Summary

Totalmente dispositivo automático centrífugo microfluídicos para a detecção de proteínas Ultrasensitive de Sangue Total

Published: April 16, 2016
doi:

Summary

This protocol demonstrates how to achieve femto molar detection sensitivity of proteins in 10 µL of whole blood within 30 min. This can be achieved by using electrospun nanofibrous mats integrated in a lab-on-a-disc, which offers high surface area as well as effective mixing and washing for enhanced signal-to-noise ratio.

Abstract

Enzyme-linked immunosorbent assay (ELISA) is a promising method to detect small amount of proteins in biological samples. The devices providing a platform for reduced sample volume and assay time as well as full automation are required for potential use in point-of-care-diagnostics. Recently, we have demonstrated ultrasensitive detection of serum proteins, C-reactive protein (CRP) and cardiac troponin I (cTnI), utilizing a lab-on-a-disc composed of TiO2 nanofibrous (NF) mats. It showed a large dynamic range with femto molar (fM) detection sensitivity, from a small volume of whole blood in 30 min. The device consists of several components for blood separation, metering, mixing, and washing that are automated for improved sensitivity from low sample volumes. Here, in the video demonstration, we show the experimental protocols and know-how for the fabrication of NFs as well as the disc, their integration and the operation in the following order: processes for preparing TiO2 NF mat; transfer-printing of TiO2 NF mat onto the disc; surface modification for immune-reactions, disc assembly and operation; on-disc detection and representative results for immunoassay. Use of this device enables multiplexed analysis with minimal consumption of samples and reagents. Given the advantages, the device should find use in a wide variety of applications, and prove beneficial in facilitating the analysis of low abundant proteins.

Introduction

Várias plataformas para o diagnóstico da doença foram desenvolvidos com base em materiais em nanoescala 1,2 tais como nanofios, 3 nanopartículas, nanotubos de 4, 5 e nanofibras (NFS) 6-8. Estes nanomateriais oferecem excelentes perspectivas na concepção de novas tecnologias para bioensaios altamente sensíveis devido às suas propriedades físico-químicas únicas. Por exemplo, nanofibras de óxido de zinco mesoporosa têm sido utilizados para a detecção sensível de femto-molar de biomarcadores de cancro da mama. 9 Recentemente, os nanomateriais à base de dióxido de titânio (TiO 2) têm sido exploradas para aplicações bioanaliticas 10 considerando-se a sua estabilidade química, 11 desnaturação de proteínas negligenciável , 12 e biocompatibilidade. 13 Além disso, os grupos hidroxilo na superfície de TiO2 facilitar a modificação química e a ligação covalente de biomoléculas. modelado 14,15 TiO 2 thin películas 16 ou 17 TiO 2 nanotubos têm sido utilizados para aumentar a sensibilidade de detecção de uma proteína-alvo, aumentando a área de superfície; No entanto, o processo de fabricação é bastante complexo e requer um equipamento caro. Por outro lado, as FN electrospun estão a receber atenção devido à sua elevada área de superfície, bem como processo de fabrico simples e de baixo custo; 18,19 ainda, a característica frágil ou solto da esteira electrospun TiO 2 NF torna difícil de manusear e integrar com dispositivos microfluídicos. 6,20 Portanto, os TiO 2 esteiras NF raramente foram utilizados em aplicações bioanalíticos, particularmente aquelas que requerem condições de lavagem agressivos.

