Application of an Amplitude-integrated EEG Monitor (Cerebral Function Monitor) to Neonates

Published: September 6, 2017 doi: 10.3791/55985

Summary

Here, we show how to apply amplitude-integrated electroencephalography to monitor cerebral function in neonates.

Abstract

Amplitude-integrated EEG (aEEG) is an easily accessible technique to monitor the electrocortical activity in preterm and term infants in neonatal intensive care units (NICUs). This method was first used to monitor newborns after asphyxia, providing information about future neurological outcomes. The aEEG is also helpful to select newborns who benefit from cooling. The aEEG monitoring of preterm infants is becoming more widespread, as various studies have shown that neurodevelopmental outcome is related to early aEEG tracings. Here, we demonstrate the application of the aEEG monitoring system and present typical patterns that depend upon gestational age and pathophysiological conditions. Furthermore, we mention pitfalls in the interpretation of the aEEG, as this method requires accurate fixation and localization of the electrodes. Additionally, the raw EEG can be used to detect neonatal seizures or to identify aEEG application problems. In conclusion, aEEG is a safe and generally well-tolerated method for the bedside monitoring of neonatal cerebral function; it can even provide information about long-term outcome.

Introduction

aEEG was originally developed as a bedside monitor for adult intensive care1. The first publications detailing its use in neonates date back to the late 1980s2,3. In early years, its clinical use was mainly for the detection of cerebral seizure activity, surveillance of antiepileptic drug treatment4, and prediction of cerebral outcome after birth asphyxia5,6,7,8,9. Infants with birth asphyxia who did not show severe suppression of background activity and seizure activity had a more favorable outcome if they were cooled8, but research on this topic is still ongoing10,11,12. Over the past 30 years, cerebral function monitoring in neonates has become more widespread in NICUs13. Nowadays, it is increasingly being used in the preterm infant population. aEEG has been proven to be a safe method for cerebral function monitoring, even in extremely preterm infants, and is generally well-accepted by NICU staff14. Several studies showed a correlation between early aEEG recordings and neurodevelopmental outcomes in preterm infants15,16,17,18,19,20.

aEEG is based upon conventional electroencephalography that is recorded with two or four scalp electrodes, depicting the amplitude of the raw EEG on a time-compressed semi-logarithmic scale1. The signal from two or four electrodes placed in the C3, P3, C4, and P4 positions of the international 10-20 system is passed through a bandpass filter, which enhances frequencies between 2 and 15 Hz. Frequencies under 2 Hz and above 15 Hz are attenuated in order to eliminate artefacts, such as sweating, movement, muscle activity, and electrical interference, as much as possible1,4. Further processing includes filtering, rectification, smoothing, semi-logarithmic amplitude compression, and time compression. Amplitudes <10 µV are displayed on a linear scale and amplitudes >10 µV on a logarithmic scale21. The lowest-detected amplitude is shown as the lower border, and the highest amplitude is shown as the upper border21. By this means, even small changes in the lower amplitude remain visible, while an overloading of the display at high amplitudes is avoided21 (Figure 1). Due to time compression, 5 - 6 cm on the time scale represents 1 h, thus making the review of brain activity for hours and even days possible1,4,13.

The visible information in the aEEG tracing is limited to changes of the amplitude. Modern devices offer the possibility of viewing the raw EEG, so the frequency and morphology of the raw EEG curve can also be considered for interpretation. This helps to distinguish between artefacts and real seizure activity during suspicious sections of the aEEG band4. Some aEEG devices can record a simultaneous video of the patient for even better identification of seizures and artifacts. Electrode impedance is monitored during the entire recording21. In two-channel aEEG devices that use four electrodes, the investigator can switch between two intraparietal curves or one transcerebral curve (Figure 2). Depending upon the manufacturer, software offers additional features, like seizure detection, burst rate analysis, electromyography, etc. It is also possible to derive an aEEG from a full-channel EEG device that offers video recording, electromyography, electrooculography, electrocardiography, etc.

The reduction of the electrophysiological information and time compression makes continuous monitoring and bedside interpretation possible without requiring specific knowledge about EEG. Because of the long recording time, even subclinical seizure activity can be detected, which would otherwise remain unnoticed4,22 because conventional EEG monitoring for very long time periods is not available to date in NICUs. It must be considered, though, that not all pathological alterations, like seizures, are found due to the small area of brain surface covered by the recording13. Therefore, aEEG is not meant to replace conventional EEG, but to complement it13.

Electrocortical activity, as reflected by the aEEG background pattern, changes according to the infant's gestational age4,23,24,25. In term infants and late preterm infants, the background pattern is mainly continuous with a lower amplitude above 5 µV4. During quiet sleep, the background pattern becomes more discontinuous26. In very preterm infants, the dominating background pattern is discontinuous: episodes of high activity (i.e., high-amplitude bursts) alternate with episodes of low-amplitude activity27. This physiological pattern must be distinguished from a burst suppression pattern, which is pathological27. With increasing gestational age, aEEG and background patterns become more continuous, and the duration of continuous activity increases27,28,29. Developing and existing pathological conditions can also be displayed by the aEEG tracing (e.g., the development of an intraventricular hemorrhage and periventricular leukomalacia is associated with acute disturbances in the background activity)30,31. Severe meningitis can cause a flat trace.

The qualitative interpretation of aEEG generally includes three categories: classification of the background pattern, sleep-wake cycling, and the presence of seizures. Several authors have made suggestions for classifications and scores that describe brain maturation16,21,24,25. The quantitative analysis of aEEG is less common, even though it is possible in modern devices, and few research groups made use of this approach32,33,34. We would like to briefly present 3 different approaches to the qualitative and semi-quantitative assessment of aEEG tracings:

Hellström-Westas:21

The assessment of the tracing is solely qualitative, and the results are not transformed into a score. The classification allows for the description of pathological conditions. Normative values for gestational ages have been published to help interpret whether a pattern is adequate for the age21: (1) background patterns: continuous normal voltage (physiological), discontinuous normal voltage (physiological in preterm infants), burst suppression pattern (pathological), continuous low voltage (pathological), and flat trace (pathological); (2) sleep-wake cycling: none, imminent, mature (physiological/pathological, depending on the infant's age); and (3) seizure activity: none, single seizures, repetitive seizures, and status epilepticus.

Burdjalov:25

The approach of this classification is the qualitative assessment of the tracing and its transformation into a score. The score rises with gestational age, and normative score values for each corresponding gestational age have been published: (1) 0 - 2 points for continuity, (2) 0 - 5 points for sleep-wake cycling, (3) 0 - 2 points for the amplitude of the lower border, (4) 0 - 4 points for the bandwidth, and (5) 0 - 13 points for the total score.

Olischar/Klebermass:16,24

Percentiles regarding the percent duration of background patterns (i.e., discontinuous normal voltage, discontinuous low voltage, and continuous normal voltage) and burst rate were developed for gestational ages. Tracings are evaluated for an age-adequate background pattern, the presence of sleep-wake cycling, and the presence of seizure activity (i.e., repetitive seizures or status epilepticus). Then, the tracings are classified into a graded score as follows: (1) normal aEEG (all categories normal), (2) moderately abnormal (1 out of 3 categories classified as abnormal), and (3) severely abnormal (2 or 3 out of 3 categories classified as abnormal). This score has been shown to have a predictive value for neurodevelopmental outcome at 3 years of corrected age.

