Summary

デジタル画像相関を使用すると、血管の組織標本のローカル株を特徴づけるために、

Published: January 24, 2016
doi:

Summary

We describe the use of digital image correlation to characterize the local surface strain field on vascular tissue samples subjected to uniaxial tensile testing. These measurements facilitate precise quantification of the sample mechanical response and the generation of constitutive stress-strain relations.

Abstract

Characterization of the mechanical behavior of biological and engineered soft tissues is a central component of fundamental biomedical research and product development. Stress-strain relationships are typically obtained from mechanical testing data to enable comparative assessment among samples and in some cases identification of constitutive mechanical properties. However, errors may be introduced through the use of average strain measures, as significant heterogeneity in the strain field may result from geometrical non-uniformity of the sample and stress concentrations induced by mounting/gripping of soft tissues within the test system. When strain field heterogeneity is significant, accurate assessment of the sample mechanical response requires measurement of local strains. This study demonstrates a novel biomechanical testing protocol for calculating local surface strains using a mechanical testing device coupled with a high resolution camera and a digital image correlation technique. A series of sample surface images are acquired and then analyzed to quantify the local surface strain of a vascular tissue specimen subjected to ramped uniaxial loading. This approach can improve accuracy in experimental vascular biomechanics and has potential for broader use among other native soft tissues, engineered soft tissues, and soft hydrogel/polymeric materials. In the video, we demonstrate how to set up the system components and perform a complete experiment on native vascular tissue.

Introduction

50年間に及ぶ研究の豊かな歴史は、血管組織の機械的特性を定量化する上で注目されています。これらの研究は、血管内デバイスの有効性/互換性を評価するための基礎を提供する、私たちはより良い血管の生理学的および病理学的行動の両方を理解することを可能にし、設計された血管の設計および製造における援助は1-6を構築します。軟組織と機械的特性の構成的モデリングの機械的応答の正確な測定は、本質的に、ほとんどの組織の種類によって示さ機械的異質、異方性、および非線形性のために挑戦しています。また、実験的な測定は、多くの場合、地元の機械的試験の過程でサンプルグリップ界面に導入された複雑さ( すなわち、曲げ、摩擦、応力集中、涙)と組織が ​​生きている動物から切除された後の機械的特性の必然的な移行によって混乱しています。 </ P>

一軸引張試験は、固体材料で作られた試料上で実行することができる最も簡単な機械的試験の中で、多くの場合、血管組織の機械的応答を評価するために使用されます。これらの実験の結果は、両方のネイティブおよび改変された組織源のために有用な予備的な情報を提供し、血管壁7-11の機械的挙動に特定の治療、疾患状態、または薬理学的化合物の効果を比較するために使用することができます。

軟組織の一軸性の機械的試験は、典型的には、最も一般的には犬の骨または7,8,12-14リングであり、比較的均一な幾何学的形状、を有するサンプルに対して行われます。しかし、これらの理想化されたジオメトリからの大きな逸脱が原因で検査システム内の組織の解剖、分離、およびクランプに関連する課題に発生する可能性があります。幾何学における任意の不均一性は、最終的には、不均一応力とひずみを生じさせますサンプルは実際のサンプル形状だけでなく、(グリップに対して)サンプルサイズと材料9,15,16の機械的性質に依存する不均一性の程度と、一軸延伸に供されるフィールド。磁場不均一性が重要である場合には、相対的なグリップ位置に基づいて、試料の歪みの計算は不正確と機械的挙動を評価するため、したがって不十分な基礎です。

ビデオ分析システムは広く、多くの場合、試料表面17,18に印加される高コントラスト色素マーカーを使用して、軟組織のひずみを測定するために使用されてきました。デジタル画像相関、変形前後の試験片の表面にグレーレベル強度値を比較することにより、全視野面の歪みを測定する光学計量技術は、ビデオと一緒に使用されてきた軟組織19-21の分析。デジタル画像相関のいくつかの利点がinterferometrに比べて存在します測定のために使用することができるICの方法。まず、非接触測定技術としては、測定システムは、試料に影響を及ぼしている方法による材料特性を変更するの交絡の影響を最小限に抑えます。第二に、それはそれほど厳格な測定環境を必要とし、他の方法よりも感度、解像度のより広い範囲を有します。第三に、全視野を捕捉する能力に恵まれて、この技術は、平均とローカル機械的応答の両方を特徴付けることができます。この方法の詳細については、読者は、Suttonの22本を参照することをお勧めします。

