Summary

الكمي المطلق للAβ<sub> 1-42</sub> في السائل الدماغي الشوكي عن طريق مرجع إجراء قياس كتلة الطيف

Published: March 21, 2017
doi:

Summary

A reference measurement procedure for the absolute quantification of Aβ1-42 in human CSF based on solid-phase extraction and liquid chromatography tandem mass spectrometry is described.

Abstract

Alzheimer’s disease (AD) is the most common neurodegenerative disease among the elderly and accounts for 60-80% of all cases of dementia. Currently, the diagnosis of AD is based on cognitive tests and mental state exams, but the peptide amyloid-beta (Aβ) in cerebrospinal fluid (CSF) is increasingly used in clinical trials and settings. As for most protein and peptide biomarkers, quantification is performed using antibody-based techniques, such as enzyme-linked immunosorbent assay (ELISA). However, intra- and inter-laboratory variability in these assays hamper its use as a diagnostic marker in clinical routine.

An antibody-independent Reference Measurement Procedure (RMP) was developed based on solid-phase extraction (SPE) and liquid chromatography (LC)-tandem mass spectrometry (MS/MS), where stable, isotope-labeled Aβ peptides were used as internal standards, enabling absolute quantification. A high-resolution quadrupole-orbitrap hybrid instrument was used for the measurements. The method allows for the quantification of CSF Aβ1-42 between 150-4,000 pg/mL.

Introduction

مرض الزهايمر (م) هو الشكل الأكثر شيوعا من الخرف ويصيب حوالي 35 مليون شخص في جميع أنحاء العالم 1. بصمات عصبية مرضية من مرض يعتقد على نطاق واسع تكمن في جوهر م المرضية هي التشابك الليفي العصبي بين الخلايا من hyperphosphorylated بروتين تاو 2 ولويحات خارج الخلية التي تتكون من (Aβ) الببتيد 3 المجمعة اميلويد بيتا. وتماشيا مع هذا، وقد تم مؤخرا شملت تقييم أمراض الترسبات في الجسم الحي من المؤشرات الحيوية في المعايير البحوث التشخيصية لمدة 4 م. لقياس السائل النخاعي من Aβ 1-42، هي عدة المناعية المتاحة وتستخدم في العديد من المختبرات السريرية 5. تركيز Aβ 1-42 في السائل النخاعي هو أقل ما يقرب من 50٪ في المرضى ميلادي مما كانت عليه في كبار السن الطبيعي معرفيا، مما يعكس ترسب الببتيد في لويحات رعين 6 و 7.

ويتم تحليل هذه المؤشرات الحيوية بشكل رئيسي باستخدام المناعية (أي التقنيات المعتمدة الضد)، ولكن قد تتأثر هذه المقايسات من الآثار مصفوفة 8. استخدام المناعية على منصات التكنولوجيا المختلفة وعدم وجود فحص التقييس 10 يجعل إدخال العالمية تركيزات قطع الصعبة 11 و 12. ومن شأن RMP التحقق من صحة تحليلي تسمح معايرة موحدة للمنصات فحص مختلفة، مما أدى مثالي في المقارنة أفضل عبر منصات التحليلية وفي تحسين السيطرة على العوامل التي تساهم في تقلب قياس العام.

الكمي المطلق للAβ 1-42 باستخدام ضعت طريقة LC-MS / MS يتغلب على العديد من القضايا المرتبطة ميتاليك القائم على الأجسام المضادةiques. سيتم استخدام طريقة، على النحو الوارد في الشرطة العسكرية الملكية من قبل اللجنة المشتركة للتتبع في طب المختبرات (JCTLM قاعدة بيانات تحديد عدد C11RMP9)، لتحديد تركيز المطلق للAβ 1-42 في المراجع معتمد (CRM) لتنسيق CSF Aβ 1 قياسات -42 عبر تقنيات ومناهج تحليلية. وينبغي أن يكون سير العمل وصفه ذات الصلة لتطوير أساليب مرشح مرجعية لالببتيدات والبروتينات داخل مناطق أخرى من الأدوية.

Protocol

ملاحظة: هذا البروتوكول يتطلب قسامات لا يقل عن 50 ميكرولتر، مع تركيز 50 ميكروغرام / مل لكل الببتيد Aβ، في بدء المواد. يجب حل الببتيد Aβ في 20٪ الأسيتونتريل (إيه سي) و 4٪ محلول الأمونيا المركزة في الماء منزوع الأيونات (ت / ت) وتخزينها في 80 درجة مئوية. <p class="jove_content" style=";text-align:righ…

Representative Results

يتم استخدام الإعداد لوحة في الشكل رقم 1 لوحة كاملة من العينات. إذا ليتم تحليلها أقل عينات غير معروفة، وتدريج الثاني، ينبغي أن توضع مجموعات الترددات اللاسلكية، ومراقبة الجودة بعد نهاية الشوط الأول من العينات غير معروف. <p class="jove_content" fo:ke…

Discussion

للتعرف على طريقة وصفها، بدلا من استخدام مصفوفة البديلة، استخدمنا نهج بديل تحليلها 13، 14، 15، 16، والتي تمكن المعايرة في السائل النخاعي البشري. ينطوي على نهج تحليلها البديل مختلفين معايير المسمى isotopi…

Disclosures

The authors have nothing to disclose.