Changes in the aEEG tracing are caused by numerous extracortical factors, such as changes in cerebral blood flow, medication (e.g., opiates, sedatives, and caffeine), acidosis, changes in carbon dioxide tension, clinical conditions (e.g., hypogylcemia, sepsis, meningitis, and patent ductus arteriosus), etc.21,32,35,36,37,38. The aEEG band itself is rather insensitive to changes of impedance, but significant changes are observed in terms of electrode distance and localization39. Artifacts may pose a problem for interpretation: absolute values of the amplitude change as a consequence of scalp edema or interelectrode distance39,40. Interferences caused by ECG, high-frequency oscillation ventilation, muscle activity, infant movement, or handling may result in a lower border increase40. In modern devices, this can be partially avoided by the simultaneous recording of raw EEG and aEEG and by marking the beginning and end of handling. Liquids (e.g., sweat or ultrasound gel) can lead to connections between electrodes, feigning a flat trace pattern. Approximately 12% of the recording time in long-term aEEGs are altered due to artifacts, 55% caused by electrical interferences and 45% being movement artifacts41.

Protocol

aEEGs are conducted as part of the clinical routine in our hospital. The presented protocol follows the guidelines of the institution's human research ethics committee. Written informed consent regarding the filming and the publication of the material was collected from both parents of all infants that appear in the video.

1. Gather the Needed Supplies

  1. Connect the eEEG device to electric power in the place where the monitoring will take place and plug the module box to the aEEG device.   
  2. Ensure that there are four electrodes for a two-channel aEEG and two electrodes for a single-channel aEEG. Choose either needle electrodes, gold cups, or hydrogel electrodes. Additionally, have one hydrogel electrode ready to serve as a reference electrode.
    NOTE: Gold cups can be disinfected and reused for up to two years. Needle and hydrogel electrodes are single-use only. Needle electrodes can be used in infants at 23 weeks of gestation without causing skin lesions or infections. Here, the best results were achieved using needle electrodes in older infants as well.
  3. Prepare the following supplies: a positioning strip provided by the manufacturer (to help place the electrodes correctly), tape appropriate for use in neonates (e.g., viscose mull), skin disinfectant appropriate for use in neonates (e.g., alcohol-based or octenidinhydrocholoride-based), swabs, skin preparation gel, a module box, and contact gel for gold cups.
    NOTE: Once disconnected from electrical power, the device will shut down, and the recording will need to be restarted. Some devices have internal batteries, however, and can be moved after being switched on or during the recording.

2. Apply the Electrodes

  1. Respecting the principals of minimal handling42,43,44, apply the electrodes during routine care or delivery room care. Wear (unsterile) gloves, a gown, a hood, and a mask, according to the institution's guidelines and the patient's infectious status.
  2. Prepare the skin for the reference electrode, as follows:
    1. Disinfect the skin. Place skin preparation gel on a cotton swab until it is moist. Apply a few gentle strokes with the cotton swab, using very little pressure. Be very cautious in extremely premature infants between 23 and 25 weeks of gestation to avoid causing lesions on the immature skin.
    2. Place the reference electrode on the back or the chest of the infant.
    3. Place the measuring device on the infant's head, and line up the same letters/signs on the infant's tragus and sagittal suture; the two arrows indicate where to place the electrodes (positions C3, P3, C4, and P4 of the 10 - 20 system).
  3. Place the electrodes on the infant's head, following the instructions, below, corresponding to the type of electrodes selected.
  4. Needle electrodes.
    1. Disinfect the area indicated by the measuring device.
    2. Stretch the skin slightly and insert the needle tangentially, just underneath the skin at the markings, with the tip of the needle pointing in the caudal direction. Use tape to keep the electrode in place.
    3. Repeat the procedure for both/all four electrodes.
      NOTE: Use needle electrodes in very premature infants, as rubbing for skin preparation is not necessary.
  5. Gold cups.
    1. Disinfect the area indicated by the measuring device.
    2. Prepare the skin in the marked areas, as described in step 2.2.
    3. Fill each cup with contact gel. Place the cup in the proper position, with the cable running towards the head end; tape it into place.
  6. Hydrogel electrodes.
    1. Disinfect the area indicated by the measuring device.
    2. Prepare the skin in the marked areas, as described in step 2.2.
    3. Place the electrodes with the cable running towards the head end. Fix the electrodes with tape in case they do not stay in place.

3. Connect the Wires to the Monitor

  1. Insert the cables into the module box, as indicated by the legend on the box.
  2. The default starting screen includes impedance monitoring for all electrodes.
  3. Make sure that all electrodes are in place and that there is no mechanical contact between electrodes. If the impedance of one or more electrodes is not satisfactory, remove the corresponding electrode and perform one or two more strokes with the cotton swab, but do not apply more pressure.
  4. Start the recording when everything is set.
    NOTE: Obligatory recording parameters are raw EEG and impedance. According to the device and clinical indication, additional options are burst suppression ratio, sharp transient intensity, and spectral edge frequency, among others. Standard parameters for review include raw EEG, amplitude-integrated EEG curve, and impedance. Depending on the device, there is the opportunity to view more features, like spectral edge frequency, or different forms of the presentation of the lower, mean, and upper border. Additional features are seizure detection and burst rate analysis.

4. Optional: Place a CPAP Hat

  1. If required, place a CPAP hat or head band on top of the aEEG electrodes.

5. Aspects to Keep in Mind During the Recording

  1. Regularly check for impedance and the dislocation of electrodes to obtain quality recordings. Also, check the infant for skin irritation to avoid lesions or infections.
  2. Mark events (e.g., handling, kangaroo care (skin-to-skin care), apnea with bradycardia, intubation, and administration of sedatives or opioids) to facilitate the identification of artifacts using the provided button on the screen of the cerebral function monitor. Leave aEEG electrodes in place during kangaroo care to resume the recording afterwards.
  3. Leave aEEG electrodes in place for intubation or other invasive measures to resume the recording later. In case the cables are not long enough to move the infant within the incubator, disconnect them from the module box and reconnect them after the procedure.

6. Review of the aEEG Tracing and Storing

  1. Review the tracing at the end of the recoding on the monitor or transfer it to an external storage device.

Representative Results

Figure 2 shows a typical view of an aEEG monitor. Continuous and discontinuous normal voltage patterns are considered physiological background patterns in term and preterm infants, respectively (Figure 3 and Figure 4). A burst suppression pattern, continuous low-voltage pattern, and a flat trace are pathological background patterns (Figure 5, Figure 6, Figure 7).

Seizures in term infants have a characteristic shape, with a sudden rise of both the lower and upper border (Figure 8). In preterm infants, however, seizures can be camouflaged by the discontinuous pattern and may only be detected by viewing the raw EEG (Figure 9).

Liquid bridging can cause an apparent flat trace (Figure 10). Usually, this happens in the two-channel aEEG (intraparietal curves). If the cross-cerebral aEEG is physiological, while the intraparietal curve shows a flat trace, the electrodes should be checked for liquids. Electrical interference, movements, and handling can lead to an apparent seizure or even to an apparent status epilepticus. If this happens, impedance and the reference electrode should be checked, and the raw EEG should be viewed (Figure 11). Another reason for an elevation of both the lower and upper border is the displacement of the reference electrode.

Figure 1
Figure 1. Formation of the aEEG Tracing.
The signal from the raw EEG (upper curve) is processed, resulting in the amplitude-integrated EEG band (lower curve). High amplitudes form the upper border, whereas low amplitudes form the lower border. While strong variation in the height of the amplitude leads to a broad aEEG band, the aEEG band is narrow if there is little variation in the height of amplitude. The scale of the y-axis is linear up to 10 µV and logarithmic above 10 µV. Please click here to view a larger version of this figure.

