. The total length and width of the system can be modulated depending on the available space and budget. The estimated cost (in Europe) is approximately 2,800 USD per meter for a 10-meter system and the minimum recommended length is 3 meter for floor-based gait analysis. A new feature of the photoelectric system has also been recently introduced, which consists in closing the corridor with two additional bars that are positioned perpendicularly to T and R bars, thereby creating a sort of grid that allows calculation of 2D footfall patterns. Additionally, only the two first-meter bars can be used for treadmill-based gait analysis, even though this would require a validation.

Before starting the measurements it is important to verify that all the bars are properly connected; this is facilitated by the red/green control LEDs disposed on each photoelectric bar. Another critical step is the definition of the starting foot, which has to be selected at the beginning of each test. In case the wrong side is selected, offline modifications can be made at any time (open the appropriate test and select Gait report, then change the foot), also after having verified the starting foot (and any other eventual doubt) on the video.

The main limitation of this protocol is the use of self-selected gait velocities (normal and faster than normal), because all the spatiotemporal gait parameters are considerable affected by the walking velocity18. An alternative option would be to impose a fixed gait speed to all subjects by means of a metronome (e.g., at 4 km/h). The validity of this approach is nevertheless uncertain as not all the patients can walk at a given velocity and/or maintain a fixed gait speed. An important limitation of the photoelectric system is the height of diodes with respect to the floor (3 mm). This instrument slightly overestimates stance time and underestimates swing time compared to floor-integrated instruments (e.g., electronic walkways or force plates) because the diodes detect rearfoot loading and forefoot unloading 3 mm above floor level (see Figure 3A in reference 7)7. Due to this limitation the system can only provide valid data for subjects who are able to raise sufficiently their feet during walking and who have a step length longer than their foot length7. This could represent a problem for the evaluation of gait variables in some seriously-impaired neurological patients.

Since this photoelectric system is very simple to operate and valid data can be quickly collected and easily organized into a comprehensive report, this is a potentially useful system for clinical assessment of spatiotemporal gait variables in patients and older adults. Clinicians could, in fact, implement these assessments in routine physical examinations with the objectives to detect gait disorders and/or to monitor patient progress following an intervention.

Subscription Required. Please recommend JoVE to your librarian.

Disclosures

The authors have nothing to disclose.

Materials

Name Company Catalog Number Comments
Optogait system (10 meters) Microgate, Bolzano, Italy www.optogait.com
Optogait software www.optogait.com/Support/Downloads
Laptop
The Optogait system contains the following equipment:
10 Light-transmitting (T) bars (1 as a first meter)
10 Light-receiving (R) bars (1 as a first meter)
18 Caps to connect the bars within a set (9 for T and 9 for R bars) and 2 special caps for the last T and R bar
1 Camera with its tripod
1 Cable for connecting the Optogait to the laptop
1 Cable for connecting the camera to the laptop
2 Power supplies (one for each set of bars)

