enfermedades neuromusculares a menudo presentan una patología de variación temporal y espacialmente heterogéneos y de múltiples facetas. El objetivo de este protocolo es la caracterización de esta patología usando métodos de resonancia magnética no invasivos.

Abstract

Quantitative magnetic resonance imaging (qMRI) describes the development and use of MRI to quantify physical, chemical, and/or biological properties of living systems. Neuromuscular diseases often exhibit a temporally varying, spatially heterogeneous, and multi-faceted pathology. The goal of this protocol is to characterize this pathology using qMRI methods. The MRI acquisition protocol begins with localizer images (used to locate the position of the body and tissue of interest within the MRI system), quality control measurements of relevant magnetic field distributions, and structural imaging for general anatomical characterization. The qMRI portion of the protocol includes measurements of the longitudinal and transverse relaxation time constants (T1 and T2, respectively). Also acquired are diffusion-tensor MRI data, in which water diffusivity is measured and used to infer pathological processes such as edema. Quantitative magnetization transfer imaging is used to characterize the relative tissue content of macromolecular and free water protons. Lastly, fat-water MRI methods are used to characterize fibro-adipose tissue replacement of muscle. In addition to describing the data acquisition and analysis procedures, this paper also discusses the potential problems associated with these methods, the analysis and interpretation of the data, MRI safety, and strategies for artifact reduction and protocol optimization.

Introduction

imágenes de resonancia magnética cuantitativa (qMRI) describe el desarrollo y el uso de MRI para cuantificar características físicas, químicas, y / o propiedades biológicas de los sistemas vivos. QMRI requiere que uno adopte un modelo biofísico para el sistema, compuesto por el tejido de interés y una secuencia de impulsos de MRI. La secuencia de pulsos está diseñado para sensibilizar a las intensidades de señal de las imágenes 'para el parámetro de interés en el modelo. Propiedades de la señal de resonancia magnética (magnitud de la señal, la frecuencia y / o fase) son medidos y analizados de acuerdo con el modelo. El objetivo es producir una estimación objetiva, cuantitativa de un parámetro físico o biológico que tiene distribuidos de forma continua, unidades físicas de medición. A menudo, las ecuaciones que describen el sistema se analizan y se acoplen a una base de pixel por pixel, produciendo una imagen cuyos valores de pixel reflejar directamente los valores de la variable. Tal imagen se conoce como un mapa paramétrico.

Un uso común de qMRI es el desarrollo y aplicación de biomarcadores. Los biomarcadores se pueden usar para investigar un mecanismo de la enfermedad, establecer un diagnóstico, determinar un pronóstico, y / o evaluar una respuesta terapéutica. Pueden adoptar la forma de las concentraciones o actividades de las moléculas endógenas o exógenas, una muestra histológica, una cantidad física, o una imagen interna. Algunos de los requisitos generales de biomarcadores son que miden objetivamente una variable distribuida de forma continua utilizando unidades físicas de medida; tener una relación clara y bien entendida con la patología de interés; son sensibles a la mejora y empeoramiento del estado clínico; y se puede medir con exactitud y precisión adecuada. biomarcadores no invasivos o mínimamente invasivos son particularmente deseables, ya que promueven la comodidad del paciente y mínimamente perturban la patología de interés.

Uno de los objetivos para el desarrollo de biomarcadores basados ​​en imágenes para enfermedades musculares es reflejar la enfermedad del músculo en formas que son Complementary a, más específico que, más espacialmente selectivo que, y / o menos invasiva que los enfoques existentes. Una ventaja particular de qMRI en este sentido es que tiene el potencial de integrar múltiples tipos de información y por lo tanto potencialmente caracterizar muchos aspectos del proceso de la enfermedad. Esta capacidad es muy importante en enfermedades musculares, que con frecuencia exhiben una patología espacialmente variables, complejo que incluye la inflamación, necrosis y / o atrofia con sustitución de la grasa, fibrosis, la interrupción de la red myofilament ( "streaming de Z-disco"), y daño de la membrana . Otra ventaja de los métodos qMRI es que las descripciones cualitativas o semi-cuantitativos de las imágenes de RM basadas en contraste reflejan no sólo la patología, sino también diferencias en los parámetros de adquisición de imágenes, hardware, y la percepción humana. Un ejemplo de este último punto fue demostrado por Wokke et al., Que mostraron que las evaluaciones semicuantitativas de la infiltración de grasa son muy variables y con frecuencia incorrecta, wgallina en comparación con la RM de grasa / agua cuantitativa (FWMRI) 1.

