Summary

Fabricação e caracterização de um Sistema Eletrônico Conformal-pele como para a gestão quantitativa, Ferida Cutânea

Published: September 02, 2015
doi:

Summary

This article presents methods to fabricate and characterize a conformal, skin-like electronic system and protocols for the use in clinical applications, particularly on cutaneous wound management.

Abstract

Recent advances in the development of electronic technologies and biomedical devices offer opportunities for non-invasive, quantitative assessment of cutaneous wound healing on the skin. Existing methods, however, still rely on visual inspections through various microscopic tools and devices that normally include high-cost, sophisticated systems and require well trained personnel for operation and data analysis. Here, we describe methods and protocols to fabricate a conformal, skin-like electronics system that enables conformal lamination to the skin surface near the wound tissues, which provides recording of high fidelity electrical signals such as skin temperature and thermal conductivity. The methods of device fabrication provide details of step-by-step preparation of the microelectronic system that is completely enclosed with elastomeric silicone materials to offer electrical isolation. The experimental study presents multifunctional, biocompatible, waterproof, reusable, and flexible/stretchable characteristics of the device for clinical applications. Protocols of clinical testing provide an overview and sequential process of cleaning, testing setup, system operation, and data acquisition with the skin-like electronics, gently mounted on hypersensitive, cutaneous wound and contralateral tissues on patients.

Introduction

No estudo clínico e de investigação biomédica, monitorização de cicatrização da ferida tem-se centrado sobre um método invasivo, que se baseia na avaliação histológica de tecido de mudança morfológica em feridas 1,2. Recentemente, avanços rápidos em tecnologias eletrônicas permitirá o desenvolvimento de ferramentas de análise de imagem e de alta precisão que pode inspecionar visualmente o processo de cicatrização através de imagens digitais 3,4 ou microscopia confocal de varredura e espectroscopia de 4,5. No entanto, estas abordagens de imagem exigem custos elevados, instrumentos ópticos de complicados e operações, e mais importante, os pacientes necessitam ser imobilizada durante o teste. Portanto, existe uma necessidade de novos dispositivos e sistemas que são quantitativa, não invasivo, de fácil utilização, de baixo custo, e multifuncional para oferecer o tratamento de feridas mais precisa.

Aqui, nós introduzir um sistema electrónico de pele-como que fornece preciso, mapeamento em tempo real de Condu temperatura e térmicasctivity e proporciona um nível preciso de aquecimento em locais ferida via conformada laminação do dispositivo de forma não invasiva. Este dispositivo apresenta uma classe de tecnologia, epidérmicas sistemas montados na pele eletrônicos que são projetados para corresponder às propriedades mecânicas e de material (espessura total, rigidez à flexão, módulos eficazes, e densidade de massa) da epiderme 6-9.

O dispositivo foi projetado em um, à prova de água-biocompatível pele-friendly, e forma reutilizável que pode ser lavado e desinfetado para aplicações clínicas em pacientes 10. O dispositivo electrónico conformada montado perto dos tecidos da ferida capta a fase de inflamação (de um processo de cicatrização de feridas), causado pelo aumento do fluxo de sangue e reacções enzimáticas para a ferida 11,12, através de gravação quantitativa de temperatura 8 e condutividade térmica 13, correlacionada com a hidratação . Estudos experimentais e computacionais determinar um projeto ótimo mecânica para accommodComeram movimentos naturais e aplicado estirpes sem fractura mecânica e capturar a física subjacente do alongamento mecânico do sistema electrónico semelhante à da pele que laminados conformalmente na superfície da pele, que proporciona a aquisição dos sinais de alta fidelidade.

Os protocolos descritos neste artigo apresentamos os métodos de microfabricação para sistemas eletrônicos de pele-like, testando preparação, incluindo a limpeza do dispositivo, a instalação do equipamento em um ambiente clínico, e aplicações clínicas para a monitorização quantitativa de temperatura e condutividade térmica em feridas cutâneas.