DOWNLOAD MATERIALS LIST

References

  1. Chow, J. W., Yablon, S. A., Horn, T. S., Stokic, D. S. Temporospatial characteristics of gait in patients with lower limb muscle hypertonia after traumatic brain injury. Brain. Inj. 24, 1575-1584 (2010).
  2. Esser, P., Dawes, H., Collett, J., Feltham, M. G., Howells, K. Assessment of spatio-temporal gait parameters using inertial measurement units in neurological populations. Gait Posture. 34, 558-560 (2011).
  3. Maffiuletti, N. A., et al. Spatiotemporal parameters of gait after total hip replacement: anterior versus posterior approach. Orthop. Clin. North Am. 40, 407-415 (2009).
  4. Webster, K. E., Wittwer, J. E., Feller, J. A. Quantitative gait analysis after medial unicompartmental knee arthroplasty for osteoarthritis. J. Arthroplasty. 18, 751-759 (2003).
  5. Chui, K. K., Lusardi, M. M. Spatial and temporal parameters of self-selected and fast walking speeds in healthy community-living adults aged 72-98 years. J. Geriatr. Phys. Ther. 33, 173-183 (2010).
  6. Hollman, J. H., McDade, E. M., Petersen, R. C. Normative spatiotemporal gait parameters in older adults. Gait Posture. 34, 111-118 (2011).
  7. Lienhard, K., Schneider, D., Maffiuletti, N. A. Validity of the Optogait photoelectric system for the assessment of spatiotemporal gait parameters. Med. Eng. Phys. 35, 500-504 (2013).
  8. Perry, J. Gait analysis, normal and pathological function. First edn, Slack Inc. , (1992).
  9. Lee, M. M., Song, C. H., Lee, K. J., Jung, S. W., Shin, D. C., Shin, S. H. Concurrent validity and test-retest reliability of the OPTOGait photoelectric cell system for the assessment of spatio-temporal parameters of the gait of young adults. J. Phys. Ther. Sci. 26, 81-85 (2014).
  10. Item-Glatthorn, J. F., Casartelli, N. C., Petrich-Munzinger, J., Munzinger, U. K., Maffiuletti, N. A. Validity of the intelligent device for energy expenditure and activity accelerometry system for quantitative gait analysis in patients with hip osteoarthritis. Arch. Phys. Med. Rehabil. 93, 2090-2093 (2012).
  11. Maffiuletti, N. A., et al. Concurrent validity and intrasession reliability of the IDEEA accelerometry system for the quantification of spatiotemporal gait parameters. Gait Posture. 27, 160-163 (2008).
  12. Reed, L. F., Urry, S. R., Wearing, S. C. Reliability of spatiotemporal and kinetic gait parameters determined by a new instrumented treadmill system. BMC Musculoskelet. Disord. 14, 249 (2013).
  13. Kressig, R. W., Beauchet, O. Guidelines for clinical applications of spatio-temporal gait analysis in older adults. Aging Clin. Exp. Res. 18, 174-176 (2006).
  14. Dubost, V., et al. Relationships between dual-task related changes in stride velocity and stride time variability in healthy older adults. Hum. Mov. Sci. 25, 372-382 (2006).
  15. Hausdorff, J. M. Gait variability: methods, modeling and meaning. J. Neuroeng. Rehabil. 2, (2005).
  16. Blin, O., Ferrandez, A. M., Serratrice, G. Quantitative analysis of gait in Parkinson patients: increased variability of stride length. J. Neurol. Sci. 98, 91-97 (1990).
  17. Webster, K. E., Merory, J. R., Wittwer, J. E. Gait variability in community dwelling adults with Alzheimer disease. Alzheimer. Dis. Assoc. Disord. 20, 37-40 (2006).
  18. Bejek, Z., Paroczai, R., Illyes, A., Kiss, R. M. The influence of walking speed on gait parameters in healthy people and in patients with osteoarthritis. Knee Surg. Sports Traumatol. Arthrosc. 14, 612-622 (2006).

Tags

Keywords: Gait Analysis Spatiotemporal Parameters Photoelectric System Gait Impairments Orthopedic Patients Neurological Patients Older Adults Clinical Assessment Walking Velocity Step Length Single Limb Support
Clinical Assessment of Spatiotemporal Gait Parameters in Patients and Older Adults
Play Video
PDF DOI DOWNLOAD MATERIALS LIST

Cite this Article

Item-Glatthorn, J. F., Maffiuletti,More

Item-Glatthorn, J. F., Maffiuletti, N. A. Clinical Assessment of Spatiotemporal Gait Parameters in Patients and Older Adults. J. Vis. Exp. (93), e51878, doi:10.3791/51878 (2014).