El protocolo descrito aquí incluye secuencias de pulso para medir la longitudinal (T1) y transversal (T 2) constantes de tiempo de relajación, transferencia de magnetización cuantitativa (Qmt) parámetros, coeficientes de difusión del agua utilizando tensor de difusión de resonancia magnética (DT-MRI) y la estructura muscular usando imágenes estructurales y FWMRI. T 1 se mide mediante el uso de una secuencia de recuperación de la inversión, en el que se invierte el vector de magnetización neta y su magnitud se muestrea como el sistema vuelve al equilibrio. T 2 se mide gracias a la reorientación repetidamente magnetización transversal usando un tren de pulsos reorientando, tales como el método de Carr-Purcell Meiboom-Gill (CPMG), y el muestreo de los spin-eco resultante. T 1 y T 2 de datos se pueden analizar utilizando métodos de ajuste de curvas no lineales que, o bien asumir un número de Exponecomponentes ntial a priori (típicamente entre uno y tres) o mediante el uso de un enfoque inverso lineal que se ajusta a los datos observados a la suma de un gran número de exponenciales en descomposición, lo que resulta en un espectro de amplitudes de señal. Este enfoque requiere una solución no negativo de mínimos cuadrados (NNLS) 3, y típicamente incluye regularización adicional para producir resultados estables. T 1 y T 2 mediciones se han utilizado ampliamente para estudiar las enfermedades musculares y lesiones 4-9. T 1 Los valores están normalmente disminuyeron en regiones infiltrado en la grasa de los músculos y elevados en las regiones inflamadas 4-6; T 2 valores son elevados en ambas regiones en grasa infiltrada e inflamadas 10.

QMT-RM caracteriza a las piscinas de protones macromoleculares de agua y sólido-como libres en los tejidos mediante la estimación de la proporción de macromolecular a los protones de agua libre (la relación entre el tamaño de la piscina, PSR); la intrínseca relajarsetasas de inflación de estas piscinas; y los tipos de cambio entre ellos. Enfoques Qmt comunes incluyen la saturación de impulsos 11 y los métodos de recuperación de la inversión 12,13 selectivos. El protocolo siguiente se describe el uso del enfoque de la saturación de impulsos, que explota el amplio ancho de línea de la señal de protón macromolecular, con relación a la estrecha anchura de línea de la señal de protón de agua. Saturando la señal macromolecular en frecuencias de resonancia suficientemente diferentes de la señal de agua, la señal de agua se reduce como resultado de la transferencia de magnetización entre las piscinas de protones del agua sólidos y gratuitas. Los datos se analizaron usando un modelo biofísico cuantitativo. QMT se ha desarrollado y aplicado en los músculos sanos 14,15, y un resumen reciente aparecido describiendo su aplicación en enfermedades musculares 16. QMT se ha utilizado para estudiar los pequeños modelos animales de inflamación muscular, en el que se ha demostrado que la inflamación disminuye el PSR 17. En la medida en MTrefleja tanto el contenido de agua y macromoleculares, los datos MT también pueden reflejar la fibrosis 18,19.

DT-MRI se utiliza para cuantificar el comportamiento de difusión anisotrópica de moléculas de agua en los tejidos con células alargadas, ordenados. En DT-MRI, la difusión del agua se mide en seis o más direcciones diferentes; estas señales se ajustaron a un modelo tensor 20. El tensor de difusión, D, se diagonalizarse para obtener tres valores propios (que son los tres principales difusividad) y tres vectores propios (que indican las direcciones correspondientes a los tres coeficientes de difusión). Estos y otros índices cuantitativos derivados de D proporcionan información acerca de la estructura del tejido y la orientación a nivel microscópico. Las propiedades de difusión del músculo, especialmente la tercera valor propio de D y el grado de anisotropía de difusión, reflejan la inflamación muscular 17 y el daño muscular debido a una lesión experimental 21, lesión por esfuerzo 22, y la enfermedad 23,24. Otras posibles influencias sobre las propiedades de difusión del músculo incluyen cambios en el diámetro de la celda 25 y cambios en la permeabilidad de la membrana.