Neste estudo, para ultrapassar estas limitações, desenvolvemos uma nova tecnologia para a transferência do electrospun NF esteiras sobre a superfície de qualquer substrato alvo, utilizando uma camada de adesivo fina polidimetilsiloxano (PDMS). Furthermore, que demonstraram com sucesso a integração de electrospun TiO 2 NF esteiras para um dispositivo de microfluidos centrífuga de policarbonato (PC). Usando este dispositivo, uma detecção sensível de alta, totalmente automatizada e integrada de proteína C-reactiva (CRP), bem como a troponina I cardíaca (TIc) foi alcançada em 30 min a partir de apenas 10 ul de sangue total. 21 Devido ao combinado vantagens das propriedades da FN e a plataforma de centrífuga, o ensaio apresentava uma vasta gama dinâmica de seis ordens de grandeza a partir de 1 pg / ml (~ 8 fm) a 100 ng / ml (~ 0,8 pM) com um limite de detecção inferior de 0,8 pg / ml (~ 6 fm) para PCR e uma faixa dinâmica de 10 pg / ml (~ 0,4 pM) e 100 ng / ml (~ 4 nM) com um limite de detecção de 37 pg / ml (~ 1,5 pM) para cTnI. Estes limites de detecção são ~ 300 e ~ 20 vezes mais baixa em comparação com os seus correspondentes resultados de ELISA convencionais. Esta técnica pode ser aplicada para a detecção de quaisquer proteínas alvo, com anticorpos apropriados. No geral, este co dispositivould contribuir grandemente para diagnóstico in-vitro e ensaios bioquímicos uma vez que pode detectar quantidades raros de proteínas-alvo com grande sensibilidade mesmo a partir de quantidades muito pequenas de amostras biológicas, por exemplo, 10 ml de sangue total. Embora apenas se demonstrou a detecção de proteína de soro utilizando ELISA no presente estudo, a tecnologia de transferência e integração de electrospun FN com dispositivos de microfluidos pode ser mais amplamente aplicado em outras reacções bioquímicas que requerem uma grande área de superfície para detecção de alta sensibilidade.

Protocol

NOTA: O sangue foi coletado de indivíduos saudáveis ​​e foi recolhido em um tubo de coleta de sangue. consentimento informado por escrito foi obtido de todos os voluntários. 1. Fabricação de TiO 2 NF Mat Preparação da solução precursora de 22 Dissolve-se 1,5 g de tetra-isopropóxido de titânio (TTIP) numa mistura de etanol (99,9%, 3 ml) e ácido acético glacial (3 ml) e misturar a solução à TA (25 …

Representative Results

Usando este protocolo, um dispositivo de microfluidos centrífuga totalmente automatizado para a detecção de proteína a partir de sangue inteiro com elevada sensibilidade foi preparado. As esteiras de TiO 2 NF foram preparados por processos de electrospinning e calcinação. A fim de fabricar o diâmetro desejado de FN, morfologia, e sua espessura, electrospinning condições tais como a taxa de fluxo, tensão, e o tempo de fiação foram optimizados. Quando as condições …

Discussion

O ensaio em TiO 2 NF disco integrada é uma técnica rápida, barata e conveniente para a detecção ultra-sensível de baixas proteínas abundantes presentes em muito baixo volume de sangue. Esta técnica tem a vantagem da utilização de pequenos volumes de amostra (10 ul) e é receptivo para a análise de múltiplas amostras em simultâneo. Isto proporciona um grande potencial como um dispositivo de multiplexação de imunoensaio. O dispositivo tem a vantagem adicional de que ele não requer passos de pré…

Disclosures

The authors have nothing to disclose.

Acknowledgements

Este trabalho foi financiado pela Fundação Nacional de Pesquisa da Coreia (NRF) subvenções (2013R1A2A2A05004314, 2012R1A1A2043747), uma bolsa da Korean Saúde Tecnologia R & D Project, Ministério da Saúde e Bem-Estar (A121994) e IBS-R020-D1 financiado pelo Governo coreano.

