Summary

Optic Nerve Sheath Point of Care Ultrasound: Image Acquisition

Published: August 18, 2023
doi:

Summary

Point-of-care ultrasound (POCUS) of the optic nerve sheath diameter (ONSD) has been shown to be useful in identifying patients with increased intracranial pressure (ICP). However, the non-standardized technique for this POCUS has hampered its use. We present a standardized image acquisition protocol for use in the acute care setting.

Abstract

The goal of this protocol is to develop a standardized method for acquiring images of the optic nerve sheath and measuring the optic nerve sheath diameter (ONSD). Diagnostic ultrasound of the ONSD to detect intracranial hypertension has traditionally faced many problems because of methodologic discrepancies. Due to inconsistencies in the measuring techniques, the potential for ONSD to become a non-invasive bedside monitoring tool for ICP has been hampered. However, establishing a transparent, consistent methodology for measuring the ONSD would support its use as a valid and reliable method of identifying intracranial hypertension. This is important as it has both high sensitivity and specificity in acute care settings. This narrative review describes ONSD POCUS image acquisition, including patient positioning, transducer selection, probe placement, the acquisition sequence, and image optimization. Further, visual aids are provided to assist in real-time during image acquisition. This method should be considered for patients for whom there are concerns regarding intracranial hypertension but who do not have an intracranial monitor in place.

Introduction

The goal of this protocol is to allow bedside providers to rapidly evaluate patients for intracranial hypertension in a non-invasive fashion using point-of-care ultrasound. In recent years, bedside decision-making and treatment have been augmented by the emergence of point-of-care ultrasound (POCUS). POCUS involves the use of ultrasound for diagnostic or procedural guidance by a patient's primary treatment provider. This article focuses on the diagnostic POCUS of the optic nerve sheath.

The rationale behind this technique is that the optic nerve and sheath communicate with the central nervous system. Specifically, the subarachnoid space extends from inside the skull around the optic nerve. Thus, when the intracranial pressure increases, the optic nerve sheath increases in size. It has been shown in studies with magnetic resonance imaging (MRI) and computed tomography (CT) that the optic nerve does not change in size with increased ICP, but the optic nerve sheath does. The vitreous of the eye provides excellent transmission of sound waves, resulting in the optic nerve being clearly visible as a hypoechoic structure inserted on the posterior eye, with the sheath visible around it. For these reasons, ultrasound of the optic nerve sheath has been used to detect elevated ICP, thus allowing the diagnosis of potentially life-threatening instances of increased ICP1.

However, despite its clinical significance, the proficiency of physicians in using ONSD POCUS is variable2,3, which limits the appropriate use of this modality4. This study aims to describe a time-efficient yet thorough image acquisition protocol for diagnostic ONSD POCUS and to illustrate the abnormal findings commonly found in clinical practice. Multiple imaging protocols have been described in the literature, and they present variations in the structural interpretation, ultrasound settings, marker placement, and scan technique5. Different marker points have varying sensitivity to changes in ICP, and the placement affects the ability to distinguish between patients with normal and high ICP. For these reasons, in this paper, we outline a standard technique to obtain optic nerve sheath images and measure them consistently.

Differences in the measurement location and technique have resulted in widely different thresholds of what is considered to be an abnormal optic nerve sheath5,6. In a recent meta-analysis, the average ONSD for patients without intracranial hypertension was 4.1 mm, and the average ONSD for those with increased ICP was 5.6 mm. A generally accepted threshold for a dilated ONSD is 5.5 mm, but a change in ONSD from a normal baseline is much more predictive if a baseline is available. In the context of severe TBI, the ONSD of both eyes seems to change together7. There are some individual cases of increased ONSD on one side, but this is rare8.

There are additional challenges in the measurement of ONSD in the severe traumatic brain injury (TBI) population. Patients with severe TBI (defined as having a Glasgow coma score of <8) are not alert enough to follow commands. This means a different approach is required to measure the nerve in this population compared to in patients being evaluated in an ophthalmology office in an elective fashion for neuritis. An A-scan can be used to measure the optic nerve in a cooperative patient that can hold their eye still for a length of time, but this is not a useful technique in TBI patients, so a B-scan is the standard technique (see step 2.7 below).

The use of this method and protocol should be considered for patients for whom there are concerns regarding intracranial hypertension but who do not have an intracranial monitor in place. Specific patient populations that may benefit are trauma patients with a clinical concern for traumatic brain injury in the prehospital or emergency room setting. Additionally, patients in the ICU with a sudden change in neurologic status are good candidates for ONSD POCUS.