Figure 2
Figure 2. Typical Display of an aEEG Monitor.
The upper half of the monitor displays the raw EEG curve (the displayed section equals 10 s). On the left display, the lower half shows the unilateral aEEG tracing (the displayed section equals approximately 3 h). On the right display, the corresponding cross-cerebral tracing is shown. The cursor indicates the section of the amplitude-integrated tracing from the raw EEG. Please click here to view a larger version of this figure.

Figure 3
Figure 3. Continuous Normal Voltage Pattern.
Continuous background pattern with sleep-wake cycling. The x-axis equals time (one square = 10 min). Please click here to view a larger version of this figure.

Figure 4
Figure 4. Discontinuous Normal Voltage Pattern.
Discontinuous background pattern with imminent sleep-wake cycling. The x-axis equals time (one square = 10 min). Please click here to view a larger version of this figure.

Figure 5
Figure 5. Burst Suppression Pattern.
Burst suppression pattern, with the lower amplitude continuously low and without alteration. The x-axis equals time (one square = 10 min).
From Bruns, N. Amplituden-integriertes EEG bei extrem unreifen Frühgeborenen in den ersten 4 Lebenswochen. http://www.diss.fu-berlin.de/diss/receive/FUDISS_thesis_000000036576 (2012). Please click here to view a larger version of this figure.

Figure 6
Figure 6. Flat Trace.
Flat trace on both sides in a term infant with severe meningoencephalitis. The x-axis equals time (one square = 10 min).
From Bruns, N. Amplituden-integriertes EEG bei extrem unreifen Frühgeborenen in den ersten 4 Lebenswochen. http://www.diss.fu-berlin.de/diss/receive/FUDISS_thesis_000000036576 (2012). Please click here to view a larger version of this figure.

Figure 7
Figure 7. Continuous Low-voltage Pattern.
Continuous low-voltage pattern without sleep-wake cycling. The x-axis equals time (one square = 10 min).
From Bruns, N. Amplituden-integriertes EEG bei extrem unreifen Frühgeborenen in den ersten 4 Lebenswochen. http://www.diss.fu-berlin.de/diss/receive/FUDISS_thesis_000000036576 (2012). Please click here to view a larger version of this figure.

Figure 8
Figure 8. Seizures in Term Infants.
Typical depiction of a seizure in the aEEG: a sudden rise of the lower and upper margin is followed by a short period of decreased activity. Repetitive seizures for approximately 3.5 h. The x-axis equals time (one square = 10 min).
From Bruns, N. Amplituden-integriertes EEG bei extrem unreifen Frühgeborenen in den ersten 4 Lebenswochen. http://www.diss.fu-berlin.de/diss/receive/FUDISS_thesis_000000036576 (2012). Please click here to view a larger version of this figure.

Figure 9
Figure 9. Seizures in Preterm Infants.
Without the raw EEG, the hypersynchronous activity in both hemispheres would remain undetected. The x-axis equals time (one square = 10 min).
From Bruns, N. Amplituden-integriertes EEG bei extrem unreifen Frühgeborenen in den ersten 4 Lebenswochen. http://www.diss.fu-berlin.de/diss/receive/FUDISS_thesis_000000036576 (2012). Please click here to view a larger version of this figure.

Figure 10
Figure 10. Apparent Flat Trace.
In the unilateral tracings, there appears to be a pathological flat trace pattern in an infant without cerebral injury. The cross-cerebral tracing shows a physiological discontinuous background pattern with short sections of continuous activity. In this case, the flat trace is an artifact caused by liquid bridging between electrodes (especially hydrogel electrodes). The x-axis equals time (one square = 10 min).
From Bruns, N. Amplituden-integriertes EEG bei extrem unreifen Frühgeborenen in den ersten 4 Lebenswochen. http://www.diss.fu-berlin.de/diss/receive/FUDISS_thesis_000000036576 (2012). Please click here to view a larger version of this figure.

Figure 11
Figure 11. Apparent Seizures.
This image shows high-frequency activity over a long period of time. Without seeing the raw EEG curve, status epilepticus is indicated. This artifact is caused by muscle activity. The x-axis equals time (one square = 10 min).
From Bruns, N. Amplituden-integriertes EEG bei extrem unreifen Frühgeborenen in den ersten 4 Lebenswochen. http://www.diss.fu-berlin.de/diss/receive/FUDISS_thesis_000000036576 (2012). Please click here to view a larger version of this figure.

Discussion

The cerebral function monitor is an easily accessible and increasingly common device used to record amplitude-integrated EEGs in NICUs13. In routine care, the application of an aEEG takes 3 - 5 min.

Critical steps within this protocol are the correct placement of the electrodes on the head and the connection of the cables to the corresponding plug of the module box. Electrode placement needs to be preceded by thorough skin disinfection and preparation, especially for the reference electrode. In our experience, the best-quality recordings are achieved when the reference electrode is placed on the back of the infant. For troubleshooting, electrodes should be checked for dislocation in case of high impedance. If electrode dislocation has been disclosed and repetition of skin preparation fails, the electrode may need to be replaced. In case of an offset of the upper border, the reference electrode needs to be optimized. A high frequency and high amplitude of both the raw EEG and amplitude-integrated EEG are caused by muscle activity or interferences (e.g., high-frequency oscillation ventilation). This part of the tracing cannot be used for interpretation. If a flat trace occurs in a cerebrally healthy infant, the cross-cerebral tracing should be viewed. If this is normal, it is likely that liquids like sweat or ultrasound gel have caused bridging between two electrodes. In case of persisting problems, manufacturers have contact persons that will help to determine a solution and will even come to the NICU to check for underlying causes. In our experience, needle electrodes are the recommended type of electrodes in very premature infants. After thorough disinfection and gentle skin preparation of the reference electrode, we did not observe a significant number of infections, serious skin lesions, or bleeding events since the beginning of the large-scale use of this technique in our center in 2008 (an average of 60 - 80 very low birth weight infants per year, 1 - 5 recordings per infant). Since 2014, we have only used needle electrodes in all neonates, as we achieve the best results using this type of electrode.

Modifications of the aEEG are not commonly performed, but electrodes could be placed in any position on the head to acquire a desired tracing (from the international 10 - 20 system). In some cases, the electrode position may need to be adjusted (e.g., due to skin lacerations after vacuum extraction or cephalohematoma)45. For classification according to amplitudes, it is important to maintain a standard interelectrode distance, as a reduction in interelectrode distance results in a reduction of amplitude39,45. In case of extreme head sizes, such extremely premature infants (i.e., 23 - 24 weeks of gestation) or term infants with augmented head circumference due to hydrocephalus, the importance of the interelectrode distance for interpretation should be kept in mind. Another modification of the traditional aEEG is the continuously monitored limited-channel EEG18,46,47. The raw EEG curve derived from the cerebral function monitor can be assessed like a conventional EEG curve. In our center, we use this approach to answer special problems regarding newborn neuropediatric patients, in close collaboration with our pediatric neurologists.

The main limitation of the aEEG is the fact that only a small area of the brain surface is covered by the tracing. Thus, alterations of electrocortical activity in different areas of the brain surface may remain unnoticed13. Due to the time compression, short-lasting changes of cerebral activity are difficult to detect without using the raw EEG curve. Further interpretation of the raw EEG curve requires knowledge about the conventional EEG or close cooperation with neurophysiologists or pediatric neurologists. Last, but not least, there are several external and internal factors that cause alterations in the aEEG band that must be kept in mind when interpreting the tracing.

