In the protocol, we present a method to manufacture a small caliber stent-graft by sandwiching a balloon expandable stent between two electrospun nanofibrous polyurethane layers.

Abstract

Stent-grafts are widely used for the treatment of various conditions such as aortic lesions, aneurysms, emboli due to coronary intervention procedures and perforations in vasculature. Such stent-grafts are manufactured by covering a stent with a polymer membrane. An ideal stent-graft should have a biocompatible stent covered by a porous, thromboresistant, and biocompatible polymer membrane which mimics the extracellular matrix thereby promoting injury site healing. The goal of this protocol is to manufacture a small caliber stent-graft by encapsulating a balloon expandable stent within two layers of electrospun polyurethane nanofibers. Electrospinning of polyurethane has been shown to assist in healing by mimicking native extracellular matrix, thereby promoting endothelialization. Electrospinning polyurethane nanofibers on a slowly rotating mandrel enabled us to precisely control the thickness of the nanofibrous membrane, which is essential to achieve a small caliber balloon expandable stent-graft. Mechanical validation by crimping and expansion of the stent-graft has shown that the nanofibrous polyurethane membrane is sufficiently flexible to crimp and expand while staying patent without showing any signs of tearing or delamination. Furthermore, stent-grafts fabricated using the methods described here are capable of being implanted using a coronary intervention procedure using standard size guide catheters.

Introduction

冠動脈インターベンションの手順は、プラークの破壊および血管壁に起因する重大な血管壁の損傷を引き起こします。これは、再狭窄、静脈グラフトの末梢塞栓症、および冠状動脈内腔1-4の不連続が生じます。これらの合併症を回避するために、有望な戦略は、血管形成部位における血管表面を覆うように、潜在的に再狭窄を阻害するであろう、血管内腔の不連続のリスクを軽減、および末梢塞栓症を防止するであろう。これまでの研究では、ステントグラフト5のための肯定的な結果とステントグラフトにベアメタルステントを比較しました。研究者は、ステントをカバーするために膜を製造するためにいくつかの材料を使用しました。これは、ポリエチレンtetraphthalate(PET)、ポリテトラフルオロエチレン(PTFE)、ポリウレタン(PU)、及びシリコン又は覆われたステント6-9を製造するため自家血管組織のような合成材料を含んでいます。ステントを被覆するために使用される理想的な移植片材料は、抗血栓性、非biodegrであるべきですadable、および過度の増殖や炎症10せずに天然の組織と統合する必要があります。ステントを被覆するために使用される移植片材料はステント移植片の治癒を促進すべきです。

ステントグラフトが広く頚動脈の擬似動脈瘤、動静脈瘻は、静脈グラフトを縮退、大動脈縮窄の治療のために使用され、巨大な脳動脈瘤の大きさです。しかし、小口径のステント移植片の開発は、ステントグラフトの展開11-14に役立つロープロファイルと柔軟性を維持する能力によって制限されます。 PUは、ロープロファイルおよび良好な柔軟性15,16を達成するための所望の形質である良好な機械的強度と弾性ポリマーです。良好なデリバリーを有することに加えて、ステントグラフトはまた、迅速な治癒および内皮化を促進すべきです。 PUは、ステント移植片は、より良い生体適合性を実証し、内皮17を強化してい覆った。研究者は持っています以前PUは内皮細胞17でそれらを播種することによりステントグラフトを覆われた内皮化しようとしました。ナノファイバーマトリックスを作成するためにPUのエレクトロスピニングは、血管移植片18,19を製造するための貴重な技術であることが示されています。天然の細胞外マトリックスの構造を模倣するナノファイバーの存在はまた、内皮細胞増殖20,21を促進することが知られています。エレクトロスピニングは、材料22の厚さを制御することができます。 PUで作られた小口径血管移植片は、表面コーティング、抗凝固剤、及び細胞増殖抑制剤としての修飾を使用して、治癒を促進するために研究されています。すべてのこれらの変更は、ホスト受け入れを媒介グラフト治癒23を促進するように設計されています。

当社グループは、動物モデル24-26で展開することができるバルーン拡張ベアメタルステントを開発しました。エレクトロポリウレタンメッシュとボールの組み合わせキューン拡張型ステントは、小口径のバルーン拡張型ステント移植片を生成することができました。現在利用可能なステント移植片のほとんどは、介入手順中に大腿動脈を介して導入されているが、わずか数商業覆われたステントは非膨張バルーン27のために必要とされるよりも大きい1フレンチサイズを導入することができます。本研究では、経皮的インターベンションの手順で標準8-9フランスのガイドカテーテルを使用して冠状動脈に送達することができるエレクトロPUの2つの層の間のバルーン拡張型ステントをカプセル化することにより、小口径血管ステントグラフトを開発しました。