Protocol

As experiências para a fabricação do dispositivo, laminação pele, caracterização e mostrado nas Figuras 1, 2, e 4, envolveu duas voluntários todos realizados no Laboratório Bio-nanoengineering interface da Universidade da Virgínia (VCU), Richmond, VA, EUA. Este estudo foi aprovado pelo Institutional Review Board VCU (número de protocolo: HM20001454) e seguiu as diretrizes de pesquisa da Human Research VCU. Os dados do aparelho e clínicas apresentadas <strong…

Representative Results

A Figura 1 apresenta uma visão geral das características do conformada, sistema eletrônico de pele semelhante, projetado para quantitativa, gestão de feridas cutâneas em pacientes. O dispositivo eletrônico multifuncional consiste em estruturas de microescala fractal 3,14 e serpentina filamentar traça 9,17 em uma fina membrana elastomérica que oferece capacidade de estiramento mecânico excepcional e bendability. O dispositivo compatível que é totalmente fechado por camada…

Discussion

This article highlights the methods and protocols to fabricate a conformal, skin-like electronics system that enables conformal lamination near the wound tissues, which offers quantitative measurement of skin temperature and thermal conductivity mapping on the skin.

The key features include the utilization of novel techniques of materials transfer printing and hard-soft materials integration to design and develop the flexible/stretchable, soft electronic device. The use of biocompatible, elec…

Disclosures

The authors have nothing to disclose.

Acknowledgements

Este trabalho foi apoiado pelo financiamento de arranque da Escola de Engenharia, Virginia Commonwealth University e alguns dispositivos eletrônicos foram preparados nas instalações de microfabricação na Virginia Wright Microelectronics Center. Nós reconhecemos os pesquisadores que fizeram contribuições para o dispositivo e os dados clínicos (Figuras 3 e 5 do presente documento), adquirida a partir do artigo publicado 10. W.-HY graças Yoshiaki Hattori para o, software de gravação de dados feito sob medida.

Materials

3" Silicon wafer University Wafer, USA N/A Use as carrier to fabricate the device
Acetone Fisher Scientific, USA A18-1 Use to clean a wafer and to remove photoresist
Isopropanol (IPA) Fisher Scientific, USA A459-1 Use to clean a wafer
AZ4620 photoresist AZ Electrionic Materials, USA N/A Use to make patterns on metals and polymers
AZ400K developer AZ Electrionic Materials, USA N/A Use to develop AZ4620 photoresist
Chromium etchant Transene, USA 1020AC Use to etch Cr layer of device
Copper etchant Transene, USA ASP-100 Use to etch Cu layer of device
Sylgard 184 Silicone Elastomer Kit (PDMS) Dow Corning, USA 39100000 Use as a substrate for 'dry' retrieval
PI2545 polyimide HD MicroSystem, USA N/A Use to encapsulate metal layer
Solaris Smooth-On, USA N/A Use as substrate and to encapsulate device
Petridish Carolina, USA 741255 Use as mold to make substrate
Water-Soluble Wave Solder Tape 5414 3M, USA AM000000217 Use to retrive device from PDMS layer
High Activity Liquid Stainless Steel Flux Worthington, USA 331929 Use to remove oxidation layer on Cu
Flexible, micro-film cable Elform, USA N/A (customized) Use to make the electrical connection between the electronic device and the data acquisition system
pH Neutral Cleaner Australian Gold, USA N/A Use as disinfectant solution to clean device in clinical testing
Solder Kester, USA 24-6337-9703 Use as material to solder hard wires
Ultraviolet lamp Cole-Parmer, USA 97600-00 Use to activate PDMS layer as hydrophilic surface
Multiplexer FixYourBoard, USA U802 Use to acquire measurements from six sensing components 
DC/AC current source Keithley, USA 6221 Use to supply current
SMD Digital Hot Air Rework Station Aoyue, China 968A+ Use to solder hard wires, to electrically connect between the device and external instruments
Infrared camera FLIR, USA 435-0001-01-00 Use to take infrared images in experiment
Digital multimeter Fluke, USA 117 Use to check electrical connection
Lock-in amplifier Stanford Research System, USA SR830 Use to perform four-point-probe-measurement
Electron beam evaporator 9 scale Vacuum Products, USA N/A (customized) Use to deposit thin films (Cu and SiO2)

