We present a method for microfluidic deposition of patterned genipin and fibronectin on PDMS substrates, allowing extended viability of vascular smooth muscle cell-dense tissues. This tissue fabrication method is combined with previous vascular muscular thin film technology to measure vascular contractility over disease-relevant time courses.
The chronic nature of vascular disease progression requires the development of experimental techniques that simulate physiologic and pathologic vascular behaviors on disease-relevant time scales. Previously, microcontact printing has been used to fabricate two-dimensional functional arterial mimics through patterning of extracellular matrix protein as guidance cues for tissue organization. Vascular muscular thin films utilized these mimics to assess functional contractility. However, the microcontact printing fabrication technique used typically incorporates hydrophobic PDMS substrates. As the tissue turns over the underlying extracellular matrix, new proteins must undergo a conformational change or denaturing in order to expose hydrophobic amino acid residues to the hydrophobic PDMS surfaces for attachment, resulting in altered matrix protein bioactivity, delamination, and death of the tissues.
Here, we present a microfluidic deposition technique for patterning of the crosslinker compound genipin. Genipin serves as an intermediary between patterned tissues and PDMS substrates, allowing cells to deposit newly-synthesized extracellular matrix protein onto a more hydrophilic surface and remain attached to the PDMS substrates. We also show that extracellular matrix proteins can be patterned directly onto deposited genipin, allowing dictation of engineered tissue structure. Tissues fabricated with this technique show high fidelity in both structural alignment and contractile function of vascular smooth muscle tissue in a vascular muscular thin film model. This technique can be extended using other cell types and provides the framework for future study of chronic tissue- and organ-level functionality.
מחלות כלי דם, כגון vasospasm המוחית 1,2, יתר לחץ דם 3, וטרשת עורקת 4, לפתח לאט, הן בדרך כלל כרוניים בטבע, וכרוכים בכוח-דור מתפקד על ידי תאי שריר חלק של כלי דם (VSMCs). אנו שואפים ללמוד בתפקוד אלה איטיים מתקדמים של כלי הדם באמצעות בשיטות מבחנה עם שליטה עדינה יותר של תנאי ניסוי מאשר בin vivo מודלים. סרטים שפותחנו בעבר בכלי דם שרירי דק (vMTFs) למדידת התכווצות פונקציונלית של רקמות מהונדסות במבחנה לב וכלי דם 5, אך שיטה זו הוגבלה למחקרים יחסית לטווח קצר. כאן, אנו מציגים טכניקת שינוי מצע שמתרחבת טכניקת vMTF הקודמת שלנו למדידות לטווח ארוך.
בעוד האנדותל הוא גם קריטי בתפקוד כלי דם כללי, lamellae עורקים מהונדסים לספק מערכת מודל שימושית להערכת שינויים בכלי דםהתכווצות במהלך התקדמות מחלה. להנדס מודל פונקציונלי של כלי דם ברקמת מחלה, גם את המבנה ותפקוד של lamella העורקים, יחידת ההתכווצות הבסיסית של כלי השיט, יש סכמה עם איכות גבוהה. lamellae עורקים הם גיליונות קונצנטריים, circumferentially מיושר של VSMCs ההתכווצות מופרדת על ידי גיליונות של אלסטין 6. הדפסת microcontact של חלבוני מטריצה תאית (ECM) על גבי מצעי polydimethylsiloxane (PDMS) כבר השתמשה בעבר כדי לספק רמזי הדרכה לארגון רקמה לחקות מיושרים רקמת לב וכלי דם 5,7-10. עם זאת, רקמות דוגמת באמצעות הדפסת microcontact יכול לאבד יושרה לאחר 3-4 ימים בתרבות, הגבלת תחולתם בלימודים כרוניים. פרוטוקול זה מספק פתרון לבעיה זו על ידי החלפת טכניקות הדפסת microcontact קודמות עם טכניקה בתצהיר microfluidic חדשה.
Genchi et al. מצעי PDMS שונה עם genipin וFound ממושך כדאיות של myocytes עד חודש בתרבות 11. כאן, אנו משתמשים בגישה דומה להארכת תרבות של תאי שריר בדוגמת כלי דם חלקים על PDMS. Genipin, נגזר hydrolytic טבעית של הפירות גרדניה, הוא מועמד רצוי לשינוי מצע בשל רעילותה הנמוכה יחסית בהשוואה לסוכני crosslinking דומים והשימוש הגובר שלה כביולוגי בתחומי תיקון רקמות 12,13 ושינוי ECM 14, 15. בפרוטוקול זה, פיברונקטין הוא מנוצל כרמז הדרכת תא, כמו בשיטות הדפסת microcontact קודמות; עם זאת, genipin מופקד על גבי מצעים לפני דפוסי פיברונקטין PDMS. לפיכך, כתאים לבזות את המטריצה בדוגמת, ECM מסונתז חדש מVSMCs המצורף יכול להיקשר למצע PDMS המצופה genipin.
