Encyclopedia of Experiments: Biological Techniques
Ein Abonnement für JoVE ist erforderlich, um diesen Inhalt ansehen zu können. Melden Sie sich an oder starten Sie Ihre kostenlose Testversion.
Transcript
First, open the software to control the mobile phases. Wash UPLC instruments with 50% solvent B and 50% solvent C at the flow rate of 0.2 milliliters per minute, for 30 minutes. Then wash, with 25% solvent A and 75% solvent B at a flow rate of 0.2 milliliters per minute, for 20 minutes, then, at a flow rate of 0.4 milliliters per minute.
Dissolve the labeled N-glycans with 25 microliters of a mixture of 100% acetonitrile and ultra-pure water at a 2:1 (volume/volume) ratio. Then, centrifuge at 134 x g for 5 minutes at 4 degrees Celsius, and use a drive pipe to load 10 microliters of the supernatant into the UPLC instruments.
Separate the labeled N-glycans at a flow rate of 0.4 milliliters per minute, with a linear gradient of 75% to 62% acetonitrile for 25 minutes. Then, perform an analytical run by Dextran Calibration Ladder/Glycopeptide column on a UPLC at 60 degrees Celsius.
Detect N-glycan fluorescence at excitation and emission wavelengths of 330 nanometers and 420 nanometers respectively.