Summary

Hjärttryck Volym Loop analys Använda Konduktans katetrar hos möss

Published: September 17, 2015
doi:

Summary

Cardiac pressure-volume loop analysis is the most comprehensive way to measure cardiac function in the intact heart. We describe a technique to perform and analyze cardiac pressure volume loops, using conductance catheters.

Abstract

Cardiac pressure-volume loop analysis is the “gold-standard” in the assessment of load-dependent and load-independent measures of ventricular systolic and diastolic function. Measures of ventricular contractility and compliance are obtained through examination of cardiac response to changes in afterload and preload. These techniques were originally developed nearly three decades ago to measure cardiac function in large mammals and humans. The application of these analyses to small mammals, such as mice, has been accomplished through the optimization of microsurgical techniques and creation of conductance catheters. Conductance catheters allow for estimation of the blood pool by exploiting the relationship between electrical conductance and volume. When properly performed, these techniques allow for testing of cardiac function in genetic mutant mouse models or in drug treatment studies. The accuracy and precision of these studies are dependent on careful attention to the calibration of instruments, systematic conduct of hemodynamic measurements and data analyses. We will review the methods of conducting pressure-volume loop experiments using a conductance catheter in mice.

Introduction

Hjärttryckvolymslinganalys ger detaljerad information om hjärtfunktion och är guldmyntfoten för funktionell bedömning 1. Medan avbildningstekniker såsom ekokardiografi eller hjärt MRI ger funktionella åtgärder, dessa åtgärder är i hög grad beroende av belastningsförhållanden. Lastoberoende mått på hjärtats kontraktilitet och avkoppling kräver dynamiska mätningar av ventrikulära trycket och volym förhållande över ett område av förbelastning och efterbelastning. Denna förståelse av tryckvolym relation härrör från den banbrytande arbete Sagawa och kollegor 2,3. De visade i ex vivo perfusion hund hjärtan att tryck volym loop härledda kontraktilitet åtgärder var oberoende av belastningsförhållanden 4.

In vivo tillämpning av dessa analyser blev möjligt med utvecklingen av konduktans katetrar på 1980-talet. Denna tekniska framsteg tillåten Kass och kollegor för att utföra tryckvolymslinganalys hos människor 5,6. Miniatyrisering av konduktans katetrar och förbättringar i kirurgiska tekniker i slutet av 1990-talet 7 gjorde analys av gnagare hjärtfunktion möjlig, vilket möjliggör genetiska och farmakologiska studier som skall utföras. Detta förskott har sedan leder till den utbredda användningen av tryck volym slinganalys och har genererat en hel del insikt i däggdjurs hjärt fysiologi.

Ett nyckelbegrepp när det gäller användningen av ledningsförmåga katetrar och tolkning av uppgifter från det är förhållandet mellan volym och ledningsförmåga. Konduktans är omvänt relaterad till spänning, som mäts med användning av en kateter med elektroder placerade proximalt, vanligen placerade under aortaklaffen, och distalt, vid LV spetsen 8. Förändringar i spänning eller ledningsförmåga mäts av förändringar i ström som flyter från proximal till distal elektrod. Även blodpoolen bidras avsevärt till konduktans, bidraget från den ventrikulära väggen, benämnd parallellkonduktansen (Vp), till uppmätt ledningsförmåga måste dras för att erhålla absoluta mätningar LV volym.

Metoderna för att utföra denna korrigering, som kallas en koksalt kalibrering, diskuteras i protokollet nedan. Det matematiska förhållandet mellan konduktans och volym, beskriven av Baan och kollegor, är att volym = 1 / α; (ρ L 2) (GG p), där α = likformigt fält korrektionsfaktor, ρ = blod resistivitet, L = avståndet mellan elektroderna, G = konduktansen och G p = icke-blod konduktans 9. Notera den enhetliga fältkorrektionsfaktorn hos möss närmar sig 1,0 på grund av små kammarvolymer 10. Tillsammans med tryckgivare ger ledningskatetern realtid samtidig tryck- och volymuppgifter.