References

  1. Dargaville, T. R., et al. Sensors and imaging for wound healing: A review. Biosens Bioelectron. 41, 30-42 (2013).
  2. Panuncialman, J., Hammerman, S., Carson, P., Falanga, V. Wound edge biopsy sites in chronic wounds heal rapidly and do not result in delayed wound healing. J Invest Dermatol. 129, S47-S47 (2009).
  3. Hess, C. T., Kirsner, R. S. . Orchestrating Wound Healing: Assessing and Preparing the Wound Bed. Adv Skin Wound Care. 16 (5), 246-257 (2003).
  4. Lange-Asschenfeldt, S., et al. Applicability of confocal laser scanning microscopy for evaluation and monitoring of cutaneous wound healing. J Biomed Opt. 17 (7), (2012).
  5. Crane, N. J., Elster, E. A. Vibrational spectroscopy: a tool being developed for the noninvasive monitoring of wound healing. J Biomed Opt. 17 (7), (2012).
  6. Jeong, J. W., et al. Materials and Optimized Designs for Human-Machine Interfaces Via Epidermal Electronics. Adv Mater. 25 (47), 6839-6846 (2013).
  7. Kim, D. H., et al. Epidermal Electronics. Science. 333 (6044), 838-843 (2011).
  8. Webb, R. C., et al. Ultrathin conformal devices for precise and continuous thermal characterization of human skin (vol 12, pg 938). Nat Mater. 12, 1078-1078 (2013).
  9. Yeo, W. H., et al. Multifunctional Epidermal Electronics Printed Directly Onto the Skin. Adv Mater. 25 (20), 2773-2778 (2013).
  10. Hattori, Y., et al. Multifunctional skin-like electronics for quantitative, clinical monitoring of cutaneous wound healing. Adv Healthc Mater. 3 (10), 1597-1607 (2014).
  11. Guo, S., DiPietro, L. A. Factors Affecting Wound Healing. J Dent Res. 89 (3), 219-229 (2010).
  12. Matzeu, G., et al. Skin temperature monitoring by a wireless sensor. Ieee Ind Elec. , 3533-3535 (2011).
  13. Cahill, D. G. Thermal-Conductivity Measurement from 30-K to 750-K – the 3-Omega Method. Rev Sci Instrum. 61 (2), 802-808 (1990).
  14. Fan, J. A., et al. Fractal Design Concepts for Stretchable Electronics. Nature Commun. 5 (3266), (2013).
  15. Song, Y. M., et al. Digital cameras with designs inspired by the arthropod eye. Nature. 497 (7447), 95-99 (2013).
  16. Jeong, J. W., et al. Capacitive Epidermal Electronics for Electrically Safe, Long-Term Electrophysiological Measurements. Adv Healthc Mater. 3 (5), 642-648 (2013).
  17. Zhang, Y. H., et al. Experimental and Theoretical Studies of Serpentine Microstructures Bonded To Prestrained Elastomers for Stretchable Electronics. Adv Funct Mater. 24 (14), 2028-2037 (2014).
  18. Carlson, A., Bowen, A. M., Huang, Y. G., Nuzzo, R. G., Rogers, J. A. Transfer Printing Techniques for Materials Assembly and Micro/Nanodevice Fabrication. Adv Mater. 24 (39), 5284-5318 (2012).
  19. Yeo, W. H., Webb, R. C., Lee, W., Jung, S., Rogers, J. A. Bio-integrated electronics and sensor systems. Proc Spie. 8725, (2013).
  20. Chung, H. J., et al. Fabrication of Releasable Single-Crystal Silicon–Metal Oxide Field-Effect Devices and Their Deterministic Assembly on Foreign Substrates. Adv Funct Mater. 21 (16), 3029-3036 (2011).
  21. Kim, H. S., et al. Unusual strategies for using indium gallium nitride grown on silicon (111) for solid-state lighting. Proc Natl Acad Sci U.S.A. 108 (25), 10072-10077 (2011).
  22. Padilla-Medina, J. A., et al. Assessment technique for acne treatments based on statistical parameters of skin thermal images. J Biomed Opt. 19 (4), 046019-046019 (2014).
check_url/53037?article_type=t

Play Video

Cite This Article
Lee, W., Kwon, O., Lee, D. S., Yeo, W. Fabrication and Characterization of a Conformal Skin-like Electronic System for Quantitative, Cutaneous Wound Management. J. Vis. Exp. (103), e53037, doi:10.3791/53037 (2015).

View Video