Fluorescence-Detection Size-Exclusion Chromatography: A Technique to Identify the Integrity of Fluorescent Membrane Proteins upon Detergent Solubilization

Published: April 30, 2023

Abstract

Source: Bird, L. E., et al. Green Fluorescent Protein-based Expression Screening of Membrane Proteins in Escherichia coli. J. Vis. Exp. (2015)

In this video, we demonstrate a fluorescence-detection size-exclusion chromatography technique to screen for green fluorescent protein (GFP)-fused membrane protein stability in Escherichia coli following detergent solubilization. Proteins exhibiting a symmetrical peak in the size-exclusion profile with little or no free GFP indicate purified proteins with minimum degradation and aggregation, which can subsequently be selected for structure-function analysis.

Protocol

1.Solubilization and Analysis of the Membrane Fraction Thaw membrane fraction (if necessary), and for each detergent to be used for solubilization, aliquot 900 μl of the membrane suspension into a 1.5 ml polyallomer micro-centrifuge tube. Add 100 μl of each freshly prepared detergent (10% w/v solutions of DDM, DM, Cymal-6 and LDAO) to a final concentration of 1% to the specified 1.5 ml tube. For solubilization of eukaryotic membrane proteins cholesterol hemisuccinate (0.2% final concentrat…

Disclosures

The authors have nothing to disclose.

Materials

Cholesterol hemisuccinate  Sigma-Aldrich  C6512
CYMAL-6 (6-Cyclohexyl-1-Hexyl-βD-Maltoside) Anatrace  C326
DDM (n-Dodecyl β-maltoside)  Generon  D97002
DM (n-Decyl β-maltoside)  Generon  D99003
LDAO (N,N-Dimethyl-ndodecylamine N-oxide)  Generon  D71111
BenchMark Fluorescent Protein Standard  Life Technologies  LC5928
Optima L-100 Ultracentrifuge running 45-Ti rotor  Beckman Coulter  393253
Optima Max-XP ultracentrifuge running TLA-55 rotor  Beckman Coulter  393315
Prominence UFLC with RF-20A Fluorescence detector  Shimadzu
Sepharose 6 10/300 GL size exclusion column  GE Healthcare  17-5172-01

Tags

check_url/21101?article_type=t

Play Video

Cite This Article
Fluorescence-Detection Size-Exclusion Chromatography: A Technique to Identify the Integrity of Fluorescent Membrane Proteins upon Detergent Solubilization. J. Vis. Exp. (Pending Publication), e21101, doi: (2023).

View Video