פרוטוקול זה משתמש בהתקן משלוח microfluidic לgenipin שני שלבים ותצהיר ECM. העיצוב של microco מחקה מכשיר microfluidicדפוסי הדפסת ntact משמשים לlamellae עורקים מהונדסים במחקרים קודמים 16. לפיכך, אנו מצפים פרוטוקול זה להניב מחקה lamellae עורקים שבהצלחה לשחזר מזדהים ביותר במבנה ותפקוד vivo התכווצות של lamellae עורקים. אנו גם מעריכים התכווצות רקמות כדי לאשר genipin שהיא תרכובת שינוי המצע מתאימה לטווח ארוך בדגמי מחלת כלי דם במבחנה.
כאן, אנו מציגים פרוטוקול שבונה על טכנולוגיה שפותחה בעבר vMTF, מאפשרים פעמים ניסוי מורחבות טיפוסיות יותר של מסלולי מחלת כלי דם כרוניים 1,23,24. כדי להשיג זאת, אנו micropattern genipin, אשר בעבר הוכח לספק functionalization לטווח הארוך של מצעי PDMS 11, תוך שימוש בטכניקה בתצהיר microfluidic ל?…
The authors have nothing to disclose.
We acknowledge financial support from the American Heart Association Scientist Development Grant, 13SDG14670062 (PWA) and the University of Minnesota Doctoral Dissertation Fellowship (ESH). We also acknowledge the microfabrication resources of the Minnesota Nano Center (MNC) and the image processing resources of the University Imaging Centers (UIC), both at the University of Minnesota. Parts of this work were carried out in the Characterization Facility, University of Minnesota, which receives partial support from NSF through the MRS program.
Coverslip staining rack | Electron Microscopy Sciences | www.emsdiasum.com/ | 72239-04 | Alternative coverslip rack may be used |
Microscope cover glass – 25 mm | Fisher Scientific, Inc. | www.fishersci.com | 12-545-102 | Alternative brand and size may be used; Microscope slides may also be substituted as substrate base |
Poly(N-iso-propylacrylamide) (PIPAAm) | Polysciences, Inc. | www.polysciences.com/ | #21458 | Sigma-Aldrich makes an alternate compound, but we have not tested it for use with this protocol; Compound gives strong odor, use proper ventilation |
1-butanol | Sigma-Aldrich | www.sigmaaldrich.com | 360465 | Hazard: flammable (store stock solution in flammable cabinet); flash point is 37 °C, avoid heating; alternative product may be used |
Spincoater | Specialty Coating Systems, Inc. | www.scscoatings.com | SCS G3P8 Model; Alternative brand and/or model may be used | |
Polydimethylsiloxane (PDMS) | Ellsworth Adhesives (Dow Corning) | www.ellsworth.com | 184 SIL ELAST KIT 0.5KG | Alternative distributor may be used |
Fluorescent microbeads | Polysciences, Inc. | www.polysciences.com/ | 17151 | Alternative brand and/or larger size may be used |
Silicon wafers | Wafer World, Inc. | www.waferworld.com | 2398 | Alternative brand and/or size may be used |
Photoresist | MicroChem Corp. | www.microchem.com | SU-8 3025 allows 20-25-µm feature height | |
Contact mask aligner | Suss MicroTec | www.suss.com | MA6 contact mask aligner; alternative brand and/or model may be used for wafer exposure | |
Developer | MicroChem Corp. | www.microchem.com | SU-8 Developer; Hazard: flammable | |
Tridecafluro-trichlorosilane | UCT Specialties, Inc. | www.unitedchem.com | T2492 | Silane for non-stick coating of patterned silicon wafers (CAUTION: Tridecafluro-trichlorosilane is a flammable and corrosive liquid. Proper personal protective equipment and local exhaust is necessary for use. ) |
Surgical biopsy punch | Integra LifeSciences Corp. | www.miltex.com | 33-31AA-P/25 | Alternative brand and/or size may be used |
Genipin | Cayman Chemical | www.caymanchem.com | 10010622 | Sigma-Aldrich (G4796-25MG) makes an alternate compound, but we have not tested it for use with this protocol |
1X phosphate buffered saline | Mediatech, Inc. | www.cellgro.com | 21-031-CV | Alternative brand may be used |
Fibronectin | Corning, Inc. | www.corning.com | 356008 | Sigma-Aldrich (F1056) makes an alternate compound, but we have not tested it for use with this protocol |
Penicillin/streptomycin | Life Technologies, Inc. | www.lifetechnologies.com | 15140-122 | Alternative brand and/or size may be used, as long as concentration is the same |
Umbillical artery smooth muscle cells | Lonza | www.lonza.com | CC-2579 | Alternative cell types may be used for alternative applications. Media should be modified accordingly |
Tyrode's solution components | Sigma-Aldrich | www.sigmaaldrich.com | various | Alternative brand may be used for mixing solution |
Stereomicroscope | Zeiss | www.zeiss.com | 4350020000000000 | SteREOLumar V12; Alternative brand/type of stereomicroscope may be used |
Temperature-controlled platform | Warner Instruments | www.warneronline.com | 641659; 640352; 641922 | |
Endothelin-1 | Sigma-Aldrich | www.sigmaaldrich.com | E7764-50UG | Alternative amount may be purchased, as long as treatment concentration is maintained |
HA-1077 | Sigma-Aldrich | www.sigmaaldrich.com | H139-10MG | Alternative amount may be purchased, as long as treatment concentration is maintained |