Reversed-Phase High-Performance Liquid Chromatography: A Robust Method for Quantitation of Fluorescently Labeled and Derivatized Sialic Acids Isolated from Mouse Liver

Published: April 30, 2023

Abstract

Source: Cao, C. et al. Determination of Sialic Acids in Liver and Milk Samples of Wild-type and CMAH Knock-out Mice. J. Vis. Exp. (2017)

In this video, we demonstrate the detection of sialic acids, N-acetylneuraminic acid and N-glycolylneuraminic acid, using high-performance liquid chromatography, HPLC, following derivatization with a suitable fluorescent labeling reagent to aid in fluorescence detection.

Protocol

1. Fluorescence Derivatization of Sialic Acids Prepare 10 mL of OPD-solution, consisting of 100 mg of o-phenylenediamine and 208 mg of sodium hydrogen sulfite in 10 mL of distilled  H₂O. Add 20 µL of OPD-solution to the sialic acid samples. Vortex vigorously for 30 s and incubate samples for 4 h at 80 °C in the dark (i.e. wrap the microtubes in aluminum foil). Let the samples cool down for 5 min, add 80 µL of distilled  H₂O and centrifuge the t…

Disclosures

The authors have nothing to disclose.

Materials

Chemicals
N-acetylneuraminic acid Sigma A0812
N-glycolylneuraminic acid Sigma 50644 1 mg aliquot should be sufficient
O-phenylenediamine Sigma 694975
Sodium hydrogen sulfite J&K Scientific Ltd 75234
HPLC Analysis
High-recovery HPLC vial Agilent Technologies #5188-2788
HPLC System Shimadzu Nexera
Fluorescence Detector for HPLC Shimadzu RF-20Axs
HPLC Column Phenomenex Hyperclone ODS 250 x 4.6 mm
LCMS-grade H2O Merck Millipore #WX00011
LCMS-grade Acetonitrile Merck Millipore #100029 Hypergrade
Ammonium hydroxide solution Fluka #44273 Puriss. p.a.