Cardiac pressure-volume loop analysis is the most comprehensive way to measure cardiac function in the intact heart. We describe a technique to perform and analyze cardiac pressure volume loops, using conductance catheters.

Abstract

Cardiac pressure-volume loop analysis is the “gold-standard” in the assessment of load-dependent and load-independent measures of ventricular systolic and diastolic function. Measures of ventricular contractility and compliance are obtained through examination of cardiac response to changes in afterload and preload. These techniques were originally developed nearly three decades ago to measure cardiac function in large mammals and humans. The application of these analyses to small mammals, such as mice, has been accomplished through the optimization of microsurgical techniques and creation of conductance catheters. Conductance catheters allow for estimation of the blood pool by exploiting the relationship between electrical conductance and volume. When properly performed, these techniques allow for testing of cardiac function in genetic mutant mouse models or in drug treatment studies. The accuracy and precision of these studies are dependent on careful attention to the calibration of instruments, systematic conduct of hemodynamic measurements and data analyses. We will review the methods of conducting pressure-volume loop experiments using a conductance catheter in mice.

Introduction

Análise de ciclo de volume pressão cardíaca fornece informações detalhadas da função cardíaca e são o padrão ouro para avaliação funcional 1. Embora as técnicas de imagem, como a ecocardiografia ou a ressonância magnética cardíaca fornecer medidas funcionais, estas medidas são altamente dependentes de condições de carga. Medidas independentes de Carga, de contratilidade cardíaca e relaxamento requerem medições dinâmicas da pressão ventricular e relação do volume em um intervalo de pré-carga e pós-carga. Esta compreensão da relação pressão-volume decorre do trabalho inovador de Sagawa e colegas 2,3. Eles demonstraram em ex vivo corações caninos perfundidos que o loop de pressão-volume derivados medidas de contratilidade foram independentes de condições de carga 4.

Na aplicação in vivo destas análises tornou-se possível, com o desenvolvimento de cateteres de condutância na década de 1980. Este avanço técnico permitiu Kass e colegas para realizar a análise de ciclo de pressão-volume em humanos 5,6. Miniaturização dos cateteres de condutância e melhorias nas técnicas cirúrgicas no final de 1990 de 7 feita a análise da função cardíaca roedor viável, permitindo estudos genéticos e farmacológicos para ser realizado. Este avanço, desde então, levar ao uso generalizado de análise de ciclo de pressão-volume e tem gerado uma grande quantidade de conhecimento sobre fisiologia cardíaca mamíferos.

Um conceito fundamental na utilização de cateteres de condutância e a interpretação dos dados obtidos a partir dele, é a relação entre o volume e a condutância. A condutância é inversamente relacionada com a tensão, que é medida através de um cateter com eléctrodos colocados proximal, geralmente colocada por baixo da válvula aórtica, e distalmente, no apex LV 8. Alterações na tensão ou condutância são medidos por mudanças na corrente que flui de proximal para distal do eletrodo. Embora a piscina de sangue contribuems significativamente a condutância, a contribuição da parede ventricular, denominado condutância paralela (V P), a condutância medido deve ser subtraído para se obter medições do volume do VE absolutos.

Os métodos para efectuar esta correcção, chamado de calibração solução salina, são discutidos no protocolo abaixo. A relação matemática entre a condutância e de volume, descrito por Baan e colegas, é que o volume = 1 / α; (ρ L 2) (GG p), onde fator de correção α = campo uniforme, ρ = resistividade sangue, L = distância entre os eletrodos, G = condutância e G p = condutância não-sangue 9. De nota, o fator de correção de campo uniforme em camundongos aproxima de 1,0 devido a pequenos volumes de câmara 10. Juntamente com transdutores de pressão, o cateter condutância fornece dados de pressão e volume simultânea em tempo real.

Pressu cardíacare-análise do volume apresenta determinadas vantagens sobre outras medidas da função cardíaca, já que permitem a medição da função ventricular independente das condições de carga e de frequência cardíaca. Índices cardíacos independente de carregamento específicos de contratilidade incluem: sistólico final do volume de relação de pressão (ESPVR), d P / d t max relação volume de -end-diastólica, elastância máxima (E máx) e pré-carga de trabalho do curso recrutável (PRSW). Uma medida independente de carga da função diastólica é a relação do volume a pressão diastólica final (EDPVR) 11. O protocolo a seguir descreve o comportamento de análise de ciclo de volume pressão cardíaca, usando tanto um carótida e uma abordagem apical. Embora a metodologia para executar estes estudos foram descritas em pormenor anteriormente 8,11, iremos analisar os passos chave para obter medições de pressão-volume preciso, incluindo tanto a correcção de solução salina e de calibração cuvete, e fornecer uma demonstração visual de Tsprocedimentos e. Pesquisas com animais realizados para este estudo foi tratado de acordo com os protocolos aprovados e os regulamentos de protecção dos animais do Comité de Conservação e Uso do Duke University Medical Center animal Institucional.

