Flujo de microfluidos Cámaras utilizando sangre reconstituida a Modelo Hemostasia y Transfusión de Plaquetas<em> In Vitro</em

Published: March 19, 2016
doi:

Summary

Platelet transfusion and hemostasis was modeled using blood reconstitution and microfluidic flow chambers to investigate the function of blood banking platelets. The data demonstrate the consequences of platelet storage lesion on hemostasis, in vitro.

Abstract

Blood platelets prepared for transfusion gradually lose hemostatic function during storage. Platelet function can be investigated using a variety of (indirect) in vitro experiments, but none of these is as comprehensive as microfluidic flow chambers. In this protocol, the reconstitution of thrombocytopenic fresh blood with stored blood bank platelets is used to simulate platelet transfusion. Next, the reconstituted sample is perfused in microfluidic flow chambers which mimic hemostasis on exposed subendothelial matrix proteins. Effects of blood donation, transport, component separation, storage and pathogen inactivation can be measured in paired experimental designs. This allows reliable comparison of the impact every manipulation in blood component preparation has on hemostasis. Our results demonstrate the impact of temperature cycling, shear rates, platelet concentration and storage duration on platelet function. In conclusion, this protocol analyzes the function of blood bank platelets and this ultimately aids in optimization of the processing chain including phlebotomy, transport, component preparation, storage and transfusion.

Introduction

Hemostasis requiere la actividad combinada y regulada de células, proteínas, iones y tejidos en un contexto espacio-temporal restringida 1. actividad no controlada puede conducir a hemorragia o trombosis y la morbilidad o la mortalidad en un espectro de trastornos relacionados con la coagulación de la sangre. Un experimento de la cámara de flujo de microfluidos es una técnica difícil que imita hemostasis in vitro. Este enfoque permite la investigación de la compleja interacción de procesos que intervienen en la hemostasia con el protagonismo de las plaquetas sanguíneas.

Después de una lesión vascular, las plaquetas se adhieren a las proteínas de la matriz subendotelial expuesta (glico) para prevenir la pérdida de sangre. Después de la adhesión, las plaquetas se activan y el agregado en respuesta a auto- y señalización paracrina que finalmente conduce a la formación de una red de plaquetas, fibrina y estabilizada por lo que resulta en una firma, la herida trombo 2 de sellado. A diferencia de la mayoría de las otras pruebas de la función plaquetaria, expetos con cámaras de flujo tienen en cuenta el parámetro físico del flujo sanguíneo y por lo tanto la influencia de la reología en las células y biomoléculas 3,4 participantes.

experimentos en cámara de flujo han proporcionado multitud de datos de la señal en la hemostasia y la trombosis mediante la variación de los parámetros clave que influyen hemostático (sub) procesos, incluyendo los perfiles de matriz adhesiva, reología y de flujo, composición celular, presencia de toxinas o drogas, fuerza iónica y muchos más. En las últimas dos décadas, experimentos con la cámara de flujo bajo rendimiento que requieren grandes volúmenes de muestra (10-100 ml) se han desarrollado para cámaras de microfluidos menudo consisten en pequeñas cámaras de placas paralelas y que incluyen la tecnología moderna para la perfusión de sangre total en condiciones controladas muro de corte 5. Microscaling ha aumentado significativamente el rendimiento de ensayo sobre todo porque la instalación del hardware ha simplificado y se requiere menos volumen (sangre), lo que hace el experimento más accesible y VersatiLe. Por ejemplo, la sangre de los animales pequeños de laboratorio ahora puede ser utilizado sin la necesidad de sacrificar los animales. De este modo muestras de sangre de ratones modificados genéticamente han ayudado en la identificación de moléculas clave promover o inhibir la hemostasis y en nuevos conocimientos básicos 6.