Less
Copy Citation Download Citation Reprints and Permissions
View Video

PLAYLIST

  • Research • Medicine
    Estimation of Urinary Nanocrystals in Humans using Calcium Fluorophore Labeling and Nanoparticle Tracking Analysis
  • Research • Medicine
    Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
  • Research • Medicine
    Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation
  • Research • Medicine
    Digital Handwriting Analysis of Characters in Chinese Patients with Mild Cognitive Impairment
  • Research • Medicine
    Segmentation and Linear Measurement for Body Composition Analysis using Slice-O-Matic and Horos
  • Research • Medicine
    Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
  • Research • Medicine
    Real-Time Magnetic Resonance Guided Focused Ultrasound for Painful Bone Metastases
  • Research • Medicine
    Isolation of Viable Adipocytes and Stromal Vascular Fraction from Human Visceral Adipose Tissue Suitable for RNA Analysis and Macrophage Phenotyping
  • Research • Medicine
    Obtaining Quality Extended Field-of-View Ultrasound Images of Skeletal Muscle to Measure Muscle Fascicle Length
  • Research • Medicine
    Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
  • Research • Medicine
    Electroretinogram Recording for Infants and Children under Anesthesia to Achieve Optimal Dark Adaptation and International Standards
  • Research • Medicine
    Measurement of Tissue Oxygenation Using Near-Infrared Spectroscopy in Patients Undergoing Hemodialysis
  • Research • Medicine
    Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
  • Research • Medicine
    Simultaneous Laryngopharyngeal and Conventional Esophageal pH Monitoring
  • Research • Medicine
    Real-Time Monitoring of Neurocritical Patients with Diffuse Optical Spectroscopies
  • Research • Neuroscience
    Evaluating Postural Control and Lower-extremity Muscle Activation in Individuals with Chronic Ankle Instability
  • Research • Medicine
    Assessment of Dependence in Activities of Daily Living Among Older Patients in an Acute Care Unit
  • Research • Medicine
    Validated LC-MS/MS Panel for Quantifying 11 Drug-Resistant TB Medications in Small Hair Samples
  • Research • Medicine
    International Expert Consensus and Recommendations for Neonatal Pneumothorax Ultrasound Diagnosis and Ultrasound-guided Thoracentesis Procedure
  • Research • Biology
    A Finite Element Approach for Locating the Center of Resistance of Maxillary Teeth
  • Research • Medicine
    Lower Limb Biomechanical Analysis of Healthy Participants
  • Research • Neuroscience
    Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
  • Research • Medicine
    Oral Health Assessment by Lay Personnel for Older Adults
  • Research • Medicine
    Determining and Controlling External Power Output During Regular Handrim Wheelchair Propulsion
  • Research • Medicine
    A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
  • Research • Medicine
    Measurement of Carotenoids in Perifovea using the Macular Pigment Reflectometer
  • Research • Medicine
    Assessment of Static Graviceptive Perception in the Roll-Plane using the Subjective Visual Vertical Paradigm
  • Research • Medicine
    Learning Modern Laryngeal Surgery in a Dissection Laboratory
  • Research • Medicine
    DIPLOMA Approach for Standardized Pathology Assessment of Distal Pancreatectomy Specimens
  • Research • Medicine
    A Computerized Functional Skills Assessment and Training Program Targeting Technology Based Everyday Functional Skills
  • Research • Medicine
    Imaging Features of Systemic Sclerosis-Associated Interstitial Lung Disease
  • Research • Medicine
    Integrating Augmented Reality Tools in Breast Cancer Related Lymphedema Prognostication and Diagnosis
  • Research • Medicine
    Ultrasonographic Assessment During Cardiopulmonary Resuscitation
  • Research • Medicine
    Measurement of the Hepatic Venous Pressure Gradient and Transjugular Liver Biopsy
  • Research • Medicine
    Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
  • Research • Medicine
    Traditional Trail Making Test Modified into Brand-new Assessment Tools: Digital and Walking Trail Making Test
  • Research • Medicine