Nonetheless, the aEEG offers the possibility for continuous cerebral function monitoring in neonates. It is easily accessible, and interpretation is not difficult. As it contains less information than a conventional EEG, it cannot replace this technique. Rather, it complements the existing means for cerebral diagnostics, such as EEG, ultrasound, and magnetic resonance imaging. There is good evidence for the prediction of outcome after birth asphyxia in term infants, and the aEEG has been established as a tool to identify infants who will benefit from cooling8. In preterm infants, there is also good evidence that long-term neurological outcome can be predicted by early aEEG recordings15,16,17,18,19,20. However, to date, this knowledge does not result in consequences for clinical decision making in this population of infants. For the future, it is likely that cerebral function monitoring will become a standard tool in NICUs, as well as in secondary centers and pediatric intensive care units.

Disclosures

The authors have nothing to disclose.

Acknowledgments

We thank our nurses for their support and contributions to the making of the video.

Materials

Name Company Catalog Number Comments
disposable subdermal needle electrodes Technomed TE/S43-438
Genuine Grass Gold Disk Electrodes Natus FE5GH-03
neonatal hydrogel sensors Natus CZA00037
positioning strips Natus OBM00047
skin markers Natus CZN00011
Nu Prep skin prepping gel Weaver and Company 10-30
contact gel Ten 20 Weaver and Company 10-20-4T
skin disinfectant
tape
cotton swab
Braintrend® aEEG Monitor aEEG Monitor

DOWNLOAD MATERIALS LIST

References

  1. Maynard, D., Prior, P. F., Scott, D. F. Device for continuous monitoring of cerebral activity in resuscitated patients. Br Med J. 4 (5682), 545-546 (1969).
  2. Greisen, G., Hellström-Westas, L., Lou, H., Rosén, I., Svenningsen, N. W. EEG depression and germinal layer haemorrhage in the newborn. Acta Paediatr Scand. 76 (3), 519-525 (1987).
  3. Greisen, G., Pryds, O., Rosén, I., Lou, H. Poor reversibility of EEG abnormality in hypotensive, preterm neonates. Acta Paed Scand. 77 (6), 785-790 (1988).
  4. Hellström-Westas, L., Rosén, I., Svenningsen, N. W. Predictive value of early continuous amplitude integrated EEG recordings on outcome after severe birth asphyxia in full term infants. Arch Dis Child Fetal Neonatal Ed. 72 (1), F34-F38 (1995).
  5. Shellhaas, R. A., Kushwaha, J. S., Plegue, M. A., Selewski, D. T., Barks, J. D. E. An evaluation of cerebral and systemic predictors of 18-month outcomes for neonates with hypoxic ischemic encephalopathy. J Child Neurol. 30 (11), 1526-1531 (2015).
  6. Gluckman, P. D., et al. Selective head cooling with mild systemic hypothermia after neonatal encephalopathy: multicentre randomised trial. Lancet. 365 (9460), 663-670 (2005).
  7. Hellström-Westas, L., Rosén, I. Continuous brain-function monitoring: state of the art in clinical practice. Semin Fetal Neonatal Med. 11 (6), 503-511 (2006).
  8. Rosén, I. The physiological basis for continuous electroencephalogram monitoring in the neonate. Clin Perinatol. 33 (3), 593-611 (2006).
  9. Davis, A. S., et al. Serial aEEG recordings in a cohort of extremely preterm infants: feasibility and safety. J Perinatol. 35 (5), 373-378 (2015).
  10. Benavente-Fernández, I., Lubián-López, S. P., Jiménez-Gómez, G., Lechuga-Sancho, A. M., Garcia-Alloza, M. Low-voltage pattern and absence of sleep-wake cycles are associated with severe hemorrhage and death in very preterm infants. Eur J Pediatr. 174 (1), 85-90 (2015).
  11. Klebermass, K., et al. Amplitude-integrated EEG pattern predicts further outcome in preterm infants. Pediatr Res. 70 (1), 102-108 (2011).
  12. Soubasi, V., et al. Early abnormal amplitude-integrated electroencephalography (aEEG) is associated with adverse short-term outcome in premature infants. Eur J Paediatr Neurol. 16 (6), 625-630 (2012).
  13. Wikström, S., et al. Early single-channel aEEG/EEG predicts outcome in very preterm infants. Acta Paediatr. 101 (7), 719-726 (2012).
  14. Welch, C., Helderman, J., Williamson, E., O'Shea, T. M. Brain wave maturation and neurodevelopmental outcome in extremely low gestational age neonates. J Perinatol. 33 (11), 867-871 (2013).
  15. Bruns, N., et al. Comparison of two common aEEG classifications for the prediction of neurodevelopmental outcome in preterm infants. Eur J Pediatr. 176 (2), 1-9 (2016).
  16. Eken, P., Toet, M. C., Groenendaal, F., de Vries, L. S. Predictive value of early neuroimaging, pulsed Doppler and neurophysiology in full term infants with hypoxic-ischaemic encephalopathy. Arch Dis Child Fetal Neonatal Ed. 73 (2), F75-F80 (1995).
  17. Shalak, L. F., Laptook, A. R., Velaphi, S. C., Perlman, J. M. Amplitude-integrated electroencephalography coupled with an early neurologic examination enhances prediction of term infants at risk for persistent encephalopathy. Pediatrics. 111 (2), 351-357 (2003).
  18. Marics, G., et al. Prevalence and etiology of false normal aEEG recordings in neonatal hypoxic-ischaemic encephalopathy. BMC Pediatr. 13 (1), 194 (2013).
  19. Azzopardi, D. V., et al. Moderate hypothermia to treat perinatal asphyxial encephalopathy. N Engl J Med. 361 (14), 1349-1358 (2009).