Protocol

1. Preparações Conductance cateter e calibração de pressão Conectar cateter de condutância para o módulo de cateter hemodinâmica. Eletronicamente calibrar as medições de pressão e volume gravando pressão predefinida eo volume ajustado no módulo cateter. Grave um traçado de 0 mm Hg e 25 mm Hg (Figura 1A) e atribuir as voltagens para ambos os traçados de pressão (Figura 1B e 1C). Da mesma forma, gravar um rastreamento volume de 5 RVU e 25 RVU <s…

Representative Results

Análise de ciclo de pressão-volume pode ser usado para medir a função cardíaca em camundongos geneticamente modificados 14,15 ou camundongos submetidos a estudos de drogas 16. Lacetes de volume pressão representativos são fornecidos a partir de trabalhos publicados anteriormente 16 investigar o efeito de beta-arrestina inclinado AT1R ligando, TRV120023. Para testar se TRV120023 afeta a função cardíaca in vivo, análise do lacete pressão-volume foi realizada em camundo…

Discussion

Nós descrevemos um método para perfoming análise do lacete pressão-volume usando um cateter de condutância em camundongos, para derivar análises abrangentes, tanto a contratilidade cardíaca e relaxamento. Suga, Sagawa e seus colegas utilizaram pressão-volume loops para definir medidas de contratilidade cardíaca, especificamente a inclinação da ESPVR, ou a elastância sistólica final (es E), e E max. Elastância, definida pela relação de pressão para o volume (P / V), varia ao longo d…

Divulgaciones

The authors have nothing to disclose.

Acknowledgements

Este trabalho é apoiado pela American Heart Association 14FTF20370058 (DMA) e NIH T32 HL007101-35 (DMA).

Materials

AnaSed (xylazine)  Lloyd Laboratories NADA no. 139-236 Anesthetic
Ketaset (ketamine) Pfizer 440842 Anesthetic
VIP3000 Matrx Medical Inc. Anesthesia machine
Ventilator Harvard Apparatus Model 683 Surgical Equipment
Tubing kit Harvard Apparatus 72-1049 Surgical Equipment
Homeothermic Blanket  Kaz Inc. 5628 Surgical Equipment
Stereo microscope Carl Zeiss Optical Inc. Stemi 2000 Surgical Equipment
Illuminator Cole–Parmer 41720 Surgical Equipment
Dumont no. 55 Dumostar Forceps  Fine Science Tools Inc 11295-51 Surgical Instruments
Graefe forceps, curved  Fine Science Tools Inc 11052-10 Surgical Instruments
Moria MC31 forceps  Fine Science Tools Inc 11370-31 Surgical Instruments
Mayo scissors  Fine Science Tools Inc 14512-15 Surgical Instruments
Iris scissors  Fine Science Tools Inc 14041-10 Surgical Instruments
Halsey needle holder  Fine Science Tools Inc 12501-13 Surgical Instruments
Olsen–Hegar needle holder  Fine Science Tools Inc 12002-12 Surgical Instruments
spring scissors Fine Science Tools Inc 15610-08 Surgical Instruments
disposable underpads Kendall/Tyco Healthcare 1038 Surgical Supplies
Sterile gauze sponges, sterile  Dukal 62208 Surgical Supplies
Cotton-tipped applicators, sterile  Solon 368 Surgical Supplies
Surgical suture,  silk, 6-0  DemeTECH FT-639-1 Surgical Supplies
1 cc Insulin syringes  Becton Dickenson 329412 Surgical Supplies
Access-9™ Hemostasis Valve Merit Medical  MAP111 Hemodynamic equipment
Sphygmomanometer Baumanometer 320 Hemodynamic equipment
Millar PV system MPVS-300/400 or MPVS Ultra (includes calibration cuvette) ADInstruments Inc Hemodynamic equipment
1.4F conductance catheter  ADInstruments Inc SPR-839 Hemodynamic equipment
PowerLab 4/30 with Chart Pro ADInstruments Inc. ML866/P Hemodynamic software
animal clipper Wahl 8787-450A Miscellaneous
Intradermic tubing PE-10 (Becton Dickinson, cat. no. ) Becton Dickenson 427401 Miscellaneous
Intradermic tubing PE-50 (Becton Dickinson, cat. no.) Becton Dickenson 427411 Miscellaneous
Needle assortment (18, 25 and 30 gauge; Thomas Scientific) Miscellaneous
0.9% (wt/vol) sodium chloride injection, USP , cat. no. ) Hospira NDC no. 0409-4888-50 Miscellaneous
Surgical tape Miscellaneous
Alconox (Alconox Inc.) for catheter cleaning ADInstruments Inc. Miscellaneous