Protocol

マンドレルコレクター上のポリウレタンの1エレクトロスピニング エレクトロスピニングのためのマンドレルを準備 オーブンを用いて155℃でメスシリンダー(約9ミリメートルの直径、深さ110 mm)の生体適合性、食品グレード、水溶性担体材料の約8ミリリットルメルト。 マンドレルの表面上に支持材料のコーティングを得るために、直径3mm及び長さ100mmのステ…

Representative Results

我々のelectrospinnerセットアップ( 図1)は、高品質のポリウレタンナノファイバー( 図2)をもたらしました。ステントグラフトは、マンドレル上にポリウレタンの内部層をエレクトロスピニングこの層の上にベアメタルステントを滑り、及びポリウレタンの第二の外側層( 図3)をエレクトロスピニングすることによって製造されます。ポリウレタ?…

Discussion

We have developed a fabrication technique for a small caliber stent-graft which can be deployed using a standard percutaneous coronary intervention (PCI) procedure. Stent-grafts currently available are limited in their ability to maintain a low profile and flexibility for deployment. Bare metal stents developed by our group in our previous studies have proven to assist in rapid healing of the stented artery24,26. Various polymers have been electrospun by other groups and polyurethane has been proven biostable …

Disclosures

The authors have nothing to disclose.

Acknowledgements

We would like to thank the Division of Engineering, Mayo Clinic for their technical support. This study was financially supported by European Regional Development Fund – FNUSA-ICRC (No. CZ.1.05/1.100/02.0123), National Institutes of Health (T32 HL007111), American Heart Association Scientist Development Grant (AHA #06-35185N), and The Grainger Innovation Fund – Grainger Foundation.

Materials

Glass syringe Air Tite 7.140-33 Syringe for spinneret
Graduated cylinder 5 mL Fisher Scientific 08-552-4G 5 mL pyrex graduated cylinder about 9mm diameter and 11 cm long
High voltage generator Bertan Accociates, Inc. 205A-30P Used to apply voltage difference across spinneret and collector
Laboratory mixer with rpm control Scilogex SCI-84010201 Available from various laboratory equipment suppliers
Polyurethane DSM BioSpan SPU Biospan Segmented Polyurethane
Rubber sheet McMaster Carr 1370N11 Used to insulate syringe during electrospinning
Stainless steel mandrel N/A N/A Manufactured 
Stainless steel needle Hamilton 91018 Used as spinneret in electrospinning
Support material EnvisionTec B04-HT-DEMOMAT Biocompatible water soluble material
Syringe Pump Harvard Apparatus 55-3333