Laboratorios de investigación especializados a menudo todavía utilizan cámaras de flujo a la medida, por ejemplo de polidimetilsiloxano (PDMS) 7 que polimeriza en moldes litografiado que puede ser blueprinted por software. La cámara resultante es barato, desechable y se puede desmontar fácilmente para el análisis post hoc. Por otra parte, básicamente cualquier diseño de embarcaciones, incluyendo las bifurcaciones o giros bruscos se puede construir en el comando. Esta ventaja es también su lado negativo porque su normalización ya era el principal problema con experimentos en cámara de flujo, y PDMS encargo cámaras no han ayudado esto. En la parte superior de este tema en particular, recubrimiento (condiciones), sondas fluorescentes, anticoagulante, tempratura y el tiempo entre el muestreo y análisis están mal estandarizados 8. La estandarización de estas variables es un reto, pero no obstante necesario para permitir la comparación de resultados entre laboratorios. Este tema es el tema principal de la Sociedad Internacional de Trombosis y Hemostasia en subcomité científico y Normalización en Biorheology 9,10.

Los concentrados de plaquetas (PC) se transfunden en pacientes que sufren de diversas enfermedades que causan trombocitopenia y / o sangrado. Pero plaquetas en PC son conocidos para desensibilizar, especialmente en función del tiempo de almacenamiento 11, un proceso de deterioro relacionado con el envejecimiento y comúnmente conocida como lesión de almacenamiento de plaquetas. A veces se afirma que tales plaquetas restaurar en circulación una vez transfundida 12, pero la evidencia de esto es escasa. Además, la funcionalidad de las plaquetas que forman un PC no se prueba de forma rutinaria debido a que la relación entre tales ensayos yla eficacia terapéutica o profiláctica está claro 13. cámaras de flujo de microfluidos ofrecen un medio para investigar la función plaquetaria en PC para optimizar la cadena de manipulaciones entre la recogida y entrega. Es una poderosa herramienta de investigación para las comparaciones directas (pares) de PC como hemos previamente publicados 14,15 y se describe aquí.

Protocol

Este protocolo sigue las directrices éticas institucionales para la investigación en muestras humanas y el consentimiento informado se obtuvo de todos los donantes involucrados. La aprobación de los experimentos descritos aquí se obtuvo de la junta de revisión institucional del Hospital de la Universidad de Amberes. Nota: Las indicaciones de temperatura son siempre la temperatura ambiente, a menos que se especifique. 1. Configuración de la cámara de flujo Pr…

Representative Results

Para demostrar la variación intra-ensayo, tres idénticas muestras de sangre entera reconstituidas fueron perfundidos simultáneamente sobre superficies revestidas de colágeno (Figura 1). Esto dio lugar a un coeficiente de variación de 8,7%. Esta estadística sugiere intra-ensayo aceptable y la variación intralaboratorio que permite una comparación fiable entre muestras relacionadas. La entrada de la cám…

Discussion

Experimentos en cámara de flujo de microfluidos son una excelente herramienta para investigar la función de las plaquetas en la sangre que fluye y se utilizan para evaluar la hemostasia in vitro en diversos contextos experimentales. A pesar de su falta de estandarización entre laboratorios 9, se demuestra que dentro de nuestro laboratorio de la variación experimental es aceptable. Esto permite comparar de forma fiable muestras pareadas () dentro de un determinado estudio. Este fue validado usando…

Disclosures

The authors have nothing to disclose.

Acknowledgements

The authors have no acknowledgements.