Hjärt pressuåtervolymanalys uppvisar särskilda fördelar jämfört med andra mått på hjärtfunktion, eftersom de möjliggör mätning av kammarfunktion oberoende av belastningsförhållanden och av hjärtfrekvens. Särskilda lastoberoende hjärtindex kontraktilitet inkluderar: slut systoliskt blodtryck volym relation (ESPVR), d P / d t max end-diastolisk volym relation, maximal elastans (E max) och förspänning rekryterslagarbete (PRSW). En lastoberoende mått på diastolisk funktion är slut diastoliska trycket volymförhållande (EDPVR) 11. Följande protokoll beskriver genomförandet av hjärttryckvolym slinganalys, med både en halspulsådern och en apikal strategi. Medan metoden att utföra dessa studier har beskrivits i detalj tidigare 8,11, går vi igenom de viktigaste åtgärder för att få exakta tryckvolymmätningar, både saltlösning och kyvett kalibrerings korrigering, och ger en visuell demonstration av these förfaranden. Forskning med djur som genomförs för denna studie hanteras i enlighet med godkända protokoll och djurskyddsbestämmelser av Duke University Medical Center Institutional Animal Care och användning kommittén.

Protocol

1. Konduktans Kateter Förberedelser och tryck Kalibrering Anslut lednings katetern till hemodynamiska katetermodulen. Elektroniskt kalibrera tryckmätningar och volym genom att spela in förinställt tryck och volym inställd på katetermodulen. Spela in en spårning av 0 mm Hg och 25 mm Hg (Figur 1A) och tilldela spänningar till både tryck kurvor (Figur 1B och 1C). På liknande sätt, spela in en volym spårning av 5 RVU och 25 RVU (figur 1D) och tilldela…

Representative Results

Tryck-volymslinganalys kan användas för att mäta hjärtfunktion hos genetiskt modifierade möss 14,15 eller möss som genomgår läkemedelsstudier 16. Representativa tryckvolym slingor tillhandahålls från tidigare publicerade arbeten 16 undersöker effekten av ß-arrestin partisk AT1R ligand, TRV120023. För att testa om TRV120023 påverkar hjärtfunktionen in vivo, var tryck volym loop analys på vildtypsmöss erhåller konventionella och nya angiotensinreceptorblockerare….

Discussion

Vi beskriver ett förfarande för perfoming tryck-volymslinganalys med användning av en konduktans kateter i möss, för att härleda omfattande analyser av både hjärtats kontraktilitet och avkoppling. Suga, Sagawa och kollegor utnyttjas tryck volym loopar att definiera åtgärder hjärtkontraktiliteten, särskilt lutningen på ESPVR eller slut systoliska elastans (E er), och Emax. Elasticiteten, som definieras av förhållandet mellan tryck och volym (P / V) varierar under hela systole. Under v…

Disclosures

The authors have nothing to disclose.

Acknowledgements

Detta arbete stöds av American Heart Association 14FTF20370058 (DMA) och NIH T32 HL007101-35 (DMA).