試料表面のひずみ場を得るために、2次元デジタル画像相関法(2D-DIC)を使用することができます。要するに、標本の画像は無負荷、様々なロードされた状態で撮影されています。最初の画像は、後続の計算のためのメッシュを形成する部分集合(M×M画素)と呼ばれる小さな正方形に分割され、2Dの歪み場。変形した試料中の各四角形の位置は、画像マッチングアルゴリズムを用いて得られます。各正方形の動きは次いで多項式フィッティングまたは有限要素補間を含む種々の方法を介して、変形勾配および株を導出するために使用することができ、変位場を得、画像ごとの画像、追跡されます。現在の原稿では、一軸引張試験および2D-DICの統合により、ネイティブの血管組織の表面歪み場の評価のための詳細な方法論を提供しています。

Protocol

注:以下の手順は、コロンビアのサウスカロライナ大学、サウスカロライナ州での制度的動物実験委員会によって承認されたプロトコルの一部として行われました。 1.組織の取得と解剖組織切開する前に、すべての手術器具を滅菌します。オートクレーブ手術用はさみ、ファイン標準の鉗子と同様に15分間15 PSIの圧力及び121℃の温度下での外科手術用ブレード。 地元の食?…

Representative Results

血管組織の傾斜一軸延伸試験から取得した機械的データは、与えられた変位速度で適用されるサンプルの変位との関係負荷対で構成されています。本研究では、一軸機械的試験と併せて、2D-DICは、種々の変形状態に直交する方向における試料の表面ひずみ場を測定するために使用されます。血管組織の粘弾性性質は、機械的なプレコンディショニングの前に荷重 – 変?…

Discussion

以前の研究は、試料株18,20,21,23,24を評価するために 、色素追跡ビデオ方法の広い範囲を使用していますが、私たちの現在の目的は、評価のための2D-DICと結合一軸引張試験に総合的な方法論を提供することです表面は、血管組織試料に菌株。サンプルは一軸負荷を受けるように、高解像度のカメラと、社内の画像解析ソフトウェアを用いて、歪み場は、所定の表面領域内で測定するこ?…

Declarações

The authors have nothing to disclose.

Acknowledgements

ソフトウェアおよび技術サポートは、相関ソリューションズ株式会社(www.correlatedsolutions.com)の礼儀でした。

Materials

Uniaxial tensile mechanical tester Enduratec 3230 AT/HR
Blue tissue marking dye http://www.ebay.com/itm/Tissue-Marking-Dye-in-Bottles-2oz-Bottle-1-ea-/201193551510?pt=LH_DefaultDomain_0&hash=item2ed811f696
Sprayer Anest-iwata CM-B Custom Micron B
Camera Point Grey GS2-GE-50S5M-C
Lens Tokina AT-X M100
Vascular tissue Caughman Inc
0.9% Sodium Chloride Injection PBS BAXTER HEALTHCARE CORP.
Vic_snap Correlated Solutions
Vic_2D Correlated Solutions
Wintest 4.1 Bose ElectroForce
Tissue adhesive  3M Vetbond  1469SB
Disinfectant  Fisher Scientific 04-355-13 Decon BDD Bacdown Detergent Disinfectant