  20. Azzopardi, D. TOBY study group. Predictive value of the amplitude integrated EEG in infants with hypoxic ischaemic encephalopathy: data from a randomised trial of therapeutic hypothermia. Arch Dis Child Fetal Neonatal Ed. 99 (1), F80-F82 (2014).
  21. Hellström-Westas, L., Rosén, I., de Vries, L. S., Greisen, G. Amplitude-integrated EEG Classification and Interpretation in Preterm and Term Infants. NeoReviews. 7 (2), e76-e87 (2006).
  22. Shah, D. K., et al. Accuracy of bedside electroencephalographic monitoring in comparison with simultaneous continuous conventional electroencephalography for seizure detection in term infants. Pediatrics. 121 (6), 1146-1154 (2008).
  23. Sisman, J., Campbell, D. E., Brion, L. P. Amplitude-integrated EEG in preterm infants: maturation of background pattern and amplitude voltage with postmenstrual age and gestational age. J Perinatol. 25 (6), 391-396 (2005).
  24. Olischar, M., et al. Reference values for amplitude-integrated electroencephalographic activity in preterm infants younger than 30 weeks' gestational age. Pediatrics. 113 (1 Pt 1), e61-e66 (2004).
  25. Burdjalov, V. F., Baumgart, S., Spitzer, A. R. Cerebral function monitoring: a new scoring system for the evaluation of brain maturation in neonates. Pediatrics. 112 (4), 855-861 (2003).
  26. Viniker, D. A., Maynard, D. E., Scott, D. F. Cerebral function monitor studies in neonates. Clin Electroencephalogr. 15 (4), 185-192 (1984).
  27. Hellström-Westas, L. Continuous electroencephalography monitoring of the preterm infant. Clin Perinatol. 33 (3), 633-647 (2006).
  28. Hayakawa, M. Background electroencephalographic (EEG) activities of very preterm infants born at less than 27 weeks gestation: a study on the degree of continuity. Arch Dis Child Fetal Neonatal Ed. 84 (3), 163 (2001).
  29. Vecchierini, M. F., d'Allest, A. M., Verpillat, P. EEG patterns in 10 extreme premature neonates with normal neurological outcome: qualitative and quantitative data. Brain Dev. 25 (5), 330-337 (2003).
  30. Hellström-Westas, L., Rosén, I., Svenningsen, N. Cerebral Function Monitoring During the First Week of Life in Extremely Small Low Birthweight (ESLBW) Infants. Neuropediatrics. 22 (01), 27-32 (1991).
  31. Connell, J., et al. Continuous four-channel EEG monitoring in the evaluation of echodense ultrasound lesions and cystic leucomalacia. Arch Dis Child. 62 (10), 1019-1024 (1987).
  32. Bruns, N., Metze, B., Bührer, C., Felderhoff-Müser, U., Hüseman, D. Electrocortical Activity at 7 Days of Life is Affected in Extremely Premature Infants with Patent Ductus Arteriosus. Klin Padiatr. 227 (5), 264-268 (2015).
  33. Thorngate, L., Foreman, S. W., Thomas, K. A. Quantification of neonatal amplitude-integrated EEG patterns. Early Hum Dev. 89 (12), 931-937 (2013).
  34. West, C. R., Harding, J. E., Williams, C. E., Gunning, M. I., Battin, M. R. Quantitative electroencephalographic patterns in normal preterm infants over the first week after birth. Early Hum Dev. 82 (1), 43-51 (2006).
  35. ter Horst, H. J., van Olffen, M., Remmelts, H. J., de Vries, H., Bos, A. F. The prognostic value of amplitude integrated EEG in neonatal sepsis and/or meningitis. Acta Paediatr. 99 (2), 194-200 (2010).
  36. Eaton, D. G., Wertheim, D., Oozeer, R., Dubowitz, L. M., Dubowitz, V. Reversible changes in cerebral activity associated with acidosis in preterm neonates. Acta Paediatr. 83 (5), 486-492 (1994).
  37. Victor, S., Appleton, R. E., Beirne, M., Marson, A. G., Weindling, A. M. Effect of carbon dioxide on background cerebral electrical activity and fractional oxygen extraction in very low birth weight infants just after birth. Pediatr Res. 58 (3), 579-585 (2005).
  38. West, C. R., et al. Early low cardiac output is associated with compromised electroencephalographic activity in very preterm infants. Pediatr Res. 59 (4 Pt 1), 610-615 (2006).
  39. Quigg, M., Leiner, D. Engineering aspects of the quantified amplitude-integrated electroencephalogram in neonatal cerebral monitoring. J Clin Neurophysiol. 26 (3), 145-149 (2009).
  40. Toet, M. C., Lemmers, P. M. A. Brain monitoring in neonates. Early Hum Dev. 85 (2), 77-84 (2009).
  41. Hagmann, C. F., Robertson, N. J., Azzopardi, D. Artifacts on electroencephalograms may influence the amplitude-integrated EEG classification: a qualitative analysis in neonatal encephalopathy. Pediatrics. 118 (6), 2552-2554 (2006).
  42. Als, H., et al. Individualized developmental care for the very low-birth-weight preterm infant. Medical and neurofunctional effects. JAMA. 272 (11), 853-858 (1994).
  43. Jacobsen, T., Grønvall, J., Petersen, S., Andersen, G. E. "Minitouch" treatment of very low-birth-weight infants. Acta Paediatr. 82 (11), 934-938 (1993).
  44. Vandenberg, K. A. Individualized developmental care for high risk newborns in the NICU: a practice guideline. Early Hum Dev. 83 (7), 433-442 (2007).
  45. Hellström-Westas, L., de Vries, L. S., Rosén, I. An Atlas of Amplitude-Integrated EEGs in the Newborn. , 2nd ed, Informa Health Care. London. (2008).
  46. Iyer, K. K., et al. Early Detection of Preterm Intraventricular Hemorrhage from Clinical Electroencephalography. Crit Care Med. 43 (10), 2219-2227 (2015).
  47. Hellström-Westas, L., Rosén, I. Electroencephalography and brain damage in preterm infants. Early Hum Dev. 81 (3), 255-261 (2005).