Por último, la atrofia muscular, sin o sin infiltración de la grasa macroscópica, es un componente patológico de muchas enfermedades musculares. La atrofia muscular puede ser evaluada mediante el uso de imágenes estructurales para medir músculo área de sección transversal o el volumen y FW-MRI para evaluar la infiltración grasa. La infiltración de grasa se puede describir cualitativamente en T1 – y T2 y 26 imágenes, pero las señales de grasa y agua se miden mejor mediante la formación de imágenes que explotan las diferentes frecuencias de resonancia de grasa y agua protones 27-29. Métodos de imagen de grasa / agua cuantitativos se han aplicado en las enfermedades musculares como la distrofia muscular 1,30,31, y pueden predecir la pérdida de la deambulación en estos pacientes 31.

<pclass = "jove_content"> El protocolo qMRI descrito aquí utiliza todas estas mediciones para caracterizar la condición muscular en las miopatías inflamatorias autoinmunes dermatomiositis (DM) y la polimiositis (PM). Otros detalles del protocolo, incluyendo su reproducibilidad, se han publicado previamente 32. El protocolo incluye secuencias de pulso estándar, así como la radiofrecuencia (RF) y objetos gradiente de campo magnético programados específicamente en nuestros sistemas. Los autores prevén que el protocolo es también aplicable en otros trastornos neuromusculares caracterizados por atrofia muscular, inflamación, y la infiltración de grasa (como las distrofias musculares).

Protocol

NOTA: Se recuerda al lector que toda la investigación en seres humanos debe ser aprobado por la junta local de Revisión Institucional (IRB) para el Uso de Sujetos Humanos en Investigación. participantes en la investigación deben ser informados del propósito, los procedimientos, los riesgos y beneficios de la investigación propuesta; la disponibilidad de tratamientos o procedimientos alternativos; la disponibilidad de una remuneración; y de sus derechos de privacidad y de retirar su consentimiento y descontinuar su participación. Ant…

Representative Results

La Figura 1 muestra imágenes anatómicas axiales representativos adquiridos a la mitad del muslo de un paciente con polimiositis. También se muestra la localización de la proyección en el plano del volumen cuña. Mapas de parámetros representativos para cada método qMRI, todo obtenido de este mismo paciente, se proporcionan en las Figuras 2 – 7. Las figuras 2A</s…

Discussion

enfermedades musculares como las distrofias musculares y las miopatías inflamatorias idiopáticas constituyen del grupo de enfermedades que son heterogéneos en la etiología y, como entidades individuales, rara en su incidencia. Por ejemplo, la distrofia muscular de Duchenne – la forma más común de distrofia muscular – tiene una incidencia de 1 de cada 3.500 varones nacidos vivos 37,38; dermatomiositis, a la que se ha aplicado este protocolo, tiene una incidencia de 1 en 100.000 39. La mayor in…

Declarações

The authors have nothing to disclose.

Acknowledgements

We acknowledge grant support from the National Institutes of Health: NIH/NIAMS R01 AR050101 (BMD), NIH/NIAMS R01 AR057091 (BMD/JHP), NIH/NIBEB K25 EB013659 (RDD), and the Vanderbilt CTSA award RR024975. We also thank the reviewers for the comments and the subject for participating in these studies.

Materials

Name of Reagent/ Equipment Company Catalog Number Comments/Description
3T human MRI system Philips Medical Systems (Best, the Netherlands) Achieva/Intera
Cardiac phased array receive coil Philips Medical Systems
Pillows, straps, bolsters, and other positioning devices
Computer with MATLAB software The Mathworks, Inc (Natick, MA) r. 2014

