Summary

פיתוח של<em> במבחנה</em> פלטפורמת Ocular כדי לבדוק עדשות מגע

Published: April 06, 2016
doi:

Summary

נוכחי במבחנת מודלים להערכת עדשות מגע (CLS) ויישומים אחרים הקשורים עין מוגבלים קשה. הפלטפורמה העינית הציגה מדמת זרימה מדמיעה פיזיולוגית, נפח מדמיע, חשיפה לאוויר ובלאי מכאני. מערכת זו היא תכליתית מאוד יכולה להיות מיושמת שונה במבחנת מנתח עם CLS.

Abstract

Currently, in vitro evaluations of contact lenses (CLs) for drug delivery are typically performed in large volume vials,1-6 which fail to mimic physiological tear volumes.7 The traditional model also lacks the natural tear flow component and the blinking reflex, both of which are defining factors of the ocular environment. The development of a novel model is described in this study, which consists of a unique 2-piece design, eyeball and eyelid piece, capable of mimicking physiological tear volume. The models are created from 3-D printed molds (Polytetrafluoroethylene or Teflon molds), which can be used to generate eye models from various polymers, such as polydimethylsiloxane (PDMS) and agar. Further modifications to the eye pieces, such as the integration of an explanted human or animal cornea or human corneal construct, will permit for more complex in vitro ocular studies. A commercial microfluidic syringe pump is integrated with the platform to emulate physiological tear secretion. Air exposure and mechanical wear are achieved using two mechanical actuators, of which one moves the eyelid piece laterally, and the other moves the eyeballeyepiece circularly. The model has been used to evaluate CLs for drug delivery and deposition of tear components on CLs.

Introduction

שני תחומים משמעותיים של עניין בתוך עדשות מגע (CL) בזירה כוללים חוסר נוחות ופיתוח יישומי CL רומן. הבהרת המנגנונים אי נוחות CL היא סוגיה הזה חמק מידיה בתחום במשך עשרות שנים. 8 התפתחות הרומן, CLS פונקציונלי, כגון מכשירים להולכת תרופות 1,3,9 ו biosensors, 10-12 הוא שטח של העניין הגובר, עם שווקים פוטנציאליים משמעותיים. בשתי הנסיבות, מתוחכם במודל חוץ גופייה יספק מידע רלוונטי על מנת לסייע בבחירת חומרי עדשה מתאימה או מאפייני עיצוב במהלך שלב הפיתוח. למרבה הצער, נוכחי במודלים במבחנה להערכת CLS ויישומים הקשורים עין שניות הם יחסית גולמיים ולא מתוחכמים. באופן מסורתי, במבחנת מחקרי CL הערכה בתצהיר שכבת דמעות או משלוח סמים מבוצעים בקבוקוני נפח סטטי, גדולים המכילים נפח נוזל קבוע, אשר greatly עולה כמויות פיסיולוגיות. יתר על כן, מודל פשוט זה חסר מרכיב זרימת דמעה הטבעי רפלקס המצמוץ, אשר שניהם מגדירים גורמי הסביבה העינית.

הפיתוח של "מודל" העין מתוחכם, רלוונטי מבחינה פיזיולוגית יחייב בגישה רב-תחומית ודורשים משמעותי אימות vivo. מסיבות אלה, במסגרת היסוד עבור דגם העין במבחנה שלנו היא תכליתית מאוד, כך המודל ניתן לשפר ללא הרף באמצעות שדרוגי מודולציות בעתיד. נכון להיום, המודל הוא מסוגל לדמות נפח מדמיע, זרימה מדמיעה, מכנים ללבוש חשיפה לאוויר. המטרה היא ליצור מודל במבחנה שיספק תוצאות משמעותיות, אשר הינה מידע צופה פני ו מחמיא in vivo ותצפיות vivo לשעבר.

