Summary

Microperf Técnica para Investigar Regulamento de microvasos Permeabilidade de Rat Mesentery

Published: September 12, 2015
doi:

Summary

The modified Landis technique enables paired measurement of the hydraulic conductivity of individual microvessels in the mesentery of normal and genetically modified rats under control and test conditions using microperfusion techniques. It provides a convenient method to evaluate mechanisms that regulate microvessel permeability and transvascular exchange under physiological conditions.

Abstract

As experiências para medir as propriedades de permeabilidade dos microvasos perfundidos individualmente fornecer uma ponte entre a investigação dos mecanismos moleculares e celulares que regulam a permeabilidade vascular em monocamadas de células endoteliais em cultura e as propriedades de leitos microvasculares inteiros câmbio funcionais. Um método para canular e perfundir microvasos venulares de mesentério de ratos e medir a condutividade hidráulica da parede de microvasos é descrito. O principal equipamento necessário inclui um microscópio intravital com um palco modificado grande que suporta micromanipuladores para posicionar três microtools diferentes: (1) uma micropipeta de vidro chanfrado de canulação e perfundem o microvasos; (2) um micro-vidro oclusor para bloquear temporariamente a perfusão e permitir a medição de transvascular movimento fluxo de água a uma pressão hidrostática medida, e (3) uma vareta de vidro fechado para estabilizar o tecido mesentérico ao sítio de canulação. A Landis micro-oclusão techniq modificadoue utiliza células vermelhas suspensas no perfusato artificial como marcadores de circulação de fluido transvascular, e também permite medições repetidas destes fluxos condições experimentais são alteradas e diferença de pressão osmótica e hidrostática colóide através dos microvasos são cuidadosamente controlados. As medições de condutividade hidráulica primeiro usando um perfusado de controlo, em seguida, após a re-canulação da mesma microvasos com os perfusatos teste permitir comparações de resposta de microvasos sob estas condições bem controladas emparelhado. As tentativas para alargar o método de microvasos no mesentério de ratos com modificações genéticas que se espera modificar a permeabilidade vascular foram severamente limitada, devido à ausência de microvasos compridas rectas e não ramificados no mesentério do rato, mas a recente disponibilidade dos ratos com modificações genéticas semelhantes usando CRISPR tecnologia / Cas9 deverá abrir novas áreas de investigação onde os métodos aqui descritos podem ser aplicados. </p>

Introduction

Microperfusão na vasculatura inclui estabelecer um fluxo de perfusato artificial de composição conhecida controlada através de uma micropipeta em um vaso sanguíneo geralmente menos de 40 um de diâmetro. O recipiente continua a ser perfundido no seu ambiente de tecido normal e é perfundido com sangue do animal até ao momento da punção. Quando usado em conjunto com uma série de imagens de vídeo ou técnicas fluorométricos, em microperfusão situ permite a medição de fluxos de água e de solutos através das paredes de microvasos em condições onde as forças de condução para estes fluxos são conhecidos e as propriedades de permeabilidade da parede vascular podem ser avaliada diretamente. Além disso, por controlo da composição do líquido que rodeia o de microvasos no tecido (perfusado e superfusate), a regulação da permeabilidade microvascular e podem ser investigadas de câmbio, permitindo que as células endoteliais que formam a parede de microvasos para ser exposta a uma variedade de econdições Xperimental (agonistas, condições de perfusão modificados, indicadores fluorescentes para medir a composição e sinalização intracelular) por períodos precisamente medidas de tempo (seg a hr). Além disso, as avaliações ultra-estruturais ou citoquímicos de estruturas moleculares celulares chave que regulam a barreira pode ser investigadas no mesmo microvasos em que a permeabilidade está diretamente medidos. A abordagem assim, forma uma ponte entre a investigação dos mecanismos celulares e moleculares para modificar a função de barreira endotelial em monocamadas de células endoteliais cultivadas e investigação em microvasos intactas. Consulte os seguintes comentários para uma avaliação mais aprofundada 1-6.