Materials

Si wafer LG SILTRON Polished Wafer, test grade Dia. (mm) = 150, orientation = <100>, dopant = boron, RES(Ohm-cm) = 1 – 30, thickness (μm) = 650 – 700
Polycarbonate (PC)  Daedong Plastic PCS#6900 Thickness (mm) = 1 and 5 
Titanium tetraisopropoxide, 98%, Sigma-Aldrich 205273
Polyvinylpyrrolidone, Mw = 1,300,000 Sigma-Aldrich 437190
Acetic acid Sigma-Aldrich 320099
Anhydrous ethanol Sigma-Aldrich 459836
Tridecafluoro-1,1,2,2-tetrahydrooctyl)-1-trichlorosilane Sigma-Aldrich 448931
PDMS and curing agent Dow Corning SYLGARD 184
GPDES Gelest Inc SIG5832.0 
Ethanol J T Baker
FE-SEM FEI Nova NanoSEM
X-ray photoelectron spectroscopy ThermoFisher K-alpha
3D modeling machine M&I CNC Lab, Korea CNC milling machine
Wax-dispensing machine Hanra Precision Eng. Co. Ltd., Korea Customized
Double-sided adhesive tape FLEXcon, USA DFM 200 clear 150 POLY H-9 V-95
Cutting plotter Graphtec Corporation, Japan Graphtec CE3000-60 MK2
Spin coater MIDAS SPIN-3000D
Furnace (calcination) R. D. WEBB COMPANY WEBB 99
Rheometer (Tack test) Thermo Scientific Haake MARS III – ORM Package
Oxygen plasma system FEMTO CUTE
Monoclonal mouse antihuman hsCRP Hytest Ltd., Finland 4C28 (clone # C5)
Monoclonal mouse anti-cTnI Hytest Ltd., Finland 4T21 (clone # 19C7)
HRP conjugated goat polyclonal anti-hsCRP Abcam plc., MA ab19175
HRP conjugated mouse monoclonal anti-cTnI Abcam plc., MA ab24460 (clone # 16A11)
hsCRP Abcam plc., MA ab111647
cTnI Fitzgerald, MA 30-AT43
Bovine Albumin Sigma-Aldrich A7906
PBS Amresco Inc E404
Blood collection tubes BD vacutainer 367844 K2 EDTA 7.2 mg plus blood
collection tubes
SuperSignal ELISA femto Invitrogen 37074
Modular multilabel plate reader Perkin Elmer Envision 2104
Disc operating machine Hanra Precision Eng. Co. Ltd., Korea Customized
Photomultiplier tube (PMT) Hamamatsu Photonics H1189-210
AutoCAD AutoDesk Version 2012 Design software
SolidWorks 3D CAD software  SOLIDWORKS Corp. Version 2013 3D Design software,
Edgecam Vero software version 2009.01.06928 Code generating software
DeskCNC Carken Co. version 2.0.2.18 CNC milling machine software