    Use of Magnetic Resonance Imaging and Biopsy Data to Guide Sampling Procedures for Prostate Cancer Biobanking
  • Research • Medicine
    A Fluorescence-based Assay for Characterization and Quantification of Lipid Droplet Formation in Human Intestinal Organoids
  • Research • Medicine
    A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
  • Research • Medicine
    Quantitative Mapping of Specific Ventilation in the Human Lung using Proton Magnetic Resonance Imaging and Oxygen as a Contrast Agent
  • Research • Neuroscience
    Portable Thermographic Screening for Detection of Acute Wallenberg's Syndrome
  • Research • Medicine
    Use of MRI-ultrasound Fusion to Achieve Targeted Prostate Biopsy
  • Research • Medicine
    Testing of all Six Semicircular Canals with Video Head Impulse Test Systems
  • Research • Medicine
    Protocol and Guidelines for Point-of-Care Lung Ultrasound in Diagnosing Neonatal Pulmonary Diseases Based on International Expert Consensus
  • Research • Neuroscience
    Bilateral Assessment of the Corticospinal Pathways of the Ankle Muscles Using Navigated Transcranial Magnetic Stimulation
  • Research • Medicine
    Targeting Gray Rami Communicantes in Selective Chemical Lumbar Sympathectomy
  • Research • Medicine
    Multi-Modal Home Sleep Monitoring in Older Adults
  • Research • Medicine
    Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
  • Research • Medicine
    Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
  • Research • Medicine
    Measurements of Motor Function and Other Clinical Outcome Parameters in Ambulant Children with Duchenne Muscular Dystrophy
  • Research • Medicine
    Assessment of the Efficacy of An Osteopathic Treatment in Infants with Biomechanical Impairments to Suckling
  • Research • Medicine
    Quantification of Levator Ani Hiatus Enlargement by Magnetic Resonance Imaging in Males and Females with Pelvic Organ Prolapse
  • Research • Medicine
    Quantitative [18F]-Naf-PET-MRI Analysis for the Evaluation of Dynamic Bone Turnover in a Patient with Facetogenic Low Back Pain
  • Research • Medicine
    Generation of Human 3D Lung Tissue Cultures (3D-LTCs) for Disease Modeling
  • Research • Medicine
    Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
  • Research • Medicine
    Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
  • Research • Medicine
    Two-Dimensional X-Ray Angiography to Examine Fine Vascular Structure Using a Silicone Rubber Injection Compound
  • Research • Medicine
    Preparation, Procedures and Evaluation of Platelet-Rich Plasma Injection in the Treatment of Knee Osteoarthritis
  • Research • Medicine
    Cardiac Magnetic Resonance Imaging at 7 Tesla
  • Research • Medicine
    Semi-quantitative Assessment Using [18F]FDG Tracer in Patients with Severe Brain Injury
  • Research • Medicine
    Handheld Metal Detector Screening for Metallic Foreign Body Ingestion in Children
  • Research • Medicine
    Conducting Maximal and Submaximal Endurance Exercise Testing to Measure Physiological and Biological Responses to Acute Exercise in Humans
  • Research • Medicine
    A Metadata Extraction Approach for Clinical Case Reports to Enable Advanced Understanding of Biomedical Concepts
  • Research • Medicine
    Autonomic Function Following Concussion in Youth Athletes: An Exploration of Heart Rate Variability Using 24-hour Recording Methodology
  • Research • Medicine
    Hydra, a Computer-Based Platform for Aiding Clinicians in Cardiovascular Analysis and Diagnosis
  • Research • Medicine
    Objective Nociceptive Assessment in Ventilated ICU Patients: A Feasibility Study Using Pupillometry and the Nociceptive Flexion Reflex
  • Research • Medicine
    'Boden Food Plate': Novel Interactive Web-based Method for the Assessment of Dietary Intake
  • Research • Medicine
    Anogenital Distance and Perineal Measurements of the Pelvic Organ Prolapse (POP) Quantification System
  • Research • Medicine
    Bedside Ultrasound for Guiding Fluid Removal in Patients with Pulmonary Edema: The Reverse-FALLS Protocol
  • Research • Medicine
    Muscle Imbalances: Testing and Training Functional Eccentric Hamstring Strength in Athletic Populations
  • Research • Medicine
    Isolation of Primary Human Decidual Cells from the Fetal Membranes of Term Placentae
