Summary

Helautomatisk Sentrifugal mikrofluid Enhet for Ultra Protein Detection fra Whole Blood

Published: April 16, 2016
doi:

Summary

This protocol demonstrates how to achieve femto molar detection sensitivity of proteins in 10 µL of whole blood within 30 min. This can be achieved by using electrospun nanofibrous mats integrated in a lab-on-a-disc, which offers high surface area as well as effective mixing and washing for enhanced signal-to-noise ratio.

Abstract

Enzyme-linked immunosorbent assay (ELISA) is a promising method to detect small amount of proteins in biological samples. The devices providing a platform for reduced sample volume and assay time as well as full automation are required for potential use in point-of-care-diagnostics. Recently, we have demonstrated ultrasensitive detection of serum proteins, C-reactive protein (CRP) and cardiac troponin I (cTnI), utilizing a lab-on-a-disc composed of TiO2 nanofibrous (NF) mats. It showed a large dynamic range with femto molar (fM) detection sensitivity, from a small volume of whole blood in 30 min. The device consists of several components for blood separation, metering, mixing, and washing that are automated for improved sensitivity from low sample volumes. Here, in the video demonstration, we show the experimental protocols and know-how for the fabrication of NFs as well as the disc, their integration and the operation in the following order: processes for preparing TiO2 NF mat; transfer-printing of TiO2 NF mat onto the disc; surface modification for immune-reactions, disc assembly and operation; on-disc detection and representative results for immunoassay. Use of this device enables multiplexed analysis with minimal consumption of samples and reagents. Given the advantages, the device should find use in a wide variety of applications, and prove beneficial in facilitating the analysis of low abundant proteins.

Introduction

Flere plattformer for sykdom diagnose har blitt utviklet basert på nanoskala materialer 1,2 eksempel nanotråder, 3 nanopartikler, 4 nanorør, 5 og nanofibers (NFS) 6-8. Disse nanomaterialer har gode utsikter i utformingen av nye teknologier for svært sensitive bioassay grunn av deres unike fysiske og kjemiske egenskaper. For eksempel har mesoporous sink oksid nanofibers blitt brukt for femto-molar sensitiv deteksjon av brystkreft biomarkører. 9 Nylig nanomaterialer basert på titandioksid (TiO2) har blitt undersøkt for Bioanalytical applikasjoner 10 vurderer sin kjemiske stabilitet, 11 ubetydelig protein denaturering , 12 og biokompatibilitet. 13 i tillegg er de hydroksylgrupper på overflaten av TiO 2 lette kjemisk modifikasjon og kovalent binding av biomolekyler. 14,15 mønstret TiO 2 thin filmer 16 eller TiO 2 nanorørene 17 er blitt benyttet for å forbedre deteksjonsfølsomheten av et mål-protein ved å øke overflatearealet; imidlertid, er fremstillingsprosessen ganske komplisert og krever kostbart utstyr. På den annen side er elektrospunnede NFS mottar oppmerksomhet på grunn av deres høyt overflateareal, så vel som enkel og billig fremstillingsprosess; 18,19 ennå, den skjøre eller løs karakteristisk for elektrospunnede TiO 2 NF matten gjør det vanskelig å håndtere og integreres med microfluidic enheter. 6,20 ble derfor TiO 2 NF matter sjelden benyttet i Bioanalytical anvendelser, spesielt de som krever sterke vaskebetingelser.

