비 투명 공사장 공중 발판에 이미징 세포 생존 - 소설 니트 티타늄 임플란트의 예를 사용하여

Published: September 07, 2016
doi:

Summary

여기에서는 불투명 티타늄 지지체에 세포 생존을 검출뿐만 아니라, 골격 불순물 일별을 검출하는 형광 기반 이미징 기술을 제시한다. 이 프로토콜은 비 투명 지지체에 세포 세포 또는 세포 상호 작용 금속을 묘화의 결점이 문제를 해결합니다.

Abstract

Intervertebral disc degeneration and disc herniation is one of the major causes of lower back pain. Depletion of extracellular matrix, culminating in nucleus pulposus (NP) extrusion leads to intervertebral disc destruction. Currently available surgical treatments reduce the pain but do not restore the mechanical functionality of the spine. In order to preserve mechanical features of the spine, total disc or nucleus replacement thus became a wide interest. However, this arthroplasty era is still in an immature state, since none of the existing products have been clinically evaluated.

This study intends to test the biocompatibility of a novel nucleus implant made of knitted titanium wires. Despite all mechanical advantages, the material has its limits for conventional optical analysis as the resulting implant is non-transparent. Here we present a strategy that describes in vitro visualization, tracking and viability testing of osteochondro-progenitor cells on the scaffold. This protocol can be used to visualize the efficiency of the cleaning protocol as well as to investigate the biocompatibility of these and other non-transparent scaffolds. Furthermore, this protocol can be used to show adherence pattern of cells as well as cell viability and proliferation rates on/in the scaffold. This in vitro biocompatibility testing assay provides a propitious tool to analyze cell-material interaction in non-transparent and opaque scaffolds.

Introduction

만성 허리 통증은 다 인성 질환이다. 퇴행성 디스크 질환의 최소 침습적 치료 방법에 대한 관심은 1950 년대 이후 성장했다. 오늘날, 척추의 분절 다중 융합은 가장 널리 사용되는 치료는 상태가 될 때까지. 이후,이 방법은 종종 영향을받는 세그먼트 1, 2의 이동성의 한계에 이르게의 전 치환술 시대의 탐사는 넓은 관심이되었다. 전체 디스크 교체 및 핵 교체에 상당한 발전 만성 허리 통증 1을 치료하는 좋은 대안이되고있다. 거대한 진전에도 불구하고, 방법 중 어느 것도 임상 적으로 평가되지 않았다. 덜 엄격한 핵 이식은 전체 디스크 교체에 유망한 대안을 나타내는 섬유륜이 3,4 그대로 것이어야한다. 그러나, 시장에 현재 존재하는 핵 이식은 종종 추체, 전위, 디스크 및 t의 수직 높이 손실의 변화와 같은 합병증과 연관된필요한 관련 기계 강성 5 그는 부족. 현재의 문제점을 해결하기 위하여, 티탄 니트 와이어로 이루어지는 신규 한 핵 이식이 성공적 6 개발되었다. 때문에 독특한 니트 구조로, 새로 개발 된 인공 지지체는 예를 들어, 고유 생체 역학적 특성, 감쇠 기능, 기공 크기, 적재 능력 및 신뢰성 (7)을 보여 주었다. 이 새로운 핵 임플란트의 생체 적합성 시험을 목표 임플란트의 불투명 특성 기인 (광) 분석 기술에 심각한 제한을 도시.

생체 적합성을 테스트하기 위하여, 세포 – 금속의 상호 작용이 중요한 역할을한다 8-10. 세포와 지지체 사이의 상호 작용은 호스트 시스템 내에 더 주입 통합 안정화 때문에 필요하다. 그러나, 점점 증식 깊이는 지지체의 기계적 성질을 변경할 수있다. inves의 목표발판 표면은 세포 부착, 증식 및 분화 또는를위한 기반을 제공하는지 여부 tigate 금속은 세포 생존에 영향을 미치는 여부가 불투명 불투명 지지체에 /에서 세포 이미징의 공통 잘 알려진 문제를 해결하는 것이 중요하다. 이 제한 여러 형광을 극복하기 위해 기반 기술을 탐구 하였다. 기업은 살아있는 세포, 세포 구획, 또는 특정 세포 상태 (11)를 시각화 형광체의 넓은 범위를 제공한다. 이 실험 형광체 가장 우리 형광 현미경을 맞추기 위해 온라인 툴 스펙트럼 시청자의 도움으로 선택되었다.