  • Research • Medicine
    Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
  • Research • Medicine
    Collecting Hair Samples for Hair Cortisol Analysis in African Americans
  • Research • Medicine
    In Vivo Morphometric Analysis of Human Cranial Nerves Using Magnetic Resonance Imaging in Menière's Disease Ears and Normal Hearing Ears
  • Research • Medicine
    Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
  • Research • Medicine
    Standardized Measurement of Nasal Membrane Transepithelial Potential Difference (NPD)
  • Research • Medicine
    Taste Exam: A Brief and Validated Test
  • Research • Medicine
    Absorption of Nasal and Bronchial Fluids: Precision Sampling of the Human Respiratory Mucosa and Laboratory Processing of Samples
  • Research • Medicine
    Methodology for Sputum Induction and Laboratory Processing
  • Research • Medicine
    Electrophysiological Measurement of Noxious-evoked Brain Activity in Neonates Using a Flat-tip Probe Coupled to Electroencephalography
  • Research • Medicine
    A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
  • Research • Medicine
    Oral Biofilm Sampling for Microbiome Analysis in Healthy Children
  • Research • Medicine
    Using Retinal Imaging to Study Dementia
  • Research • Medicine
    Application of an Amplitude-integrated EEG Monitor (Cerebral Function Monitor) to Neonates
  • Research • Medicine
    3D Ultrasound Imaging: Fast and Cost-effective Morphometry of Musculoskeletal Tissue
  • Research • Medicine
    The 4-vessel Sampling Approach to Integrative Studies of Human Placental Physiology In Vivo
  • Research • Medicine
    A Component-resolved Diagnostic Approach for a Study on Grass Pollen Allergens in Chinese Southerners with Allergic Rhinitis and/or Asthma
  • Research • Medicine
    A Novel Method: Super-selective Adrenal Venous Sampling
  • Research • Medicine
    A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
  • Research • Medicine
    Non-invasive Assessments of Subjective and Objective Recovery Characteristics Following an Exhaustive Jump Protocol
  • Research • Medicine
    Experimental Protocol of a Three-minute, All-out Arm Crank Exercise Test in Spinal-cord Injured and Able-bodied Individuals
  • Research • Medicine
    Phosphorus-31 Magnetic Resonance Spectroscopy: A Tool for Measuring In Vivo Mitochondrial Oxidative Phosphorylation Capacity in Human Skeletal Muscle
  • Research • Medicine
    Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
  • Research • Medicine
    Assessment of Child Anthropometry in a Large Epidemiologic Study
  • Research • Medicine
    Video Movement Analysis Using Smartphones (ViMAS): A Pilot Study
  • Research • Medicine
    Network Analysis of Foramen Ovale Electrode Recordings in Drug-resistant Temporal Lobe Epilepsy Patients
  • Research • Medicine
    A Model to Simulate Clinically Relevant Hypoxia in Humans
  • Research • Medicine
    Interictal High Frequency Oscillations Detected with Simultaneous Magnetoencephalography and Electroencephalography as Biomarker of Pediatric Epilepsy
  • Research • Medicine
    Induction and Assessment of Exertional Skeletal Muscle Damage in Humans
  • Research • Medicine
    A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device
  • Research • Medicine
    Drug-Induced Sleep Endoscopy (DISE) with Target Controlled Infusion (TCI) and Bispectral Analysis in Obstructive Sleep Apnea
  • Research • Medicine
    Integrated Compensatory Responses in a Human Model of Hemorrhage
  • Research • Medicine
    Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
  • Research • Medicine
    Impression Cytology of the Lid Wiper Area
  • Research • Behavior
    A Protocol of Manual Tests to Measure Sensation and Pain in Humans
  • Research • Medicine
    Unbiased Deep Sequencing of RNA Viruses from Clinical Samples
  • Research • Medicine
    A Choroid Plexus Epithelial Cell-based Model of the Human Blood-Cerebrospinal Fluid Barrier to Study Bacterial Infection from the Basolateral Side
  • Research • Medicine
    Isolation and Profiling of MicroRNA-containing Exosomes from Human Bile
  • Research • Medicine
    Generation of Microtumors Using 3D Human Biogel Culture System and Patient-derived Glioblastoma Cells for Kinomic Profiling and Drug Response Testing
