Summary

والاستشعار القائمة على أبتمر لUnchelated الجادولينيوم (III)

Published: January 09, 2017
doi:

Summary

The use of polydeoxynucleotide (44-mer aptamer) molecules for sensing unchelated gadolinium(III) ion in an aqueous solution is described. The presence of the ion is detected via an increase in the fluorescence emission of the sensor.

Abstract

A method for determining the presence of unchelated trivalent gadolinium ion (Gd3+) in aqueous solution is demonstrated. Gd3+ is often present in samples of gadolinium-based contrast agents as a result of incomplete reactions between the ligand and the ion, or as a dissociation product. Since the ion is toxic, its detection is of critical importance. Herein, the design and usage of an aptamer-based sensor (Gd-sensor) for Gd3+ are described. The sensor produces a fluorescence change in response to increasing concentrations of the ion, and has a limit of detection in the nanomolar range (~100 nM with a signal-to-noise ratio of 3). The assay may be run in an aqueous buffer at ambient pH (~7 – 7.4) in a 384-well microplate. The sensor is relatively unreactive toward other physiologically relevant metal ions such as sodium, potassium, and calcium ions, although it is not specific for Gd3+ over other trivalent lanthanides such as europium(III) and terbium(III). Nevertheless, the lanthanides are not commonly found in contrast agents or the biological systems, and the sensor may therefore be used to selectively determine unchelated Gd3+ in aqueous conditions.

Introduction

وقد أدت الأهمية المتزايدة للتصوير بالرنين المغناطيسي (MRI) في التشخيص السريري، والذي يحد من حساسية الكامنة في أسلوب في النمو السريع للبحث في تطوير عوامل رواية القائم على الجادولينيوم المقابل (GBCAs) 1. GBCAs هي الجزيئات التي تدار لتحسين جودة الصورة، وأنها عادة ما يكون التركيب الكيميائي للأيون الثلاثي الجادولينيوم (ش ج 3+) منسقة ليجند polydentate. هذا complexation له أهمية حاسمة كما unchelated ش ج 3+ غير سامة. وقد تورط في تطوير تليف النظامية كلوي في بعض المرضى الذين يعانون من أمراض الكلى أو الفشل 2. ونتيجة لذلك، والكشف عن أيون حر مائي غير فعال في ضمان سلامة GBCAs. وجود unchelated ش ج 3+ في حلول GBCA غالبا ما يكون نتيجة رد فعل غير مكتمل بين يجند وأيون، تفكك المجمع، أو displacemenر من قبل غيرها من الكاتيونات المعدنية البيولوجية 3.

ومن بين العديد من التقنيات المستخدمة حاليا لتحديد وجود ش ج 3+، أولئك الذين يعتمدون على اللوني و / أو الطيف رتبة أعلى من حيث التنوع وتطبيق 4. بين مواطن القوة لديهم هي حساسية عالية ودقة، والقدرة على تحليل مختلف المصفوفات عينة (بما في ذلك المصل البشري 5 والبول والشعر 7 مياه الصرف الصحي، وتركيبات وكيل النقيض 8)، والكمي في وقت واحد من عدة ش ج 3+ المجمعات (قائمة من الدراسات قبل 2013 الموضح في مراجعة شاملة من قبل Telgmann وآخرون.) (4). لكن العائق الوحيد هو أن العديد من هذه الأساليب تتطلب أدوات القياس (مثل إضافة بالحث مطياف البلازما الشامل) 4 أن بعض المختبرات قد لا يكون الوصول إليها. في سياق اكتشاف GBCA الرواية في البحث ومستويات إثبات صحة مفهوم، عelatively أكثر ملاءمة، والسريع، وفعالة من حيث التكلفة طريقة القائم الطيفية (مثل امتصاص الأشعة فوق البنفسجية فيس أو مضان) قد يكون بمثابة بديل قيمة. مع هذه التطبيقات في الاعتبار، وقد وضعت الفلورسنت القائم على أبتمر الاستشعار عن مائي ش ج 3+ 9.

