Summary

Samtidig opptak av Elektroretinografi og visuelle fremkalt respons hos bedøvede rotter

Published: July 01, 2016
doi:

Summary

This protocol describes simultaneous measurement of electroretinogram and visual evoked potentials in anesthetized rats.

Abstract

The electroretinogram (ERG) and visual evoked potential (VEP) are commonly used to assess the integrity of the visual pathway. The ERG measures the electrical responses of the retina to light stimulation, while the VEP measures the corresponding functional integrity of the visual pathways from the retina to the primary visual cortex following the same light event. The ERG waveform can be broken down into components that reflect responses from different retinal neuronal and glial cell classes. The early components of the VEP waveform represent the integrity of the optic nerve and higher cortical centers. These recordings can be conducted in isolation or together, depending on the application. The methodology described in this paper allows simultaneous assessment of retinal and cortical visual evoked electrophysiology from both eyes and both hemispheres. This is a useful way to more comprehensively assess retinal function and the upstream effects that changes in retinal function can have on visual evoked cortical function.

Introduction

Måling av elektroretinogrammet (ERG) og visuell fremkalt potensial (VEP) gir nyttige kvantitative vurderinger av integriteten av visuelle veien. ERG måler de elektriske responser i netthinnen til lys stimulering, mens den VEP måler den tilsvarende funksjonell integritet av synsbanene fra retina til den primære visuelle cortex å følge den samme lys hendelse. Dette manuskriptet beskriver en protokoll for registrering og analyse av ERG og VEP responser i et vanlig laboratoriemodell, rotten.

ERG gir en indeks av den funksjonelle integriteten til en rekke sentrale retinal celle klasser ved å tallfeste netthinnen brutto elektriske respons til et lysglimt. En koordinert serie av ioniske flussmidler initieres av lys begynn og offset, gi målbare forandringer i spenning som kan måles ved hjelp av overflate elektroder plassert utenfor øyet. Den resulterende bølgeform representerer kombinasjonen av en i seg selvrier av veldefinerte komponenter, ulik amplitude, timing og frekvens. En betydelig mengde forskning har vist at disse komponentene er relativt godt konservert på tvers av mange virveldyr retinae og at komponentene kan separeres fra hverandre. Ved judiciously velge stimulus (flash stimulus, bakgrunn, interstimulus intervall) forhold og velge spesifikke funksjoner i den sammensatte bølgeform å analysere kan man være sikker på å returnere et mål på en bestemt gruppe av netthinnens celler 1,2. Disse egenskapene ligger til grunn for verktøyet, og dermed den utbredte anvendelser av ERG som et non-invasiv måling av netthinnen funksjon. Dette manuskriptet fokuserer på metodikk for å måle ERG og analysere sine funksjoner for å returnere informasjon om noen av de store celle klasser i netthinnen, nemlig fotoreseptorer (PIII komponent), bipolare celler (PII komponent) og gangliecelle (den positive scotopic terskel svar eller pSTR).

<p class= "Jove_content"> The VEP gir en analyse av kortikale respons på lys; første stammer fra netthinnen og deretter kommunisert serielt via synsnerven, synskanalen, thalamus (lateral geniculate kjernen, LGN) og optisk stråling til området V1 av cortex tre. I gnagere, majoriteten (90 – 95%) av optisk nervefibre fra hvert øye decussate 4 og innerverer kontralaterale midten av hjernen. I motsetning til ERG, er det foreløpig ikke mulig å tillegge forskjellige komponenter av VEP til spesifikke celle klasser, 5 endrer dermed hvor som helst langs den visuelle pathway kan påvirke VEP bølgeform. Ikke desto mindre er det VEP et nyttig ikke-invasiv måling av visuell ytelse og visuell vei integritet. Den VEP, når det brukes sammen med ERG, kan gi en mer fullstendig vurdering av det visuelle system (dvs. netthinnen / visuelle veien).

ERG og VEP opptak kan gjennomføres isolert eller i kombinasjon, avhengig av apparakasjon. Metodikken er beskrevet i denne artikkelen tillater samtidig vurdering av retinal og kortikal visuell fremkalt elektrofysiologi fra begge øynene og begge halvkuler i bedøvet rotter. Dette er en nyttig måte å mer grundig vurdere retinal funksjon og oppstrøms effekter som endringer i netthinnefunksjon kan ha på visuell fremkalt kortikal funksjon.