  • Research • Medicine
    Ultrasound Assessment of Endothelial Function: A Technical Guideline of the Flow-mediated Dilation Test
  • Research • Medicine
    Using a Laminating Technique to Perform Confocal Microscopy of the Human Sclera
  • Research • Medicine
    Intravenous Endotoxin Challenge in Healthy Humans: An Experimental Platform to Investigate and Modulate Systemic Inflammation
  • Research • Medicine
    Modeling and Simulations of Olfactory Drug Delivery with Passive and Active Controls of Nasally Inhaled Pharmaceutical Aerosols
  • Research • Medicine
    Exosomal miRNA Analysis in Non-small Cell Lung Cancer (NSCLC) Patients' Plasma Through qPCR: A Feasible Liquid Biopsy Tool
  • Research • Medicine
    A Multimodal Imaging- and Stimulation-based Method of Evaluating Connectivity-related Brain Excitability in Patients with Epilepsy
  • Research • Medicine
    Measuring Cardiac Autonomic Nervous System (ANS) Activity in Toddlers - Resting and Developmental Challenges
  • Research • Medicine
    Using Saccadometry with Deep Brain Stimulation to Study Normal and Pathological Brain Function
  • Research • Medicine
    Quantitative Fundus Autofluorescence for the Evaluation of Retinal Diseases
  • Research • Medicine
    Diagnosis of Musculus Gastrocnemius Tightness - Key Factors for the Clinical Examination
  • Research • Medicine
    Stereo-Electro-Encephalo-Graphy (SEEG) With Robotic Assistance in the Presurgical Evaluation of Medical Refractory Epilepsy: A Technical Note
  • Research • Medicine
    Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
  • Research • Medicine
    Transcutaneous Microcirculatory Imaging in Preterm Neonates
  • Research • Medicine
    Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
  • Research • Medicine
    Design, Fabrication, and Administration of the Hand Active Sensation Test (HASTe)
  • Research • Medicine
    MRI-guided dmPFC-rTMS as a Treatment for Treatment-resistant Major Depressive Disorder
  • Research • Medicine
    Functional Human Liver Preservation and Recovery by Means of Subnormothermic Machine Perfusion
  • Research • Medicine
    A Multicenter MRI Protocol for the Evaluation and Quantification of Deep Vein Thrombosis
  • Research • Medicine
    Determining The Electromyographic Fatigue Threshold Following a Single Visit Exercise Test
  • Research • Medicine
    Use of Electromagnetic Navigational Transthoracic Needle Aspiration (E-TTNA) for Sampling of Lung Nodules
  • Research • Medicine
    Trabecular Meshwork Response to Pressure Elevation in the Living Human Eye
  • Research • Medicine
    In Vivo, Percutaneous, Needle Based, Optical Coherence Tomography of Renal Masses
  • Research • Medicine
    Establishment of Human Epithelial Enteroids and Colonoids from Whole Tissue and Biopsy
  • Research • Medicine
    Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
  • Research • Medicine
    Preparation and Respirometric Assessment of Mitochondria Isolated from Skeletal Muscle Tissue Obtained by Percutaneous Needle Biopsy
  • Research • Medicine
    A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
  • Research • Medicine
    Isolation and Immortalization of Patient-derived Cell Lines from Muscle Biopsy for Disease Modeling
  • Research • Medicine
    State of the Art Cranial Ultrasound Imaging in Neonates
  • Research • Medicine
    Measurement of Dynamic Scapular Kinematics Using an Acromion Marker Cluster to Minimize Skin Movement Artifact
  • Research • Medicine
    The Supraclavicular Fossa Ultrasound View for Central Venous Catheter Placement and Catheter Change Over Guidewire
  • Research • Medicine
    Ultrasound Assessment of Endothelial-Dependent Flow-Mediated Vasodilation of the Brachial Artery in Clinical Research
  • Research • Medicine
    Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
  • Research • Medicine
    A Neuroscientific Approach to the Examination of Concussions in Student-Athletes
  • Research • Medicine
    DTI of the Visual Pathway - White Matter Tracts and Cerebral Lesions
  • Research • Medicine
    Collection, Isolation, and Flow Cytometric Analysis of Human Endocervical Samples
  • Research • Medicine