온 세포의 추적을 허용하는 osteochondro-전구 세포는 1) 형광체 (녹색 형광 단백질 / GFP) 표지 : 불투명 니트 티타늄 지지체의 /에 부착 세포 거동의 분석을 위해 개발 된 전략들은 다음을 포함 지지체, 2) 측정 가능성 (미토chondrial 활동) 세포, 3) 시각화 세포 – 세포와 지지체에서 세포 물질의 상호 작용. 절차는 쉽게 다른 부착 세포 및 기타 비 투명 또는 불투명 지지체에 전사 될 수 있다는 장점을 갖는다. 또한, 생존 및 증식 패턴은 며칠에 걸쳐 모니터링 할 수있다 따라서, 이것은 지지체 재료 또는 세포의 제한된 양으로 사용될 수있다.

본 연구는 세포 생존율을 측정 불투명 니트 티타늄 지지체의 ON / osteochondro-전구 세포의 성장 패턴을 시각화 현재 프로토콜의 성공적 사용을 설명한다. 또한, 개발 된 프로토콜 골격 불순물을 결정하고 세척 프로토콜을 확인하기 위해 사용될 수있다.

Protocol

주 : 불멸화 된 인간 간엽 기질 전구체 세포 (SCP-1 세포)를 실험에 사용 하였다. SCP-1 세포 교수 마티아스 Schieker (12)에 의해 제공되었다. SCP-1 세포의 1. 확장 전에 SCP-1 세포와 협력에 제대로 70 % 에탄올 (v / v)로 입고 장갑 작업 영역 (지정 바이오 안전성 캐비닛 I)를 청소합니다. 세정 바이오 캐비닛의 표 1에 나타낸 바와 같이 필요한 성분을 ?…

Representative Results

예비 결과는 설명 새로운 핵 이식뿐만 아니라 좋은 댐핑 기능을 가지고뿐만 아니라 SCP-1 세포와 생체 적합성 인 것으로 나타났다. 임플란트의 제조 공정 동안, 강한 부식성 및 독성 물질 (윤활제, 매염제, 전기 연마 용액)와 접촉한다. 간접 형광 염색하는 기술을 사용하여, 우리는 잔여 불순물을 시각화하고, 결과적으로, 골격에 물질 부하의 상당한 감소를 도시하는 클리닝 …

Discussion

골격 표면하여 임플란트를 기능적 내구성을 결정 생체 내에서 주변 조직과의 상호 작용에서 중요한 역할을한다. 따라서, 지지체의 생체 적합성의 지지체 상에 플레이 팅 세포를 이용한 시험 관내 분석법 (SCP1 세포주)에 의해 연구된다.

얇고 광학적으로 투명 지지체와 잘 작동 현미경 기술은 가난 생체 적합성을 연구하는 불투명 발판 적합합니다. 이것은 불투명 …

Disclosures

The authors have nothing to disclose.

Acknowledgements

이 프로젝트는 부분적으로 Zentrales Innovationsprogramm Mittelstand에 (ZIM) 데 Bundesministeriums 대 Wirtschaft 싶게 에네르기 -KF3010902AJ4에 의해 투자된다. 발행 수수료는 BG 외상 병원 튀빙겐, 독일에 의해 덮여있다.

Materials

6/24/48 well plates, T25/ T75 culture flask Greiner Bio-One GmbH *
* 24 well plates Greiner Bio-One GmbH CELLSTAR 662 160
* 48 well plates Corning Incorporated USA 3548
* 6 well plates Falcon 353046
* T25 Greiner Bio-One GmbH 690 175
* T75 Greiner Bio-One GmbH 658 175
Acetic acid, purum ≥ 99,0 % Carl Roth 3738.4
Acetone Carl Roth 5025.1
Axioplan-2  Carl Zeiss, Germany
Biological safety cabinets Thermo Scientific safe 2020
Calcein acetoxymethyl ester (calcein AM) Sigma 17783
Cell Culture Incubtator Binder, Tuttlingen, Germany 9040-0078
Filter unit (0.22µm) Millipore, IRL SLGP033RS
Centrifuges 5810 R And 5417 R Thermo Fisher Scientific, NY Megafuge 40R
Dimethylsulfoxid (DMSO) Carl Roth 4720.2
Dulbecco’s PBS without Ca & Mg Sigma H15-002
Ethanol 99 %  SAV liquid prod. GmBH 475956
Ethidium homodimer Sigma 46043
EVOS Fluorescence imaging system Life technologies AMF4300
Fetal Bovine Serum (FCS) Gibco 10270-106
Hemocytometer Hausser Scientific, PA, USA
Hoechst 33342 Sigma 14533-100MG
Knitted titanium nucleus implant Buck co & KG,Germany
MEM Alpha Modification with Glutamine w/o nucleoside Sigma E15-832
Omega microplate Reader BMG Labtech,Germany FLUOstar Omega
Penicillin/Streptomycin Sigma P11-010
Resazurin sodium salt Sigma 199303-1G
Sulforhodamine B sodium salt Sigma S1402-1G
Test tube rotator Labinco B.V.,The Netherlands Model LD-76
TRIS (hydroxymethyl) aminomethan Carl Roth AE15.1
Triton Carl Roth 3051.2
Trypan Blue 0.5 % Carl Roth CN76.1
Trypsin/EDTA Sigma L11-004