وأبتمر (ش ج-أبتمر) هو 44 قاعدة طويلة واحد الذين تقطعت بهم السبل جزيء الحمض النووي مع تسلسل معين من القواعد التي كانت معزولة من خلال عملية التطور المنهجي للبروابط كتبها تخصيب الأسي (سيليكس) 9. لتكييف أبتمر إلى جهاز استشعار الفلورسنت، ويرد على fluorophore إلى محطة 5 'من حبلا، وهو ما يتم تهجين مع حبلا التبريد (QS) عبر 13 قواعد تكميلية (الشكل 1). والموسومة QS مع جزيء فاكهه تقضي الظلام في 'محطة 3. في غياب ش ج 3+، أجهزة الاستشعار (ش ج استشعار)، تتكون من نسبة 1: الخلد 2 ش ج-أبتمر وQS على التوالي، وسيكون الحد الأدنى من الانبعاثات مضان ر المناسبس نقل الطاقة من fluorophore إلى فاكهه تقضي. وإضافة مائي ش ج 3+ تحل محل QS من ش ج-أبتمر، مما أدى إلى زيادة في الانبعاثات مضان.

شكل 1
الشكل 1. استشعار (ش ج استشعار) التي تتكون من أبتمر 44 قاعدة طويلة (ش ج-أبتمر) الموسومة مع فلوريسئين (أ fluorophore) و13 قاعدة تبريد طويل حبلا (QS) الموسومة مع dabcyl (فاكهه تقضي الظلام) . في غياب unchelated ش ج 3+، مضان من أجهزة الاستشعار هو الحد الأدنى. مع اضافة كلمة المدير العام 3+، تهجير QS يحدث ولوحظ حدوث زيادة في الانبعاثات مضان. الرجاء انقر هنا لعرض نسخة أكبر من هذا الرقم.

هناك في الوقت الحاضر، احد يشيع استخدامها أسلوب قائم على التحليل الطيفي للكشف عنجي مائي ش ج 3+. يستخدم هذا الاختبار البرتقال زايلينول جزيء، والتي تخضع لعملية التحول في أقصى الطول الموجي امتصاص 433-573 نانومتر على عملية إزالة معدن ثقيل إلى أيون 10. نسبة هذين الامتصاصية ماكسيما يمكن أن تستخدم لقياس كمية من unchelated ش ج 3+. أجهزة الاستشعار أبتمر هو بديل (قد تكون أيضا مكملة) لفحص زايلينول البرتقال، وطرق لهما ظروف التفاعل المختلفة (مثل درجة الحموضة وتكوين المحاليل المستخدمة)، الإختياريات الهدف، ويتراوح خطية من تقدير، وطرائق الكشف 9.

Protocol

ملاحظة: يتم استخدام الجزيئي المياه البيولوجيا الصف في كافة الاستعدادات العازلة والحل. جميع أنابيب القابل للتصرف (microcentrifuge لوPCR) ونصائح الماصة هي DNase- وريبونوكلياز خالية. يرجى الاطلاع على ورقة بيانات سلامة المواد (MSDS) لجميع المواد الكيميائية قبل استخدامها. ينصح بشدة اس…

Representative Results

ويرد التغيير مضان نموذجي من الحل ش ج استشعار بحضور unchelated ش ج 3+ في الشكل 2. قد يكون رسم انبعاث كما تغير مضان أضعاف (الشكل 2A) أو مضان الخام القراءة (الشكل 2B) في وحدات التعسفي (AFU). كلا المؤامرات تسفر عن منحنيات المعايرة متش?…

Discussion

باستخدام كلمة المدير العام استشعار القائم على أبتمر، وزيادة في الانبعاثات مضان أن يتناسب مع تركيز unchelated ش ج 3+ لوحظ. لتقليل كمية العينة المستخدمة، يمكن تشغيل الفحص في صفيحة 384 بشكل جيد مع حجم العينة ما مجموعه 45 ميكرولتر لكل بئر. في هذا التصميم، واختيار فلوريسئين …

Divulgazioni

The authors have nothing to disclose.

Acknowledgements

We would like to gratefully acknowledge Dr. Milan N Stojanovic from Columbia University, New York, NY for valuable scientific input. This work is supported by funding from the California State University East Bay (CSUEB) and the CSUEB Faculty Support Grant-Individual Researcher. O.E., T.C., and A.L. were supported by the CSUEB Center for Student Research (CSR) Fellowship.

