Summary

Farelerde İletkenlik Balon Kullanımı kalp Basınç Hacim Döngüsü Analizi

Published: September 17, 2015
doi:

Summary

Cardiac pressure-volume loop analysis is the most comprehensive way to measure cardiac function in the intact heart. We describe a technique to perform and analyze cardiac pressure volume loops, using conductance catheters.

Abstract

Cardiac pressure-volume loop analysis is the “gold-standard” in the assessment of load-dependent and load-independent measures of ventricular systolic and diastolic function. Measures of ventricular contractility and compliance are obtained through examination of cardiac response to changes in afterload and preload. These techniques were originally developed nearly three decades ago to measure cardiac function in large mammals and humans. The application of these analyses to small mammals, such as mice, has been accomplished through the optimization of microsurgical techniques and creation of conductance catheters. Conductance catheters allow for estimation of the blood pool by exploiting the relationship between electrical conductance and volume. When properly performed, these techniques allow for testing of cardiac function in genetic mutant mouse models or in drug treatment studies. The accuracy and precision of these studies are dependent on careful attention to the calibration of instruments, systematic conduct of hemodynamic measurements and data analyses. We will review the methods of conducting pressure-volume loop experiments using a conductance catheter in mice.

Introduction

Kardiyak basınç volüm analizi kalp fonksiyonu ayrıntılı bilgi sağlar ve fonksiyonel değerlendirme 1 altın standarttır. Böyle ekokardiyografi veya kardiyak MR gibi görüntüleme teknikleri, fonksiyonel tedbirleri sağlarken, bu önlemlerin yükleme koşulları son derece bağımlıdırlar. Kalp kontraktilite ve gevşeme Yük bağımsız önlemler ön ve son yüklerin aralığında ventriküler basınç ve hacim ilişkisi dinamik ölçümlerini gerektirmektedir. Basınç-hacim ilişkisi Bu anlayış Sagawa ve arkadaşları 2,3 çığır açan çalışmalarından kaynaklanmaktadır. Bunlar basınç-hacim döngü türetilmiş kontraktilite önlemleri yükleme koşulları 4 bağımsız olduğunu ex vivo olarak perfüze köpek kalplerinde gösterdi.

Bu analizlerin in vivo uygulamasında 1980'lerde iletkenliği kateter geliştirilmesi ile mümkün oldu. Ka izin Bu teknik ilerlemess ve arkadaşları insanlarda 5,6 basınç-volüm halkası analizi gerçekleştirmek için. Geç 1990 7 iletkenlik kateter ve cerrahi tekniklerdeki gelişmeler minyatürleştirilmesi yapılacak genetik ve farmakolojik çalışmalar için izin kemirgen kalp fonksiyonunun analizi mümkün yaptı. Bu peşin basınç-volüm döngü analizi yaygın kullanımı kurşun beri vardır ve memeli kalp fizyolojisi içgörü büyük bir üretti.

Iletkenlik kateter kullanımı ve ondan elde edilen verilerin yorumlanması bir anahtar kavram hacmi ve iletkenlik arasındaki ilişkidir. İletkenlik ters proksimale yerleştirilen elektrotlar, genellikle LV apeks 8'de, aort kapak altına yerleştirilir ve distalde bir kateter kullanılarak ölçülür gerilime ile ilgilidir. Voltaj veya iletkenlikteki değişimler proksimal uzak elektrot akan akım değişiklikleri ile ölçülür. Kan havuzu katkıda rağmenönemli ölçüde iletkenliğe s, ventriküler duvar katkısı, paralel iletkenlik (V p) olarak adlandırılan, ölçülen iletkenlik mutlak LV volüm ölçümleri elde etmek için çıkarılır olması gerekir.