Referências

  1. Holzapfel, G. A. Biomechanics of soft tissue. The handbook of materials behavior models. 3, 1049-1063 (2001).
  2. Vito, R. P., & Dixon, S. A. Blood vessel constitutive models-1995-2002. Annu Rev Biomed Eng. 5, 413-439 (2003).
  3. Dodson, R. B., Martin, J. T., Hunter, K. S., & Ferguson, V. L. Determination of hyperelastic properties for umbilical artery in preeclampsia from uniaxial extension tests. Eur J Obstet Gynecol Reprod Biol. 169, 207-212 (2013).
  4. Chuong, C. J., & Fung, Y. C. On residual stresses in arteries. J Biomech Eng. 108, 189-192 (1986).
  5. Borschel, G. H. et al. Tissue engineering of recellularized small-diameter vascular grafts. Tissue Eng. 11, 778-786 (2005).
  6. Wagenseil, J. E., & Mecham, R. P. Vascular extracellular matrix and arterial mechanics. Physiol Rev. 89, 957-989 (2009).
  7. Holzapfel, G. A. Determination of material models for arterial walls from uniaxial extension tests and histological structure. J Theor Biol. 238, 290-302 (2006).
  8. Tanaka, T. T., & Fung, Y.-C. Elastic and inelastic properties of the canine aorta and their variation along the aortic tree. J Biomech. 7, 357-370 (1974).
  9. Sokolis, D. Passive mechanical properties and structure of the aorta: segmental analysis. Acta physiologica. 190, 277-289 (2007).
  10. Twal, W. et al. Cellularized Microcarriers as Adhesive Building Blocks for Fabrication of Tubular Tissue Constructs. Ann Biomed Eng., 1-12 (2013).
  11. Shazly, T. et al. On the Uniaxial Ring Test of Tissue Engineered Constructs. Exp Mech., 1-11 (2014).
  12. Kim, J., & Baek, S. Circumferential variations of mechanical behavior of the porcine thoracic aorta during the inflation test. J Biomech. 44, 1941-1947 (2011).
  13. Li, L. et al. Determination of material parameters of the two-dimensional Holzapfel-Weizsacker type model based on uniaxial extension data of arterial walls. Comput Methods Biomech Biomed Engin. 16, 358-367 (2013).
  14. Li, L. et al. Determination of the material parameters of four-fibre family model based on uniaxial extension data of arterial walls. Comput Methods Biomech Biomed Engin. 17, 695-703 (2014).
  15. Hoeltzel, D. A., Altman, P., Buzard, K., & Choe, K.-i. Strip extensiometry for comparison of the mechanical response of bovine, rabbit, and human corneas. J Biomech Eng. 114, 202-215 (1992).
  16. Guo, X., & Kassab, G. S. Variation of mechanical properties along the length of the aorta in C57bl/6 mice. Am J Physiol Heart Circ Physiol. 285, H2614-H2622 (2003).
  17. Smutz, W., Drexler, M., Berglund, L., Growney, E., & An, K. Accuracy of a video strain measurement system. J Biomech. 29, 813-817 (1996).
  18. Genovese, K., Lee, Y. U., Lee, A. Y., & Humphrey, J. D. An improved panoramic digital image correlation method for vascular strain analysis and material characterization. J Mech Behav Biomed Mater. 27, 132-142 (2013).
  19. Wang, C. C., Deng, J.-M., Ateshian, G. A., & Hung, C. T. An automated approach for direct measurement of two-dimensional strain distributions within articular cartilage under unconfined compression. J Biomech Eng. 124, 557-567 (2002).
  20. Ning, J. et al. Deformation measurements and material property estimation of mouse carotid artery using a microstructure-based constitutive model. J Biomech Eng. 132, 121010 (2010).
  21. Sutton, M. A. et al. Strain field measurements on mouse carotid arteries using microscopic three-dimensional digital image correlation. J Biomed Mater Res A. 84, 178-190 (2008).
  22. Sutton, M. A., Orteu, J. J., & Schreier, H. Image correlation for shape, motion and deformation measurements: basic concepts, theory and applications. Springer Science & Business Media, (2009).
  23. Verhulp, E., van Rietbergen, B., & Huiskes, R. A three-dimensional digital image correlation technique for strain measurements in microstructures. J Biomech. 37, 1313-1320 (2004).
  24. Wang, C. C., Deng, J.-M., Ateshian, G. A., & Hung, C. T. An automated approach for direct measurement of two-dimensional strain distributions within articular cartilage under unconfined compression. Journal of Biomechanical Engineering. 124, 557-567 (2002).
  25. Franck, C., Hong, S., Maskarinec, S., Tirrell, D., & Ravichandran, G. Three-dimensional full-field measurements of large deformations in soft materials using confocal microscopy and digital volume correlation. Exp Mech. 47, 427-438 (2007).
  26. Garcia, A. et al. Experimental study and constitutive modelling of the passive mechanical properties of the porcine carotid artery and its relation to histological analysis: Implications in animal cardiovascular device trials. Med Eng Phys. 33, 665-676 (2011).
  27. Miller, K. How to test very soft biological tissues in extension? J Biomech. 34, 651-657 (2001).
  28. Sutton, M. A. in Springer handbook of experimental solid mechanics. 565-600, Springer, (2008).
  29. Han, H. C., & Fung, Y. C. Longitudinal strain of canine and porcine aortas. J Biomech. 28, 637-641 (1995).
  30. Sokolis, D. P. A passive strain-energy function for elastic and muscular arteries: correlation of material parameters with histological data. Med Biol Eng Comput. 48, 507-518 (2010).
  31. Zhou, B., Wolf, L., Rachev, A., & Shazly, T. A structure-motivated model of the passive mechanical response of the primary porcine renal artery. J Mech Med Biol. (2013).
  32. Zhou, B., Rachev, A., & Shazly, T. The biaxial active mechanical properties of the porcine primary renal artery. J Mech Behav Biomed Mater. 48, 28-37 (2015).
  33. Sommer, G., & Holzapfel, G. A. 3D constitutive modeling of the biaxial mechanical response of intact and layer-dissected human carotid arteries. J Mech Behav Biomed Mater. 5, 116-128 (2012).
check_url/pt/53625?article_type=t

Play Video

Citar este artigo
Zhou, B., Ravindran, S., Ferdous, J., Kidane, A., Sutton, M. A., Shazly, T. Using Digital Image Correlation to Characterize Local Strains on Vascular Tissue Specimens. J. Vis. Exp. (107), e53625, doi:10.3791/53625 (2016).

View Video