Amphotericin-B Mediated Perforated Patch-Clamp Analysis: An Electrophysiological Technique to Study Ionic Currents in Urinary Bladder Cells

Published: April 30, 2023

Abstract

Source: Malysz, J. et al. Preparation and Utilization of Freshly Isolated Human Detrusor Smooth Muscle Cells for Characterization of 9-Phenanthrol-Sensitive Cation Currents. J. Vis. Exp. (2020)

This video demonstrates the characterization of cationic currents from transient receptor potential melastatin type 4 (TRPM4) ion channels in detrusor smooth muscle cells using amphotericin-B mediated perforated patch-clamp technique. The technique helps evaluate the biophysical and pharmacological properties of ion channels.

Protocol

1. Recording Voltage-step Induced Cation Currents from DSM Cells Using Amphotericin-B Perforated Whole-cell Voltage Patch-Clamp Technique Pipette 0.25–1 mL of cell suspension onto a glass-bottom chamber sitting on a stage of an inverted microscope and allow the cells to adhere to the glass bottom. After incubation for at least 45 min, remove digestion solution (DS), from the bath and replace with the E solution (Table 1) by superfu…

Representative Results

Figure 1: Schematic illustration of steps involved in giga-seal formation and amphotericin-B perforation of human DSM cells. Illustrated are spatial positions of an amphotericin-B containing pipette and a DSM cell along with associated responses for membrane tests obtained in the commercial acquisition software (Table of Materials) by altering voltage steps (eith…

Disclosures

The authors have nothing to disclose.

Materials

5 ml polystyrene round-bottom tube Falcon 352054 Tubes for DS containing enzymes used in digestion steps
Amphotericin-B Fisher BP928-250 Used for patch/cell perforation
Amphotericin-B European Pharmacopoeia Reference Standards 5 Used for patch/cell perforation
Amphotericin-B Sigma-Aldrich A9528-100MG Used for patch/cell perforation
Analog vortex mixer VWR 58816-121
Aspartic acid Sigma-Aldrich A9006 Intracellular pipette solution
Bovine serum albumin Sigma-Aldrich A7906 DS
CaCl2 Sigma-Aldrich C1016 Extracellular solution and DS
Capillary Glass Sutter BF150-110-7.5 Capillary for preparation of pulled patch electrodes
Cesium hydroxide hydrate Sigma-Aldrich C8518 Intracellular pipette solution
Clampex ver. 10 software includes data acqusition (Clampex) and analysis (Clampfit) programs Axon Instruments/ Molecular Devices pCLAMP-10 Commerical software and part of patch-clamp rig setup
CsCl Sigma-Aldrich 203025 Extracellular and intracellular solutions
Dental wax Miltex Dental Wax Technologies, Inc. 18058351
Dimethyl sulfoxide (DMSO) Sigma-Aldrich D2650 Solvent
EGTA Sigma-Aldrich E3889 Ca2+ chelator, used in intracellular pipette solution
Flaming/Brown micropipette puller Sutter P-97 Required to pull electrodes with very fine tips
Floating foam tube rack/holder VWR Scientific 82017-634 Used for holding tubes with enzymes for temperature control
Glucose Sigma G8270
Glutamic acid (Na salt) Sigma-Aldrich G1626 DS
HEPES Sigma-Aldrich H3375 pH Buffer
KCl Fisher Scientific BP366-1 Extracellular solution
Low Noise Data Acquisition System Axon Instruments/ Molecular Devices Digidata 1440A Part of patch-clamp rig setup
Magnetic stirrer VWR 01-442-684
MgCl2 (hexahydrate) Sigma-Aldrich M2670 Extracellular and intracellular solutions
MicroForge Narishige MF-830 Used for fire-polishing electrodes
NaCl Sigma-Aldrich S7653 Extracellular and intracellular solutions
NaOH Sigma-Aldrich S8045
Nifedipine Sigma-Aldrich N7634 L-type voltage-gated Ca2+ channel blocker
Nikon inverted microscope, TS100 with T1-SM stage with 5x, 10x, 20x, and 40x objectives Nikon Discontinued Part of Patch-clamp rig setup
Non-metalic syringe needle, MicroFil WPI MF-34G-5 Filling of intracellular pipette solution
Pasteur pipette FisherBrand 13-678-20A Tips are broken off and fire-polished and used for titration of enzymatically treated tissues to release single DSM cells from pieces
Patch-clamp amplifier Axon Instruments/ Molecular Devices Axon Axopatch 200B Part of patch-clamp rig setup
PC computer DELL Custom configuration Part of patch-clamp rig setup
pH Meter Aspera Instruments PH700
Polyethylene tubing Intramedic 427-436 Tubing for superfusion of extracellular bath connected to glass-bottom recording chamber
Tetraethylammonium chloride Sigma-Aldrich T2265 Ion channel blocker of Kv and BK channels added to the extracellular bath solution
Weighting scale Mettler Toledo XS64
ZeissAxiovert 40C inverted microscope with 10x and 40x objectives Carl-Zeiss Discontinued Part of patch-clamp rig setup