Uma limitação de microperfusão é que ela só pode ser utilizada em leitos microvasculares que são finas, transparentes e possuem integridade estrutural suficiente para permitir a canulação com uma micropipeta de vidro. Enquanto as primeiras investigações usado microvasos rã no mesentério e fino cutânea pectoris muscle 7,8, de longe, a preparação mais utilizada em modelos de mamíferos é o mesentério de rato 9-15. A maioria das investigações concentraram-se em alterações agudas na permeabilidade vascular estudado ao longo de períodos de 1-4 horas, mas as investigações mais recentes foram estendidos para medições em navios individuais 24-72 horas após uma perfusão inicial 12,16. A tecnologia CRISPR recentemente desenvolvido, que promete tornar mais modelos de ratos geneticamente modificados disponível para estudar regulação da permeabilidade vascular 17 deverá permitir que os métodos descritos na presente comunicação a ser aplicado em microvasos venulares do mesentério nestes novos modelos importantes de ratos.

O método requer um microscópio invertido equipado com um estágio de microscópio construído sob encomenda suficientemente grande para conter tanto a preparação dos animais e, pelo menos, três micromanipuladores usadas para microferramentas posição próxima do vaso e perfundidos para alinhar uma micropipeta de perfusão com o naviolúmen. Por exemplo, uma plataforma personalizada para um estágio de microscópio XY (cerca de 90 × 60 cm) pode ser fabricada a partir de uma chapa de aço de 1 cm de espessura com um revestimento à prova de ferrugem. O palco está ligado a uma tabela de índice de engenharia ou duas lâminas de cauda de andorinha montados em ângulos retos e apoiados em pilares de teflon ou transferências bola para o movimento no plano horizontal. Um equipamento típico (ver Figura 2) tem muito em comum com o microscópio e microposicionamento equipamento utilizado para uma série de experimentos microcirculação intravital tais como aqueles para medir o fluxo único dos vasos sanguíneos e hematócrito, entrega local de oxigênio pelo sangue perfusão microvasos, regulação da liso vascular tônus ​​muscular, eo acúmulo microvascular local dos marcadores fluorescentes injetadas em toda a circulação 18-26.

O aspecto fundamental da técnica é a medição do volume do caudal (J v) através de uma área de superfície definido (S) da parede de microvasos. Cumpriresta via a técnica Landis modificado aqui descrito um microscópio invertido simples é adequada. Uma pequena câmara de vídeo está montado na porta e o sinal de imagem de vídeo, com uma base de tempo adicional, é apresentada num monitor de video e registadas na forma digital num computador ou como um sinal digital ou analógico de um gravador de vídeo. Uma vez que a microvasculatura é canulada a porção do microvasos visível para a câmara pode ser alterado movendo o palco e manipuladores como uma unidade, sem perturbar a canulação.

Medição dos fluxos transvascular pode também ser combinado com investigações mais detalhadas, utilizando um microscópio de fluorescência com filtros adequados sofisticados, tais como sondas utilizadas para as medições de permeabilidade soluto, fluorescente monitorização relação de cálcio citoplasmático ou outros mecanismos celulares, e imagiologia confocal 6,12,13, 27. Uma das principais vantagens de todas as abordagens microperfusão é a capacidade de fazer medidas repetidas, no mesmo navio, Sob a mudança controlada de força motriz, tais como pressões hidrostática e oncótica, ou mudança induzida em respostas dos vasos para condições inflamatórias. O design mais comum é uma comparação pareada da condutividade hidráulica medida (L p) no mesmo navio com o primeiro recipiente de perfusão através de uma micropipeta preenchido com um perfusato controle e suspensão de eritrócitos para estabelecer um estado permeabilidade da linha de base, em seguida, com uma segunda pipeta com o agente de teste adicionado ao perfusado. Várias cânulas são possíveis com o ciclo repetido após reperfusão, com a pipeta de controle.

