Summary

Centrifuga dispositivo a microfluidi capillare a base per la Formazione Size-controllabile di Monodisperse microgocce

Published: February 22, 2016
doi:

Summary

Here, we demonstrate a simple production method for size-controllable, monodisperse, water-in-oil (W/O) microdroplets using a capillary-based centrifugal microfluidic device. This method requires only a small sample volume and enables high-yield production. We expect this method will be useful for rapid biochemical and cellular analyses.

Abstract

Qui, dimostriamo un metodo semplice per la produzione rapida di, monodisperse, W microgocce / O size-controllabile mediante un dispositivo di microfluidica centrifuga capillare-based. microgocce W / O sono stati recentemente utilizzati in potenti metodi che consentono miniaturizzazione esperimenti chimici. Pertanto, lo sviluppo di un metodo versatile per cedere monodisperse W è necessario / O microgocce. Abbiamo sviluppato un metodo per generare monodisperse W microgocce / O basato su un dispositivo assialsimmetrici centrifuga capillare basata co-scorre microfluidica. Siamo riusciti a controllare la dimensione delle microgoccioline regolando l'orifizio capillare. Il nostro metodo richiede un'attrezzatura che è più facile da usare rispetto ad altre tecniche di microfluidica, richiede solo una piccola quantità (0,1-1 ml) di soluzione campione per incapsulamento, e consente la produzione di centinaia di migliaia numero di microgocce / O W al secondo . Ci aspettiamo che questo metodo vi aiuterà studi biologici che richiedono prezioso s biologicoamples conservando il volume dei campioni per una rapida biochimica analisi quantitativa e studi biologici.

Introduction

W / O microgocce 1-5 avere molte applicazioni importanti per lo studio della biochimica e bioingegneria, tra cui la sintesi proteica 6, proteine ​​cristallizzazione 7, emulsione PCR 8,9, incapsulamento della pila 10, e la costruzione di sistemi simili alle cellule artificiali 5,6. Per produrre microgocce W / O per queste applicazioni, criteri importanti sono il controllo della dimensione e monodispersibility delle microgocce W / O. Dispositivi microfluidici per fare monodisperse, dimensioni controllabili W / O microgocce 11 sono basati sul metodo di co-flowing 12,13, metodo flow-focalizzazione 14,15, e il metodo T-svincolo 16 in microcanali. Sebbene questi metodi producono altamente monodisperse microgocce / O W, il processo di microfabbricazione richiede una gestione complicata e tecniche specializzate per la preparazione di microcanali, e richiede anche una grande quantità di soluzione campione (almeno diverse centinaia81; l) a causa del volume morto inevitabili nelle pompe a siringa e tubi che conducono la soluzione campione microcanali. Così, un metodo di facile utilizzo e basso volume morto per generare monodisperse W è necessario / O microgocce.

Questo documento, insieme a video di procedure sperimentali, descrive un dispositivo di centrifuga capillare basata assialsimmetrici co-scorre microfluidica 17 per la generazione di cellule di dimensioni, monodisperse W / O microgocce (Figura 1). Questo semplice metodo raggiunge monodispersity dimensioni e controllabilità dimensioni. Si richiede solo un mini-centrifuga da tavolo e un dispositivo di co-scorre assialsimmetrici capillare basata microfluidica fissa in un microtubo di campionamento. Il nostro metodo necessita solo un volume molto piccolo (0,1 mL), e non perde volume significativo del campione.

Protocol

1. Realizzazione di un dispositivo a microfluidi capillare basata Impostare dei titolari Nota: Il disegno del supporto è presentato in Figura 2A. Tagliare ciascuno dei quattro dischi dei titolari (Figura 2A (i) – (iv)) da 2 mm di spessore piatto di plastica poliacetale utilizzando una fresatrice. Utilizzare le seguenti dimensioni per ciascuna delle quattro dischi del titolare: (i) il disco 1 di diametro 8,5 millimetri, foro capillare (CH) di …

Representative Results

In questo studio, presentiamo un metodo semplice per la generazione di microgocce W / O cellulari dimensioni utilizzando un dispositivo microfluidico centrifuga capillare basata (Figura 1). Il dispositivo microfluidico era composto da un supporto capillare (Figura 2B), due capillari di vetro (vetro capillari interno ed esterno in Figura 3C), e un microtubo contenente un olio compreso tensioattivo. Abbiamo iniettato 0,1 ml di soluzione di…

Discussion

L'utilizzo di questo dispositivo, il monodisperse W microgocce / O sono stati generati da Plateau-Rayleigh instabilità di un jet-flusso 17. L'esame microscopico non ha rivelato la presenza di goccioline satellitari. Nella fabbricazione del dispositivo, tre passi critici sono essenziali per generare successo monodisperse microgocce / O W. Innanzitutto, per fornire un flusso rettilineo di tensioattivo contenente olio e soluzione acquosa, i fori capillari di quattro dischi devono essere disposte in un m…

Disclosures

The authors have nothing to disclose.

Acknowledgements

This work was supported by the PRESTO “Design and Control of Cellular Functions” research area of the Japan Science and Technology Agency (JST), a Grant-in-Aid for Scientific Research of Innovative Areas “Molecular Robotics” (Project No. 24104002) from the Ministry of Education, Culture, Sports, Science and Technology (MEXT), Japan, Grant-in-Aid for Young Scientists (A) (Project No. 24680033) and Scientific Research (B) (Project No. 26280097) from the Japan Society for the Promotion of Science (JSPS), and the Creative Design for Bioscience and Biotechnology course of the School of Bioscience and Biotechnology at Tokyo Tech.