References

  1. Zhang, Y., et al. Nanomaterials for Ultrasensitive Protein Detection. Adv. Mater. 25 (28), 3802-3819 (2013).
  2. Hu, W., Li, C. M. Nanomaterial-based advanced immunoassays. Wiley Interdiscip. Rev. Nanomed. Nanobiotechnol. 3 (2), 119-133 (2011).
  3. Yang-Kyu, C., Chang-Hoon, K. Silicon Nanowire Biosensor for Cancer Markers. Biosensors and Cancer. , 164-183 (2012).
  4. Baltazar, R., Vistas, C. R., Ferreira, G. M. Biosensing Applications Using Nanoparticles. Nanocomposite Particles for Bio-Applications. , 265-282 (2011).
  5. Roy, P., Berger, S., Schmuki, P. TiO2 Nanotubes: Synthesis and Applications. Angew. Chem. Int. Ed. 50 (13), 2904-2939 (2011).
  6. Yang, D., et al. Electrospun Nanofibrous Membranes: A Novel Solid Substrate for Microfluidic Immunoassays for HIV. Adv. Mater. 20 (24), 4770-4775 (2008).
  7. Chantasirichot, S., Ishihara, K. Electrospun phospholipid polymer substrate for enhanced performance in immunoassay system. Biosens. Bioelectron. 38 (1), 209-214 (2012).
  8. Zhang, N., et al. Electrospun TiO2 Nanofiber-Based Cell Capture Assay for Detecting Circulating Tumor Cells from Colorectal and Gastric Cancer Patients. Adv. Mater. 24 (20), 2756-2760 (2012).
  9. Ali, M. A., Mondal, K., Singh, C., Dhar Malhotra, B., Sharma, A. Anti-epidermal growth factor receptor conjugated mesoporous zinc oxide nanofibers for breast cancer diagnostics. Nanoscale. 7 (16), 7234-7245 (2015).
  10. Mondal, K., Ali, M. A., Agrawal, V. V., Malhotra, B. D., Sharma, A. Highly Sensitive Biofunctionalized Mesoporous Electrospun TiO2 Nanofiber Based Interface for Biosensing. ACS Appl. Mater. Interfaces. 6 (4), 2516-2527 (2014).
  11. Tu, W., Dong, Y., Lei, J., Ju, H. Low-Potential Photoelectrochemical Biosensing Using Porphyrin-Functionalized TiO2 Nanoparticles. Anal. Chem. 82 (20), 8711-8716 (2010).
  12. Liu, S., Chen, A. Coadsorption of Horseradish Peroxidase with Thionine on TiO2 Nanotubes for Biosensing. Langmuir. 21 (18), 8409-8413 (2005).
  13. Portan, D. V., Kroustalli, A. A., Deligianni, D. D., Papanicolaou, G. C. On the biocompatibility between TiO2 nanotubes layer and human osteoblasts. J.Biomed.Mater.Res. Part A. 100 (10), 2546-2553 (2012).
  14. Dettin, M., et al. Covalent surface modification of titanium oxide with different adhesive peptides: Surface characterization and osteoblast-like cell adhesion. J. Biomed. Mater. Res. Part A. 90 (1), 35-45 (2009).
  15. Kim, W. -. J., et al. Enhanced Protein Immobilization Efficiency on a TiO2 Surface Modified with a Hydroxyl Functional Group. Langmuir. 25 (19), 11692-11697 (2009).
  16. Son, K. J., Ahn, S. H., Kim, J. H., Koh, W. -. G. Graft Copolymer-Templated Mesoporous TiO2 Films Micropatterned with Poly(ethylene glycol) Hydrogel: Novel Platform for Highly Sensitive Protein Microarrays. ACS Appl. Mater. Interfaces. 3 (2), 573-581 (2011).
  17. Kar, P., Pandey, A., Greer, J. J., Shankar, K. Ultrahigh sensitivity assays for human cardiac troponin I using TiO2 nanotube arrays. Lab Chip. 12 (4), 821-828 (2012).
  18. Agarwal, S., Wendorff, J. H., Greiner, A. Use of electrospinning technique for biomedical applications. Polymer. 49 (26), 5603-5621 (2008).
  19. Ding, B., Wang, M., Wang, X., Yu, J., Sun, G. Electrospun nanomaterials for ultrasensitive sensors. Mater. Today. 13 (11), 16-27 (2010).
  20. Liu, Y., Yang, D., Yu, T., Jiang, X. Incorporation of electrospun nanofibrous PVDF membranes into a microfluidic chip assembled by PDMS and scotch tape for immunoassays. ELECTROPHORESIS. 30 (18), 3269-3275 (2009).
  21. Lee, W. S., Sunkara, V., Han, J. -. R., Park, Y. -. S., Cho, Y. -. K. Electrospun TiO2 nanofiber integrated lab-on-a-disc for ultrasensitive protein detection from whole blood. Lab Chip. 15 (2), 478-485 (2015).
  22. Li, D., Xia, Y. Fabrication of Titania Nanofibers by Electrospinning. Nano Lett. 3 (4), 555-560 (2003).
  23. Lombard, M. . SolidWorks 2013 BIBLE. , (2013).
  24. Tickoo, S. . EdgeCAM 11.0 for Manufacturers. , (2007).
  25. Zhu, R., et al. Improved adhesion of interconnected TiO2 nanofiber network on conductive substrate and its application in polymer photovoltaic devices. Appl. Phys. Lett. 93 (1), 013102 (2008).
  26. Song, M. Y., Ahn, Y. R., Jo, S. M., Kim, D. Y., Ahn, J. -. P. TiO2 single-crystalline nanorod electrode for quasi-solid-state dye-sensitized solar cells. Appl. Phys. Lett. 87 (11), 113113 (2005).
  27. Katsuhiro, O., et al. Electrospinning processed nanofibrous TiO2 membranes for photovoltaic applications. Nanotechnology. 17 (4), 1026-1031 (2006).
check_url/54143?article_type=t

Play Video

Cite This Article
Park, Y., Sunkara, V., Kim, Y., Lee, W. S., Han, J., Cho, Y. Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood. J. Vis. Exp. (110), e54143, doi:10.3791/54143 (2016).

View Video