Dynamic measurement of chloride ions is presented. Transition time of an Ag/AgCl electrode, during a chronopotentiometric technique, can give the concentration of chloride ions in electrolyte. This method does not require a stable conventional reference electrode.

Abstract

This protocol describes the dynamic measurement of chloride ions using the transition time of a silver silver chloride (Ag/AgCl) electrode. Silver silver chloride electrode is used extensively for potentiometric measurement of chloride ions concentration in electrolyte. In this measurement, long-term and continuous monitoring is limited due to the inherent drift and the requirement of a stable reference electrode. We utilized the chronopotentiometric approach to minimize drift and avoid the use of a conventional reference electrode. A galvanostatic pulse is applied to an Ag/AgCl electrode which initiates a faradic reaction depleting the Clˉ ions near the electrode surface. The transition time, which is the time to completely deplete the ions near the electrode surface, is a function of the ion concentration, given by the Nernst equation. The square root of the transition time is in linear relation to the chloride ion concentration. Drift of the response over two weeks is negligible (59 µM/day) when measuring 1 mM [Clˉ]using a current pulse of 10 Am-2. This is a dynamic measurement where the moment of transition time determines the response and thus is independent of the absolute potential. Any metal wire can be used as a pseudo-reference electrode, making this approach feasible for long-term measurement inside concrete structures.

Introduction

ويرد استشعار أيون كلوريد تقوم على قياس الفترة الانتقالية من القطب حج / أجكل. والهدف من ذلك هو تجنب الانجرافات المتأصلة خلال المراقبة المستمرة على المدى الطويل من أيونات الكلوريد في بالكهرباء. ويستخدم قياس Chronopotentiometric، وهو نهج القياس الحيوي، من القطب حج / أجكل لهذا الغرض. هنا يتم قياس معدل التغير في إمكانية قيام القطب حج / أجكل خلال التحفيز (نبض galvanostatic). وتتجلى ميزة هذا النهج الذي التملص القطب إشارة السائل تقاطع وبدلا من استخدام أي سلك معدني باعتبارها القطب الزائفة إشارة، وبالتالي السماح للكشف عن CL تركيز أيونات لفترة طويلة الأجل (سنة) وفي تطبيقات الموقع، مثل قياس داخل الهياكل الخرسانية.

أيونات الكلوريد في المنشآت الخرسانية هي واحدة من الأسباب الرئيسية لتدهور 1،2. أنه يبدأ تآكل تأليب في حديد التسليح لالنتائج ND في الفشل النهائي للهيكل 3. لذلك، وقياس أيونات الكلوريد في الخرسانة أمر لا مفر منه للتنبؤ دورة حياة الخدمة والصيانة من 4،5 هيكل. تم الإبلاغ عن مبادئ الاستشعار المختلفة لقياس أيون الكلوريد في الخرسانة مثل الكهروكيميائية 6،7، 8،9 البصرية والكهرومغناطيسية 10،11. ومع ذلك، وسائل بصرية والكهرومغناطيسية ديك الاجهزة الضخمة، من الصعب دمج كنظام مستقل ولديك مشاكل مع الانتقائية 12. في تقنية الكهروكيميائية، وقياس الجهدية من القطب حج / أجكل هو حالة 6،7،13 نهج الفن. وعلى الرغم من نتائج واعدة، وهذا النهج يقتصر على قياس مختبر النطاق منذ الانجرافات في إمكانية المرجعية ونشر النتائج انخفاض المحتملة في البيانات الخاطئة 14،15. نهج الفترة الانتقالية تقوم على قياس الكهروكيميائية الحيوي (ماركا) يمكن أن تخفف من مشكلة نظرا لاحتمالاتالانجراف 16.