O presente protocolo demonstra a canulação e microperfusão de um navio venular em mesentério de rato para gravar fluxos de água através da parede de microvasos e medir a G P da parede do vaso, um índice útil da permeabilidade da via comum para a água e os solutos através da intacta barreira endotelial. O procedimento é chamado de techniq Landis modificadoue porque o princípio original Landis de utilizar o movimento relativo de células vermelhas como uma medida de troca de fluido transvascular após perfusão é bloqueado é preservada 28, mas a gama de condições experimentais (por exemplo, as diferenças de pressão oncótica hidrostáticas e albumina através da parede de microvasos) disponível depois de microperfusão é muito maior do que no sangue perfusão microvasos canulados 8,29.

Protocol

Declaração de Ética: Todos os procedimentos foram revistos e aprovados pelo Comitê de Cuidado e Uso do animal Institucional. 1. Fabricação preliminar de Micropipetas, contenção e bloqueadores Puxar vários tubos capilares de vidro de borosilicato limpas ao meio usando um extractor de electrónica ajustada de modo que, quando puxado, a parte alongada do tubo é de cerca de 1 cm de comprimento e as duas metades são um tanto simétrica. Certifique-se de que a vela é coerent…

Representative Results

A Figura 4 mostra os resultados da medição do curso de tempo das alterações no G P num rato venular microvasos canulada sucessivamente com quatro perfusatos. 33 A magnitude de L p calculado a uma pressão constante foi usada como uma medida de mudanças na microvasos permeabilidade da parede, primeiro no estado de controlo com um perfusato contendo 1% de albumina de soro bovino, em seguida, quando o recipiente foi exposta ao agente inflamatório bradicinina (BK) atra…

Discussion

Detalhes dos cálculos de L p. Embora o movimento fluido transvascular ocorre enquanto o navio é perfundido livremente, essa troca é pequena demais para ser medido durante a perfusão livre porque é tipicamente menor do que 0,01% da taxa de perfusão navio. No entanto, quando a perfusão é transitoriamente interrompido por oclusão da micro-vasos, o fluxo transvascular (isto é, por filtração) é medido a partir do movimento das células marcador vermelho no lúmen que a coluna de líq…

Declarações

The authors have nothing to disclose.

Acknowledgements

Este trabalho foi apoiado pelo National Institutes of Health subvenções HL44485 e HL28607.