Acknowledgements

This work was performed on behalf of the International Federation of Clinical Chemistry Working Group on CSF Proteins, which has the following composition: Kaj Blennow (Chair) and Henrik Zetterberg, Gothenburg University, Sweden; Les Shaw and Magdalena Korecka, University of Philadelphia, PA, USA; Ingrid Zegers, Institute for Reference Materials and Measurements, Geel, Belgium; Piotr Lewczuk, Universitätsklinikum Erlangen, Germany; and Rand Jenkins, PPD Bioanalytical Laboratory, Richmond, VA, USA; Randall Bateman, Washington University, MO, USA; and H Vanderstichele, Innogenetics NV, Ghent, Belgium. This work was also part of the Global Consortium for Biomarker Standardization CSF Reference Methods Working Group, which is led by Maria C. Carrillo, Ph.D., Senior Director, Medical & Scientific Relations, Alzheimer’s Association. The study was supported by The Swedish Research Council (grants #14002, #521-2011-4709, and #2013-2546); the Knut and Alice Wallenberg Foundation; Demensförbundet; and Emil and Wera Cornell’s, Aina Wallström and Mary-Ann Sjöblom’s, Gun and Bertil Stohne’s, Magnus Bergvall’s, and Gamla Tjänarinnor’s Foundations.

Materials

Eppendorf Protein LoBind microcentrifuge tubes, 0.5 mL Eppendorf 022431081
Eppendorf Protein LoBind microcentrifuge tubes, 1.5 mL Eppendorf 022431081
Eppendorf Protein LoBind microcentrifuge tubes, 2 mL Eppendorf 022431102
Eppendorf Protein LoBind microcentrifuge tubes, 5 mL Eppendorf 0030 108.310
Eppendorf Protein LoBind Deepwell Plates 96 Eppendorf 951032905
Micronic 0.75 mL polypropylene tubes, V-bottom Micronic MPW32071BC3 Tubes used for collecting SPE eluate, but other collection devices can be used such as 96-well plates.
Micronic Split TPE Capcluster for 96 ind. tubes Micronic MP53026 Caps for Micronic tubes.
Micronic Loborack-96 white w high cover bar coded Micronic MPW51015BC3 Holder for Micronic tubes.
Biohit Optifit tips 1.2 mL Sartorius 791210 Used to reach all the way to the bottom of the LoBind deepwell plates (96-well). Other narrow tips might work as well.
Waters Oasis MCX 96-well µElution Plate Waters 186001830BA 
Waters Plate manifold reservoir tray Waters WAT058942
Waters Extraction Plate Manifold for Oasis 96-Well Plates Waters 186001831
Ammonium hydroxide solution, puriss. p.a., reag. ISO, reag. Ph. Eur., ~25% NH3 basis Sigma-Aldrich 30501-1L-D
Acetonitrile, Far UV HPLC Gradient grade, 2.5 L Fisher Scientific A/0627/17X
ortho-Phosphoric acid 85%, ACS,ISO,Reag. Ph Eur Merck Millipore 1005731000
Guanidine Hydrochloride, 500 g Thermo Scientific 24110
Thermo Scientific Q-exactive Hybrid Quadrupole Orbitrap Mass Spectrometer Thermo Scientific IQLAAEGAAPFALGMAZR With bundled Dionex Ultimate 3000 LC & autosampler.
Dionex ProSwift RP-4H Monolith Column (1.0 x 250mm) Dionex 066640
Bovine Serum Albumin, lyophilized powder, suitable for (for molecular biology), Non-acetylated Sigma-Aldrich B6917
Beta-Amyloid (1-42), Ultra Pure, TFA rPeptide A-1002-2
15N Beta-Amyloid (1-42), Uniformly labeled rPeptide A-1102-2
13C Beta-Amyloid (1-42), Uniformly labeled rPeptide A-1106-2
Microplate shakers, TiMix 2 Edmund Bühler 6110 000