Yöntemler bu düzeltme yapmak için, aşağıdaki protokolde ele alınmaktadır, tuzlu kalibrasyonu denir. Baan ve meslektaşları tarafından açıklanan iletkenlik ve hacim arasındaki matematiksel ilişki, bu hacim = 1 / α olduğunu; (ρ L 2) (GG p), burada α = üniforma alan düzeltme faktörü, ρ = kan direnç, L = elektrotlar arasındaki mesafe, G = iletkenlik ve G p = olmayan kan iletkenliği 9. Not, farelerde üniforma alan düzeltme faktörü nedeniyle küçük odacık hacimleri 10 1.0 yaklaşır. Basınç dönüştürücüler ile birleştiğinde, iletkenlik kateter gerçek zamanlı eşzamanlı basınç ve hacim verileri sağlar.

Kardiyak pressuonlar yükleme koşulları ve kalp hızından bağımsız ventrikül fonksiyonu ölçümü için izin verdiği yeniden hacim analizi, kalp fonksiyonunun diğer tedbirler üzerinde özel avantajlar sunuyor. Sonu sistolik basınç hacim ilişkisi (ESPVR) d P / d t max -end-diyastolik volüm ilişkisi, maksimal elastisite (E max) ve askere almalarla inme çalışmaları (PRSW) preload: kontraktilitenin Özgül yük bağımsız kalp endeksleri içerir. Diyastolik fonksiyon bir yük bağımsız önlem diyastol sonu basınç hacim ilişkisi (EDPVR) 11 olduğunu. Aşağıdaki protokol karotis ve apikal yaklaşım her ikisini de kullanarak, kardiyak basınç volüm halkası analizinin yürütülmesini açıklar. Bu çalışmaları gerçekleştirmek için metodoloji detaylı, daha önce 8,11 olarak tarif edilmiş olmasına rağmen, biz de tuzlu ve küvet kalibrasyon düzeltme de dahil olmak üzere hassas basınç-hacim ölçümleri elde etmek önemli adımlar gözden ve thes görsel gösteri sağlayacake prosedürleri. Bu çalışma için yürütülen hayvanlarla Araştırma protokollerle ve onaylanmış Duke Üniversitesi Tıp Merkezi'nin Kurumsal Hayvan Bakım ve Kullanım Kurulu hayvan refahı düzenlemelerine göre ele alınmıştır.

Protocol

1. iletkenlik Kateter Hazırlıklar ve Basınç Kalibrasyon Hemodinamik kateter modülüne iletkenlik kateteri bağlayın. Elektronik kateter modülünde ayarlanmış önceden ayarlanmış basınç ve hacim kaydederek basınç ve hacim ölçümlerini kalibre. 0 mm Hg ve 25 mm Hg (Şekil 1A) bir izleme kaydedin ve her iki basınç trasların (Şekil 1B ve 1C) ile gerilimleri atayın. Benzer şekilde, 5 RVU ve 25 RVU (Şekil 1D) bir hacim izlem…

Representative Results

Basınç-hacim döngü yöntemi, ilaç çalışmaları 16 uygulanan genetik olarak tadil edilmiş fareler 14,15 veya farelerde kalp fonksiyonu ölçmek için kullanılabilir. Temsili basınç hacim döngüler ß-arrestin etkisini araştıran daha önce yayınlanmış çalışmaları 16 sağlanmaktadır TRV120023, AT1R ligand önyargılı. TRV120023 in vivo kardiyak fonksiyonları etkiler olup olmadığını sınamak için, basınç-volüm döngü analizi konvansiyonel v…

Discussion

Biz kardiyak kontraktilite ve rahatlama hem de kapsamlı analizler elde etmek, farelerde bir iletkenlik kateter kullanılarak basınç-hacim döngü analizi perfoming için bir yöntem açıklanmaktadır. Suga, Sagawa ve arkadaşları basınç-volüm kardiyak kontraktilite önlemleri özellikle ESPVR eğimini veya sistol sonu elastans (E es) ve E max tanımlamak için döngüler kullanılır. Hacminin (P / V) basınç oranı ile tanımlanır elastisite, sistol süresi boyunca değişir. Her sistol …

Divulgaciones

The authors have nothing to disclose.

Acknowledgements

Bu çalışma, Amerikan Kalp Derneği 14FTF20370058 (DMA) ve NIH T32 HL007101-35 (DMA) tarafından desteklenmektedir.