Materials

MICROSCOPE, TABLE AND STAGE
inverted microscope (metallurgical type) with trinocular head for video: example Olympus CK-40 try to place eyepieces higher relative to stage–you have to look through eyepieces while reaching around to top of stage over intervening micromanipulators
inverted microscope (metallurgical type) with trinocular head for video: example Leica DMIL try to place eyepieces higher relative to stage–you have to look through eyepieces while reaching around to top of stage over intervening micromanipulators
narrow diameter, long working distance objective: example Nikon Nikon E Plan 10×/0.25 LWD
stage platform–1/2 inch or 1 cm sheet steel welding shop this should be heavy to reduce vibration
Unislide x-y table: dove tail slides Velmex AXY4006W1
VIDEO
CCD video camera: example Pulnix TM-7CN (no longer available) no color needed
video capture system with audio–generic
video playback system (completely still frame, single frame motion)
small microphone
MICROMANIPULATORS, HOLDERS
micromanipulator, XYZ (3) Prior/Stoelting (no longer available) look for fine Z, and larger range of travel in coarse drives for ease of positioning
hydraulic probe drive, one way FHC 50-12-1C need to buy either manual drive or electronic drive
manual drum drive  FHC 50-12-9-02
or hydraulic drive, 3 way Siskiyou Corporation MX610 (1-way) or MX630 (3-way) great for short arms, water filled and must be sent back for refill ~every 2 years
connectors/rods/holders Siskiyou Corporation MXC-2.5, MXB etc.
pin vise Starrett 162C to hold restrainer
pipette holder World Prescision Instruments MPH3
water manometer ~120 cm
MICROSCOPE TRAY
clear Plexiglas for microscope tray for animal
3/4 inch polished quartz disc ~1/4 inch tall Quartz Scientific Inc. custom  (or polished plexiglass, glass); make sure the height is less than working distance of objective
Plexiglas glue (Weld-on 4: CAUTION CARCINOGEN)
medical adhesive for tissue well NuSil MED-1037
All-purpose silicone rubber heat mat, 5" L x 2" W Cole Parmer EW-03125-20 heater for microscope tray–needs cord and controller–240V version available
Power Cord Adapter for Kapton Heaters and Kits, 6 ft, 120 VAC Cole Parmer EW-03122-75
STACO 3PN1010B Variable-Voltage Controller, 10 A; 120 V In, 0-140 V Out Cole Parmer EW-01575-00
PIPET MANUFACTURE
vertical pipette puller Sutter Instrument Company P-30 with nichrome filament
1.5 mm OD thin wall capillary tubing Sutter Instrument Company B150-110-10
pipette grinder air stone and dissection microscope–see reference in text or purchase a package from Sutter Instruments or World Precision Instruments
RX Honing Machine, System II RX Honing Machine Corporation MAC-10700 Rx System II Machine alternative for air stone, use with a dissecting microscope mounted at an angle
   with ceramic sharpening disc RX Honing Machine Corporation use "as is" or attach lapping film
lapping film sheets, 0.3 or 0.5 um 3M part no. 051144 80827 268X Imperial lapping film sheets with adhesive back–can be purchased from Amazon