References

  1. Bridwell, K. H., Anderson, P. A., Boden, S. D., Vaccaro, A. R., Wang, J. C. What’s new in spine surgery. J Bone Joint Surg Am. 95, 1144-1150 (2013).
  2. Adams, M. A., Dolan, P. Intervertebral disc degeneration: evidence for two distinct phenotypes. J Anat. 221, 497-506 (2012).
  3. Schizas, C., Kulik, G., Kosmopoulos, V. Disc degeneration: current surgical options. Eur Cell Mater. 20, 306-315 (2010).
  4. Lewis, G. Nucleus pulposus replacement and regeneration/repair technologies: present status and future prospects. J Biomed Mater Res B Appl Biomater. 100, 1702-1720 (2012).
  5. Cunningham, B. W. Basic scientific considerations in total disc arthroplasty. Spine J. 4, 219-230 (2004).
  6. Buck, A. E., Kaps, H. -. P. Implant for surgical use in humans or vertebrates. US8728164 B2. Google Patents. , (2014).
  7. Kettler, A., Kaps, H. P., Haegele, B., Wilke, H. J. Biomechanical behavior of a new nucleus prosthesis made of knitted titanium filaments. SAS J. 1, 125-130 (2007).
  8. Nerurkar, N. L., Elliott, D. M., Mauck, R. L. Mechanical design criteria for intervertebral disc tissue engineering. J Biomech. 43, 1017-1030 (2010).
  9. Elias, C. N., Lima, J. H. C., Valiev, R., Meyers, M. A. Biomedical applications of titanium and its alloys. JOM. 60, 46-49 (2008).
  10. Hallab, N., Link, H. D., McAfee, P. C. Biomaterial optimization in total disc arthroplasty. Spine (Phila Pa 1976). 28, 139-152 (2003).
  11. Gustafsdottir, S. M. Multiplex cytological profiling assay to measure diverse cellular states. PLoS One. 8, e80999 (2013).
  12. Bocker, W., et al. Introducing a single-cell-derived human mesenchymal stem cell line expressing hTERT after lentiviral gene transfer. J Cell Mol Med. 12, 1347-1359 (2008).
  13. Ehnert, S., et al. Transforming growth factor beta1 inhibits bone morphogenic protein (BMP)-2 and BMP-7 signaling via upregulation of Ski-related novel protein N (SnoN): possible mechanism for the failure of BMP therapy. BMC Med. 10, 101 (2012).
  14. Morgan, S. P., Rose, F. R., Matcher, S. J. . Optical Techniques in Regenerative Medicine. , (2013).
  15. Vielreicher, M., et al. Taking a deep look: modern microscopy technologies to optimize the design and functionality of biocompatible scaffolds for tissue engineering in regenerative medicine. J R Soc Interface. 10, 20130263 (2013).
  16. Curtis, A., Wilkinson, C. Topographical control of cells. Biomaterials. 18, 1573-1583 (1997).
  17. Niu, G., et al. Fluorescent imaging of endothelial cells in bioengineered blood vessels: the impact of crosslinking of the scaffold. J Tissue Eng Regen Med. , (2014).
  18. Chan, B. P., Leong, K. W. Scaffolding in tissue engineering: general approaches and tissue-specific considerations. Eur Spine J. 17, 467-479 (2008).
  19. Navarro, M., Michiardi, A., Castano, O., Planell, J. A. Biomaterials in orthopaedics. J R Soc Interface. 5, 1137-1158 (2008).
  20. Priyadarshani, P., Li, Y., Yao, L. Advances in biological therapy for nucleus pulposus regeneration. Osteoarthritis Cartilage. , (2015).
  21. . Thermofisher Fluorescence Spectraviewer Available from: https://www.thermofisher.com/de/de/home/life-science/cell-analysis/labeling-chemistry/fluorescence-spectraviewer.html (2016)
check_url/kr/54537?article_type=t

Play Video

Cite This Article
Tendulkar, G., Grau, P., Ziegler, P., Buck, Sr., A., Buck, Jr., A., Badke, A., Kaps, H., Ehnert, S., Nussler, A. K. Imaging Cell Viability on Non-transparent Scaffolds — Using the Example of a Novel Knitted Titanium Implant. J. Vis. Exp. (115), e54537, doi:10.3791/54537 (2016).

View Video