    Fundus Photography as a Convenient Tool to Study Microvascular Responses to Cardiovascular Disease Risk Factors in Epidemiological Studies
  • Research • Medicine
    A Multi-Modal Approach to Assessing Recovery in Youth Athletes Following Concussion
  • Research • Medicine
    Clinical Assessment of Spatiotemporal Gait Parameters in Patients and Older Adults
  • Research • Medicine
    Multi-electrode Array Recordings of Human Epileptic Postoperative Cortical Tissue
  • Research • Medicine
    Collection and Extraction of Saliva DNA for Next Generation Sequencing
  • Research • Medicine
    Fast and Accurate Exhaled Breath Ammonia Measurement
  • Research • Medicine
    Developing Neuroimaging Phenotypes of the Default Mode Network in PTSD: Integrating the Resting State, Working Memory, and Structural Connectivity
  • Research • Medicine
    Two Methods for Establishing Primary Human Endometrial Stromal Cells from Hysterectomy Specimens
  • Research • Medicine
    Assessment of Vascular Function in Patients With Chronic Kidney Disease
  • Research • Medicine
    Coordinate Mapping of Hyolaryngeal Mechanics in Swallowing
  • Research • Medicine
    Network Analysis of the Default Mode Network Using Functional Connectivity MRI in Temporal Lobe Epilepsy
  • Research • Medicine
    EEG Mu Rhythm in Typical and Atypical Development
  • Research • Medicine
    The Multiple Sclerosis Performance Test (MSPT): An iPad-Based Disability Assessment Tool
  • Research • Medicine
    Isolation and Functional Characterization of Human Ventricular Cardiomyocytes from Fresh Surgical Samples
  • Research • Medicine
    Dynamic Visual Tests to Identify and Quantify Visual Damage and Repair Following Demyelination in Optic Neuritis Patients
  • Research • Medicine
    Primary Culture of Human Vestibular Schwannomas
  • Research • Medicine
    Utility of Dissociated Intrinsic Hand Muscle Atrophy in the Diagnosis of Amyotrophic Lateral Sclerosis
  • Research • Medicine
    Lesion Explorer: A Video-guided, Standardized Protocol for Accurate and Reliable MRI-derived Volumetrics in Alzheimer's Disease and Normal Elderly
  • Research • Medicine
    Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
  • Research • Medicine
    Isolation, Culture, and Imaging of Human Fetal Pancreatic Cell Clusters
  • Research • Medicine
    3D-Neuronavigation In Vivo Through a Patient's Brain During a Spontaneous Migraine Headache
  • Research • Medicine
    A Novel Application of Musculoskeletal Ultrasound Imaging
  • Research • Medicine
    Computerized Dynamic Posturography for Postural Control Assessment in Patients with Intermittent Claudication
  • Research • Medicine
    Collecting Saliva and Measuring Salivary Cortisol and Alpha-amylase in Frail Community Residing Older Adults via Family Caregivers
  • Research • Medicine
    Diffusion Tensor Magnetic Resonance Imaging in the Analysis of Neurodegenerative Diseases
  • Research • Medicine
    Transcriptomic Analysis of Human Retinal Surgical Specimens Using jouRNAl
  • Research • Medicine
    Improved Protocol For Laser Microdissection Of Human Pancreatic Islets From Surgical Specimens
  • Research • Medicine
    Evaluation of Respiratory Muscle Activation Using Respiratory Motor Control Assessment (RMCA) in Individuals with Chronic Spinal Cord Injury
  • Research • Medicine
    Minimal Erythema Dose (MED) Testing
  • Research • Medicine
    Measuring Cardiac Autonomic Nervous System (ANS) Activity in Children
  • Research • Medicine
    Collecting And Measuring Wound Exudate Biochemical Mediators In Surgical Wounds
  • Research • Medicine
    A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
  • Research • Medicine
    Using a Chemical Biopsy for Graft Quality Assessment
  • Research • Medicine
    Characterizing Exon Skipping Efficiency in DMD Patient Samples in Clinical Trials of Antisense Oligonucleotides
  • Research • Medicine
    In Vitro Assessment of Cardiac Function Using Skinned Cardiomyocytes
  • Research • Medicine
    Normothermic Ex Situ Heart Perfusion in Working Mode: Assessment of Cardiac Function and Metabolism
  • Research • Medicine
    Evaluation of Vascular Control Mechanisms Utilizing Video Microscopy of Isolated Resistance Arteries of Rats
  • Research • Medicine
