Summary

Desenvolvimento de um<em> In Vitro</em> Ocular Platform para testar as lentes de contato

Published: April 06, 2016
doi:

Summary

Atuais modelos in vitro para avaliação de lentes de contacto (CLS) e outras aplicações relacionadas com o olho são severamente limitados. A plataforma apresentada ocular simula o fluxo lacrimal fisiológico, o volume lacrimal, de exposição ao ar e desgaste mecânico. Este sistema é altamente versátil e pode ser aplicado a várias análises in vitro com o CLS.

Abstract

Currently, in vitro evaluations of contact lenses (CLs) for drug delivery are typically performed in large volume vials,1-6 which fail to mimic physiological tear volumes.7 The traditional model also lacks the natural tear flow component and the blinking reflex, both of which are defining factors of the ocular environment. The development of a novel model is described in this study, which consists of a unique 2-piece design, eyeball and eyelid piece, capable of mimicking physiological tear volume. The models are created from 3-D printed molds (Polytetrafluoroethylene or Teflon molds), which can be used to generate eye models from various polymers, such as polydimethylsiloxane (PDMS) and agar. Further modifications to the eye pieces, such as the integration of an explanted human or animal cornea or human corneal construct, will permit for more complex in vitro ocular studies. A commercial microfluidic syringe pump is integrated with the platform to emulate physiological tear secretion. Air exposure and mechanical wear are achieved using two mechanical actuators, of which one moves the eyelid piece laterally, and the other moves the eyeballeyepiece circularly. The model has been used to evaluate CLs for drug delivery and deposition of tear components on CLs.

Introduction

Duas áreas significativas de interesse dentro da arena lente de contato (CL) incluem desconforto e o desenvolvimento de aplicações CL novos. Elucidar os mecanismos subjacentes CL desconforto é um problema que iludiu o campo durante décadas. 8 O desenvolvimento do romance, o CLS funcionais, tais como dispositivos de entrega de drogas 1,3,9 e biossensores, 10-12 é uma área de interesse crescente, com mercados potenciais substanciais. Em ambas as circunstâncias, um sofisticado modelo in vitro poderá fornecer informações relevantes para ajudar com a seleção de materiais de lentes ou características de design apropriado durante a fase de desenvolvimento. Infelizmente, atual em modelos in vitro para avaliar o CLS e outras aplicações relacionadas oculares são relativamente bruto e sem sofisticação. Tradicionalmente, os estudos in vitro que avaliam CL deposição do filme lacrimal ou entrega da droga são realizadas em frascos de volume estáticos, contendo um grande volume de fluido fixo, que excelenTLY excede valores fisiológicos. Além disso, este modelo simples falta a componente de fluxo natural das lágrimas e o reflexo de piscar, ambas as quais são factores de definir o meio ocular.

O desenvolvimento de um "modelo" sofisticado, fisiologicamente relevante olho exigirá uma abordagem multidisciplinar e exigem substancial na validação vivo. Por estas razões, o quadro fundamental para o nosso modelo in vitro olho é altamente versátil, de modo que o modelo pode ser melhorado continuamente por meio de upgrades e modulações futuras. Até à data, o modelo é capaz de simular o volume lacrimal, do fluxo lacrimal, o desgaste mecânico e exposição ao ar. O objectivo é o de criar um modelo in vitro que irá fornecer resultados significativos, que é complementar ao preditivo e in vivo e ex vivo observações.

Protocol

Todos os experimentos foram preenchidos em conformidade e cumprimento de todas as orientações pertinentes delineadas pelo comité de ética pesquisa animal da Universidade de Waterloo de. Os olhos de bovinos são generosamente doados a partir de um matadouro local. 1. Modelo de Olho Concepção e produção de moldes de 13 Design os modelos de olho de acordo com as dimensões médias de fisiológicas olhos humanos adultos. 13 Deixe um espaço de 250 mm entre …