References

  1. Querfurth, H. W., LaFerla, F. M. Alzheimer’s disease. N Engl J Med. 362 (4), 329-344 (2010).
  2. Braak, H., Braak, E. Evolution of the neuropathology of Alzheimer’s disease. Acta Neurol Scand Suppl. 165, 3-12 (1996).
  3. Blennow, K., de Leon, M. J., Zetterberg, H. Alzheimer’s disease. Lancet. 368 (9533), 387-403 (2006).
  4. Dubois, B., et al. Advancing research diagnostic criteria for Alzheimer’s disease: the IWG-2 criteria. Lancet Neurol. 13 (6), 614-629 (2014).
  5. Blennow, K., Hampel, H., Weiner, M., Zetterberg, H. Cerebrospinal fluid and plasma biomarkers in Alzheimer disease. Nat Rev Neurol. 6 (3), 131-144 (2010).
  6. Buchhave, P., et al. Cerebrospinal fluid levels of beta-amyloid 1-42, but not of tau, are fully changed already 5 to 10 years before the onset of Alzheimer dementia. Arch Gen Psychiatry. 69 (1), 98-106 (2012).
  7. Olsson, B., et al. CSF and blood biomarkers for the diagnosis of Alzheimer’s disease: a systematic review and meta-analysis. The Lancet Neurology. 15 (7), 673-684 (2016).
  8. Bjerke, M., et al. Confounding factors influencing amyloid Beta concentration in cerebrospinal fluid. Int J Alzheimers Dis. , (2010).
  9. Dumurgier, J., et al. Intersite variability of CSF Alzheimer’s disease biomarkers in clinical setting. Alzheimers Dement. 9 (4), 406-413 (2013).
  10. Mattsson, N., et al. CSF biomarker variability in the Alzheimer’s Association quality control program. Alzheimers Dement. 9 (3), 251-261 (2013).
  11. Kang, J. H., Korecka, M., Toledo, J. B., Trojanowski, J. Q., Shaw, L. M. Clinical Utility and Analytical Challenges in Measurement of Cerebrospinal Fluid Amyloid-beta1-42 and tau Proteins as Alzheimer Disease Biomarkers. Clin Chem. 59 (6), 903-916 (2013).
  12. Mattsson, N., et al. Reference measurement procedures for Alzheimer’s disease cerebrospinal fluid biomarkers: definitions and approaches with focus on amyloid beta42. Biomark Med. 6 (4), 409-417 (2012).
  13. Ahmadkhaniha, R., Shafiee, A., Rastkari, N., Kobarfard, F. Accurate quantification of endogenous androgenic steroids in cattle’s meat by gas chromatography mass spectrometry using a surrogate analyte approach. Anal Chim Acta. 631 (1), 80-86 (2009).
  14. Jemal, M., Schuster, A., Whigan, D. B. Liquid chromatography/tandem mass spectrometry methods for quantitation of mevalonic acid in human plasma and urine: method validation, demonstration of using a surrogate analyte, and demonstration of unacceptable matrix effect in spite of use of a stable isotope analog internal standard. Rapid Commun Mass Spectrom. 17 (15), 1723-1734 (2003).
  15. Leinenbach, A., et al. Mass spectrometry-based candidate reference measurement procedure for quantification of amyloid-beta in cerebrospinal fluid. Clin Chem. 60 (7), 987-994 (2014).
  16. Li, W., Cohen, L. H. Quantitation of endogenous analytes in biofluid without a true blank matrix. Anal Chem. 75 (21), 5854-5859 (2003).
  17. Perret-Liaudet, A., et al. Cerebrospinal fluid collection tubes: a critical issue for Alzheimer disease diagnosis. Clin Chem. 58 (4), 787-789 (2012).
  18. Lewczuk, P., et al. Effect of sample collection tubes on cerebrospinal fluid concentrations of tau proteins and amyloid beta peptides. Clin Chem. 52 (2), 332-334 (2006).
  19. Pannee, J., et al. Reference measurement procedure for CSF amyloid beta (Aβ)1-42 and the CSF Aβ1-42 /Aβ1-40 ratio – a cross-validation study against amyloid PET. J Neurochem. 139 (4), 651-658 (2016).
  20. Thienpont, L. M., Van Houcke, S. K. Traceability to a common standard for protein measurements by immunoassay for in-vitro diagnostic purposes. Clin Chim Acta. 411 (23-24), 2058-2061 (2010).
  21. Vesper, H. W., Thienpont, L. M. Traceability in laboratory medicine. Clin Chem. 55 (6), 1067-1075 (2009).
check_url/kr/55386?article_type=t

Play Video

Cite This Article
Pannee, J., Blennow, K., Zetterberg, H., Portelius, E. Absolute Quantification of Aβ1-42 in CSF Using a Mass Spectrometric Reference Measurement Procedure. J. Vis. Exp. (121), e55386, doi:10.3791/55386 (2017).

View Video