References

  1. Elsner, M., et al. Coronary stent grafts covered by a polytetrafluoroethylene membrane. Am. J. Cardiol. 84 (3), 335-338 (1999).
  2. Störger, H., Haase, J. Polytetrafluoroethylene-Covered Stents: Indications, Advantages, and Limitations. J. Interv. Cardiol. 12 (6), 451-456 (1999).
  3. Moreno, P. R., et al. Macrophage infiltration predicts restenosis after coronary intervention in patients with unstable angina. Circulation. 94 (12), 3098-3102 (1996).
  4. Briguori, C., Sarais, C., Colombo, A. The polytetrafluoroethylene-covered stent: a device with multiple potential advantages. Int. J. Cardiovasc. Interv. 4 (3), 145-149 (2001).
  5. Qureshi, M. A., Martin, Z., Greenberg, R. K. Endovascular management of patients with Takayasu arteritis: stents versus stent grafts. Semin. Vasc. Surg. 24 (1), 44-52 (2011).
  6. Ahmadi, R., Schillinger, M., Maca, T., Minar, E. Femoropopliteal arteries: immediate and long-term results with a Dacron-covered stent-graft. Radiology. 223 (2), 345-350 (2002).
  7. Geremia, G., et al. Experimental arteriovenous fistulas: treatment with silicone-covered metallic stents. AJNR. Am. J. Neuroradiol. 18 (2), 271-277 (1997).
  8. Saatci, I., et al. Treatment of internal carotid artery aneurysms with a covered stent: experience in 24 patients with mid-term follow-up results. AJNR. Am. J. Neuroradiol. 25 (10), 1742-1749 (2004).
  9. Stefanadis, C., et al. Stents Wrapped in Autologous Vein: An Experimental Study1. J. Am. Coll. Cardiol. 28 (4), 1039-1046 (1996).
  10. Palmaz, J. C. Review of polymeric graft materials for endovascular applications. J. Vasc. Interv. Radiol. 9, 7-13 (1998).
  11. Bruckheimer, E., Dagan, T., Amir, G., Birk, E. Covered Cheatham-Platinum stents for serial dilation of severe native aortic coarctation. Catheter Cardiovasc. Interv. 74 (1), 117-123 (2009).
  12. Tzifa, A., et al. Covered Cheatham-platinum stents for aortic coarctation: early and intermediate-term results. J. Am. Coll. Cardiol. 47 (7), 1457-1463 (2006).
  13. Kuraishi, K., et al. Development of nanofiber-covered stents using electrospinning: in vitro and acute phase in vivo experiments. J. Biomed. Mater. Res. Part B Appl. Biomater. 88 (1), 230-239 (2009).
  14. Pant, S., Bressloff, N. W., Limbert, G. Geometry parameterization and multidisciplinary constrained optimization of coronary stents. Biomech. Model Mechanobiol. 11 (1-2), 61-82 (2012).
  15. Muller-Hulsbeck, S., et al. Experience on endothelial cell adhesion on vascular stents and stent-grafts: first in vitro results. Invest. Radiol. 37 (6), 314-320 (2002).
  16. Sarkar, S., Salacinski, H. J., Hamilton, G., Seifalian, A. M. The mechanical properties of infrainguinal vascular bypass grafts: their role in influencing patency. Eur. J. Vasc. Endovasc. Surg. 31 (6), 627-636 (2006).
  17. Shirota, T., Yasui, H., Shimokawa, H., Matsuda, T. Fabrication of endothelial progenitor cell (EPC)-seeded intravascular stent devices and in vitro endothelialization on hybrid vascular tissue. Biomaterials. 24 (13), 2295-2302 (2003).
  18. Grasl, C., et al. Electrospun polyurethane vascular grafts: in vitro mechanical behavior and endothelial adhesion molecule expression. J. Biomed. Mater. Res. A. 93 (2), 716-723 (2010).
  19. Kidoaki, S., Kwon, I. K., Matsuda, T. Structural features and mechanical properties of in situ-bonded meshes of segmented polyurethane electrospun from mixed solvents. J. Biomed. Mater. Res. Part B Appl. Biomater. 76 (1), 219-229 (2006).
  20. Stegemann, J. P., Kaszuba, S. N., Rowe, S. L. Review: advances in vascular tissue engineering using protein-based biomaterials. Tissue Eng. 13 (11), 2601-2613 (2007).
  21. Sankaran, K. K., Subramanian, A., Krishnan, U. M., Sethuraman, S. Nanoarchitecture of scaffolds and endothelial cells in engineering small diameter vascular grafts. Biotechnol. J. 10 (1), 96-108 (2015).
  22. Gibson, P., Schreuder-Gibson, H., Rivin, D. Transport properties of porous membranes based on electrospun nanofibers. Colloid Surf., A. 187, 469-481 (2001).
  23. Zdrahala, R. J. Small caliber vascular grafts. Part II: Polyurethanes revisited. J. Biomater. Appl. 11 (1), 37-61 (1996).
  24. Uthamaraj, S., et al. Design and validation of a novel ferromagnetic bare metal stent capable of capturing and retaining endothelial cells. Ann. Biomed. Eng. 42 (12), 2416-2424 (2014).
  25. Tefft, B. J., et al. Cell Labeling and Targeting with Superparamagnetic Iron Oxide Nanoparticles. J. Vis. Exp. (105), e53099 (2015).
  26. Uthamaraj, S., et al. Ferromagnetic Bare Metal Stent for Endothelial Cell Capture and Retention. J. Vis. Exp. (103), e53100 (2015).
  27. de Giovanni, J. V. Covered stents in the treatment of aortic coarctation. J. Interv. Cardiol. 14 (2), 187-190 (2001).
  28. Hans, F. J., et al. Treatment of wide-necked aneurysms with balloon-expandable polyurethane-covered stentgrafts: experience in an animal model. Acta. Neurochir. (Wien). 147 (8), 871-876 (2005).
  29. Hasan, A., et al. Electrospun scaffolds for tissue engineering of vascular grafts. Acta. Biomater. 10 (1), 11-25 (2014).
check_url/54731?article_type=t

Play Video

Cite This Article
Uthamaraj, S., Tefft, B. J., Jana, S., Hlinomaz, O., Kalra, M., Lerman, A., Dragomir-Daescu, D., Sandhu, G. S. Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents. J. Vis. Exp. (116), e54731, doi:10.3791/54731 (2016).

View Video