Representative Results

Os moldes olho sintetizados obtidos a partir da loja de máquinas de impressão e 3-D são apresentados na Figura 1. Esses moldes podem ser usados ​​com uma variedade de polímeros, tais como PDMS e de agarose, para produzir oculares com as propriedades desejadas. O conjunto de gesto da plataforma de modelo de olho com uma bomba de seringa de microfluidos é mostrado na Figura 2. A plataforma simula o desgaste mecânico através da rotação da peça…

Discussion

Há três passos críticos no âmbito do protocolo que requerem atenção especial: concepção e produção de moldes (seção 1.1), a montagem da plataforma (seção 2.2.1-2.2.3), e acompanhamento do funcionamento experimental (seção 2.2.4-2.2.7 ). Em termos de concepção e produção de moldes (seção 1.1), a peça globo ocular deve ser projetado de acordo com as dimensões de uma córnea humana. No entanto, pode exigir vários protótipos do molde antes de uma peça do globo ocular pode ser criado que se encaixa…

Disclosures

The authors have nothing to disclose.

Acknowledgements

Os autores gostariam de reconhecer a nossa fonte de financiamento NSERC 20/20 Rede para o Desenvolvimento da Advanced Materials oftalmológica.

Materials

Arduino Uno R3 (Atmega328 – assembled) Adafruit 50 Board
Stepper motor Adafruit 324 Motor and Motor shield
Equal Leg Coupler 1.6mm 1/16" VWR CA11009-280 50 pcs of tube connector
Tubing PT/SIL 1/16"x1/8" VWR 16211-316 Case of 50feet
PDMS Dow Corning Sylgard 184 Solar Cell Encapsulation
Agarose, Type 1-A, low EEO Sigma-Aldrich A0169-25G
PHD UltraTM Harvard Apparatus 703006 MicroFluidic Pump
Bovine cornea Cargill, Guelph/ON
Soldidworks Dassault Systemes Software
3-D printing University of Waterloo – 3D Print Centre
Dissection tools Fine Science Tools General dissection tools
Medium 199 Sigma-Aldrich Culture medium storage for cornea
Fetal bovine serum Thermo Fisher Add to culture medium, 3% total volume