Referências

  1. Curry, F. R. Permeability measurements in an individually perfused capillary: the ‘squid axon’ of the microcirculation. Experimental physiology. 93, 444-446 (2008).
  2. Curry, F. R., Adamson, R. H. Vascular permeability modulation at the cell, microvessel, or whole organ level: towards closing gaps in our knowledge. Cardiovasc Res. 87, 218-229 (2010).
  3. Curry, F. R., Adamson, R. H. Tonic regulation of vascular permeability. Acta physiologica. 207, 628-649 (2013).
  4. Michel, C. C. Fluid exchange in the microcirculation. The Journal of physiology. 557, 701-702 (2004).
  5. Tarbell, J. M., Simon, S. I., Curry, F. R. Mechanosensing at the vascular interface. Annual review of biomedical engineering. 16, 505-532 (2014).
  6. Sarelius, I. H., Kuebel, J. M., Wang, J., Huxley, V. H. Macromolecule permeability of in situ and excised rodent skeletal muscle arterioles and venules. American journal of physiology. Heart and circulatory physiology. 290, H474-H480 (2006).
  7. Curry, F. E., Frokjaer-Jensen, J. Water flow across the walls of single muscle capillaries in the frog, Rana pipiens. The Journal of physiology. 350, 293-307 (1984).
  8. Michel, C. C., Mason, J. C., Curry, F. E., Tooke, J. E., Hunter, P. J. A development of the Landis technique for measuring the filtration coefficient of individual capillaries in the frog mesentery. Q J Exp Physiol Cogn Med Sci. 59, 283-309 (1974).
  9. Adamson, R. H., Zeng, M., Adamson, G. N., Lenz, J. F., Curry, F. E. PAF- and bradykinin-induced hyperpermeability of rat venules is independent of actin-myosin contraction. American journal of physiology, Heart and circulatory physiology. 285, H406-H417 (2003).
  10. Huxley, V. H., Rumbaut, R. E. The microvasculature as a dynamic regulator of volume and solute exchange. Clinical and experimental pharmacology, & physiology. 27, 847-854 (2000).
  11. Rumbaut, R. E., Wang, J., Huxley, V. H. Differential effects of L-NAME on rat venular hydraulic conductivity. American journal of physiology, Heart and circulatory physiology. , 279-H2023 (2000).
  12. Yuan, D., He, P. Vascular remodeling alters adhesion protein and cytoskeleton reactions to inflammatory stimuli resulting in enhanced permeability increases in rat venules. Journal of applied physiology. 113, 1110-1120 (2012).
  13. Zhou, X., He, P. Temporal and spatial correlation of platelet-activating factor-induced increases in endothelial [Ca(2)(+)]i, nitric oxide, and gap formation in intact venules. American journal of physiology, Heart and circulatory physiology. 301, H1788-H1797 (2011).
  14. Adamson, R. H., et al. Oncotic pressures opposing filtration across non-fenestrated rat microvessels. The Journal of physiology. 557, 889-907 (2004).
  15. Adamson, R. H., et al. Epac/Rap1 pathway regulates microvascular hyperpermeability induced by PAF in rat mesentery. American journal of physiology, Heart and circulatory physiology. 294, H1188-H1196 (2008).
  16. Curry, F. E., Zeng, M., Adamson, R. H. Thrombin increases permeability only in venules exposed to inflammatory conditions. American journal of physiology, Heart and circulatory physiology. 294, H1188-H1196 (2003).
  17. Sander, J. D., Joung, J. K. CRISPR-Cas systems for editing, regulating and targeting genomes. Nature. 32, 347-355 (2014).
  18. Bagher, P., Davis, M. J., Segal, S. S. Intravital macrozoom imaging and automated analysis of endothelial cell calcium signals coincident with arteriolar dilation in Cx40(BAC) -GCaMP2 transgenic mice. Microcirculation. 18, 331-338 (2011).
  19. Duza, T., Sarelius, I. H. Increase in endothelial cell Ca(2+) in response to mouse cremaster muscle contraction. The Journal of physiology. 555, 459-469 (2004).
  20. Oshiro, H., et al. L-type calcium channel blockers modulate the microvascular hyperpermeability induced by platelet-activating factor in vivo. Journal of vascular surgery. 22, 732-739 (1995).
  21. Chen, W., et al. Atrial natriuretic peptide-mediated inhibition of microcirculatory endothelial Ca2+ and permeability response to histamine involves cGMP-dependent protein kinase I and TRPC6 channels. Arteriosclerosis, thrombosis, and vascular biology. 33, 2121-2129 (2013).
  22. Harris, N. R., Whitt, S. P., Zilberberg, J., Alexander, J. S., Rumbaut, R. E. Extravascular transport of fluorescently labeled albumins in the rat mesentery. Microcirculation. 9, 177-187 (2002).
  23. Yuan, W., Li, G., Zeng, M., Fu, B. M. Modulation of the blood-brain barrier permeability by plasma glycoprotein orosomucoid. Microvascular research. 80, 148-157 (2010).