Referências

  1. Wokke, B. H., et al. Comparison of Dixon and T1-weighted MR methods to assess the degree of fat infiltration in duchenne muscular dystrophy patients. J Magn Reson Imaging. 38 (3), 619-624 (2013).
  2. Carr, H., Purcell, E. Effects of diffusion on free precession in NMR experiments. Phys Rev. 94, 630-638 (1954).
  3. Whittall, K. P., MacKay, A. L. Quantitative interpretation of NMR relaxation data. Journal of Magnetic Resonance. 84 (1), 134-152 (1989).
  4. Park, J. H., et al. Dermatomyositis: correlative MR imaging and P-31 MR spectroscopy for quantitative characterization of inflammatory disease. Radiology. 177 (2), 473-479 (1990).
  5. Park, J. H., et al. Magnetic resonance imaging and p-31 magnetic resonance spectroscopy provide unique quantitative data useful in the longitudinal management of patients with dermatomyositis. Arthritis & Rheumatism. 37 (5), 736-746 (1994).
  6. Park, J. H., et al. Use of magnetic resonance imaging and p-31 magnetic resonance spectroscopy to detect and quantify muscle dysfunction in the amyopathic and myopathic variants of dermatomyositis. Arthritis & Rheumatism. 38 (1), 68-77 (1995).
  7. Huang, Y., et al. Quantitative MR relaxometry study of muscle composition and function in Duchenne muscular dystrophy. J Magn Reson Imaging. 4 (1), 59-64 (1994).
  8. Kim, H. K., et al. T2 mapping in Duchenne muscular dystrophy: distribution of disease activity and correlation with clinical assessments. Radiology. 255 (3), 899-908 (2010).
  9. Arpan, I., et al. T2 mapping provides multiple approaches for the characterization of muscle involvement in neuromuscular diseases: a cross-sectional study of lower leg muscles in 5-15-year-old boys with Duchenne muscular dystrophy. NMR in Biomedicine. 26 (3), 320-328 (2013).
  10. Fan, R. H., Does, M. D. Compartmental relaxation and diffusion tensor imaging measurements in vivo in λ-carrageenan-induced edema in rat skeletal muscle. NMR in Biomedicine. 21 (6), 566-573 (2008).
  11. Sled, J. G., Pike, G. B. Quantitative interpretation of magnetization transfer in spoiled gradient echo MRI sequences. J Magn Reson. 145 (1), 24-36 (2000).
  12. Gochberg, D. F., Gore, J. C. Quantitative magnetization transfer imaging via selective inversion recovery with short repetition times. Magn Reson Med. 57 (2), 437-441 (2007).
  13. Li, K., et al. Optimized inversion recovery sequences for quantitative T1 and magnetization transfer imaging. Magn Reson Med. 64 (2), 491-500 (2010).
  14. Louie, E. A., Gochberg, D. F., Does, M. D., Damon, B. M. Magnetization transfer and T2 measurements of isolated muscle: effect of pH. Magn Reson Med. 61 (3), 560-569 (2009).
  15. Sinclair, C. D. J., et al. Quantitative magnetization transfer in in vivo healthy human skeletal muscle at 3 T. Magn Reson Med. 64 (6), 1739-1748 (2010).
  16. Sinclair, C., et al. Multi-parameter quantitation of coincident fat and water skeletal muscle pathology. Proc 21st Ann Meeting ISMRM. , (2013).
  17. Bryant, N., et al. Multi-parametric MRI characterization of inflammation in murine skeletal muscle. NMR Biomed. 27 (6), 716-725 (2014).
  18. Aisen, A. M., Doi, K., Swanson, S. D. Detection of liver fibrosis with magnetic cross-relaxation. Magn Reson Med. 31 (5), 551-556 (1994).
  19. Kim, H., et al. Induced hepatic fibrosis in rats: hepatic steatosis, macromolecule content, perfusion parameters, and their correlations-preliminary MR imaging in rats. Radiology. 247 (3), 696-705 (2008).
  20. Basser, P. J., Mattiello, J., LeBihan, D. MR diffusion tensor spectroscopy and imaging. Biophys J. 66 (1), 259-267 (1994).
  21. Heemskerk, A., Strijkers, G., Drost, M., van Bochove, G., Nicolay, K. Skeletal muscle degeneration and regeneration following femoral artery ligation in the mouse: diffusion tensor imaging monitoring. Radiology. 243 (2), 413-421 (2007).
  22. Zaraiskaya, T., Kumbhare, D., Noseworthy, M. D. Diffusion tensor imaging in evaluation of human skeletal muscle injury. J Magn Reson Imaging. 24 (2), 402-408 (2006).