References

  1. Phan, C. M., Subbaraman, L. N., Jones, L. In vitro drug release of natamycin from beta-cyclodextrin and 2-hydroxypropyl-beta-cyclodextrin-functionalized contact lens materials. J Biomater Sci Polym Ed. 25, 1907-1919 (2014).
  2. Peng, C. C., Kim, J., Chauhan, A. Extended delivery of hydrophilic drugs from silicone-hydrogel contact lenses containing vitamin E diffusion barriers. Biomaterials. 31, 4032-4047 (2010).
  3. Hui, A., Willcox, M., Jones, L. In vitro and in vivo evaluation of novel ciprofloxacin-releasing silicone hydrogel contact lenses. Invest Ophthalmol Vis Sci. 55, 4896-4904 (2014).
  4. Boone, A., Hui, A., Jones, L. Uptake and release of dexamethasone phosphate from silicone hydrogel and group I, II, and IV hydrogel contact lenses. Eye Contact Lens. 35, 260-267 (2009).
  5. Lorentz, H., Heynen, M., Trieu, D., Hagedorn, S. J., Jones, L. The impact of tear film components on in vitro lipid uptake. Optom Vis Sci. 89, 856-867 (2012).
  6. Hall, B., Phan, C. M., Subbaraman, L., Jones, L. W., Forrest, J. Extraction versus in situ techniques for measuring surface-adsorbed lysozyme. Optom Vis Sci. 91, 1062-1070 (2014).
  7. Mishima, S., Gasset, A., Klyce, S. D., Baum, J. L. Determination of tear volume and tear flow. Invest Ophthalmol Vis Sci. 5, 264-276 (1966).
  8. Nichols, J. J., et al. The TFOS international workshop on contact lens discomfort: executive summary. Invest Ophthalmol Vis Sci. 54, 7-13 (2013).
  9. Peng, C. C., Burke, M. T., Carbia, B. E., Plummer, C., Chauhan, A. Extended drug delivery by contact lenses for glaucoma therapy. J Control Release. 162, 152-158 (2012).
  10. Faschinger, C., Mossbock, G. Continuous 24 h monitoring of changes in intraocular pressure with the wireless contact lens sensor Triggerfish. First results in patients. Der Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft. 107, 918-922 (2010).
  11. Shaw, A. J., Davis, B. A., Collins, M. J., Carney, L. G. A technique to measure eyelid pressure using piezoresistive sensors. IEEE transactions on bio-medical engineering. 56, 2512-2517 (2009).
  12. Liao, Y. T., Yao, H. F., Lingley, A., Parviz, B., Otis, B. P. A 3-mu W CMOS glucose sensor for wireless contact-lens tear glucose monitoring. Ieee J Solid-St Circ. 47, 335-344 (2012).
  13. Coster, D. J. . Cornea. , (2002).
  14. Parekh, M., et al. A simplified technique for in situ excision of cornea and evisceration of retinal tissue from human ocular globe. Journal of visualized experiments : JoVE. , e3765 (2012).
  15. Way, S. Gear and pinion. US patent. , (1942).
  16. Lorentz, H., et al. Contact lens physical properties and lipid deposition in a novel characterized artificial tear solution. Molecular vision. 17, 3392-3405 (2011).
  17. Furukawa, R. E., Polse, K. A. Changes in tear flow accompanying aging. American journal of optometry and physiological optics. 55, 69-74 (1978).
  18. Bajgrowicz, M., Phan, C. M., Subbaraman, L., Jones, L. Release of ciprofloxacin and moxifloxacin from daily disposable contact lenses from an in vitro eye model. Invest Ophthalmol Vis Sci. , (2015).
  19. Luensmann, D., Zhang, F., Subbaraman, L., Sheardown, H., Jones, L. Localization of lysozyme sorption to conventional and silicone hydrogel contact lenses using confocal microscopy. Current eye research. 34, 683-697 (2009).
  20. Tieppo, A., Pate, K. M., Byrne, M. E. In vitro controlled release of an anti-inflammatory from daily disposable therapeutic contact lenses under physiological ocular tear flow. Eur J Pharm Biopharm. 81, 170-177 (2012).
  21. Ali, M., et al. Zero-order therapeutic release from imprinted hydrogel contact lenses within in vitro physiological ocular tear flow. J Control Release. 124, 154-162 (2007).
  22. White, C. J., McBride, M. K., Pate, K. M., Tieppo, A., Byrne, M. E. Extended release of high molecular weight hydroxypropyl methylcellulose from molecularly imprinted, extended wear silicone hydrogel contact lenses. Biomaterials. 32, 5698-5705 (2011).
  23. Kaczmarek, J. C., Tieppo, A., White, C. J., Byrne, M. E. Adjusting biomaterial composition to achieve controlled multiple-day release of dexamethasone from an extended-wear silicone hydrogel contact lens. J Biomater Sci Polym Ed. 25, 88-100 (2014).
  24. Mohammadi, S., Postnikoff, C., Wright, A. M., Gorbet, M. Design and development of an in vitro tear replenishment system. Ann Biomed Eng. 42, 1923-1931 (2014).
  25. Lorentz, H., Heynen, M., Khan, W., Trieu, D., Jones, L. The impact of intermittent air exposure on lipid deposition. Optom Vis Sci. 89, 1574-1581 (2012).
  26. Peng, C. C., Fajardo, N. P., Razunguzwa, T., Radke, C. J. In vitro spoilation of silicone-hydrogel soft contact lenses in a model-blink cell. Optom Vis Sci. 92, 768-780 (2015).
  27. Liu, P., et al. Dissolution studies of poorly soluble drug nanosuspensions in non-sink conditions. AAPS PharmSciTech. 14, 748-756 (2013).
check_url/53907?article_type=t

Play Video

Cite This Article
Phan, C., Walther, H., Gao, H., Rossy, J., Subbaraman, L. N., Jones, L. Development of an In Vitro Ocular Platform to Test Contact Lenses. J. Vis. Exp. (110), e53907, doi:10.3791/53907 (2016).

View Video