Protocol

כל הניסויים הושלמו בהתאם ועמיד בכל ההנחיות הרלוונטיות שהתוו ועדת האתיקה במחקר בבעלי החיים של האוניברסיטה ווטרלו. עיני השור שנתרמות בנדיבות מן מטבחיים מקומי. 1. עין דגם עיצוב והפקה של תבניות <…

Representative Results

תבניות העין המסונתזות מתקבל המכונה בחנות ומן 3-D הדפסה מוצגות באיור 1. תבניות אלה יכולים לשמש עם מגוון רחב של פולימרים, כגון PDMS ו agarose, לייצר eyepieces עם התכונות הרצויות. עצרת הסימן של פלטפורמת מודל עין עם משאבת מזרק microfluidic מוצגת באיור 2. ?…

Discussion

ישנם שלושה שלבים קריטיים בתוך הפרוטוקול מחייבות התייחסות מיוחדת: עיצוב וייצור של תבניות (סעיף 1.1), הרכבת פלטפורמה (סעיף 2.2.1-2.2.3), וניטור בטווח הניסוי (סעיף 2.2.4-2.2.7 ). במונחים של העיצוב והייצור של תבניות (סעיף 1.1), יצירת גלגל העין צריכה להיות מתוכננת על פי המידות של קרנית אדם…

Disclosures

The authors have nothing to disclose.

Acknowledgements

המחברים מבקשים להודות מקור מימון שלנו NSERC 20/20 הרשת לפיתוח חומרים עיניים מתקדמים.

Materials

Arduino Uno R3 (Atmega328 – assembled) Adafruit 50 Board
Stepper motor Adafruit 324 Motor and Motor shield
Equal Leg Coupler 1.6mm 1/16" VWR CA11009-280 50 pcs of tube connector
Tubing PT/SIL 1/16"x1/8" VWR 16211-316 Case of 50feet
PDMS Dow Corning Sylgard 184 Solar Cell Encapsulation
Agarose, Type 1-A, low EEO Sigma-Aldrich A0169-25G
PHD UltraTM Harvard Apparatus 703006 MicroFluidic Pump
Bovine cornea Cargill, Guelph/ON
Soldidworks Dassault Systemes Software
3-D printing University of Waterloo – 3D Print Centre
Dissection tools Fine Science Tools General dissection tools
Medium 199 Sigma-Aldrich Culture medium storage for cornea
Fetal bovine serum Thermo Fisher Add to culture medium, 3% total volume