  24. Sugiura, Y., Morikawa, T., Takenouchi, T., Suematsu, M., Kajimura, M. Cilostazol strengthens the endothelial barrier of postcapillary venules from the rat mesentery in situ. Phlebology / Venous Forum of the Royal Society of Medicine. 29, 594-599 (2014).
  25. Guo, M., et al. Fibrinogen-gamma C-terminal fragments induce endothelial barrier dysfunction and microvascular leak via integrin-mediated and RhoA-dependent mechanism. Arteriosclerosis, thrombosis, and vascular biology. 29, 394-400 (2009).
  26. Dewar, A. M., Clark, R. A., Singer, A. J., Frame, M. D. Curcumin mediates both dilation and constriction of peripheral arterioles via adrenergic receptors. The Journal of investigative dermatology. 131, 1754-1760 (2011).
  27. Lee, J. F., et al. Balance of S1P1 and S1P2 signaling regulates peripheral microvascular permeability in rat cremaster muscle vasculature. American journal of physiology, Heart and circulatory physiology. 296, H33-H42 (2009).
  28. Landis, E. M. Microinjection studies of capillary permeability. II. The relation between capillary pressure and the rate at which fluid passes through the walls of single capillaries. Am J Physiol. 82, 217-238 (1927).
  29. Curry, F. E., Huxley, V. H., Sarelius, I. H., Linden, R. J. . Techniques in cardiovascular physiology Part 1. P3/1, 1-34 (1983).
  30. Vurek, G. G., Bennett, C. M., Jamison, R. L., Troy, J. L. An air-driven micropipette sharpener). J Appl Physiol. 22, 191-192 (1967).
  31. Curry, F. E., Clark, J. F., Adamson, R. H. Erythrocyte-derived sphingosine-1-phosphate stabilizes basal hydraulic conductivity and solute permeability in rat microvessels. American journal of physiology, Heart and circulatory physiology. 303, H825-H834 (2012).
  32. Bagher, P., Polo-Parada, L., Segal, S. S. Microiontophoresis and micromanipulation for intravital fluorescence imaging of the microcirculation. Journal of visualized experiments : JoVE. , (2011).
  33. Adamson, R. H., et al. Attenuation by sphingosine-1-phosphate of rat microvessel acute permeability response to bradykinin is rapidly reversible. American journal of physiology, Heart and circulatory physiology. 302, H1929-H1935 (2012).
  34. Bates, D. O. Vascular endothelial growth factors and vascular permeability. Cardiovasc Res. 87, 262-271 (2010).
  35. Adamson, R. H., et al. Rho and rho kinase modulation of barrier properties: cultured endothelial cells and intact microvessels of rats and mice. The Journal of physiology. 539, 295-308 (2002).
  36. Curry, F. R., et al. Atrial natriuretic peptide modulation of albumin clearance and contrast agent permeability in mouse skeletal muscle and skin: role in regulation of plasma volume. The Journal of physiology. 588, 325-339 (2010).
  37. Neal, C. R., Bates, D. O. Measurement of hydraulic conductivity of single perfused Rana mesenteric microvessels between periods of controlled shear stress. The Journal of physiology. 543, 947-957 (2002).
  38. Adamson, R. H., et al. Albumin modulates S1P delivery from red blood cells in perfused microvessels: mechanism of the protein effect. American journal of physiology, Heart and circulatory physiology. 306, H1011-H1017 (2014).
  39. Huxley, V. H., Wang, J. J., Sarelius, I. H. Adaptation of coronary microvascular exchange in arterioles and venules to exercise training and a role for sex in determining permeability responses. American journal of physiology, Heart and circulatory physiology. 293, H1196-H1205 (2007).
  40. Huxley, V. H., Williams, D. A. Basal and adenosine-mediated protein flux from isolated coronary arterioles. Am J Physiol. 271, H1099-H1108 (1996).
  41. Davis, M. J., Gore, R. W. Double-barrel pipette system for microinjection. Am J Physiol. 253, H965-H967 (1987).
  42. Adamson, R. H., et al. Sphingosine-1-phosphate modulation of basal permeability and acute inflammatory responses in rat venular microvessels. Cardiovasc Res. 88, 344-351 (2010).
  43. Zeng, Y., Adamson, R. H., Curry, F. R., Tarbell, J. M. Sphingosine-1-phosphate protects endothelial glycocalyx by inhibiting syndecan-1 shedding. American journal of physiology, Heart and circulatory physiology. , H306-H363 (2014).
check_url/pt/53210?article_type=t

Play Video

Citar este artigo
Curry, F. E., Clark, J. F., Adamson, R. H. Microperfusion Technique to Investigate Regulation of Microvessel Permeability in Rat Mesentery. J. Vis. Exp. (103), e53210, doi:10.3791/53210 (2015).

View Video