I denne studien, for å overvinne disse begrensninger har vi utviklet en ny teknologi for overføring av elektrospunnede NF matter på overflaten av enhver target-substrat ved å benytte en tynn polydimetylsiloksan (PDMS) klebesjikt. furthermore, har vi lykkes viste integreringen av elektrospunnede TiO to NF-matter på en sentrifugal-mikrofluidanordning fremstilt av polykarbonat (PC). Ved hjelp av denne anordning, ble en høy-følsom, helautomatisk, og integrert påvisning av C-reaktivt protein (CRP), så vel som hjerte troponin I (cTnI) oppnådd i løpet av 30 min fra bare 10 ul av fullblod. 21 På grunn av den kombinerte fordelene med egenskapene til NFS og sentrifugal plattformen, analysen viste et bredt dynamisk område av seks størrelsesordener fra 1 pg / ml (~ 8 FM) til 100 ng / ml (~ 0,8 pM) med en nedre deteksjonsgrense på 0,8 pg / ml (~ 6 FM) for CRP og et dynamisk område fra 10 pg / ml (~ 0,4 pm) til 100 ng / ml (~ 4 nM) med en påvisningsgrense på 37 pg / ml (~ 1,5 pM) for cTnI. Disse deteksjonsgrenser er ~ 300 ~ og 20-ganger lavere enn de tilsvarende konvensjonelle ELISA-resultater. Denne teknikken kan anvendes for påvisning av enhver target-proteiner, med passende antistoffer. Samlet sett denne enheten could bidra sterkt til in vitro-diagnostikk og biokjemiske analyser, siden det kan detektere sjeldne mengder av target-proteiner med stor følsomhet, selv fra meget små mengder av biologiske prøver, for eksempel, 10 pl fullblod. Selv om vi bare vist serumproteindeteksjon ved hjelp av ELISA i denne studien, kan overføring og integrering teknologi av elektrospunnede NFS med microfluidic enhetene være mer bredt anvendt i andre biokjemiske reaksjoner som krever et stort overflateareal for høy deteksjonsfølsomhet.

Protocol

MERK: Blod ble tatt fra friske individer og ble samlet i en blodprøverøret. Skriftlig informert samtykke ble innhentet fra alle frivillige. 1. Fabrikasjon av TiO 2 NF Mat Utarbeidelse av forløper løsning 22 Oppløs 1,5 g titantetraisopropoksyd (TTIP) i en blanding av etanol (99,9%, 3 ml) og iseddik (3 ml) og blande løsningen ved romtemperatur (25 ° C) i 30 min på en magnetrører. Oppløs 11 vekt% po…

Representative Results

Ved hjelp av denne protokollen, en helautomatisk sentrifugal mikrofluidanordning for protein deteksjon fra fullblod med høy følsomhet ble utarbeidet. De TiO 2 NF matter ble utarbeidet av prosesser av electro og kalsinering. For å fremstille NFS av ønsket diameter, morfologi og tykkelse, electro tilstander så som strømningsmengde, spenning, og spinning tid ble optimalisert. Når forholdene ikke var optimalisert, kvaliteten på NFS dannet var dårlig. Spesielt ble NFS ikke…

Discussion

Analysen på TiO 2 NF integrert plate er en hurtig, billig og praktisk teknikk for ultra påvisning av lave rikelig proteiner tilstede i meget lave volumet av blod. Denne teknikken har den fordelen av å bruke små prøvevolumer (10 ul), og er mottagelig for analyse av flere prøver samtidig. Dette gir et stort potensial som en multiplexing immunoassay enhet. Enheten har den ekstra fordelen at den ikke krever prøve forbehandlingstrinn som plasma separasjon, som er nødvendig i konvensjonelle ELISA. Dessuten …

Disclosures

The authors have nothing to disclose.

Acknowledgements

Dette arbeidet ble støttet av National Research Foundation of Korea (NRF) tilskudd (2013R1A2A2A05004314, 2012R1A1A2043747), et stipend fra den koreanske Health Technology R & D Project, Ministry of Health & Welfare (A121994) og IBS-R020-D1 finansiert av den koreanske regjeringen.