في ماركا، يتم قياس استجابة النظام لحافز تطبيق 17-19. على سبيل المثال من هذا النظام هو chronopotentiometry. هنا يتم استخدام نبض الحالية المطبقة كحافز المستنفدة أيونات بالقرب من سطح القطب ويتم قياس الاستجابة المحتملة المقابلة. تيار انوديك في القطب حج / أجكل يبدأ رد فعل faradaic (حج + CL الشكل 1 أجكل + E) مما أدى إلى استنزاف أيونات الكلور بالقرب من سطح القطب. التغيير المحتمل هو وظيفة من التيار التطبيقية وتركيز أيونات (الانتقائية) في 12،20 بالكهرباء. لحظة هذه الأيونات تستنزف تماما بالقرب من القطب سطح معدل التغير من الارتفاعات المحتملة بسرعة، وهو ما يعطي نقطة انعطاف 21. نقطة انعطاف على منحنى استجابة لمرة والمحتملين (chronopotentiogram) ويظهر في الوقت التي تمر بمرحلة انتقالية ويمكن تحديدها منالحد الأقصى للمشتقات الأولى من الاستجابة المحتملة 22. الفترة الانتقالية هو سمة من تركيز أيون. وقد استخدم هذا الأسلوب لتحديد أيونات مختلفة تركيز 17 ودرجة الحموضة من الشوارد 23،24. في حالة وجود القطب حج / أجكل باعتباره القطب العمل (التي الحالي يتم تطبيق) وأيونات أوزون يكون أيونات الكلوريد (17). وبالتالي قياس الوقت انتقالها ستحدد تركيزه.

Protocol

1. رقاقة تلفيق ملاحظة: رقاقة يتكون من القطب حج / أجكل العمل (نحن)، وهو حج / أجكل الكهربائي شبه المرجعية (الزائفة-RE) والبلاتين مكافحة الكهربائي على شريحة زجاجية. يترسب معدن الفضة على شريحة زجاجية، وذلك باستخدام غرف الأبحاث معيار العمل?…

Representative Results

ملفقة القطب حج / أجكل على شريحة زجاجية (Figure1) باستخدام عملية غرف الأبحاث القياسية. وقد استخدم الإعداد قياس chronopotentiometric (الشكل 2)، وقياس استجابة باستخدام potentiostat. لمراقبة تأثير تركيز أيون الكلور في الفترة الانتقالية، يتم قياس المحاليل…

Discussion

الساعة الانتقالية لحظة انعطاف. أنها مستقلة من الناحية النظرية من أي المحتملين إشارة، فإن القطب المرجع. لذا فإن أي سلك معدني يمكن استخدام القطب شبه مرجعية لقياس الفترة الانتقالية. وعلى النقيض من قياس الجهدية القائمة أيونات الكلوريد في الخرسانة يتيح هذا الأ?…

Divulgations

The authors have nothing to disclose.

Acknowledgements

This work is a part of the STW project “Integral solution for sustainable construction (IS2C, Fleur van Rossem for her support during the chip fabrication, Justyna Wiedemair for the chip design and Allison Bidulock for her support during the manuscript preparation.

Materials

Platinum wire (≥99.99% trace metals) Sigma Aldrich, the Netherlands EP1330-1EA
Potassium chloride (BioXtra, ≥99.0%) Sigma Aldrich, the Netherlands P9333-500G
Potassium hydroxide (90% pure reagent grade) Sigma Aldrich, the Netherlands 484016-1KG
Ferric chloride  Sigma Aldrich, the Netherlands 451649-1G
potassium nitrate (> 99% reagent grade)  Sigma Aldrich, the Netherlands P6083-500G
Ag/AgCl liquid junction reference electrode  BASi, USA model MF-2079
VSP potentiostat Biologic Science Instruments, France VSP 300
Steel wire Microlab TU Delft
Silver wire  Sigma Aldrich, the Netherlands