Materials

2-mm-thick polyacetal plastic plate Tool Nikkyo Technos, Co., Ltd. (Japan) 244-6432-08
Milling machine Tool Roland DG Co., Ltd. (Japan) MDX-40A
End Mill RSE230-0.5*2.5 Tool NS Tool Co., Ltd. (Japan) 01-00644-00501
M2*40 screws Tool Jujo Synthetic Chemistry Labo. (Japan) 0001-024
Glass Capillry Puller Tool Narishige (Japan) PC-10
Microforge Tool Narishige (Japan) MF-900
Inner Glass Capillary Tool Narishige (Japan) G-1
Outer Glass Capillary Tool World Precision Instruments Inc. (USA) 1B200-6
1.5 ml Sample tube Tool INA OPTIKA CO.,LTD (Japan) ST-0150F
Hexadecane Reagent Wako Pure Chemical Industries Ltd. (Japan) 080-03685 
Sorbitan monooleate (Span 80) Reagent Tokyo Chemical Industry Co., Ltd. (Japan) S0060
Milli Q system Reagent Merck Millipore Corporation (Germany) ZRQSVP030
Swinging-out-type Mini-centrifuge Tool Hitech Co., Ltd. (Japan) ATT101
Digital Microscope Tool KEYENCE Corporation (Japan) VHX-2001

References

  1. Song, H., Chen, D. L., Ismagilov, R. F. Reactions in droplets in microfluidic channels. Angew. Chem., Int. Ed. 45 (44), 7336-7356 (2006).
  2. Huebner, A., et al. Microdroplets: a sea of applications?. Lab Chip. 8, 1244-1254 (2008).
  3. Taly, V., Kelly, B. T., Griffiths, A. D. Droplets as microreactors for highthroughput biology. ChemBioChem. 8 (3), 263-272 (2007).
  4. Teh, S. Y., Lin, R., Hung, L. H., Lee, A. P. . Droplet microfluidics. Lab Chip . 8, 198-220 (2008).
  5. Takinoue, M., Takeuchi, S. Droplet microfluidics for the study of artificial cells. Anal. Bioanal. Chem. 400 (6), 1705-1716 (2011).
  6. Hase, M., Yamada, A., Hamada, T., Baigl, D., Yoshikawa, K. Manipulation of cell-sized phospholipid-coated microdroplets and their use as biochemical microreactors. Langmuir. 23 (2), 348-352 (2007).
  7. Zheng, B., Tice, J. D., Roach, L. S., Ismagilov, R. F. A Droplet-Based, Composite PDMS/Glass Capillary Microfluidic System for Evaluating Protein Crystallization Conditions by Microbatch and Vapor-Diffusion Methods with On-Chip X-Ray Diffraction. Angew. Chem., Int. Ed. 43 (19), 2508-2511 (2004).
  8. Nakano, M., et al. Single-molecule PCR using water-in-oil emulsion. J. Biotechnol. 102 (2), 117-124 (2003).
  9. Diehl, F., et al. BEAMing: single-molecule PCR on microparticles in water-in-oil emulsions. Nat. Methods. 3, 551-559 (2006).
  10. He, M., et al. Selective encapsulation of single cells and subcellular organelles into picoliter- and femtoliter-volume droplets. Anal. Chem. 77 (6), 1539-1544 (2005).
  11. Baroud, C., Gallaire, F., Dangla, R. Dynamics of microfluidic droplets. Lab Chip. 10, 2032-2045 (2010).
  12. Utada, A. S., Nieves, A. F., Stone, H. A., Weitz, D. A. Dripping to jetting transitions in coflowing liquid streams. Phys. Rev. Lett. 99 (9), 094502 (2007).
  13. Cramer, C., Fischer, P., Windhab, E. J. Drop formation in a co-flowing ambient fluid. Chem. Eng. Sci. 59 (15), 3045-3058 (2004).
  14. Anna, S. L., Bontoux, N., Stone, H. A. Formation of dispersions using "flow-focusing" in microchannels. Appl. Phys. Lett. 82, 364-366 (2003).
  15. Takeuchi, S., Garstecki, P., Weibel, D. B., Whitesides, G. M. An axisymmetric flow-focusing microfluidic device. Adv. Mater. 17 (8), 1067-1072 (2005).
  16. Thorsen, T., Roberts, R. W., Arnold, F. H., Quake, S. R. Dynamic pattern formation in a vesicle-generating microfluidic device. Phys. Rev. Lett. 86 (18), 4163-4166 (2001).
  17. Yamashita, H., et al. Generation of monodisperse cell-sized microdroplets using a centrifuge-based axisymmetric co-flowing microfluidic device. J. Biosci. Biotech. 119 (4), 492-495 (2015).
  18. Maeda, K., Onoe, H., Takinoue, M., Takeuchi, S. Controlled synthesis of 3D multi-compartmental particles with centrifuge-based microdroplet formation from a multi-barrelled capillary. Adv. Mater. 24 (10), 1340-1346 (2012).
check_url/53860?article_type=t

Play Video

Cite This Article
Morita, M., Yamashita, H., Hayakawa, M., Onoe, H., Takinoue, M. Capillary-based Centrifugal Microfluidic Device for Size-controllable Formation of Monodisperse Microdroplets. J. Vis. Exp. (108), e53860, doi:10.3791/53860 (2016).

View Video