Materials

Si wafer LG SILTRON Polished Wafer, test grade Dia. (mm) = 150, orientation = <100>, dopant = boron, RES(Ohm-cm) = 1 – 30, thickness (μm) = 650 – 700
Polycarbonate (PC)  Daedong Plastic PCS#6900 Thickness (mm) = 1 and 5 
Titanium tetraisopropoxide, 98%, Sigma-Aldrich 205273
Polyvinylpyrrolidone, Mw = 1,300,000 Sigma-Aldrich 437190
Acetic acid Sigma-Aldrich 320099
Anhydrous ethanol Sigma-Aldrich 459836
Tridecafluoro-1,1,2,2-tetrahydrooctyl)-1-trichlorosilane Sigma-Aldrich 448931
PDMS and curing agent Dow Corning SYLGARD 184
GPDES Gelest Inc SIG5832.0 
Ethanol J T Baker
FE-SEM FEI Nova NanoSEM
X-ray photoelectron spectroscopy ThermoFisher K-alpha
3D modeling machine M&I CNC Lab, Korea CNC milling machine
Wax-dispensing machine Hanra Precision Eng. Co. Ltd., Korea Customized
Double-sided adhesive tape FLEXcon, USA DFM 200 clear 150 POLY H-9 V-95
Cutting plotter Graphtec Corporation, Japan Graphtec CE3000-60 MK2
Spin coater MIDAS SPIN-3000D
Furnace (calcination) R. D. WEBB COMPANY WEBB 99
Rheometer (Tack test) Thermo Scientific Haake MARS III – ORM Package
Oxygen plasma system FEMTO CUTE
Monoclonal mouse antihuman hsCRP Hytest Ltd., Finland 4C28 (clone # C5)
Monoclonal mouse anti-cTnI Hytest Ltd., Finland 4T21 (clone # 19C7)
HRP conjugated goat polyclonal anti-hsCRP Abcam plc., MA ab19175
HRP conjugated mouse monoclonal anti-cTnI Abcam plc., MA ab24460 (clone # 16A11)
hsCRP Abcam plc., MA ab111647
cTnI Fitzgerald, MA 30-AT43
Bovine Albumin Sigma-Aldrich A7906
PBS Amresco Inc E404
Blood collection tubes BD vacutainer 367844 K2 EDTA 7.2 mg plus blood
collection tubes
SuperSignal ELISA femto Invitrogen 37074
Modular multilabel plate reader Perkin Elmer Envision 2104
Disc operating machine Hanra Precision Eng. Co. Ltd., Korea Customized
Photomultiplier tube (PMT) Hamamatsu Photonics H1189-210
AutoCAD AutoDesk Version 2012 Design software
SolidWorks 3D CAD software  SOLIDWORKS Corp. Version 2013 3D Design software,
Edgecam Vero software version 2009.01.06928 Code generating software
DeskCNC Carken Co. version 2.0.2.18 CNC milling machine software