Protocol

Alle eksperimentelle prosedyrer ble utført i henhold til den australske anbefalingen for omsorg og bruk av dyr for vitenskapelige formål, fastsatt av National Health and Medical Research Council i Australia. Etikk clearance ble hentet fra University of Melbourne, teknisk- naturvitenskapelige fakultet, dyreetikk komité (godkjenningsnummer 0911322,1). 1. Pre-implantasjon av kronisk VEP Elektroder Merk: Hvis samtidige ERG og VEP signaler skal samles dyr må implantert med VEP elektroder minst en uke før s…

Representative Results

ERG en bølge (> -1,38 log cd.sm -2), B-bølger (> – 4,99 log cd.sm -2) MT-rapporter (<- 4,99 log cd.sm -2) og veps (> – 0.52 log cd.sm -2) ble registrert samtidig (figur 1 og 3). Ved svært dim blinker, er en positiv STR (pSTR) sett på ca 110 msek etter blitsen, og en negativ STR (nSTR) på ca 220 msek (figur 1 og 2). En ERG med et stort b bølger, topper mellom 50 til 100 millisekunder etter star…

Discussion

ERG og VEP er objektive mål på visuell funksjon fra netthinnen og cortex, henholdsvis. Fordelen med samtidig opptak er at en mer omfattende oversikt over hele visuelle veien er råd. Spesielt kan utfyllende informasjon fra sine samtidige vurdering gi en klarere avgrensning av skadestedet i det visuelle pathway (f.eks for lidelser med overlappende ERG ennå tydelig VEP manifestasjoner 18, når optikusnevropati kan sameksistere med primær cerebral atrofi 19, 20, eller når VEP tap kan bli…

Disclosures

The authors have nothing to disclose.

Acknowledgements

Funding for this project was provided by the National Health and Medical Research Council (NHMRC) 1046203 (BVB, AJV) and Melbourne Neuroscience Institute Fellowship (CTN).

Materials

Alligator clip generic brand HM3022 Stainless steel 26 mm clip for connecting VEP screw electrodes to cables
Bioamplifier ADInstruments ML 135 For amplifying ERG and VEP signals
Carboxymethylcellulose sodium 1.0% Allergan CAS 0009000-11-7 Viscous fluid for improving signal quality of the active ERG electrode
Carprofen 0.5% Pfizer Animal Health Group CAS 53716-49-7 Proprietary name: Rimadyl injectable (50 mg/mL). For post-surgery analgesia, diluted to 0.5% (5 mg/mL) in normal saline
Chlorhexadine 0.5% Orion Laboratories 27411, 80085 For disinfecting surgical instruments
Circulating water bath Lauda-Königshoffen MGW Lauda For maintaining body temperature of the anesthetized animal during surgery and electrophysiological recordings
Dental amalgam DeguDent GmbH 64020024 For encasing the electrode-skull assembly to make it more robust
Dental burr Storz Instruments, Bausch and Lomb #E0824A A miniature drill head of ~0.7mm diameter for making a small hole in the skull over each hemisphere to implant VEP screws
Drill Bosch Dremel 300 series An automatic drill for trepanning
Electrode lead Grass Telefactor  F-E2-30 Platinum cables for connecting silver wire electrodes to the amplifier
Faraday Cage custom-made Ensures light proof to maintain dark adaptation. Encloses the Ganzfeld setup to improve signal to noise ratio
Gauze swabs Multigate Medical Products Pty Ltd 57-100B For drying the surgical incision and exposed skull surface during surgery