Materials

BD vacutainer tube with EDTA  Becton, Dickinson and Company 368856
BD vacutainer tube with Heparin Becton, Dickinson and Company 368480
BD vacutainer tube with Sodium Citrate Becton, Dickinson and Company 366575
Hirudin Blood tube Roche 6675751 001
BD vacutainer Eclipse Becton, Dickinson and Company 368650 Blood collection needle with preattached holder
Pipette tips 100-1000 Greiner bio-one 740290
Pipette tips 2-200 Greiner bio-one 739280
Pipette tips 1-10 Eppendorf A08928
Tube 5mL Simport 11691380
Conical tube 15mL Greiner bio-one 1888271
Conical tube 50mL Greiner bio-one 227261
10 mL Syringe BD 309604
Precision wipes Kimtech 5511
Vena8 Fluoro+ Biochips Cellix 188V8CF-400-100-02P10 Named in figure S1 A as 'Biochip'
Vena8 Tubing Cellix TUBING-TYGON-B1IC-B1OC-ROLL 100F Named in figure S1 B as 'Disposable tubing'
Vena8 Needles Cellix SS-P-B1IC-B1OC-PACK200 Named in figure S1 B as 'Pin'
Connectors for single inlet cables of biochips Cellix CONNECTORS-B1IC-PACK100
Multiflow8 connect Cellix MF8-CONNECT-BIC3-N-THROMBOSIS Named in figure S1 B as 'Reusable tubing' and 'Splitter'
Humidified box Cellix HUMID-BOX
Software microfluidic pump Cellix N/A Venaflux Assay
Horm Collagen Takeda/Nycomed 1130630 Native equine tendon collagen (type I)
Isotonic glucose solution to dilute collagen is supplemented
HEPES buffered saline (HBS) in house preparation in house preparation 10mM 4-(2-hydroxyethyl)-1-piperazineethanesulfonic acid (HEPES) buffered saline (0.9% (w/v) NaCl, pH 7.4
Blocking buffer in house preparation in house preparation 1.0% (w/v) bovine serum albumin and 0.1% (w/v) glucose in HBS
Calcein AM Molecular probes C1430
Bleach 10% in house preparation in house preparation
0.1M NaOH in house preparation in house preparation
Denaturated alcohol Fiers T0011.5
Mirus Evo Nanopump Cellix 188-MIRUS-PUMP-EVO with Multiflow8. Named in figure S1 A as 'Pump' and 'Manifold'
Microscope Zeiss Axio Observer Z1 equipped with a colibri-LED and high resolution CCD camera
Software microscope  Zeiss N/A ZEN 2012
Hematology analyzer Sysmex N/A
Table Top Centrifuge Eppendorf 521-0095
Platelet incubater Helmer PF-48i
Incubation water bath GFL 1013
Pipette Brand A03429
Tube Roller Ratek BTR5-12V
Sterile docking device Terumo BCT TSCD
Tubing Sealer Terumo BCT AC-155
Vortex VWR 58816-121