References

  1. Zhang, Y., et al. Nanomaterials for Ultrasensitive Protein Detection. Adv. Mater. 25 (28), 3802-3819 (2013).
  2. Hu, W., Li, C. M. Nanomaterial-based advanced immunoassays. Wiley Interdiscip. Rev. Nanomed. Nanobiotechnol. 3 (2), 119-133 (2011).
  3. Yang-Kyu, C., Chang-Hoon, K. Silicon Nanowire Biosensor for Cancer Markers. Biosensors and Cancer. , 164-183 (2012).
  4. Baltazar, R., Vistas, C. R., Ferreira, G. M. Biosensing Applications Using Nanoparticles. Nanocomposite Particles for Bio-Applications. , 265-282 (2011).
  5. Roy, P., Berger, S., Schmuki, P. TiO2 Nanotubes: Synthesis and Applications. Angew. Chem. Int. Ed. 50 (13), 2904-2939 (2011).
  6. Yang, D., et al. Electrospun Nanofibrous Membranes: A Novel Solid Substrate for Microfluidic Immunoassays for HIV. Adv. Mater. 20 (24), 4770-4775 (2008).
  7. Chantasirichot, S., Ishihara, K. Electrospun phospholipid polymer substrate for enhanced performance in immunoassay system. Biosens. Bioelectron. 38 (1), 209-214 (2012).
  8. Zhang, N., et al. Electrospun TiO2 Nanofiber-Based Cell Capture Assay for Detecting Circulating Tumor Cells from Colorectal and Gastric Cancer Patients. Adv. Mater. 24 (20), 2756-2760 (2012).
  9. Ali, M. A., Mondal, K., Singh, C., Dhar Malhotra, B., Sharma, A. Anti-epidermal growth factor receptor conjugated mesoporous zinc oxide nanofibers for breast cancer diagnostics. Nanoscale. 7 (16), 7234-7245 (2015).
  10. Mondal, K., Ali, M. A., Agrawal, V. V., Malhotra, B. D., Sharma, A. Highly Sensitive Biofunctionalized Mesoporous Electrospun TiO2 Nanofiber Based Interface for Biosensing. ACS Appl. Mater. Interfaces. 6 (4), 2516-2527 (2014).
  11. Tu, W., Dong, Y., Lei, J., Ju, H. Low-Potential Photoelectrochemical Biosensing Using Porphyrin-Functionalized TiO2 Nanoparticles. Anal. Chem. 82 (20), 8711-8716 (2010).
  12. Liu, S., Chen, A. Coadsorption of Horseradish Peroxidase with Thionine on TiO2 Nanotubes for Biosensing. Langmuir. 21 (18), 8409-8413 (2005).
  13. Portan, D. V., Kroustalli, A. A., Deligianni, D. D., Papanicolaou, G. C. On the biocompatibility between TiO2 nanotubes layer and human osteoblasts. J.Biomed.Mater.Res. Part A. 100 (10), 2546-2553 (2012).
  14. Dettin, M., et al. Covalent surface modification of titanium oxide with different adhesive peptides: Surface characterization and osteoblast-like cell adhesion. J. Biomed. Mater. Res. Part A. 90 (1), 35-45 (2009).
  15. Kim, W. -. J., et al. Enhanced Protein Immobilization Efficiency on a TiO2 Surface Modified with a Hydroxyl Functional Group. Langmuir. 25 (19), 11692-11697 (2009).
  16. Son, K. J., Ahn, S. H., Kim, J. H., Koh, W. -. G. Graft Copolymer-Templated Mesoporous TiO2 Films Micropatterned with Poly(ethylene glycol) Hydrogel: Novel Platform for Highly Sensitive Protein Microarrays. ACS Appl. Mater. Interfaces. 3 (2), 573-581 (2011).
  17. Kar, P., Pandey, A., Greer, J. J., Shankar, K. Ultrahigh sensitivity assays for human cardiac troponin I using TiO2 nanotube arrays. Lab Chip. 12 (4), 821-828 (2012).
  18. Agarwal, S., Wendorff, J. H., Greiner, A. Use of electrospinning technique for biomedical applications. Polymer. 49 (26), 5603-5621 (2008).
  19. Ding, B., Wang, M., Wang, X., Yu, J., Sun, G. Electrospun nanomaterials for ultrasensitive sensors. Mater. Today. 13 (11), 16-27 (2010).
  20. Liu, Y., Yang, D., Yu, T., Jiang, X. Incorporation of electrospun nanofibrous PVDF membranes into a microfluidic chip assembled by PDMS and scotch tape for immunoassays. ELECTROPHORESIS. 30 (18), 3269-3275 (2009).
  21. Lee, W. S., Sunkara, V., Han, J. -. R., Park, Y. -. S., Cho, Y. -. K. Electrospun TiO2 nanofiber integrated lab-on-a-disc for ultrasensitive protein detection from whole blood. Lab Chip. 15 (2), 478-485 (2015).
  22. Li, D., Xia, Y. Fabrication of Titania Nanofibers by Electrospinning. Nano Lett. 3 (4), 555-560 (2003).
  23. Lombard, M. . SolidWorks 2013 BIBLE. , (2013).
  24. Tickoo, S. . EdgeCAM 11.0 for Manufacturers. , (2007).
  25. Zhu, R., et al. Improved adhesion of interconnected TiO2 nanofiber network on conductive substrate and its application in polymer photovoltaic devices. Appl. Phys. Lett. 93 (1), 013102 (2008).
  26. Song, M. Y., Ahn, Y. R., Jo, S. M., Kim, D. Y., Ahn, J. -. P. TiO2 single-crystalline nanorod electrode for quasi-solid-state dye-sensitized solar cells. Appl. Phys. Lett. 87 (11), 113113 (2005).
  27. Katsuhiro, O., et al. Electrospinning processed nanofibrous TiO2 membranes for photovoltaic applications. Nanotechnology. 17 (4), 1026-1031 (2006).
check_url/54143?article_type=t

Play Video

Cite This Article
Park, Y., Sunkara, V., Kim, Y., Lee, W. S., Han, J., Cho, Y. Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood. J. Vis. Exp. (110), e54143, doi:10.3791/54143 (2016).

View Video