Ganzfeld integrating sphere Photometric Solutions International Custom designed light stimulator: 36 mm diameter, 13 cm aperture size
Velcro VELCRO Australia Pty Ltd VELCRO Brand Reusable Wrap Hook-and-loop fastener to secure the electrodes and the animal on the recording platform
Isoflurane 99.9% Abbott Australasia Pty Ltd CAS 26675-46-7 Proprietary Name: Isoflo(TM) Inhalation anaaesthetic. Pharmaceutical-grade inhalation anesthetic mixed with oxygen gas for VEP electrode implant surgery
Ketamine  Troy Laboratories Ilium Ketamil Proprietary name: Ketamil Injection, Brand: Ilium. Pharmaceutical-grade anesthetic for electrophysiological recording
Luxeon LEDs Phillips Lighting Co. For light stimulation twenty 5 watt and one 1 watt LEDs.
Micromanipulator Harvard Apparatus BS4 50-2625 Holds the ERG active electrode during recordings
Needle electrode Grass Telefactor  F-E2-30 Subcutaneously inserted in the tail to serve as the ground electrode for both the ERG and VEP
Phenylephrine 2.5% minims  Bausch and Lomb CAS 61-76-7 Instilled with Tropicamide to achieve maximal dilation for ERG recording
Povidone iodine 10% Sanofi-Aventis CAS 25655-41-8 Proprietory name: Betadine, Antiseptic to prepare the shaved skin for surgery 10%, 500 mL
Powerlab data acquisition system ADInstruments ML 785 Controls the LEDs
Proxymetacaine 0.5% Alcon Laboratories  CAS 5875-06-9 For corneal anaesthesia during ERG recordings
Saline solution Gelflex Non-injectable, for electroplating silver wire electrodes
Scope Software ADInstruments version 3.7.6 Simultaneously triggers the stimulus via the Powerlab system and collects data
Silver (fine round wire) A&E metal 0.3 mm Used to make active and inactive ERG electrodes, and the inactive VEP electrode
Stainless streel screws  MicroFasterners 0.7 mm shaft diameter, 3 mm in length to be implanted over the primary visual cortex and serve as the active VEP electrodes
Stereotaxic frame David Kopf Model 900 A small animal stereotaxic instrument for locating the primary visual cortices according to Paxinos & Watson's 2007 rat brain atlas coordinates
Surgical blade Swann-Morton Ltd. 0206 For incising the area of skin overlaying the primary visual cortex to implant the VEP electrodes
Suture Shanghai Pudong Jinhuan Medical Products Co.,Ltd 3-0 silk braided suture non-absorbable, for skin retraction during VEP electrode implantation surgery
Tobramycine eye ointment 0.3% Alcon Laboratories  CAS 32986-56-4 Proprietary name: Tobrex. Prophylactic antibiotic ointment applied around the skin wound after surgery
Tropicamide 0.5% Alcon Laboratories  CAS 1508-75-4 Proprietary name: 0.5% Mydriacyl eye drop, Instilled to achieve mydriasis for ERG recording
Xylazine Troy Laboratories Ilium Xylazil-100 Pharmaceutical-grade anesthetic for electrophysiological recording
Pipette tip  Eppendorf Pty Ltd 0030 073.169 Eppendorf epTIPS 100 – 5000 mL, for custom-made electrodes
Microsoft Office Excel Microsoft version 2010 spreadsheet software for data analysis
Lethabarb Euthanazia Injection Virbac (Australia) Pty Ltd LETHA450 325 mg/mL pentobarbital sodium for rapid euthanazia