  23. Qi, J., Olsen, N. J., Price, R. R., Winston, J. A., Park, J. H. Diffusion-weighted imaging of inflammatory myopathies: polymyositis and dermatomyositis. J Magn Reson Imaging. 27 (1), 212-217 (2008).
  24. McMillan, A. B., Shi, D., Pratt, S. J., Lovering, R. M. Diffusion tensor MRI to assess damage in healthy and dystrophic skeletal muscle after lengthening contractions. J Biomed Biotech. , (2011).
  25. Scheel, M., et al. Fiber type characterization in skeletal muscle by diffusion tensor imaging. NMR Biomed. 26 (10), 1220-1224 (2013).
  26. Kaufman, L. D., Gruber, B. L., Gerstman, D. P., Kaell, A. T. Preliminary observations on the role of magnetic resonance imaging for polymyositis and dermatomyositis. Annalsrheumatic Dis. 46 (8), 569-572 (1987).
  27. Dixon, W. T. Simple proton spectroscopic imaging. Radiology. 153 (1), 189-194 (1984).
  28. Glover, G. H. Multipoint Dixon technique for water and fat proton and susceptibility imaging. J Magn Reson Imaging. 1 (5), 521-530 (1991).
  29. Berglund, J., Kullberg, J. Three-dimensional water/fat separation and T2* estimation based on whole-image optimization–application in breathhold liver imaging at 1.5 T. Magn Reson Med. 67 (6), 1684-1693 (2012).
  30. Gloor, M., et al. Quantification of fat infiltration in oculopharyngeal muscular dystrophy: Comparison of three MR imaging methods. J Magn Reson Imaging. 33 (1), 203-210 (2011).
  31. Fischmann, A., et al. Quantitative MRI and loss of free ambulation in Duchenne muscular dystrophy. J Neurol. 260 (4), 969-974 (2013).
  32. Li, K., et al. Multi-parametric MRI characterization of healthy human thigh muscles at 3.0 T – relaxation, magnetization transfer, fat/water, and diffusion tensor imaging. NMR Biomed. 27 (9), 1070-1084 (2014).
  33. Morrison, C., Stanisz, G., Henkelman, R. M. Modeling magnetization transfer for biological-like systems using a semi-solid pool with a super-Lorentzian lineshape and dipolar reservoir. J Magn Reson Series B. 108 (2), 103-113 (1995).
  34. Li, J. G., Graham, S. J., Henkelman, R. M. A flexible magnetization transfer line shape derived from tissue experimental data. Magn Reson Med. 37 (6), 866-871 (1997).
  35. Mangin, J. F., Poupon, C., Clark, C., Le Bihan, D., Bloch, I. Distortion correction and robust tensor estimation for MR diffusion imaging. Med Image Anal. 6 (3), 191-198 (2002).
  36. Moser, H. Duchenne muscular dystrophy: pathogenetic aspects and genetic prevention. Hum Genet. 66 (1), 17-40 (1984).
  37. van Essen, A. J., Busch, H. F., te Meerman, G. J., ten Kate, L. P. Birth and population prevalence of Duchenne muscular dystrophy in The Netherlands. Hum Genet. 88 (3), 258-266 (1992).
  38. Bendewald, M. J., Wetter, D. A., Li, X., Davis, M. P. Incidence of dermatomyositis and clinically amyopathic dermatomyositis: A population-based study in olmsted county, minnesota. Arch Dermatol. 146 (1), 26-30 (2010).
  39. Carlier, P. G. Global T2 versus water T2 in NMR imaging of fatty infiltrated muscles: different methodology, different information and different implications. Neuromuscul Disord. 24 (5), 390-392 (2014).
  40. Foley, J. M., Jayaraman, R. C., Prior, B. M., Pivarnik, J. M., Meyer, R. A. MR measurements of muscle damage and adaptation after eccentric exercise. J Appl Physiol. 87 (6), 2311-2318 (1999).
  41. Garrood, P., et al. MR imaging in Duchenne muscular dystrophy: quantification of T1-weighted signal, contrast uptake, and the effects of exercise. J Magn Reson Imaging. 30 (5), 1130-1138 (2009).
  42. Bratton, C. B., Hopkins, A. L., Weinberg, J. W. Nuclear magnetic resonance studies of living muscle. Science. 147, 738-739 (1965).
  43. Fleckenstein, J. L., Canby, R. C., Parkey, R. W., Peshock, R. M. Acute effects of exercise on MR imaging of skeletal muscle in normal volunteers. AJR Am J Roentgenol. 151 (2), 231-237 (1988).
  44. Williams, S., Heemskerk, A., Welch, E., Damon, B., Park, J. The quantitative effects of inclusion of fat on muscle diffusion tensor MRI measurements. J Magn Reson Imaging. 38 (5), 1292-1297 (2013).
  45. Hernando, D., et al. Removal of olefinic fat chemical shift artifact in diffusion MRI. Magn Reson Med. 65 (3), 692-701 (2011).