Materials

AnaSed (xylazine)  Lloyd Laboratories NADA no. 139-236 Anesthetic
Ketaset (ketamine) Pfizer 440842 Anesthetic
VIP3000 Matrx Medical Inc. Anesthesia machine
Ventilator Harvard Apparatus Model 683 Surgical Equipment
Tubing kit Harvard Apparatus 72-1049 Surgical Equipment
Homeothermic Blanket  Kaz Inc. 5628 Surgical Equipment
Stereo microscope Carl Zeiss Optical Inc. Stemi 2000 Surgical Equipment
Illuminator Cole–Parmer 41720 Surgical Equipment
Dumont no. 55 Dumostar Forceps  Fine Science Tools Inc 11295-51 Surgical Instruments
Graefe forceps, curved  Fine Science Tools Inc 11052-10 Surgical Instruments
Moria MC31 forceps  Fine Science Tools Inc 11370-31 Surgical Instruments
Mayo scissors  Fine Science Tools Inc 14512-15 Surgical Instruments
Iris scissors  Fine Science Tools Inc 14041-10 Surgical Instruments
Halsey needle holder  Fine Science Tools Inc 12501-13 Surgical Instruments
Olsen–Hegar needle holder  Fine Science Tools Inc 12002-12 Surgical Instruments
spring scissors Fine Science Tools Inc 15610-08 Surgical Instruments
disposable underpads Kendall/Tyco Healthcare 1038 Surgical Supplies
Sterile gauze sponges, sterile  Dukal 62208 Surgical Supplies
Cotton-tipped applicators, sterile  Solon 368 Surgical Supplies
Surgical suture,  silk, 6-0  DemeTECH FT-639-1 Surgical Supplies
1 cc Insulin syringes  Becton Dickenson 329412 Surgical Supplies
Access-9™ Hemostasis Valve Merit Medical  MAP111 Hemodynamic equipment
Sphygmomanometer Baumanometer 320 Hemodynamic equipment
Millar PV system MPVS-300/400 or MPVS Ultra (includes calibration cuvette) ADInstruments Inc Hemodynamic equipment
1.4F conductance catheter  ADInstruments Inc SPR-839 Hemodynamic equipment
PowerLab 4/30 with Chart Pro ADInstruments Inc. ML866/P Hemodynamic software
animal clipper Wahl 8787-450A Miscellaneous
Intradermic tubing PE-10 (Becton Dickinson, cat. no. ) Becton Dickenson 427401 Miscellaneous
Intradermic tubing PE-50 (Becton Dickinson, cat. no.) Becton Dickenson 427411 Miscellaneous
Needle assortment (18, 25 and 30 gauge; Thomas Scientific) Miscellaneous
0.9% (wt/vol) sodium chloride injection, USP , cat. no. ) Hospira NDC no. 0409-4888-50 Miscellaneous
Surgical tape Miscellaneous
Alconox (Alconox Inc.) for catheter cleaning ADInstruments Inc. Miscellaneous