Materials

AnaSed (xylazine)  Lloyd Laboratories NADA no. 139-236 Anesthetic
Ketaset (ketamine) Pfizer 440842 Anesthetic
VIP3000 Matrx Medical Inc. Anesthesia machine
Ventilator Harvard Apparatus Model 683 Surgical Equipment
Tubing kit Harvard Apparatus 72-1049 Surgical Equipment
Homeothermic Blanket  Kaz Inc. 5628 Surgical Equipment
Stereo microscope Carl Zeiss Optical Inc. Stemi 2000 Surgical Equipment
Illuminator Cole–Parmer 41720 Surgical Equipment
Dumont no. 55 Dumostar Forceps  Fine Science Tools Inc 11295-51 Surgical Instruments
Graefe forceps, curved  Fine Science Tools Inc 11052-10 Surgical Instruments
Moria MC31 forceps  Fine Science Tools Inc 11370-31 Surgical Instruments
Mayo scissors  Fine Science Tools Inc 14512-15 Surgical Instruments
Iris scissors  Fine Science Tools Inc 14041-10 Surgical Instruments
Halsey needle holder  Fine Science Tools Inc 12501-13 Surgical Instruments
Olsen–Hegar needle holder  Fine Science Tools Inc 12002-12 Surgical Instruments
spring scissors Fine Science Tools Inc 15610-08 Surgical Instruments
disposable underpads Kendall/Tyco Healthcare 1038 Surgical Supplies
Sterile gauze sponges, sterile  Dukal 62208 Surgical Supplies
Cotton-tipped applicators, sterile  Solon 368 Surgical Supplies
Surgical suture,  silk, 6-0  DemeTECH FT-639-1 Surgical Supplies
1 cc Insulin syringes  Becton Dickenson 329412 Surgical Supplies
Access-9™ Hemostasis Valve Merit Medical  MAP111 Hemodynamic equipment
Sphygmomanometer Baumanometer 320 Hemodynamic equipment
Millar PV system MPVS-300/400 or MPVS Ultra (includes calibration cuvette) ADInstruments Inc Hemodynamic equipment
1.4F conductance catheter  ADInstruments Inc SPR-839 Hemodynamic equipment
PowerLab 4/30 with Chart Pro ADInstruments Inc. ML866/P Hemodynamic software
animal clipper Wahl 8787-450A Miscellaneous
Intradermic tubing PE-10 (Becton Dickinson, cat. no. ) Becton Dickenson 427401 Miscellaneous
Intradermic tubing PE-50 (Becton Dickinson, cat. no.) Becton Dickenson 427411 Miscellaneous
Needle assortment (18, 25 and 30 gauge; Thomas Scientific) Miscellaneous
0.9% (wt/vol) sodium chloride injection, USP , cat. no. ) Hospira NDC no. 0409-4888-50 Miscellaneous
Surgical tape Miscellaneous
Alconox (Alconox Inc.) for catheter cleaning ADInstruments Inc. Miscellaneous