  46. Willcocks, R. J., et al. Longitudinal measurements of MRI-T2 in boys with Duchenne muscular dystrophy: effects of age and disease progression. Neuromuscul Disord. 24 (5), 393-401 (2014).
  47. Poon, C. S., Henkelman, R. M. Practical T2 quantitation for clinical applications. J Magn Reson Imaging. 2 (5), 541-553 (1992).
  48. Does, M. D., Gore, J. C. Complications of nonlinear echo time spacing for measurement of T2. NMR Biomed. 13 (1), 1-7 (2000).
  49. Poon, C. S., Henkelman, R. M. 180° refocusing pulses which are insensitive to static and radiofrequency field inhomogeneity. J Magn Reson. 99 (1), 45-55 (1992).
  50. Hollingsworth, K. G., de Sousa, P. L., Straub, V., Carlier, P. G. Towards harmonization of protocols for MRI outcome measures in skeletal muscle studies: consensus recommendations from two TREAT-NMD NMR workshops, 2 May 2010, Stockholm, Sweden, 1-2 October 2009, Paris, France. Neuromuscul Disord. 22, S54-S67 (2010).
  51. Underhill, H. R., Rostomily, R. C., Mikheev, A. M., Yuan, C., Yarnykh, V. L. Fast bound pool fraction imaging of the in vivo rat brain: Association with myelin content and validation in the C6 glioma model. Neuroimage. 54 (3), 2052-2065 (2011).
  52. Smith, S. A., et al. Quantitative magnetization transfer characteristics of the human cervical spinal cord in vivo: application to adrenomyeloneuropathy. Magn Reson Med. 61 (1), 22-27 (2009).
  53. Li, K. D. R., Dortch, R. D., Gochberg, D. F., Smith, S. A., Damon, B. M., Park, J. H. Quantitative magnetization transfer with fat component in human muscles. Proc. 20th Ann Meeting ISMRM. , (2012).
  54. Damon, B. M. Effects of image noise in muscle diffusion tensor (DT)-MRI assessed using numerical simulations. Magn Reson Med. 60 (4), 934-944 (2008).
  55. Damon, B. M., Buck, A. K. W., Ding, Z. Diffusion-tensor MRI-based skeletal muscle fiber tracking. Imaging Med. 3 (6), 675-687 (2011).
  56. Froeling, M., Nederveen, A. J., Nicolay, K., Strijkers, G. J. DTI of human skeletal muscle: the effects of diffusion encoding parameters, signal-to-noise ratio and T2 on tensor indices and fiber tracts. NMR in Biomedicine. 26 (11), 1339-1352 (2013).
  57. Basser, P. J., Pajevic, S. Statistical artifacts in diffusion tensor MRI (DT-MRI) caused by background noise. Magn Reson Med. 44 (1), 41-50 (2000).
  58. Anderson, A. W. Theoretical analysis of the effects of noise on diffusion tensor imaging. Magn Reson Med. 46 (6), 1174-1188 (2001).
  59. Saupe, N., White, L. M., Stainsby, J., Tomlinson, G., Sussman, M. S. Diffusion tensor imaging and fiber tractography of skeletal muscle: optimization of B value for imaging at 1.5 T. AJR Am J Roentgenol. 192 (6), W282-W290 (2009).
  60. Levin, D. I., Gilles, B., Madler, B., Pai, D. K. Extracting skeletal muscle fiber fields from noisy diffusion tensor data. Med Image Anal. 15 (3), 340-353 (2011).
  61. Sinha, U., Sinha, S., Hodgson, J. A., Edgerton, R. V. Human soleus muscle architecture at different ankle joint angles from magnetic resonance diffusion tensor imaging. J Appl Physiol. 110 (3), 807-819 (2011).
  62. Jones, D. K., Cercignani, M. Twenty-five pitfalls in the analysis of diffusion MRI data. NMR Biomed. 23 (7), 803-820 (2010).
  63. Hamilton, G., et al. In vivo characterization of the liver fat 1H MR spectrum. NMR Biomed. 24 (7), 784-790 (2011).
  64. Hernando, D., Kellman, P., Haldar, J. P., Liang, Z. P. Robust water/fat separation in the presence of large field inhomogeneities using a graph cut algorithm. Magn Reson Med. 63 (1), 79-90 (2010).
  65. Hernando, D., Hines, C. D., Yu, H., Reeder, S. B. Addressing phase errors in fat-water imaging using a mixed magnitude/complex fitting method. Magn Reson Med. 67 (3), 638-644 (2012).
check_url/pt/52352?article_type=t

Play Video

Citar este artigo
Damon, B. M., Li, K., Dortch, R. D., Welch, E. B., Park, J. H., Buck, A. K. W., Towse, T. F., Does, M. D., Gochberg, D. F., Bryant, N. D. Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease. J. Vis. Exp. (118), e52352, doi:10.3791/52352 (2016).

View Video