Materials

Gd-aptamer IDTDNA Input sequence and fluorophore modification in the order form A fluorophore with a different emission wavelength may be used. The aptamer may also be ordered from another company.
Quenching strand IDTDNA Input sequence and quencher modification in the order form A different quencher for optimal energy transfer from the fluorophore may be used. The aptamer may also be ordered from another company. 
Molecular biology grade water No specific manufacturer, both DEPC or non-DEPC treated work equally well
Gadolinium(III) chloride anhydrous Strem 936416 Toxic
HEPES Fisher Scientific BP310-500
Magnesium chloride anhydrous MP Biomedicals 0520984480 – 100 g
Sodium Chloride Acros Organics 327300025
Potassium chloride Fisher Scientific P333-500
Sodium hydroxide, pellets Fisher Scientific BP359 Corrosive
Hydrochloric acid Fisher Scientific SA49 Toxic and corrosive
384-well low flange black flat bottom polystyrene NBS plates Corning 3575 Plates which are suitable for fluorescence reading are required. 
Nalgene Rapid-Flow sterile disposable bottle top filter Thermo Scientific  5680020 The bottle top is fitted with  0.2 micron PES membrane
Disposable sterile bottles 250 mL Corning 430281 A larger or smaller bottle may be used
1.5 mL microcentrifuge tubes No specific manufacturer, as long as they are DNAse and RNAse-free
0.2 mL PCR tubes No specific manufacturer, as long as they are DNAse and RNAse-free
Micropipets No specific manufacturer
Pipet tips (non filter) of appropriate sizes No specific manufacturer, as long as they are DNAse and RNAse-free
Name of Equipment
Plate reader Biotek Synergy H1 Plate readers from other manufacturers would work equally well

Riferimenti

  1. Shen, C., New, E. J. Promising strategies for Gd-based responsive magnetic resonance imaging contrast agents. Curr. Opin. Chem. Biol. 17 (2), 158-166 (2013).
  2. Cheong, B. Y. C., Muthupillai, R. Nephrogenic systemic fibrosis: a concise review for cardiologists. Tex. Heart Inst. J. 37 (5), 508-515 (2010).
  3. Hao, D., Ai, T., Goerner, F., Hu, X., Runge, V. M., Tweedle, M. MRI contrast agents: basic chemistry and safety. J Magn. Reson. Imaging. 36 (5), 1060-1071 (2012).
  4. Telgmann, L., Sperling, M., Karst, U. Determination of gadolinium-based MRI contrast agents in biological and environmental samples: a review. Anal. Chim. Acta. 764, 1-16 (2013).
  5. Frenzel, T., Lengsfeld, P., Schirmer, H., Hütter, J., Weinmann, H. -. J. Stability of gadolinium-based magnetic resonance imaging contrast agents in human serum at 37 degrees C. Invest. Radiol. 43 (12), 817-828 (2008).
  6. Loreti, V., Bettmer, J. Determination of the MRI contrast agent Gd-DTPA by SEC-ICP-MS. Anal. Bioanal. Chem. 379 (7), 1050-1054 (2004).
  7. Telgmann, L., et al. Speciation and isotope dilution analysis of gadolinium-based contrast agents in wastewater. Environ. Sci. Technol. 46 (21), 11929-11936 (2012).
  8. Cleveland, D., et al. Chromatographic methods for the quantification of free and chelated gadolinium species in MRI contrast agent formulations. Anal. Bioanal. Chem. 398 (7), 2987-2995 (2010).
  9. Edogun, O., Nguyen, N. H., Halim, M. Fluorescent single-stranded DNA-based assay for detecting unchelated gadolinium(III) ions in aqueous solution. Anal. Bioanal. Chem. 408 (15), 4121-4131 (2016).
  10. Barge, A., Cravotto, G., Gianolio, E., Fedeli, F. How to determine free Gd and free ligand in solution of Gd chelates. A technical note. Contrast Med. Mol. Imaging. 1 (5), 184-188 (2006).
  11. Johansson, M. K. Choosing reporter-quencher pairs for efficient quenching through formation of intramolecular dimers. Methods Mol. Biol. 335, 17-29 (2006).
  12. Sherry, A. D., Caravan, P., Lenkinski, R. E. A primer on gadolinium chemistry. J. Magn. Reson. Imaging. 30 (6), 1240-1248 (2009).
  13. Shakhverdov, T. A. A cross-relaxation mechanism of fluorescence quenching in complexes of lanthanide ions with organic ligands. Opt. Spectrosc. 95 (4), 571-580 (2003).
  14. Brittain, H. G. Submicrogram determination of lanthanides through quenching of calcein blue fluorescence. Anal. Chem. 59 (8), 1122-1125 (1987).
check_url/it/55216?article_type=t

Play Video

Citazione di questo articolo
Edogun, O., Chan, T. Y., Nguyen, N. H., Luu, A., Halim, M. An Aptamer-based Sensor for Unchelated Gadolinium(III). J. Vis. Exp. (119), e55216, doi:10.3791/55216 (2017).

View Video