Tags

Keywords: AEEG Cerebral Function Monitor Neonates Preterm Infants Term Infants Neonatal Intensive Care Unit Bedside Monitoring Single-channel AEEG Two-channel AEEG Electrodes Reference Electrode Skin Preparation Impedance Gold Cup Electrodes Hydrogel Electrodes Cable Placement Recording
Application of an Amplitude-integrated EEG Monitor (Cerebral Function Monitor) to Neonates
Play Video
PDF DOI DOWNLOAD MATERIALS LIST

Cite this Article

Bruns, N., Blumenthal, S., Meyer,More

Bruns, N., Blumenthal, S., Meyer, I., Klose-Verschuur, S., Felderhoff-Müser, U., Müller, H. Application of an Amplitude-integrated EEG Monitor (Cerebral Function Monitor) to Neonates. J. Vis. Exp. (127), e55985, doi:10.3791/55985 (2017).

Less
Copy Citation Download Citation Reprints and Permissions
View Video

PLAYLIST

  • Research • Medicine
    Estimation of Urinary Nanocrystals in Humans using Calcium Fluorophore Labeling and Nanoparticle Tracking Analysis
  • Research • Medicine
    Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
  • Research • Medicine
    Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation
  • Research • Medicine
    Digital Handwriting Analysis of Characters in Chinese Patients with Mild Cognitive Impairment
  • Research • Medicine
    Segmentation and Linear Measurement for Body Composition Analysis using Slice-O-Matic and Horos
  • Research • Medicine
    Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
  • Research • Medicine
    Real-Time Magnetic Resonance Guided Focused Ultrasound for Painful Bone Metastases
  • Research • Medicine
    Isolation of Viable Adipocytes and Stromal Vascular Fraction from Human Visceral Adipose Tissue Suitable for RNA Analysis and Macrophage Phenotyping
  • Research • Medicine
    Obtaining Quality Extended Field-of-View Ultrasound Images of Skeletal Muscle to Measure Muscle Fascicle Length
  • Research • Medicine
    Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
  • Research • Medicine
    Electroretinogram Recording for Infants and Children under Anesthesia to Achieve Optimal Dark Adaptation and International Standards
  • Research • Medicine
    Measurement of Tissue Oxygenation Using Near-Infrared Spectroscopy in Patients Undergoing Hemodialysis
  • Research • Medicine
    Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
  • Research • Medicine
    Simultaneous Laryngopharyngeal and Conventional Esophageal pH Monitoring
  • Research • Medicine
    Real-Time Monitoring of Neurocritical Patients with Diffuse Optical Spectroscopies
  • Research • Neuroscience
    Evaluating Postural Control and Lower-extremity Muscle Activation in Individuals with Chronic Ankle Instability
  • Research • Medicine
    Assessment of Dependence in Activities of Daily Living Among Older Patients in an Acute Care Unit
  • Research • Medicine
    Validated LC-MS/MS Panel for Quantifying 11 Drug-Resistant TB Medications in Small Hair Samples
  • Research • Medicine
    International Expert Consensus and Recommendations for Neonatal Pneumothorax Ultrasound Diagnosis and Ultrasound-guided Thoracentesis Procedure
  • Research • Biology
    A Finite Element Approach for Locating the Center of Resistance of Maxillary Teeth
  • Research • Medicine
    Lower Limb Biomechanical Analysis of Healthy Participants
  • Research • Neuroscience
    Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
  • Research • Medicine
    Oral Health Assessment by Lay Personnel for Older Adults
  • Research • Medicine
    Determining and Controlling External Power Output During Regular Handrim Wheelchair Propulsion
  • Research • Medicine
    A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
  • Research • Medicine
    Measurement of Carotenoids in Perifovea using the Macular Pigment Reflectometer
  • Research • Medicine
    Assessment of Static Graviceptive Perception in the Roll-Plane using the Subjective Visual Vertical Paradigm
  • Research • Medicine
    Learning Modern Laryngeal Surgery in a Dissection Laboratory
  • Research • Medicine
    DIPLOMA Approach for Standardized Pathology Assessment of Distal Pancreatectomy Specimens
  • Research • Medicine
    A Computerized Functional Skills Assessment and Training Program Targeting Technology Based Everyday Functional Skills
  • Research • Medicine
    Imaging Features of Systemic Sclerosis-Associated Interstitial Lung Disease
  • Research • Medicine
    Integrating Augmented Reality Tools in Breast Cancer Related Lymphedema Prognostication and Diagnosis
  • Research • Medicine
    Ultrasonographic Assessment During Cardiopulmonary Resuscitation
  • Research • Medicine
    Measurement of the Hepatic Venous Pressure Gradient and Transjugular Liver Biopsy
  • Research • Medicine
    Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
  • Research • Medicine
    Traditional Trail Making Test Modified into Brand-new Assessment Tools: Digital and Walking Trail Making Test
  • Research • Medicine
    Use of Magnetic Resonance Imaging and Biopsy Data to Guide Sampling Procedures for Prostate Cancer Biobanking
  • Research • Medicine
    A Fluorescence-based Assay for Characterization and Quantification of Lipid Droplet Formation in Human Intestinal Organoids
  • Research • Medicine
    A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
  • Research • Medicine
    Quantitative Mapping of Specific Ventilation in the Human Lung using Proton Magnetic Resonance Imaging and Oxygen as a Contrast Agent
  • Research • Neuroscience
    Portable Thermographic Screening for Detection of Acute Wallenberg's Syndrome
  • Research • Medicine
    Use of MRI-ultrasound Fusion to Achieve Targeted Prostate Biopsy
  • Research • Medicine
    Testing of all Six Semicircular Canals with Video Head Impulse Test Systems
  • Research • Medicine
    Protocol and Guidelines for Point-of-Care Lung Ultrasound in Diagnosing Neonatal Pulmonary Diseases Based on International Expert Consensus
  • Research • Neuroscience
    Bilateral Assessment of the Corticospinal Pathways of the Ankle Muscles Using Navigated Transcranial Magnetic Stimulation
  • Research • Medicine
    Targeting Gray Rami Communicantes in Selective Chemical Lumbar Sympathectomy
  • Research • Medicine
    Multi-Modal Home Sleep Monitoring in Older Adults
  • Research • Medicine
    Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
  • Research • Medicine
    Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
  • Research • Medicine
    Measurements of Motor Function and Other Clinical Outcome Parameters in Ambulant Children with Duchenne Muscular Dystrophy
  • Research • Medicine
    Assessment of the Efficacy of An Osteopathic Treatment in Infants with Biomechanical Impairments to Suckling
  • Research • Medicine
    Quantification of Levator Ani Hiatus Enlargement by Magnetic Resonance Imaging in Males and Females with Pelvic Organ Prolapse
  • Research • Medicine
    Quantitative [18F]-Naf-PET-MRI Analysis for the Evaluation of Dynamic Bone Turnover in a Patient with Facetogenic Low Back Pain
  • Research • Medicine
    Generation of Human 3D Lung Tissue Cultures (3D-LTCs) for Disease Modeling
  • Research • Medicine
    Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
  • Research • Medicine
    Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
  • Research • Medicine
    Two-Dimensional X-Ray Angiography to Examine Fine Vascular Structure Using a Silicone Rubber Injection Compound
  • Research • Medicine
    Preparation, Procedures and Evaluation of Platelet-Rich Plasma Injection in the Treatment of Knee Osteoarthritis
  • Research • Medicine
    Cardiac Magnetic Resonance Imaging at 7 Tesla
  • Research • Medicine
    Semi-quantitative Assessment Using [18F]FDG Tracer in Patients with Severe Brain Injury
  • Research • Medicine
    Handheld Metal Detector Screening for Metallic Foreign Body Ingestion in Children
  • Research • Medicine
    Conducting Maximal and Submaximal Endurance Exercise Testing to Measure Physiological and Biological Responses to Acute Exercise in Humans
  • Research • Medicine
    A Metadata Extraction Approach for Clinical Case Reports to Enable Advanced Understanding of Biomedical Concepts
  • Research • Medicine
    Autonomic Function Following Concussion in Youth Athletes: An Exploration of Heart Rate Variability Using 24-hour Recording Methodology
  • Research • Medicine
    Hydra, a Computer-Based Platform for Aiding Clinicians in Cardiovascular Analysis and Diagnosis
  • Research • Medicine
    Objective Nociceptive Assessment in Ventilated ICU Patients: A Feasibility Study Using Pupillometry and the Nociceptive Flexion Reflex
  • Research • Medicine
    'Boden Food Plate': Novel Interactive Web-based Method for the Assessment of Dietary Intake
  • Research • Medicine