References

  1. Broos, K., Feys, H. B., De Meyer, S. F., Vanhoorelbeke, K., Deckmyn, H. Platelets at work in primary hemostasis. Blood Rev. 25, 155-167 (2011).
  2. Stalker, T. J., et al. Hierarchical organization in the hemostatic response and its relationship to the platelet-signaling network. Blood. 121, 1875-1885 (2013).
  3. Sakariassen, K. S., Bolhuis, P. A., Sixma, J. J. Human blood platelet adhesion to artery subendothelium is mediated by factor VIII-Von Willebrand factor bound to the subendothelium. Nature. 279, 636-638 (1979).
  4. Savage, B., Saldivar, E., Ruggeri, Z. M. Initiation of platelet adhesion by arrest onto fibrinogen or translocation on von Willebrand factor. Cell. 84, 289-297 (1996).
  5. Westein, E., de Witt, S., Lamers, M., Cosemans, J. M., Heemskerk, J. W. Monitoring in vitro thrombus formation with novel microfluidic devices. Platelets. 23, 501-509 (2012).
  6. Varga-Szabo, D., et al. The calcium sensor STIM1 is an essential mediator of arterial thrombosis and ischemic brain infarction. J Exp Med. 205, 1583-1591 (2008).
  7. Westein, E., et al. Atherosclerotic geometries exacerbate pathological thrombus formation poststenosis in a von Willebrand factor-dependent manner. Proc Natl Acad Sci USA. , (2013).
  8. Grabowski, E. F., Yam, K., Gerace, M. Evaluation of hemostasis in flowing blood. Am J Hematol. 87 (Suppl 1), S51-S55 (2012).
  9. Roest, M., et al. Flow chamber-based assays to measure thrombus formation in vitro: requirements for standardization. J Thromb Haemost. 9, 2322-2324 (2011).
  10. Heemskerk, J. W. M., et al. Collagen surfaces to measure thrombus formation under flow: possibilities for standardization. J Thromb Haemost. 9, 856-858 (2011).
  11. Shrivastava, M. The platelet storage lesion. Transfus Apher Sci. 41, 105-113 (2009).
  12. Miyaji, R., et al. Decreased platelet aggregation of platelet concentrate during storage recovers in the body after transfusion. Transfusion. 44, 891-899 (2004).
  13. Goodrich, R. P., et al. Correlation of in vitro platelet quality measurements with in vivo platelet viability in human subjects. Vox Sang. 90, 279-285 (2006).
  14. Van Aelst, B., et al. Riboflavin and amotosalen photochemical treatments of platelet concentrates reduce thrombus formation kinetics in vitro. Vox Sang. 108, 328-339 (2015).
  15. Van Aelst, B., Devloo, R., Vandekerckhove, P., Compernolle, V., Feys, H. B. Ultraviolet c light pathogen inactivation treatment of platelet concentrates preserves integrin activation but affects thrombus formation kinetics on collagen in vitro. Transfusion. , (2015).
  16. de Witt, S. M., et al. Identification of platelet function defects by multi-parameter assessment of thrombus formation. Nat Commun. 5, 4257 (2014).
  17. Cazenave, J. P., et al. Preparation of washed platelet suspensions from human and rodent blood. Methods Mol Biol. 272, 13-28 (2004).
  18. Jackson, S. P., Nesbitt, W. S., Westein, E. Dynamics of platelet thrombus formation. J Thromb Haemost. 7 (Suppl 1), 17-20 (2009).
  19. Diedrich, B., Remberger, M., Shanwell, A., Svahn, B. M., Ringden, O. A prospective randomized trial of a prophylactic platelet transfusion trigger of 10 x 10(9) per L versus 30 x 10(9) per L in allogeneic hematopoietic progenitor cell transplant recipients. Transfusion. 45, 1064-1072 (2005).
  20. Neeves, K. B., et al. Sources of variability in platelet accumulation on type 1 fibrillar collagen in microfluidic flow assays. PloS one . 8, e54680 (2013).
  21. Born, G. V. Aggregation of blood platelets by adenosine diphosphate and its reversal. Nature. 194, 927-929 (1962).
  22. Cattaneo, M., et al. Recommendations for the Standardization of Light Transmission Aggregometry: A Consensus of the Working Party from the Platelet Physiology Subcommittee of SSC/ISTH. J Thromb Haemost. 11, 1183-1189 (2013).
  23. Deckmyn, H., Feys, H. B. Assays for quality control of platelets for transfusion. ISBT Sci Series. 8, 221-224 (2013).
  24. Andre, P., et al. Anticoagulants (thrombin inhibitors) and aspirin synergize with P2Y12 receptor antagonism in thrombosis. Circulation. 108, 2697-2703 (2003).
  25. Casari, C., et al. von Willebrand factor mutation promotes thrombocytopathy by inhibiting integrin alphaIIbbeta3. J Clin Invest. 123, 5071-5081 (2013).
  26. Gitz, E., et al. Improved platelet survival after cold storage by prevention of Glycoprotein Ibα clustering in lipid rafts. Haematologica. , (2012).
  27. Maurer-Spurej, E., Labrie, A., Brown, K. Routine Quality Testing of Blood Platelet Transfusions with Dynamic Light Scattering. Part Part Syst Char. 25, 99-104 (2008).
  28. Van Kruchten, R., Cosemans, J. M., Heemskerk, J. W. Measurement of whole blood thrombus formation using parallel-plate flow chambers – a practical guide. Platelets. 23, 229-242 (2012).
  29. Zwaginga, J. J., et al. Flow-based assays for global assessment of hemostasis. Part 2: current methods and considerations for the future. J Thromb Haemost. 4, 2716-2717 (2006).
  30. Zwaginga, J. J., et al. Flow-based assays for global assessment of hemostasis. Part 1: Biorheologic considerations. J Thromb Haemost. 4, 2486-2487 (2006).
check_url/cn/53823?article_type=t

Play Video

Cite This Article
Van Aelst, B., Feys, H. B., Devloo, R., Vandekerckhove, P., Compernolle, V. Microfluidic Flow Chambers Using Reconstituted Blood to Model Hemostasis and Platelet Transfusion In Vitro. J. Vis. Exp. (109), e53823, doi:10.3791/53823 (2016).