References

  1. Phan, C. M., Subbaraman, L. N., Jones, L. In vitro drug release of natamycin from beta-cyclodextrin and 2-hydroxypropyl-beta-cyclodextrin-functionalized contact lens materials. J Biomater Sci Polym Ed. 25, 1907-1919 (2014).
  2. Peng, C. C., Kim, J., Chauhan, A. Extended delivery of hydrophilic drugs from silicone-hydrogel contact lenses containing vitamin E diffusion barriers. Biomaterials. 31, 4032-4047 (2010).
  3. Hui, A., Willcox, M., Jones, L. In vitro and in vivo evaluation of novel ciprofloxacin-releasing silicone hydrogel contact lenses. Invest Ophthalmol Vis Sci. 55, 4896-4904 (2014).
  4. Boone, A., Hui, A., Jones, L. Uptake and release of dexamethasone phosphate from silicone hydrogel and group I, II, and IV hydrogel contact lenses. Eye Contact Lens. 35, 260-267 (2009).
  5. Lorentz, H., Heynen, M., Trieu, D., Hagedorn, S. J., Jones, L. The impact of tear film components on in vitro lipid uptake. Optom Vis Sci. 89, 856-867 (2012).
  6. Hall, B., Phan, C. M., Subbaraman, L., Jones, L. W., Forrest, J. Extraction versus in situ techniques for measuring surface-adsorbed lysozyme. Optom Vis Sci. 91, 1062-1070 (2014).
  7. Mishima, S., Gasset, A., Klyce, S. D., Baum, J. L. Determination of tear volume and tear flow. Invest Ophthalmol Vis Sci. 5, 264-276 (1966).
  8. Nichols, J. J., et al. The TFOS international workshop on contact lens discomfort: executive summary. Invest Ophthalmol Vis Sci. 54, 7-13 (2013).
  9. Peng, C. C., Burke, M. T., Carbia, B. E., Plummer, C., Chauhan, A. Extended drug delivery by contact lenses for glaucoma therapy. J Control Release. 162, 152-158 (2012).
  10. Faschinger, C., Mossbock, G. Continuous 24 h monitoring of changes in intraocular pressure with the wireless contact lens sensor Triggerfish. First results in patients. Der Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft. 107, 918-922 (2010).
  11. Shaw, A. J., Davis, B. A., Collins, M. J., Carney, L. G. A technique to measure eyelid pressure using piezoresistive sensors. IEEE transactions on bio-medical engineering. 56, 2512-2517 (2009).
  12. Liao, Y. T., Yao, H. F., Lingley, A., Parviz, B., Otis, B. P. A 3-mu W CMOS glucose sensor for wireless contact-lens tear glucose monitoring. Ieee J Solid-St Circ. 47, 335-344 (2012).
  13. Coster, D. J. . Cornea. , (2002).
  14. Parekh, M., et al. A simplified technique for in situ excision of cornea and evisceration of retinal tissue from human ocular globe. Journal of visualized experiments : JoVE. , e3765 (2012).
  15. Way, S. Gear and pinion. US patent. , (1942).
  16. Lorentz, H., et al. Contact lens physical properties and lipid deposition in a novel characterized artificial tear solution. Molecular vision. 17, 3392-3405 (2011).
  17. Furukawa, R. E., Polse, K. A. Changes in tear flow accompanying aging. American journal of optometry and physiological optics. 55, 69-74 (1978).
  18. Bajgrowicz, M., Phan, C. M., Subbaraman, L., Jones, L. Release of ciprofloxacin and moxifloxacin from daily disposable contact lenses from an in vitro eye model. Invest Ophthalmol Vis Sci. , (2015).
  19. Luensmann, D., Zhang, F., Subbaraman, L., Sheardown, H., Jones, L. Localization of lysozyme sorption to conventional and silicone hydrogel contact lenses using confocal microscopy. Current eye research. 34, 683-697 (2009).
  20. Tieppo, A., Pate, K. M., Byrne, M. E. In vitro controlled release of an anti-inflammatory from daily disposable therapeutic contact lenses under physiological ocular tear flow. Eur J Pharm Biopharm. 81, 170-177 (2012).
  21. Ali, M., et al. Zero-order therapeutic release from imprinted hydrogel contact lenses within in vitro physiological ocular tear flow. J Control Release. 124, 154-162 (2007).
  22. White, C. J., McBride, M. K., Pate, K. M., Tieppo, A., Byrne, M. E. Extended release of high molecular weight hydroxypropyl methylcellulose from molecularly imprinted, extended wear silicone hydrogel contact lenses. Biomaterials. 32, 5698-5705 (2011).
  23. Kaczmarek, J. C., Tieppo, A., White, C. J., Byrne, M. E. Adjusting biomaterial composition to achieve controlled multiple-day release of dexamethasone from an extended-wear silicone hydrogel contact lens. J Biomater Sci Polym Ed. 25, 88-100 (2014).
  24. Mohammadi, S., Postnikoff, C., Wright, A. M., Gorbet, M. Design and development of an in vitro tear replenishment system. Ann Biomed Eng. 42, 1923-1931 (2014).
  25. Lorentz, H., Heynen, M., Khan, W., Trieu, D., Jones, L. The impact of intermittent air exposure on lipid deposition. Optom Vis Sci. 89, 1574-1581 (2012).
  26. Peng, C. C., Fajardo, N. P., Razunguzwa, T., Radke, C. J. In vitro spoilation of silicone-hydrogel soft contact lenses in a model-blink cell. Optom Vis Sci. 92, 768-780 (2015).
  27. Liu, P., et al. Dissolution studies of poorly soluble drug nanosuspensions in non-sink conditions. AAPS PharmSciTech. 14, 748-756 (2013).

Play Video

Cite This Article
Phan, C., Walther, H., Gao, H., Rossy, J., Subbaraman, L. N., Jones, L. Development of an In Vitro Ocular Platform to Test Contact Lenses. J. Vis. Exp. (110), e53907, doi:10.3791/53907 (2016).

View Video