References

  1. Page, C., Treadaway, K. Aspects of the electrochemistry of steel in concrete. Nature. 297, 109-115 (1982).
  2. Koleva, D. A., Hu, J., van Breugel, K., Boshkov, N., de Wit, H. Conventional and pulse cathodic protection of reinforced concrete: electrochemical approach and microstructural investigations. ECS Transactions. 1, 287-298 (2006).
  3. Montemor, M., Simoes, A., Ferreira, M. Chloride-induced corrosion on reinforcing steel: from the fundamentals to the monitoring techniques. Cement and Concrete Composites. 25, 491-502 (2003).
  4. Wegen, G., Polder, R. B., Breugel, K. V. Guideline for service life design of structural concrete: A performance based approach with regard to chloride induced corrosion. Heron. 57 (3), (2012).
  5. Yoon, I., Koenders, E. Theoretical time evolution of critical chloride content in concrete. Structural Durability & Health Monitoring. 5, 275-294 (2010).
  6. Du, R. G., Hu, R. G., Huang, R. S., Lin, C. J. In situ measurement of Cl-concentrations and pH at the reinforcing steel/concrete interface by combination sensors. Analytical Chemistry. 78, 3179-3185 (2006).
  7. Angst, U., Elsener, B., Larsen, C. K., Vennesland, &. #. 2. 1. 6. ;. Potentiometric determination of the chloride ion activity in cement based materials. Journal of Applied Electrochemistry. 40, 561-573 (2010).
  8. Laferrière, F., Inaudi, D., Kronenberg, P., Smith, I. F. A new system for early chloride detection in concrete. Smart Materials and Structures. 17, 045017 (2008).
  9. Tang, J. L., Wang, J. N. Measurement of chloride-ion concentration with long-period grating technology. Smart Materials and Structures. 16, 665 (2007).
  10. Kohri, M., Ueda, T., Mizuguchi, H. Application of a near-infrared spectroscopic technique to estimate the chloride ion content in mortar deteriorated by chloride attack and carbonation. Journal of Advanced Concrete Technology. 8, 15-25 (2010).
  11. Tripathi, S. R., Inoue, H., Hasegawa, T., Kawase, K. Non-destructive Inspection of Chloride Ion in Concrete Structures Using Attenuated Total Reflection of Millimeter Waves. Journal of Infrared, Millimeter, and Terahertz Waves. 34, 181-186 (2013).
  12. Abbas, Y., Olthuis, W., van den Berg, A. A chronopotentiometric approach for measuring chloride ion concentration. Sensors and Actuators B: Chemical. 188, 433-439 (2013).
  13. Climent-Llorca, M. A., Viqueira-Pérez, E., Lòpez-Atalaya, M. M. Embeddable Ag/AgCl sensors for in-situ monitoring chloride contents in concrete. Cement and Concrete Research. 26, 1157-1161 (1996).
  14. Myrdal, R. . The electrochemistry and characteristics of embeddable reference electrodes for concrete. , (2014).
  15. Angst, U., Vennesland, &. #. 2. 1. 6. ;., Myrdal, R. Diffusion potentials as source of error in electrochemical measurements in concrete. Materials and Structures. 42, 365-375 (2009).
  16. Abbas, Y., de Graaf, D. B., Olthuis, W., van den Berg, A. No more conventional reference electrode: Transition time for determining chloride ion concentration. Analytica Chimica Acta. 821, 81-88 (2014).
  17. Meyer, R. E., Posey, F. A., Lantz, P. M. Chronopotentiometry of the Ag− AgCl system and analysis for the chloride ion. Journal of Electroanalytical Chemistry and Interfacial Electrochemistry. 19, 99-109 (1968).
  18. Olthuis, W., Langereis, G., Bergveld, P. The metrits of differential measuring in time and space. Biocybernetics and Biomedical Engineering. 21, 5-26 (2001).
  19. Bakker, E., Bhakthavatsalam, V., Gemene, K. L. Beyond potentiometry: robust electrochemical ion sensor concepts in view of remote chemical sensing. Talanta. 75, 629-635 (2008).
  20. Olthuis, W., Bomer, J., Bergveld, P., Bos, M., Van der Linden, W. Iridium oxide as actuator material for the ISFET-based sensor-actuator system. Sensors and Actuators B: Chemical. 5, 47-52 (1991).
  21. Bergveld, P., Eijkel, J., Olthuis, W. Detection of protein concentrations with chronopotentiometry. Biosensors and Bioelectronics. 12, 905-916 (1997).
  22. Iwamoto, R. Derivative chronopotentiometry. Analytical Chemistry. 31, 1062-1065 (1959).
  23. Olthuis, W., Bergveld, P. Simplified design of the coulometric sensor-actuator system by the application of a time-dependent actuator current. Sensors and Actuators B: Chemical. 7, 479-483 (1992).
  24. Olthuis, W., Bergveld, P. Integrated coulometric sensor-actuator devices. Microchimica Acta. 121, 191-223 (1995).
  25. Bard, A. J., Faulkner, L. R. . Electrochemical methods: fundamentals and applications. , (2001).
  26. Bakker, E., Bühlmann, P., Pretsch, E. Polymer Membrane Ion-Selective Electrodes-What are the Limits? . Electroanalysis. 11, 915-933 (1999).
check_url/fr/53312?article_type=t

Play Video

Citer Cet Article
Abbas, Y., de Graaf, D. B., Olthuis, W., van den Berg, A. Dynamic Electrochemical Measurement of Chloride Ions. J. Vis. Exp. (108), e53312, doi:10.3791/53312 (2016).

View Video