    Anogenital Distance and Perineal Measurements of the Pelvic Organ Prolapse (POP) Quantification System
  • Research • Medicine
    Bedside Ultrasound for Guiding Fluid Removal in Patients with Pulmonary Edema: The Reverse-FALLS Protocol
  • Research • Medicine
    Muscle Imbalances: Testing and Training Functional Eccentric Hamstring Strength in Athletic Populations
  • Research • Medicine
    Isolation of Primary Human Decidual Cells from the Fetal Membranes of Term Placentae
  • Research • Medicine
    Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
  • Research • Medicine
    Collecting Hair Samples for Hair Cortisol Analysis in African Americans
  • Research • Medicine
    In Vivo Morphometric Analysis of Human Cranial Nerves Using Magnetic Resonance Imaging in Menière's Disease Ears and Normal Hearing Ears
  • Research • Medicine
    Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
  • Research • Medicine
    Standardized Measurement of Nasal Membrane Transepithelial Potential Difference (NPD)
  • Research • Medicine
    Taste Exam: A Brief and Validated Test
  • Research • Medicine
    Absorption of Nasal and Bronchial Fluids: Precision Sampling of the Human Respiratory Mucosa and Laboratory Processing of Samples
  • Research • Medicine
    Methodology for Sputum Induction and Laboratory Processing
  • Research • Medicine
    Electrophysiological Measurement of Noxious-evoked Brain Activity in Neonates Using a Flat-tip Probe Coupled to Electroencephalography
  • Research • Medicine
    A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
  • Research • Medicine
    Oral Biofilm Sampling for Microbiome Analysis in Healthy Children
  • Research • Medicine
    Using Retinal Imaging to Study Dementia
  • Research • Medicine
    Application of an Amplitude-integrated EEG Monitor (Cerebral Function Monitor) to Neonates
  • Research • Medicine
    3D Ultrasound Imaging: Fast and Cost-effective Morphometry of Musculoskeletal Tissue
  • Research • Medicine
    The 4-vessel Sampling Approach to Integrative Studies of Human Placental Physiology In Vivo
  • Research • Medicine
    A Component-resolved Diagnostic Approach for a Study on Grass Pollen Allergens in Chinese Southerners with Allergic Rhinitis and/or Asthma
  • Research • Medicine
    A Novel Method: Super-selective Adrenal Venous Sampling
  • Research • Medicine
    A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
  • Research • Medicine
    Non-invasive Assessments of Subjective and Objective Recovery Characteristics Following an Exhaustive Jump Protocol
  • Research • Medicine
    Experimental Protocol of a Three-minute, All-out Arm Crank Exercise Test in Spinal-cord Injured and Able-bodied Individuals
  • Research • Medicine
    Phosphorus-31 Magnetic Resonance Spectroscopy: A Tool for Measuring In Vivo Mitochondrial Oxidative Phosphorylation Capacity in Human Skeletal Muscle
  • Research • Medicine
    Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
  • Research • Medicine
    Assessment of Child Anthropometry in a Large Epidemiologic Study
  • Research • Medicine
    Video Movement Analysis Using Smartphones (ViMAS): A Pilot Study
  • Research • Medicine
    Network Analysis of Foramen Ovale Electrode Recordings in Drug-resistant Temporal Lobe Epilepsy Patients
  • Research • Medicine
    A Model to Simulate Clinically Relevant Hypoxia in Humans
  • Research • Medicine
    Interictal High Frequency Oscillations Detected with Simultaneous Magnetoencephalography and Electroencephalography as Biomarker of Pediatric Epilepsy
  • Research • Medicine
    Induction and Assessment of Exertional Skeletal Muscle Damage in Humans
  • Research • Medicine
    A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device
  • Research • Medicine
    Drug-Induced Sleep Endoscopy (DISE) with Target Controlled Infusion (TCI) and Bispectral Analysis in Obstructive Sleep Apnea
  • Research • Medicine
    Integrated Compensatory Responses in a Human Model of Hemorrhage
  • Research • Medicine
    Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
  • Research • Medicine
    Impression Cytology of the Lid Wiper Area
  • Research • Behavior
    A Protocol of Manual Tests to Measure Sensation and Pain in Humans
  • Research • Medicine
    Unbiased Deep Sequencing of RNA Viruses from Clinical Samples
  • Research • Medicine
    A Choroid Plexus Epithelial Cell-based Model of the Human Blood-Cerebrospinal Fluid Barrier to Study Bacterial Infection from the Basolateral Side
  • Research • Medicine
    Isolation and Profiling of MicroRNA-containing Exosomes from Human Bile
  • Research • Medicine
    Generation of Microtumors Using 3D Human Biogel Culture System and Patient-derived Glioblastoma Cells for Kinomic Profiling and Drug Response Testing
  • Research • Medicine
    Ultrasound Assessment of Endothelial Function: A Technical Guideline of the Flow-mediated Dilation Test
  • Research • Medicine
    Using a Laminating Technique to Perform Confocal Microscopy of the Human Sclera
  • Research • Medicine
    Intravenous Endotoxin Challenge in Healthy Humans: An Experimental Platform to Investigate and Modulate Systemic Inflammation
  • Research • Medicine
    Modeling and Simulations of Olfactory Drug Delivery with Passive and Active Controls of Nasally Inhaled Pharmaceutical Aerosols
  • Research • Medicine
    Exosomal miRNA Analysis in Non-small Cell Lung Cancer (NSCLC) Patients' Plasma Through qPCR: A Feasible Liquid Biopsy Tool
  • Research • Medicine
    A Multimodal Imaging- and Stimulation-based Method of Evaluating Connectivity-related Brain Excitability in Patients with Epilepsy
  • Research • Medicine
    Measuring Cardiac Autonomic Nervous System (ANS) Activity in Toddlers - Resting and Developmental Challenges
  • Research • Medicine
    Using Saccadometry with Deep Brain Stimulation to Study Normal and Pathological Brain Function
  • Research • Medicine
    Quantitative Fundus Autofluorescence for the Evaluation of Retinal Diseases
  • Research • Medicine
    Diagnosis of Musculus Gastrocnemius Tightness - Key Factors for the Clinical Examination
  • Research • Medicine
    Stereo-Electro-Encephalo-Graphy (SEEG) With Robotic Assistance in the Presurgical Evaluation of Medical Refractory Epilepsy: A Technical Note
  • Research • Medicine
    Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
  • Research • Medicine
    Transcutaneous Microcirculatory Imaging in Preterm Neonates
  • Research • Medicine
    Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
  • Research • Medicine
    Design, Fabrication, and Administration of the Hand Active Sensation Test (HASTe)
  • Research • Medicine
    MRI-guided dmPFC-rTMS as a Treatment for Treatment-resistant Major Depressive Disorder
  • Research • Medicine
    Functional Human Liver Preservation and Recovery by Means of Subnormothermic Machine Perfusion
  • Research • Medicine
    A Multicenter MRI Protocol for the Evaluation and Quantification of Deep Vein Thrombosis
  • Research • Medicine
    Determining The Electromyographic Fatigue Threshold Following a Single Visit Exercise Test
  • Research • Medicine
    Use of Electromagnetic Navigational Transthoracic Needle Aspiration (E-TTNA) for Sampling of Lung Nodules
  • Research • Medicine
    Trabecular Meshwork Response to Pressure Elevation in the Living Human Eye
  • Research • Medicine
    In Vivo, Percutaneous, Needle Based, Optical Coherence Tomography of Renal Masses
  • Research • Medicine
    Establishment of Human Epithelial Enteroids and Colonoids from Whole Tissue and Biopsy
  • Research • Medicine
    Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
  • Research • Medicine
    Preparation and Respirometric Assessment of Mitochondria Isolated from Skeletal Muscle Tissue Obtained by Percutaneous Needle Biopsy
  • Research • Medicine
    A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
  • Research • Medicine
    Isolation and Immortalization of Patient-derived Cell Lines from Muscle Biopsy for Disease Modeling
  • Research • Medicine
    State of the Art Cranial Ultrasound Imaging in Neonates
  • Research • Medicine
    Measurement of Dynamic Scapular Kinematics Using an Acromion Marker Cluster to Minimize Skin Movement Artifact
  • Research • Medicine
    The Supraclavicular Fossa Ultrasound View for Central Venous Catheter Placement and Catheter Change Over Guidewire
  • Research • Medicine
    Ultrasound Assessment of Endothelial-Dependent Flow-Mediated Vasodilation of the Brachial Artery in Clinical Research
  • Research • Medicine
    Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
  • Research • Medicine
    A Neuroscientific Approach to the Examination of Concussions in Student-Athletes
  • Research • Medicine