    Bronchoalveolar Lavage (BAL) for Research; Obtaining Adequate Sample Yield
  • Research • Medicine
    Non-invasive Optical Measurement of Cerebral Metabolism and Hemodynamics in Infants
  • Research • Medicine
    Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
  • Research • Medicine
    Driving Simulation in the Clinic: Testing Visual Exploratory Behavior in Daily Life Activities in Patients with Visual Field Defects
  • Research • Medicine
    Isolation, Characterization and Comparative Differentiation of Human Dental Pulp Stem Cells Derived from Permanent Teeth by Using Two Different Methods
  • Research • Medicine
    Portable Intermodal Preferential Looking (IPL): Investigating Language Comprehension in Typically Developing Toddlers and Young Children with Autism
  • Research • Medicine
    Intraoperative Detection of Subtle Endometriosis: A Novel Paradigm for Detection and Treatment of Pelvic Pain Associated with the Loss of Peritoneal Integrity
  • Research • Medicine
    The Use of Primary Human Fibroblasts for Monitoring Mitochondrial Phenotypes in the Field of Parkinson's Disease
  • Research • Medicine
    Collection Protocol for Human Pancreas
  • Research • Medicine
    The α-test: Rapid Cell-free CD4 Enumeration Using Whole Saliva
  • Research • Medicine
    The Measurement and Treatment of Suppression in Amblyopia
  • Research • Medicine
    Corneal Donor Tissue Preparation for Endothelial Keratoplasty
  • Research • Medicine
    Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography (FDG-PET/CT)
  • Research • Medicine
    Eye Tracking Young Children with Autism
  • Research • Medicine
    Doppler Optical Coherence Tomography of Retinal Circulation
  • Research • Medicine
    Utilizing Transcranial Magnetic Stimulation to Study the Human Neuromuscular System
  • Research • Medicine
    Detection and Genogrouping of Noroviruses from Children's Stools By Taqman One-step RT-PCR
  • Research • Medicine
    Method to Measure Tone of Axial and Proximal Muscle
  • Research • Medicine
    The Trier Social Stress Test Protocol for Inducing Psychological Stress
  • Research • Medicine
    Probing the Brain in Autism Using fMRI and Diffusion Tensor Imaging
  • Research • Medicine
    Multifocal Electroretinograms
  • Research • Medicine
    Isolation of Human Islets from Partially Pancreatectomized Patients
  • Research • Medicine
    Examining the Characteristics of Episodic Memory using Event-related Potentials in Patients with Alzheimer's Disease
  • Research • Medicine
    Magnetic Resonance Imaging Quantification of Pulmonary Perfusion using Calibrated Arterial Spin Labeling
  • Research • Medicine
    Manual Muscle Testing: A Method of Measuring Extremity Muscle Strength Applied to Critically Ill Patients
  • Research • Medicine
    Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
  • Research • Medicine
    A Protocol for Comprehensive Assessment of Bulbar Dysfunction in Amyotrophic Lateral Sclerosis (ALS)
  • Research • Medicine
    An Investigation of the Effects of Sports-related Concussion in Youth Using Functional Magnetic Resonance Imaging and the Head Impact Telemetry System
  • Research • Medicine
    Corneal Confocal Microscopy: A Novel Non-invasive Technique to Quantify Small Fibre Pathology in Peripheral Neuropathies
  • Research • Medicine
    Methods to Quantify Pharmacologically Induced Alterations in Motor Function in Human Incomplete SCI
  • Research • Medicine
    Multispectral Real-time Fluorescence Imaging for Intraoperative Detection of the Sentinel Lymph Node in Gynecologic Oncology
  • Research • Medicine
    Technique to Collect Fungiform (Taste) Papillae from Human Tongue
  • Research • Medicine
    Assessing Endothelial Vasodilator Function with the Endo-PAT 2000
  • Research • Medicine
    Making Sense of Listening: The IMAP Test Battery
  • Research • Medicine
    An Experimental Paradigm for the Prediction of Post-Operative Pain (PPOP)
  • Research • Biology
    Bioelectric Analyses of an Osseointegrated Intelligent Implant Design System for Amputees
  • Research • Biology
    Demonstration of Cutaneous Allodynia in Association with Chronic Pelvic Pain
  • Get cutting-edge science videos from JoVE sent straight to your inbox every month.

    Waiting X
    Simple Hit Counter