References

  1. Clark, J. E., Marber, M. S. Advancements in pressure-volume catheter technology – stress remodelling after infarction. Exp Physiol. 98 (3), 614-621 (2013).
  2. Sunagawa, K., Maughan, W. L., Burkhoff, D., Sagawa, K. Left ventricular interaction with arterial load studied in isolated canine ventricle. Am J Physiol. 245 (Pt 1), H773-H780 (1983).
  3. Suga, H., Sagawa, K., Demer, L. Determinants of instantaneous pressure in canine left ventricle. Time and volume specification. Circ Res. 46 (2), 256-263 (1980).
  4. Suga, H., Sagawa, K., Shoukas, A. A. Load independence of the instantaneous pressure-volume ratio of the canine left ventricle and effects of epinephrine and heart rate on the ratio. Circ Res. 32 (3), 314-322 (1973).
  5. Kass, D. A., et al. Improved left ventricular mechanics from acute VDD pacing in patients with dilated cardiomyopathy and ventricular conduction delay. Circulation. 99 (12), 1567-1573 (1999).
  6. Kass, D. A., et al. Diastolic Compliance of Hypertrophied Ventricle Is Not Acutely Altered by Pharmacological Agents Influencing Active Processes. Annals of Internal Medicine. 119 (6), 466-473 (1993).
  7. Georgakopoulos, D., et al. In vivo murine left ventricular pressure-volume relations by miniaturized conductance micromanometry. Am J Physiol. 274 (4 pt 2), H1416-H1422 (1998).
  8. Cingolani, O. H., Kass, D. A. Pressure-volume relation analysis of mouse ventricular function. Am J Physiol Heart Circ Physiol. 301 (6), H2198-H2206 (2011).
  9. Baan, J., et al. Continuous Measurement of Left-Ventricular Volume in Animals and Humans by Conductance Catheter. Circulation. 70 (5), 812-823 (1984).
  10. Pearce, J. A., Porterfield, J. E., Larson, E. R., Valvano, J. W., Feldman, M. D. Accuracy considerations in catheter based estimation of left ventricular volume. Conf Proc IEEE Eng Med Biol Soc. 2010, 3556-3558 (2010).
  11. Pacher, P., Nagayama, T., Mukhopadhyay, P., Batkai, S., Kass, D. A. Measurement of cardiac function using pressure-volume conductance catheter technique in mice and rats. Nature Protocols. 3 (9), 1422-1434 (1038).
  12. Hanusch, C., Hoeger, S., Beck, G. C. Anaesthesia of small rodents during magnetic resonance imaging. Methods. 43 (1), 68-78 (2007).
  13. Georgakopoulos, D., Kass, D. A. Estimation of parallel conductance by dual-frequency conductance catheter in mice. Am J Physiol Heart Circ Physiol. 279 (1), H443-H450 (2000).
  14. Esposito, G., et al. Increased myocardial contractility and enhanced exercise function in transgenic mice overexpressing either adenylyl cyclase 5 or 8. Basic Res Cardiol. 103 (1), 22-30 (2008).
  15. Kohout, T. A., et al. Augmentation of cardiac contractility mediated by the human beta(3)-adrenergic receptor overexpressed in the hearts of transgenic mice. Circulation. 104 (20), 2485-2491 (2001).
  16. Kim, K. S., et al. beta-Arrestin-biased AT1R stimulation promotes cell survival during acute cardiac injury. Am J Physiol Heart Circ Physiol. 303 (8), H1001-H1010 (2012).
  17. Suga, H., Sagawa, K. Mathematical Interrelationship between Instantaneous Ventricular Pressure-Volume Ratio and Myocardial Force-Velocity Relation. Annals of Biomedical Engineering. 1 (2), 160-181 (1972).
  18. Suga, H. Ventricular energetics. Physiol Rev. 70 (2), 247-277 (1990).
  19. Kass, D. A., et al. Influence of contractile state on curvilinearity of in situ end-systolic pressure-volume relations. Circulation. 79 (1), 167-178 (1989).
  20. Little, W. C. The left ventricular dP/dtmax-end-diastolic volume relation in closed-chest dogs. Circ Res. 56 (6), 808-815 (1985).
  21. Glower, D. D., et al. Linearity of the Frank-Starling relationship in the intact heart: the concept of preload recruitable stroke work. Circulation. 71 (5), 994-1009 (1985).
  22. Sharir, T., et al. Ventricular systolic assessment in patients with dilated cardiomyopathy by preload-adjusted maximal power. Validation and noninvasive application. Circulation. 89 (5), 2045-2053 (1994).
  23. Baan, J., Van der Velde, E. T. Sensitivity of left ventricular end-systolic pressure-volume relation to type of loading intervention in dogs. Circ Res. 62 (6), 1247-1258 (1988).
  24. Wei, C. L., et al. Volume catheter parallel conductance varies between end-systole and end-diastole. IEEE Trans Biomed Eng. 54 (8), 1480-1489 (2007).
  25. Porterfield, J. E., et al. Dynamic correction for parallel conductance, GP, and gain factor, alpha, in invasive murine left ventricular volume measurements. J Appl Physiol (1985). 107 (6), 1693-1703 (2009).
check_url/52942?article_type=t

Play Video

Cite This Article
Abraham, D., Mao, L. Cardiac Pressure-Volume Loop Analysis Using Conductance Catheters in Mice. J. Vis. Exp. (103), e52942, doi:10.3791/52942 (2015).

View Video