Protocol

This procedure is in accordance with the ethical standards of the institutional committee on human experimentation and with the Helsinki Declaration of 1975. Ultrasound is considered minimal risk procedure, so written consent from the patient is not required. 1. Probe selection NOTE: ONSD POCUS can be performed with a multitude of different transducers. To evaluate abnormalities in suspected cases of increased ICP, choose a linear high-frequency (5-14…

Representative Results

There are several findings that can be seen on the POCUS of the optic nerve sheath and several pitfalls that can occur in the measurements. As can be seen in Figure 5, the nerve typically heads back from the eye at an angle when the eye is in a neutral position. In patients undergoing elective ultrasound of the eye for optic neuritis, for example, the patient is typically asked to move the eye and straighten the nerve to facilitate the measurement of the diameter along the axis of the nerve….

Discussion

The most critical steps of this protocol are ensuring the correct settings are used, such as ophthalmic or small parts, to minimize the energy transmitted to the eye. Additionally, ensuring that the optic nerve and not a blooming artifact is being measured is important.

Within diagnostic POCUS, the evaluation of the ONSD is well-suited to investigating patients with traumatic brain injury to evaluate if there is increased ICP. Although invasive intracranial monitoring remains the gold standard…

Divulgaciones

The authors have nothing to disclose.

Acknowledgements

This effort was partially sponsored by the US Government under Other Transactions (W81XWH-15-9-0001/W81XWH-19-9-0015) and the Medical Technology Enterprise Consortium (MTEC) under 19-08-MuLTI-0079.

Materials

Butterfly iQ+ with USB-C 2.0 Butterfly n/a Used to obtain one of the images
Clarius L7HD Portable Ultrasound Machine Clarius n/a Used to obtain one of the images
Ultrasound Gel Parker n/a Used to obtain all images
Transparent dressing 3M 9534HP Used to protect eye

Referencias

  1. Hansen, H. C., Helmke, K. Validation of the optic nerve sheath response to changing cerebrospinal fluid pressure: ultrasound findings during intrathecal infusion tests. Journal of Neurosurgery. 87 (1), 34-40 (1997).
  2. Bauerle, J., Lochner, P., Kaps, M., Nedelmann, M. Intra- and interobsever reliability of sonographic assessment of the optic nerve sheath diameter in healthy adults. Journal of Neuroimaging. 22 (1), 42-45 (2012).
  3. Lochner, P., et al. Intra- and interobserver reliability of transorbital sonographic assessment of the optic nerve sheath diameter and optic nerve diameter in healthy adults. Journal of Ultrasound. 19 (1), 41-45 (2016).
  4. Shokoohi, H., et al. Optic nerve sheath diameter measured by point-of-care ultrasound and MRI. Journal of Neuroimaging. 30 (6), 793-799 (2020).
  5. Montorfano, L., et al. Mean value of B-mode optic nerve sheath diameter as an indicator of increased intracranial pressure: A systematic review and meta-analysis. The Ultrasound Journal. 13 (1), 35 (2021).
  6. Jeon, J. P., et al. Correlation of optic nerve sheath diameter with directly measured intracranial pressure in Korean adults using bedside ultrasonography. PLoS One. 12 (9), e0183170 (2017).
  7. Maissan, I. M., et al. Ultrasonographic measured optic nerve sheath diameter as an accurate and quick monitor for changes in intracranial pressure. Journal of Neurosurgery. 123 (3), 743-747 (2015).
  8. Sedwick, L. A., Burde, R. M. Unilateral and asymmetric optic disk swelling with intracranial abnormalities. American Journal of Ophthalmology. 96 (4), 484-487 (1983).
  9. Chen, L. M., et al. Ultrasonic measurement of optic nerve sheath diameter: a non-invasive surrogate approach for dynamic, real-time evaluation of intracranial pressure. British Journal of Ophthalmology. 103 (4), 437-441 (2019).
  10. Copetti, R., Cattarossi, L. Optic nerve ultrasound: artifacts and real images. Intensive Care Medicine. 35 (8), 1488-1489 (2009).
  11. Aspide, R., et al. A proposal for a new protocol for sonographic assessment of the optic nerve sheath diameter: The CLOSED protocol. Neurocrit Care. 32 (1), 327-332 (2020).
  12. Robba, C., et al. Ultrasound non-invasive measurement of intracranial pressure in neurointensive care: A prospective observational study. PLoS Medicine. 14 (7), e1002356 (2017).
  13. Anas, I. Transorbital sonographic measurement of normal optic sheath nerve diameter in Nigerian adult population. The Malaysian Journal of Medical Sciences. 21 (5), 24-29 (2014).
check_url/es/64929?article_type=t

Play Video

Citar este artículo
Montgomery, S. P., Moore, B., Hampton, S. M., Macy, G., Li, W., Bronshteyn, Y. S. Optic Nerve Sheath Point of Care Ultrasound: Image Acquisition. J. Vis. Exp. (198), e64929, doi:10.3791/64929 (2023).

View Video