References

  1. Nguyen, C. T. O., Vingrys, A. J., Bui, B. V. Dietary omega-3 fatty acids and ganglion cell function. Invest Ophthalmol Vis Sci. 49, 3586-3594 (2008).
  2. Weymouth, A. E., Vingrys, A. J. Rodent electroretinography: methods for extraction and interpretation of rod and cone responses. Prog Retin Eye Res. 27, 1-44 (2008).
  3. Tsai, T. I., Bui, B. V., Vingrys, A. J. Effect of acute intraocular pressure challenge on rat retinal and cortical function. Invest Ophthalmol Vis Sci. 55, 1067-1077 (2014).
  4. Cowey, A., Franzini, C. The retinal origin of uncrossed optic nerve fibres in rats and their role in visual discrimination. Exp Brain Res. 35, 443-455 (1979).
  5. Weinstein, G. W., Odom, J. V., Cavender, S. Visually evoked potentials and electroretinography in neurologic evaluation. Neurol Clin. 9, 225-242 (1991).
  6. Odom, J. V., et al. Visual evoked potentials standard (2004). Doc Ophthalmol. 108, 115-123 (2004).
  7. Ridder, W. H., Nusinowitz, S., Heckenlively, J. R. Causes of cataract development in anesthetized mice. Exp Eye Res. 75, 365-370 (2002).
  8. Nixon, P. J., Bui, B. V., Armitage, J. A., Vingrys, A. J. The contribution of cone responses to rat electroretinograms. Clin Experiment Ophthalmol. 29, 193-196 (2001).
  9. Bui, B. V., et al. Using the electroretinogram to understand how intraocular pressure elevation affects the rat retina. J Ophthalmol. 2013, 262467 (2013).
  10. Nguyen, C. T., Vingrys, A. J., Bui, B. V. Dietary omega-3 fatty acids and ganglion cell function. Invest Ophthalmol Vis Sci. 49, 3586-3594 (2008).
  11. Hood, D. C., Birch, D. G. A quantitative measure of the electrical activity of human rod photoreceptors using electroretinography. Vis Neurosci. 5, 379-387 (1990).
  12. Birch, D. G., Hood, D. C., Locke, K. G., Hoffman, D. R., Tzekov, R. T. Quantitative electroretinogram measures of phototransduction in cone and rod photoreceptors – Normal aging, progression with disease, and test-retest variability. Arch Ophthalmol. 120, 1045-1051 (2002).
  13. Bui, B. V., Vingrys, A. J. Development of receptoral responses in pigmented and albino guinea-pigs (Cavia porcellus). Doc Ophthalmol. 99, 151-170 (1999).
  14. Robson, J. G., Saszik, S. M., Ahmed, J., Frishman, L. J. Rod and cone contributions to the a-wave of the electroretinogram of the macaque. J Physiol. 547, 509-530 (2003).
  15. Severns, M. L., Johnson, M. A. The care and fitting of Naka-Rushton functions to electroretinographic intensity-response data. Doc Ophthalmol. 85, 135-150 (1993).
  16. Bui, B. V., Fortune, B. Origin of electroretinogram amplitude growth during light adaptation in pigmented rats. Vis Neurosci. 23, 155-167 (2006).
  17. Bui, B. V., Fortune, B. Ganglion cell contributions to the rat full-field electroretinogram. J Physiol. 555, 153-173 (2004).
  18. Tremblay, F., Laroche, R. G., Debecker, I. The Electroretinographic Diagnosis of the Incomplete Form of Congenital Stationary Night Blindness. Vision Res. 35, 2383-2393 (1995).
  19. Bayer, A. U., Keller, O. N., Ferrari, F., Maag, K. P. Association of glaucoma with neurodegenerative diseases with apoptotic cell death: Alzheimer’s disease and Parkinson’s disease. Am J Ophthalmol. 133, 135-137 (2002).
  20. Wostyn, P., Audenaert, K., De Deyn, P. P. An abnormal high trans-lamina cribrosa pressure difference: A missing link between Alzheimer’s disease and glaucoma. Clinical Neurology and Neurosurgery. 110, 753-754 (2008).
  21. Yucel, Y. H., Zhang, Q. A., Weinreb, R. N., Kaufman, P. L., Gupta, N. Effects of retinal ganglion cell loss on magno-, parvo-, koniocellular pathways in the lateral geniculate nucleus and visual cortex in glaucoma. Prog Retin Eye Res. 22, 465-481 (2003).
  22. Gupta, N., Yucel, Y. H. What changes can we expect in the brain of glaucoma patients. Survey of Ophthalmology. 52, 122-126 (2007).
  23. Kong, Y. X., et al. Impact of aging and diet restriction on retinal function during and after acute intraocular pressure injury. Neurobiol Aging. 33, 1115-1125 (2012).
  24. Bui, B. V., Sinclair, A. J., Vingrys, A. J. Electroretinograms of albino and pigmented guinea-pigs (Cavia porcellus). Aust N Z J Ophthalmol. 26, 98-100 (1998).
  25. Jobling, A. I., Wan, R., Gentle, A., Bui, B. V., McBrien, N. A. Retinal and choroidal TGF-beta in the tree shrew model of myopia: isoform expression, activation and effects on function. Exp Eye Res. 88, 458-466 (2009).
  26. Robson, J. G., Frishman, L. J. Dissecting the dark-adapted electroretinogram. Doc Ophthalmol. 95, 187-215 (1998).
  27. Robson, J. G., Frishman, L. J. The rod-driven a-wave of the dark-adapted mammalian electroretinogram. Prog Retin Eye Res. 39, 1-22 (2014).
  28. Hudnell, H. K., Boyes, W. K. The comparability of rat and human visual-evoked potentials. Neurosci Biobehav Rev. 15, 159-164 (1991).
  29. Charng, J., et al. Conscious wireless electroretinogram and visual evoked potentials in rats. PLoS One. 8, e74172 (2013).
  30. Hetzler, B. E., Berger, L. K. Ketamine-Induced Modification of Photic Evoked-Potentials in the Superior Colliculus of Hooded Rats. Neuropharmacology. 23, 473-476 (1984).
check_url/54158?article_type=t

Play Video

Cite This Article
Nguyen, C. T., Tsai, T. I., He, Z., Vingrys, A. J., Lee, P. Y., Bui, B. V. Simultaneous Recording of Electroretinography and Visual Evoked Potentials in Anesthetized Rats. J. Vis. Exp. (113), e54158, doi:10.3791/54158 (2016).

View Video