    DTI of the Visual Pathway - White Matter Tracts and Cerebral Lesions
  • Research • Medicine
    Collection, Isolation, and Flow Cytometric Analysis of Human Endocervical Samples
  • Research • Medicine
    Fundus Photography as a Convenient Tool to Study Microvascular Responses to Cardiovascular Disease Risk Factors in Epidemiological Studies
  • Research • Medicine
    A Multi-Modal Approach to Assessing Recovery in Youth Athletes Following Concussion
  • Research • Medicine
    Clinical Assessment of Spatiotemporal Gait Parameters in Patients and Older Adults
  • Research • Medicine
    Multi-electrode Array Recordings of Human Epileptic Postoperative Cortical Tissue
  • Research • Medicine
    Collection and Extraction of Saliva DNA for Next Generation Sequencing
  • Research • Medicine
    Fast and Accurate Exhaled Breath Ammonia Measurement
  • Research • Medicine
    Developing Neuroimaging Phenotypes of the Default Mode Network in PTSD: Integrating the Resting State, Working Memory, and Structural Connectivity
  • Research • Medicine
    Two Methods for Establishing Primary Human Endometrial Stromal Cells from Hysterectomy Specimens
  • Research • Medicine
    Assessment of Vascular Function in Patients With Chronic Kidney Disease
  • Research • Medicine
    Coordinate Mapping of Hyolaryngeal Mechanics in Swallowing
  • Research • Medicine
    Network Analysis of the Default Mode Network Using Functional Connectivity MRI in Temporal Lobe Epilepsy
  • Research • Medicine
    EEG Mu Rhythm in Typical and Atypical Development
  • Research • Medicine
    The Multiple Sclerosis Performance Test (MSPT): An iPad-Based Disability Assessment Tool
  • Research • Medicine
    Isolation and Functional Characterization of Human Ventricular Cardiomyocytes from Fresh Surgical Samples
  • Research • Medicine
    Dynamic Visual Tests to Identify and Quantify Visual Damage and Repair Following Demyelination in Optic Neuritis Patients
  • Research • Medicine
    Primary Culture of Human Vestibular Schwannomas
  • Research • Medicine
    Utility of Dissociated Intrinsic Hand Muscle Atrophy in the Diagnosis of Amyotrophic Lateral Sclerosis
  • Research • Medicine
    Lesion Explorer: A Video-guided, Standardized Protocol for Accurate and Reliable MRI-derived Volumetrics in Alzheimer's Disease and Normal Elderly
  • Research • Medicine
    Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
  • Research • Medicine
    Isolation, Culture, and Imaging of Human Fetal Pancreatic Cell Clusters
  • Research • Medicine
    3D-Neuronavigation In Vivo Through a Patient's Brain During a Spontaneous Migraine Headache
  • Research • Medicine
    A Novel Application of Musculoskeletal Ultrasound Imaging
  • Research • Medicine
    Computerized Dynamic Posturography for Postural Control Assessment in Patients with Intermittent Claudication
  • Research • Medicine
    Collecting Saliva and Measuring Salivary Cortisol and Alpha-amylase in Frail Community Residing Older Adults via Family Caregivers
  • Research • Medicine
    Diffusion Tensor Magnetic Resonance Imaging in the Analysis of Neurodegenerative Diseases
  • Research • Medicine
    Transcriptomic Analysis of Human Retinal Surgical Specimens Using jouRNAl
  • Research • Medicine
    Improved Protocol For Laser Microdissection Of Human Pancreatic Islets From Surgical Specimens
  • Research • Medicine
    Evaluation of Respiratory Muscle Activation Using Respiratory Motor Control Assessment (RMCA) in Individuals with Chronic Spinal Cord Injury
  • Research • Medicine
    Minimal Erythema Dose (MED) Testing
  • Research • Medicine
    Measuring Cardiac Autonomic Nervous System (ANS) Activity in Children
  • Research • Medicine
    Collecting And Measuring Wound Exudate Biochemical Mediators In Surgical Wounds
  • Research • Medicine
    A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
  • Research • Medicine
    Using a Chemical Biopsy for Graft Quality Assessment
  • Research • Medicine
    Characterizing Exon Skipping Efficiency in DMD Patient Samples in Clinical Trials of Antisense Oligonucleotides
  • Research • Medicine
    In Vitro Assessment of Cardiac Function Using Skinned Cardiomyocytes
  • Research • Medicine
    Normothermic Ex Situ Heart Perfusion in Working Mode: Assessment of Cardiac Function and Metabolism
  • Research • Medicine
    Evaluation of Vascular Control Mechanisms Utilizing Video Microscopy of Isolated Resistance Arteries of Rats
  • Research • Medicine
    Bronchoalveolar Lavage (BAL) for Research; Obtaining Adequate Sample Yield
  • Research • Medicine
    Non-invasive Optical Measurement of Cerebral Metabolism and Hemodynamics in Infants
  • Research • Medicine
    Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
  • Research • Medicine
    Driving Simulation in the Clinic: Testing Visual Exploratory Behavior in Daily Life Activities in Patients with Visual Field Defects
  • Research • Medicine
    Isolation, Characterization and Comparative Differentiation of Human Dental Pulp Stem Cells Derived from Permanent Teeth by Using Two Different Methods
  • Research • Medicine
    Portable Intermodal Preferential Looking (IPL): Investigating Language Comprehension in Typically Developing Toddlers and Young Children with Autism
  • Research • Medicine
    Intraoperative Detection of Subtle Endometriosis: A Novel Paradigm for Detection and Treatment of Pelvic Pain Associated with the Loss of Peritoneal Integrity
  • Research • Medicine
    The Use of Primary Human Fibroblasts for Monitoring Mitochondrial Phenotypes in the Field of Parkinson's Disease
  • Research • Medicine
    Collection Protocol for Human Pancreas
  • Research • Medicine
    The α-test: Rapid Cell-free CD4 Enumeration Using Whole Saliva
  • Research • Medicine
    The Measurement and Treatment of Suppression in Amblyopia
  • Research • Medicine
    Corneal Donor Tissue Preparation for Endothelial Keratoplasty
  • Research • Medicine
    Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography (FDG-PET/CT)
  • Research • Medicine
    Eye Tracking Young Children with Autism
  • Research • Medicine
    Doppler Optical Coherence Tomography of Retinal Circulation
  • Research • Medicine
    Utilizing Transcranial Magnetic Stimulation to Study the Human Neuromuscular System
  • Research • Medicine
    Detection and Genogrouping of Noroviruses from Children's Stools By Taqman One-step RT-PCR
  • Research • Medicine
    Method to Measure Tone of Axial and Proximal Muscle
  • Research • Medicine
    The Trier Social Stress Test Protocol for Inducing Psychological Stress
  • Research • Medicine
    Probing the Brain in Autism Using fMRI and Diffusion Tensor Imaging
  • Research • Medicine
    Multifocal Electroretinograms
  • Research • Medicine
    Isolation of Human Islets from Partially Pancreatectomized Patients
  • Research • Medicine
    Examining the Characteristics of Episodic Memory using Event-related Potentials in Patients with Alzheimer's Disease
  • Research • Medicine
    Magnetic Resonance Imaging Quantification of Pulmonary Perfusion using Calibrated Arterial Spin Labeling
  • Research • Medicine
    Manual Muscle Testing: A Method of Measuring Extremity Muscle Strength Applied to Critically Ill Patients
  • Research • Medicine
    Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
  • Research • Medicine
    A Protocol for Comprehensive Assessment of Bulbar Dysfunction in Amyotrophic Lateral Sclerosis (ALS)
  • Research • Medicine
    An Investigation of the Effects of Sports-related Concussion in Youth Using Functional Magnetic Resonance Imaging and the Head Impact Telemetry System
  • Research • Medicine
    Corneal Confocal Microscopy: A Novel Non-invasive Technique to Quantify Small Fibre Pathology in Peripheral Neuropathies
  • Research • Medicine
    Methods to Quantify Pharmacologically Induced Alterations in Motor Function in Human Incomplete SCI
  • Research • Medicine
    Multispectral Real-time Fluorescence Imaging for Intraoperative Detection of the Sentinel Lymph Node in Gynecologic Oncology
  • Research • Medicine
    Technique to Collect Fungiform (Taste) Papillae from Human Tongue
  • Research • Medicine
    Assessing Endothelial Vasodilator Function with the Endo-PAT 2000
  • Research • Medicine
    Making Sense of Listening: The IMAP Test Battery
  • Research • Medicine
    An Experimental Paradigm for the Prediction of Post-Operative Pain (PPOP)
  • Research • Biology
    Bioelectric Analyses of an Osseointegrated Intelligent Implant Design System for Amputees
  • Research • Biology
    Demonstration of Cutaneous Allodynia in Association with Chronic Pelvic Pain
  • Get cutting-edge science videos from JoVE sent straight to your inbox every month.

    Waiting X
    Simple Hit Counter