Referencias

  1. Clark, J. E., Marber, M. S. Advancements in pressure-volume catheter technology – stress remodelling after infarction. Exp Physiol. 98 (3), 614-621 (2013).
  2. Sunagawa, K., Maughan, W. L., Burkhoff, D., Sagawa, K. Left ventricular interaction with arterial load studied in isolated canine ventricle. Am J Physiol. 245 (Pt 1), H773-H780 (1983).
  3. Suga, H., Sagawa, K., Demer, L. Determinants of instantaneous pressure in canine left ventricle. Time and volume specification. Circ Res. 46 (2), 256-263 (1980).
  4. Suga, H., Sagawa, K., Shoukas, A. A. Load independence of the instantaneous pressure-volume ratio of the canine left ventricle and effects of epinephrine and heart rate on the ratio. Circ Res. 32 (3), 314-322 (1973).
  5. Kass, D. A., et al. Improved left ventricular mechanics from acute VDD pacing in patients with dilated cardiomyopathy and ventricular conduction delay. Circulation. 99 (12), 1567-1573 (1999).
  6. Kass, D. A., et al. Diastolic Compliance of Hypertrophied Ventricle Is Not Acutely Altered by Pharmacological Agents Influencing Active Processes. Annals of Internal Medicine. 119 (6), 466-473 (1993).
  7. Georgakopoulos, D., et al. In vivo murine left ventricular pressure-volume relations by miniaturized conductance micromanometry. Am J Physiol. 274 (4 pt 2), H1416-H1422 (1998).
  8. Cingolani, O. H., Kass, D. A. Pressure-volume relation analysis of mouse ventricular function. Am J Physiol Heart Circ Physiol. 301 (6), H2198-H2206 (2011).
  9. Baan, J., et al. Continuous Measurement of Left-Ventricular Volume in Animals and Humans by Conductance Catheter. Circulation. 70 (5), 812-823 (1984).
  10. Pearce, J. A., Porterfield, J. E., Larson, E. R., Valvano, J. W., Feldman, M. D. Accuracy considerations in catheter based estimation of left ventricular volume. Conf Proc IEEE Eng Med Biol Soc. 2010, 3556-3558 (2010).
  11. Pacher, P., Nagayama, T., Mukhopadhyay, P., Batkai, S., Kass, D. A. Measurement of cardiac function using pressure-volume conductance catheter technique in mice and rats. Nature Protocols. 3 (9), 1422-1434 (1038).
  12. Hanusch, C., Hoeger, S., Beck, G. C. Anaesthesia of small rodents during magnetic resonance imaging. Methods. 43 (1), 68-78 (2007).
  13. Georgakopoulos, D., Kass, D. A. Estimation of parallel conductance by dual-frequency conductance catheter in mice. Am J Physiol Heart Circ Physiol. 279 (1), H443-H450 (2000).
  14. Esposito, G., et al. Increased myocardial contractility and enhanced exercise function in transgenic mice overexpressing either adenylyl cyclase 5 or 8. Basic Res Cardiol. 103 (1), 22-30 (2008).
  15. Kohout, T. A., et al. Augmentation of cardiac contractility mediated by the human beta(3)-adrenergic receptor overexpressed in the hearts of transgenic mice. Circulation. 104 (20), 2485-2491 (2001).
  16. Kim, K. S., et al. beta-Arrestin-biased AT1R stimulation promotes cell survival during acute cardiac injury. Am J Physiol Heart Circ Physiol. 303 (8), H1001-H1010 (2012).
  17. Suga, H., Sagawa, K. Mathematical Interrelationship between Instantaneous Ventricular Pressure-Volume Ratio and Myocardial Force-Velocity Relation. Annals of Biomedical Engineering. 1 (2), 160-181 (1972).
  18. Suga, H. Ventricular energetics. Physiol Rev. 70 (2), 247-277 (1990).
  19. Kass, D. A., et al. Influence of contractile state on curvilinearity of in situ end-systolic pressure-volume relations. Circulation. 79 (1), 167-178 (1989).
  20. Little, W. C. The left ventricular dP/dtmax-end-diastolic volume relation in closed-chest dogs. Circ Res. 56 (6), 808-815 (1985).
  21. Glower, D. D., et al. Linearity of the Frank-Starling relationship in the intact heart: the concept of preload recruitable stroke work. Circulation. 71 (5), 994-1009 (1985).
  22. Sharir, T., et al. Ventricular systolic assessment in patients with dilated cardiomyopathy by preload-adjusted maximal power. Validation and noninvasive application. Circulation. 89 (5), 2045-2053 (1994).
  23. Baan, J., Van der Velde, E. T. Sensitivity of left ventricular end-systolic pressure-volume relation to type of loading intervention in dogs. Circ Res. 62 (6), 1247-1258 (1988).
  24. Wei, C. L., et al. Volume catheter parallel conductance varies between end-systole and end-diastole. IEEE Trans Biomed Eng. 54 (8), 1480-1489 (2007).
  25. Porterfield, J. E., et al. Dynamic correction for parallel conductance, GP, and gain factor, alpha, in invasive murine left ventricular volume measurements. J Appl Physiol (1985). 107 (6), 1693-1703 (2009).

Play Video

Citar este artículo
Abraham, D., Mao, L. Cardiac Pressure-Volume Loop Analysis Using Conductance Catheters in Mice. J. Vis. Exp. (103), e52942, doi:10.3791/52942 (2015).

View Video