Referencias

  1. Clark, J. E., Marber, M. S. Advancements in pressure-volume catheter technology – stress remodelling after infarction. Exp Physiol. 98 (3), 614-621 (2013).
  2. Sunagawa, K., Maughan, W. L., Burkhoff, D., Sagawa, K. Left ventricular interaction with arterial load studied in isolated canine ventricle. Am J Physiol. 245 (Pt 1), H773-H780 (1983).
  3. Suga, H., Sagawa, K., Demer, L. Determinants of instantaneous pressure in canine left ventricle. Time and volume specification. Circ Res. 46 (2), 256-263 (1980).
  4. Suga, H., Sagawa, K., Shoukas, A. A. Load independence of the instantaneous pressure-volume ratio of the canine left ventricle and effects of epinephrine and heart rate on the ratio. Circ Res. 32 (3), 314-322 (1973).
  5. Kass, D. A., et al. Improved left ventricular mechanics from acute VDD pacing in patients with dilated cardiomyopathy and ventricular conduction delay. Circulation. 99 (12), 1567-1573 (1999).
  6. Kass, D. A., et al. Diastolic Compliance of Hypertrophied Ventricle Is Not Acutely Altered by Pharmacological Agents Influencing Active Processes. Annals of Internal Medicine. 119 (6), 466-473 (1993).
  7. Georgakopoulos, D., et al. In vivo murine left ventricular pressure-volume relations by miniaturized conductance micromanometry. Am J Physiol. 274 (4 pt 2), H1416-H1422 (1998).
  8. Cingolani, O. H., Kass, D. A. Pressure-volume relation analysis of mouse ventricular function. Am J Physiol Heart Circ Physiol. 301 (6), H2198-H2206 (2011).
  9. Baan, J., et al. Continuous Measurement of Left-Ventricular Volume in Animals and Humans by Conductance Catheter. Circulation. 70 (5), 812-823 (1984).
  10. Pearce, J. A., Porterfield, J. E., Larson, E. R., Valvano, J. W., Feldman, M. D. Accuracy considerations in catheter based estimation of left ventricular volume. Conf Proc IEEE Eng Med Biol Soc. 2010, 3556-3558 (2010).
  11. Pacher, P., Nagayama, T., Mukhopadhyay, P., Batkai, S., Kass, D. A. Measurement of cardiac function using pressure-volume conductance catheter technique in mice and rats. Nature Protocols. 3 (9), 1422-1434 (1038).
  12. Hanusch, C., Hoeger, S., Beck, G. C. Anaesthesia of small rodents during magnetic resonance imaging. Methods. 43 (1), 68-78 (2007).
  13. Georgakopoulos, D., Kass, D. A. Estimation of parallel conductance by dual-frequency conductance catheter in mice. Am J Physiol Heart Circ Physiol. 279 (1), H443-H450 (2000).
  14. Esposito, G., et al. Increased myocardial contractility and enhanced exercise function in transgenic mice overexpressing either adenylyl cyclase 5 or 8. Basic Res Cardiol. 103 (1), 22-30 (2008).
  15. Kohout, T. A., et al. Augmentation of cardiac contractility mediated by the human beta(3)-adrenergic receptor overexpressed in the hearts of transgenic mice. Circulation. 104 (20), 2485-2491 (2001).
  16. Kim, K. S., et al. beta-Arrestin-biased AT1R stimulation promotes cell survival during acute cardiac injury. Am J Physiol Heart Circ Physiol. 303 (8), H1001-H1010 (2012).
  17. Suga, H., Sagawa, K. Mathematical Interrelationship between Instantaneous Ventricular Pressure-Volume Ratio and Myocardial Force-Velocity Relation. Annals of Biomedical Engineering. 1 (2), 160-181 (1972).
  18. Suga, H. Ventricular energetics. Physiol Rev. 70 (2), 247-277 (1990).
  19. Kass, D. A., et al. Influence of contractile state on curvilinearity of in situ end-systolic pressure-volume relations. Circulation. 79 (1), 167-178 (1989).
  20. Little, W. C. The left ventricular dP/dtmax-end-diastolic volume relation in closed-chest dogs. Circ Res. 56 (6), 808-815 (1985).
  21. Glower, D. D., et al. Linearity of the Frank-Starling relationship in the intact heart: the concept of preload recruitable stroke work. Circulation. 71 (5), 994-1009 (1985).
  22. Sharir, T., et al. Ventricular systolic assessment in patients with dilated cardiomyopathy by preload-adjusted maximal power. Validation and noninvasive application. Circulation. 89 (5), 2045-2053 (1994).
  23. Baan, J., Van der Velde, E. T. Sensitivity of left ventricular end-systolic pressure-volume relation to type of loading intervention in dogs. Circ Res. 62 (6), 1247-1258 (1988).
  24. Wei, C. L., et al. Volume catheter parallel conductance varies between end-systole and end-diastole. IEEE Trans Biomed Eng. 54 (8), 1480-1489 (2007).
  25. Porterfield, J. E., et al. Dynamic correction for parallel conductance, GP, and gain factor, alpha, in invasive murine left ventricular volume measurements. J Appl Physiol (1985). 107 (6), 1693-1703 (2009).